Published in: Clinical Orthopaedics and Related Research® 7/2014

01-07-2014 | Clinical Research

Current Modes of Failure in TKA: Infection, Instability, and Stiffness Predominate

Authors: David H. Le, MD, Stuart B. Goodman, MD, PhD, William J. Maloney, MD, James I. Huddleston, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 7/2014

Login to get access

Abstract

Background

Historically, polyethylene wear and its sequelae (osteolysis, late instability, aseptic loosening) were common causes for revision total knee arthroplasty (TKA). Recently, polyethylene manufacturing has become more consistent; furthermore, a clearer understanding of the importance of oxidation on polyethylene performance led to packaging of the polyethylene bearings in an inert environment. This improved the quality and consistency of polyethylene used in TKA, raising the question of whether different failure modes now predominate after TKA.

Questions/purposes

The purpose of this study was to determine the current reasons for (1) early and (2) late failures after TKA at one high-volume arthroplasty center.

Methods

We reviewed all first-time revision TKAs performed between 2001 and 2011 at one institution, yielding a group of 253 revision TKAs in 251 patients. Mean age at the time of revision was 64 years (SD 10 years). Mean time to revision was 35 months (SD 23 months). Preoperative evaluations, laboratory data, radiographs, and intraoperative findings were used to determine causes for revision. Early failure was defined as revision within 2 years of the index procedure. The primary failure mechanism was determined by the operating surgeon.

Results

Early failure accounted for 46% (116 of 253) of all revisions with infection (28 of 116 [24%]), instability (30 of 116 [26%]), and stiffness (21 of 116 [18%]) being the leading causes. Late failure accounted for 54% (137 of 253) of all revisions with the most common causes including infection (34 of 137 [25%]), instability (24 of 137 [18%]), and stiffness (19 of 253 [14%]). Polyethylene wear was implicated as the failure mechanism in 2% of early cases (two of 116) and 9% of late cases (13 of 137).

Conclusions

In contrast to previous studies, wear-related implant failure in TKA was relatively uncommon in this series. Changes in polyethylene manufacturing, sterilization, and storage may have accounted for some of this difference; however, longer-term followup will be required to verify this finding. Infection, instability, and stiffness represent the most common causes of early and late failure. Strategies to improve outcomes in TKA should be aimed at infection prophylaxis and treatment, surgical technique, and patient selection.

