Published in: Clinical Orthopaedics and Related Research® 2/2013

01-02-2013 | Symposium: Papers Presented at the Annual Meetings of The Hip Society

The 2012 Frank Stinchfield Award: Decreasing Patient Activity With Aging: Implications for Crosslinked Polyethylene Wear

Authors: Andrew K. Battenberg, BS, Jeffrey S. Hopkins, MD, Andrew D. Kupiec, Thomas P. Schmalzried, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 2/2013

Login to get access

Abstract

Background

Patient activity influences polyethylene wear. However, it is unclear how individual activity changes with patient aging after THA.

Questions/purposes

We quantified changes in individual gait cycles and gait speed, assessed age-related differences in these parameters, and determined their relationship to polyethylene wear.

Methods

A microprocessor was worn on the ankle to quantify the activity of 14 healthy patients with a well-functioning THA at two time periods: early (within 3.5 years of implantation) and late (10–13 postoperative years). Wear was measured on serial radiographs using edge detection-based software.

Results

Mean activity decreased by 16% from the early to the late period: 2.04 million gait cycles/year to 1.71 million gait cycles/year. Mean gait speed decreased by 9%: 15.4 cycles/minute to 14.0 cycles/minute. The activity of the 10 patients who were younger than 65 years at surgery decreased by 14% (2.34 million gait cycles/year to 2.02 million gait cycles/year), while the four patients 65 years or older at surgery decreased by 28% (1.29 million gait cycles/year to 0.94 million gait cycles/year). Gait speed was 26% slower for patients 65 years or older than for patients younger than 65 years. The mean linear penetration rate decreased by 42% from the first 5 years (early wear rate) to the next 8 years (late wear rate, 5–13 years): 0.043 mm/year to 0.025 mm/year.

