Skip to main content
Top
Published in: Clinical Orthopaedics and Related Research® 5/2010

01-05-2010 | Symposium: Current Issues in Knee Reconstruction

Computer Assistance Increases Precision of Component Placement in Total Knee Arthroplasty with Articular Deformity

Authors: Daniel Hernández-Vaquero, MD, PhD, Abelardo Suarez-Vazquez, MD, PhD, Manuel A. Sandoval-Garcia, MD, PhD, Alfonso Noriega-Fernandez, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 5/2010

Login to get access

Abstract

Background

The accuracy of computer navigation applied to total knee arthroplasty (TKA) in knees with severe deformity has not been studied.

Questions/purposes

The purpose of this study was to compare the radiographic alignment achieved in total knee replacements performed with and without navigation and to search for differences in the final alignment of two groups of patients (with and without previous joint deformities) using the same system of surgical navigation.

Methods

The first series comprised 40 arthroplasties with minimal preoperative deformity. In 20 of them, surgical navigation was used, whereas the other 20 were performed with conventional jig-based technique. We compared the femoral angle, tibial angle, and femorotibial angle (FTA) by performing a post-TKA CT of the entire limb. In the second series, 40 additional TKAs were studied; in this case, however, they presented preoperative deformities greater than 10º in the frontal plane.

Results

The positioning of the femoral and tibial component was more accurate in the group treated with surgical navigation and FTA improvement was statistically significant. When comparing the results of both series, FTA precision was always higher when using computer-assisted surgery. As for optimal FTA, data showed the use of surgical navigation improved the results both in the group with preoperative deformity greater than 10° in the frontal plane and in the group with minimal preoperative knee deformity.

Conclusions

Surgical navigation obtains better radiographic results in the positioning of the femoral and tibial components and in the final axis of the limb in arthroplasties performed on both deformed and more normally aligned knees.