Level of Evidence

Level III, therapeutic study. See Instructions for Authors for a complete description of levels of evidence.
Literature
1.
go back to reference Berzins A, Jacobs JJ, Berger R, Ed C, Natarajan R, Andriacchi T, Galante JO. Surface damage in machine ram-extruded and net-shape molded retrieved polyethylene tibial inserts in total knee replacements. J Bone Joint Surg Am. 2002;84:1534–1540.PubMed Berzins A, Jacobs JJ, Berger R, Ed C, Natarajan R, Andriacchi T, Galante JO. Surface damage in machine ram-extruded and net-shape molded retrieved polyethylene tibial inserts in total knee replacements. J Bone Joint Surg Am. 2002;84:1534–1540.PubMed
2.
go back to reference Bozic KJ, Kurtz SM, Lau E, Ong K, Chiu V, Vail TP, Rubash HE, Berry DJ. The epidemiology of revision total knee arthroplasty in the United States. Clin Orthop Relat Res. 2010;468:45–51.PubMedCentralPubMedCrossRef Bozic KJ, Kurtz SM, Lau E, Ong K, Chiu V, Vail TP, Rubash HE, Berry DJ. The epidemiology of revision total knee arthroplasty in the United States. Clin Orthop Relat Res. 2010;468:45–51.PubMedCentralPubMedCrossRef
3.
go back to reference Callaghan JJ, O’Rourke MR, Saleh KJ. Why knees fail: lessons learned. J Arthroplasty. 2004;19:31–34.PubMedCrossRef Callaghan JJ, O’Rourke MR, Saleh KJ. Why knees fail: lessons learned. J Arthroplasty. 2004;19:31–34.PubMedCrossRef
4.
go back to reference Collier MB, Engh CA Jr, McAuley JP, Ginn SD, Engh GA. Osteolysis after total knee arthroplasty: influence of tibial baseplate surface finish and sterilization of polyethylene insert: findings at five to ten years postoperatively. J Bone Joint Surg Am. 2005;87:2702–2708.PubMedCrossRef Collier MB, Engh CA Jr, McAuley JP, Ginn SD, Engh GA. Osteolysis after total knee arthroplasty: influence of tibial baseplate surface finish and sterilization of polyethylene insert: findings at five to ten years postoperatively. J Bone Joint Surg Am. 2005;87:2702–2708.PubMedCrossRef
5.
go back to reference Fehring TK, Odum S, Griffin WL, Mason J, Nadaud M. Early failures in total knee arthroplasty. Clin Orthop Relat Res. 2001;392:315–318.PubMedCrossRef Fehring TK, Odum S, Griffin WL, Mason J, Nadaud M. Early failures in total knee arthroplasty. Clin Orthop Relat Res. 2001;392:315–318.PubMedCrossRef
6.
go back to reference Gioe TJ, Killeen KK, Grimm K, Mehle S, Scheltema K. Why are total knee replacements revised? Analysis of early revision in a community knee implant registry. Clin Orthop Relat Res. 2004;428:100–106.PubMedCrossRef Gioe TJ, Killeen KK, Grimm K, Mehle S, Scheltema K. Why are total knee replacements revised? Analysis of early revision in a community knee implant registry. Clin Orthop Relat Res. 2004;428:100–106.PubMedCrossRef
7.
go back to reference Griffin WL, Fehring TK, Pomeroy DL, Gruen TA, Murphy JA. Sterilization and wear-related failure in first- and second-generation press-fit condylar total knee arthroplasty. Clin Orthop Relat Res. 2007;464:16–20.PubMed Griffin WL, Fehring TK, Pomeroy DL, Gruen TA, Murphy JA. Sterilization and wear-related failure in first- and second-generation press-fit condylar total knee arthroplasty. Clin Orthop Relat Res. 2007;464:16–20.PubMed
8.
go back to reference Hossain F, Patel S, Haddad FS. Midterm assessment of causes and results of revision total knee arthroplasty. Clin Orthop Relat Res. 2010;468:1221–1228.PubMedCentralPubMedCrossRef Hossain F, Patel S, Haddad FS. Midterm assessment of causes and results of revision total knee arthroplasty. Clin Orthop Relat Res. 2010;468:1221–1228.PubMedCentralPubMedCrossRef
9.
go back to reference Kasahara Y, Kimura S, Nishiike O, Uchida J. What are the causes of revision total knee arthroplasty in Japan? Clin Orthop Relat Res. 2013;471:1533–538.PubMedCentralPubMedCrossRef Kasahara Y, Kimura S, Nishiike O, Uchida J. What are the causes of revision total knee arthroplasty in Japan? Clin Orthop Relat Res. 2013;471:1533–538.PubMedCentralPubMedCrossRef
10.
go back to reference Kurtz S, Ong K, Lau E, Mowat F, Halpern M. Projections of primary and revision hip and knee arthroplasty in the United States from 2005–2030. J Bone Joint Surg Am. 2007;89:780–785.PubMedCrossRef Kurtz S, Ong K, Lau E, Mowat F, Halpern M. Projections of primary and revision hip and knee arthroplasty in the United States from 2005–2030. J Bone Joint Surg Am. 2007;89:780–785.PubMedCrossRef
11.
go back to reference Kurtz, SM, Ong KL, Schmier J, Mowat F, Saleh K, Dybvik E, Karrholm J, Garellick G, Havelin LI, Furnes O, Malchau H, Lau E. Future clinical and economic impact of revision total hip and knee arthroplasty. J Bone Joint Surg Am. 2007;89(Suppl 3):144–151.PubMedCrossRef Kurtz, SM, Ong KL, Schmier J, Mowat F, Saleh K, Dybvik E, Karrholm J, Garellick G, Havelin LI, Furnes O, Malchau H, Lau E. Future clinical and economic impact of revision total hip and knee arthroplasty. J Bone Joint Surg Am. 2007;89(Suppl 3):144–151.PubMedCrossRef
12.
go back to reference Medel FJ, Kurtz SM, Hozack WJ, Parvizi J, Purtill JJ, Sharkey PF, MacDonald K, Kraay MJ, Goldberg V, Rimnac CM. Gamma inert sterilization: a solution to polyethylene oxidation? J Bone Joint Surg Am. 2009;91:839–849.PubMedCentralPubMedCrossRef Medel FJ, Kurtz SM, Hozack WJ, Parvizi J, Purtill JJ, Sharkey PF, MacDonald K, Kraay MJ, Goldberg V, Rimnac CM. Gamma inert sterilization: a solution to polyethylene oxidation? J Bone Joint Surg Am. 2009;91:839–849.PubMedCentralPubMedCrossRef
13.
go back to reference Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM. Why are total knee arthroplasties failing today? Clin Orthop Relat Res. 2002;404:7–13.PubMedCrossRef Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM. Why are total knee arthroplasties failing today? Clin Orthop Relat Res. 2002;404:7–13.PubMedCrossRef
15.
go back to reference Vessely MB, Whaley AL, Harmsen WS, Schleck CD, Berry DJ. Long-term survivorship and failure modes of 1000 cemented condylar total knee arthroplasties. Clin Orthop Relat Res. 2006;452:28–34.PubMedCrossRef Vessely MB, Whaley AL, Harmsen WS, Schleck CD, Berry DJ. Long-term survivorship and failure modes of 1000 cemented condylar total knee arthroplasties. Clin Orthop Relat Res. 2006;452:28–34.PubMedCrossRef
17.
go back to reference Weinstein AM, Rome BN, Reichmann WM, Collins JE, Burbine SA, Thornhill TS, Wright J, Katz JN, Losina E. Estimating the burden of total knee replacement in the United States. J Bone Joint Surg Am. 2013;95:385–392.PubMedCentralPubMedCrossRef Weinstein AM, Rome BN, Reichmann WM, Collins JE, Burbine SA, Thornhill TS, Wright J, Katz JN, Losina E. Estimating the burden of total knee replacement in the United States. J Bone Joint Surg Am. 2013;95:385–392.PubMedCentralPubMedCrossRef
Metadata
Title
Current Modes of Failure in TKA: Infection, Instability, and Stiffness Predominate
Authors
David H. Le, MD
Stuart B. Goodman, MD, PhD
William J. Maloney, MD
James I. Huddleston, MD
Publication date
01-07-2014
Publisher
Springer US
Published in
Clinical Orthopaedics and Related Research® / Issue 7/2014
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-014-3540-y

Other articles of this Issue 7/2014

Clinical Orthopaedics and Related Research® 7/2014 Go to the issue

Symposium: Traumatic Elbow Instability and its Sequelae

Anconeus Interposition Arthroplasty: Mid- to Long-term Results