Conclusions

The greatest patient activity and wear occurred during the first 5 years. Walking speed and gait cycles both decreased with aging, resulting in deceasing wear over time.
Literature
1.
go back to reference Bitsch RG, Loidolt T, Heisel C, Ball S, Schmalzried TP. Reduction of osteolysis with use of Marathon cross-linked polyethylene: a concise follow-up, at a minimum of five years, of a previous report. J Bone Joint Surg Am. 2008;90:1487–1491.PubMedCrossRef Bitsch RG, Loidolt T, Heisel C, Ball S, Schmalzried TP. Reduction of osteolysis with use of Marathon cross-linked polyethylene: a concise follow-up, at a minimum of five years, of a previous report. J Bone Joint Surg Am. 2008;90:1487–1491.PubMedCrossRef
2.
go back to reference Charnley J. The long-term results of low-friction arthroplasty of the hip performed as a primary intervention. J Bone Joint Surg Br. 1972;54:61–76.PubMed Charnley J. The long-term results of low-friction arthroplasty of the hip performed as a primary intervention. J Bone Joint Surg Br. 1972;54:61–76.PubMed
3.
go back to reference Clarke IC, Manley MT. How do alternative bearing surfaces influence wear behavior? Implant Wear Symposium 2007: Engineering Work Group. J Am Acad Orthop Surg. 2008;16(suppl 1):S86–S93.PubMed Clarke IC, Manley MT. How do alternative bearing surfaces influence wear behavior? Implant Wear Symposium 2007: Engineering Work Group. J Am Acad Orthop Surg. 2008;16(suppl 1):S86–S93.PubMed
4.
go back to reference Dumbleton JH, Manley MT, Edidin AA. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty. 2002;17:649–661.PubMedCrossRef Dumbleton JH, Manley MT, Edidin AA. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty. 2002;17:649–661.PubMedCrossRef
5.
go back to reference Kurtz SM, Gawel HA, Patel JD. History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:2262–2277.PubMedCrossRef Kurtz SM, Gawel HA, Patel JD. History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:2262–2277.PubMedCrossRef
6.
go back to reference Lachiewicz PF, Heckman DS, Soileau ES, Mangla J, Martell JM. Femoral head size and wear of highly cross-linked polyethylene at 5 to 8 years. Clin Orthop Relat Res. 2009;467:3290–3296.PubMedCrossRef Lachiewicz PF, Heckman DS, Soileau ES, Mangla J, Martell JM. Femoral head size and wear of highly cross-linked polyethylene at 5 to 8 years. Clin Orthop Relat Res. 2009;467:3290–3296.PubMedCrossRef
7.
go back to reference Leung SB, Egawa H, Stepniewski A, Beykirch S, Engh CA Jr, Engh CA Sr. Incidence and volume of pelvic osteolysis at early follow-up with highly cross-linked and noncross-linked polyethylene. J Arthroplasty. 2007;22:134–139.PubMedCrossRef Leung SB, Egawa H, Stepniewski A, Beykirch S, Engh CA Jr, Engh CA Sr. Incidence and volume of pelvic osteolysis at early follow-up with highly cross-linked and noncross-linked polyethylene. J Arthroplasty. 2007;22:134–139.PubMedCrossRef
8.
go back to reference Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am. 1997;79:1635–1641.PubMed Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am. 1997;79:1635–1641.PubMed
9.
go back to reference McKellop H, Shen FW, DiMaio W, Lancaster JG. Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls. Clin Orthop Relat Res. 1999;369:73–82.PubMedCrossRef McKellop H, Shen FW, DiMaio W, Lancaster JG. Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls. Clin Orthop Relat Res. 1999;369:73–82.PubMedCrossRef
10.
go back to reference McKellop HA. Laboratory wear testing. In: Callaghan JJ, Rosenberg AG, Rubash H, eds. The Adult Hip. Philadelphia, PA: JB Lippincott; 1998. McKellop HA. Laboratory wear testing. In: Callaghan JJ, Rosenberg AG, Rubash H, eds. The Adult Hip. Philadelphia, PA: JB Lippincott; 1998.
11.
go back to reference Schmalzried TP, Shepherd EF, Dorey FJ, Jackson WO, dela Rosa M, Favae F, McKellop HA, McClung CD, Martell J, Moreland JR, Amstutz HC. Wear is a function of use, not time. Clin Orthop Relat Res. 2000;381:36–46.PubMedCrossRef Schmalzried TP, Shepherd EF, Dorey FJ, Jackson WO, dela Rosa M, Favae F, McKellop HA, McClung CD, Martell J, Moreland JR, Amstutz HC. Wear is a function of use, not time. Clin Orthop Relat Res. 2000;381:36–46.PubMedCrossRef
12.
go back to reference Schmalzried TP, Szuszczewicz ES, Northfield MR, Akizuki KH, Frankel RE, Belcher G, Amstutz HC. Quantitative assessment of walking activity after total hip or knee replacement. J Bone Joint Surg Am. 1998;80:54–59.PubMedCrossRef Schmalzried TP, Szuszczewicz ES, Northfield MR, Akizuki KH, Frankel RE, Belcher G, Amstutz HC. Quantitative assessment of walking activity after total hip or knee replacement. J Bone Joint Surg Am. 1998;80:54–59.PubMedCrossRef
13.
go back to reference Shepherd EF, Toloza E, McClung CD, Schmalzried TP. Step activity monitor (SAM): increased accuracy in quantifying ambulatory activity. J Orthop Res. 1999;17:703–708.PubMedCrossRef Shepherd EF, Toloza E, McClung CD, Schmalzried TP. Step activity monitor (SAM): increased accuracy in quantifying ambulatory activity. J Orthop Res. 1999;17:703–708.PubMedCrossRef
14.
go back to reference Silva M, McClung CD, dela Rosa MA, Dorey FJ, Schmalzried TP. Activity sampling in the assessment of patients with total joint arthroplasty. J Arthroplasty. 2005;20:487–491.PubMedCrossRef Silva M, McClung CD, dela Rosa MA, Dorey FJ, Schmalzried TP. Activity sampling in the assessment of patients with total joint arthroplasty. J Arthroplasty. 2005;20:487–491.PubMedCrossRef
15.
go back to reference Silva M, Shepherd EF, Jackson WO, Dorey FJ, Schmalzried TP. Average patient walking activity approaches 2 million cycles per year: pedometers under-record walking activity. J Arthroplasty. 2002;17:693–697.PubMedCrossRef Silva M, Shepherd EF, Jackson WO, Dorey FJ, Schmalzried TP. Average patient walking activity approaches 2 million cycles per year: pedometers under-record walking activity. J Arthroplasty. 2002;17:693–697.PubMedCrossRef
Metadata
Title
The 2012 Frank Stinchfield Award: Decreasing Patient Activity With Aging: Implications for Crosslinked Polyethylene Wear
Authors
Andrew K. Battenberg, BS
Jeffrey S. Hopkins, MD
Andrew D. Kupiec
Thomas P. Schmalzried, MD
Publication date
01-02-2013
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 2/2013
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-012-2497-y

Other articles of this Issue 2/2013

Clinical Orthopaedics and Related Research® 2/2013 Go to the issue

Symposium: Papers Presented at the Annual Meetings of The Hip Society

The Shape of the Proximal Femur Influences Acetabular Wear Patterns Over Time

Symposium: Papers Presented at the Annual Meetings of The Hip Society

Patients Report Improvement in Quality of Life and Satisfaction After Hip Resurfacing Arthroplasty