Level of Evidence

Level II, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Literature
1.
go back to reference Ahlback S. Osteoarthrosis of the knee. A radiographic investigation. Acta Radiol Diagn. 1968;Suppl 277:7–72. Ahlback S. Osteoarthrosis of the knee. A radiographic investigation. Acta Radiol Diagn. 1968;Suppl 277:7–72.
2.
go back to reference Bargren JH, Blaha JD, Freeman MA. Alignment in total knee arthroplasty. Correlated biomechanical and clinical observations. Clin Orthop Relat Res. 1983;173:178–183.PubMed Bargren JH, Blaha JD, Freeman MA. Alignment in total knee arthroplasty. Correlated biomechanical and clinical observations. Clin Orthop Relat Res. 1983;173:178–183.PubMed
3.
go back to reference Bäthis H, Shafizadeh S, Paffrath T, Simanski C, Grifka J, Lüring C. Are computer assisted total knee replacements more accurately placed? A meta-analysis of comparative studies. Orthopäde. 2006;35:1056–1065.CrossRefPubMed Bäthis H, Shafizadeh S, Paffrath T, Simanski C, Grifka J, Lüring C. Are computer assisted total knee replacements more accurately placed? A meta-analysis of comparative studies. Orthopäde. 2006;35:1056–1065.CrossRefPubMed
4.
go back to reference Bauwens K, Wich M, Ekkernkamp A, Stengel D, Matthes G, Gebhard F, Hanson B. Navigated total knee replacement. A meta-analysis. J Bone Joint Surg Am. 2007;89:261–269.CrossRefPubMed Bauwens K, Wich M, Ekkernkamp A, Stengel D, Matthes G, Gebhard F, Hanson B. Navigated total knee replacement. A meta-analysis. J Bone Joint Surg Am. 2007;89:261–269.CrossRefPubMed
5.
go back to reference Bottros J, Klika AK, Lee HH, Polousky J, Barsoum WK. The use of navigation in total knee arthroplasty for patients with extra-articular deformity. J Arthroplasty. 2008;23:74–78.CrossRefPubMed Bottros J, Klika AK, Lee HH, Polousky J, Barsoum WK. The use of navigation in total knee arthroplasty for patients with extra-articular deformity. J Arthroplasty. 2008;23:74–78.CrossRefPubMed
6.
go back to reference Carter RE III, Rush PF, Smid JA, Smith WL. Experience with computer-assisted navigation for total knee arthroplasty in a community setting. J Arthroplasty. 2008;23:707–713.CrossRefPubMed Carter RE III, Rush PF, Smid JA, Smith WL. Experience with computer-assisted navigation for total knee arthroplasty in a community setting. J Arthroplasty. 2008;23:707–713.CrossRefPubMed
7.
go back to reference Haaker RG, Stockheim M, Kamp M, Breitenfelder J, Ottersbach A. computer-assisted navigation increases precision of component placement in total knee arthroplasty. Clin Orthop Relat Res. 2005;433:152–159.CrossRefPubMed Haaker RG, Stockheim M, Kamp M, Breitenfelder J, Ottersbach A. computer-assisted navigation increases precision of component placement in total knee arthroplasty. Clin Orthop Relat Res. 2005;433:152–159.CrossRefPubMed
8.
go back to reference Hernandez-Vaquero D, Suarez-Vazquez A. Complications of fixed infrared emitters in computer-assisted total knee arthroplasties. BMC Musculoskelet Disord. 2007;8:71–73.CrossRefPubMed Hernandez-Vaquero D, Suarez-Vazquez A. Complications of fixed infrared emitters in computer-assisted total knee arthroplasties. BMC Musculoskelet Disord. 2007;8:71–73.CrossRefPubMed
9.
go back to reference Jung HJ, Jung YB, Song KS, Park SJ Lee JS. Fractures associated with computer-navigated total knee arthroplasty. A report of two cases. J Bone Joint Surg Am. 2007;89:2280–2284.CrossRefPubMed Jung HJ, Jung YB, Song KS, Park SJ Lee JS. Fractures associated with computer-navigated total knee arthroplasty. A report of two cases. J Bone Joint Surg Am. 2007;89:2280–2284.CrossRefPubMed
10.
go back to reference Kim YH, Kim JS, Choi Y, Kwon OR. Computer-assisted surgical navigation does not improve the alignment and orientation of the components in total knee arthroplasty. J Bone Joint Surg Am. 2009;91:14–19.CrossRefPubMed Kim YH, Kim JS, Choi Y, Kwon OR. Computer-assisted surgical navigation does not improve the alignment and orientation of the components in total knee arthroplasty. J Bone Joint Surg Am. 2009;91:14–19.CrossRefPubMed
11.
go back to reference Klein GR, Austin MS, Smith EB, Hozack WJ. Total knee arthroplasty using computer-assisted navigation in patients with deformities of the femur and tibia. J Arthroplasty. 2006;21:284–288.CrossRefPubMed Klein GR, Austin MS, Smith EB, Hozack WJ. Total knee arthroplasty using computer-assisted navigation in patients with deformities of the femur and tibia. J Arthroplasty. 2006;21:284–288.CrossRefPubMed
12.
go back to reference Mahaluxmivala J, Bankes MJK, Nicolai P, Aldam CH, Allen W. The effect of surgeon experience on component positioning in 673 press fit condylar posterior cruciate-sacrificing total knee arthroplasties. J Arthroplasty. 2001;16:635–640.CrossRefPubMed Mahaluxmivala J, Bankes MJK, Nicolai P, Aldam CH, Allen W. The effect of surgeon experience on component positioning in 673 press fit condylar posterior cruciate-sacrificing total knee arthroplasties. J Arthroplasty. 2001;16:635–640.CrossRefPubMed
13.
go back to reference Mason JB, Fehring TK, Estok R, Banel D, Fahrbach K. Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplasty. 2007;22:1097–1106.CrossRefPubMed Mason JB, Fehring TK, Estok R, Banel D, Fahrbach K. Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplasty. 2007;22:1097–1106.CrossRefPubMed
14.
go back to reference Mihalko WM, Boyle J, Clark LD, Krackow KA. The variability of intramedullary alignment of the femoral component during total knee arthroplasty. J Arthroplasty. 2005;20:25–28.CrossRefPubMed Mihalko WM, Boyle J, Clark LD, Krackow KA. The variability of intramedullary alignment of the femoral component during total knee arthroplasty. J Arthroplasty. 2005;20:25–28.CrossRefPubMed
15.
go back to reference Ritter MA, Faris PM, Keating EM, Meding JB. Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop Relat Res. 1994;299:153–156.PubMed Ritter MA, Faris PM, Keating EM, Meding JB. Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop Relat Res. 1994;299:153–156.PubMed
16.
go back to reference Stulberg SD, Yaffe MA, Koo SS. Computer-assisted surgery versus manual total knee arthroplasty: a case-controlled study. J Bone Joint Surg Am. 2006;88:47–54.CrossRefPubMed Stulberg SD, Yaffe MA, Koo SS. Computer-assisted surgery versus manual total knee arthroplasty: a case-controlled study. J Bone Joint Surg Am. 2006;88:47–54.CrossRefPubMed
17.
go back to reference Yaffe MA, Koo SS, Stulberg SD. Radiographic and navigation measurements of TKA limb alignment do not correlate. Clin Orthop Relat Res. 2008;466:2736–2744.CrossRefPubMed Yaffe MA, Koo SS, Stulberg SD. Radiographic and navigation measurements of TKA limb alignment do not correlate. Clin Orthop Relat Res. 2008;466:2736–2744.CrossRefPubMed
18.
go back to reference Yau WP, Leung A, Chiu KY, Tang WM, Ng TP. Intraobserver errors in obtaining visually selected anatomic landmarks during registration process in nonimage-based navigation-assisted total knee arthroplasty. A cadaveric experiment. J Arthroplasty. 2005;20:591–596.CrossRefPubMed Yau WP, Leung A, Chiu KY, Tang WM, Ng TP. Intraobserver errors in obtaining visually selected anatomic landmarks during registration process in nonimage-based navigation-assisted total knee arthroplasty. A cadaveric experiment. J Arthroplasty. 2005;20:591–596.CrossRefPubMed
Metadata
Title
Computer Assistance Increases Precision of Component Placement in Total Knee Arthroplasty with Articular Deformity
Authors
Daniel Hernández-Vaquero, MD, PhD
Abelardo Suarez-Vazquez, MD, PhD
Manuel A. Sandoval-Garcia, MD, PhD
Alfonso Noriega-Fernandez, MD
Publication date
01-05-2010
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 5/2010
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-009-1175-1

Other articles of this Issue 5/2010

Clinical Orthopaedics and Related Research® 5/2010 Go to the issue