Skip to main content
Top
Published in: Clinical Orthopaedics and Related Research® 2/2010

01-02-2010 | Symposium: Papers Presented at the Hip Society Meetings 2009

Failure of the Durom Metasul® Acetabular Component

Authors: William T. Long, MD, Manish Dastane, MD, Michael J. Harris, MD, Zhinian Wan, MD, Lawrence D. Dorr, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 2/2010

Login to get access

Abstract

Large-diameter metal-on-metal articulations reportedly provide better stability and range of motion than smaller diameter bearings. We therefore asked whether a large-diameter (44- to 50-mm) metal-on-metal articulation (Durom®) would eliminate dislocation and provide similar functional scores and clinical and radiographic failure rates as those with 28-mm articulation. We prospectively followed 181 patients (207 hips) who had a large-diameter articulation implanted between May 2006 and November 2007. We compared these patients with a historical control of 54 patients who had a small-diameter (28-mm Metasul®) articulation. All patients had a Harris hip score and a self-assessment of outcome and radiographic followup. The minimum followup was 1 year (mean, 1.6 years; range, 1–2 years). During the followup period, we performed revisions on 29 patients (30 hips [15%]) with 21 of 29 (72%) having radiographic criteria of loosening. Thirteen retrieved cups and acetabular tissue were examined histologically. Twenty-eight of 151 unrevised patients had radiographic impending failure; 12 without revision had clinical failure. Eight patients (nine hips) had both clinical failure and impending radiographic failure. Cup inclination was 41.3° ± 5.4° and anteversion was 20.2° ± 7°. The revision rate and quality of clinical results were unacceptable as compared with our historical controls. We do not recommend use of the Durom® implant.
Level of Evidence: Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Literature
1.
go back to reference Ackland MK, Bourne WB, Uhthoff HK. Anteversion of the acetabular cup. Measurement of angle after total hip replacement. J Bone Joint Surg Br. 1986;68:409–413.PubMed Ackland MK, Bourne WB, Uhthoff HK. Anteversion of the acetabular cup. Measurement of angle after total hip replacement. J Bone Joint Surg Br. 1986;68:409–413.PubMed
2.
go back to reference Blumenfeld TJ, Bargar WL. Early aseptic loosening of a modern acetabular component secondary to a change in manufacturing. J Arthroplasty. 2006;21:689–695.CrossRefPubMed Blumenfeld TJ, Bargar WL. Early aseptic loosening of a modern acetabular component secondary to a change in manufacturing. J Arthroplasty. 2006;21:689–695.CrossRefPubMed
3.
go back to reference Bourne RB, Rorabeck CH, Burkart BC, Kirk PG. Ingrowth surfaces. Plasma spray coating to titanium alloy hip replacements. Clin Orthop Relat Res. 1994;298:37–46.PubMed Bourne RB, Rorabeck CH, Burkart BC, Kirk PG. Ingrowth surfaces. Plasma spray coating to titanium alloy hip replacements. Clin Orthop Relat Res. 1994;298:37–46.PubMed
4.
go back to reference Callaghan JJ, Salvati EA, Pellicci PM, Wilson PD Jr, Ranawat CS. Results of revision for mechanical failure after cemented total hip replacement, 1979 to 1982. A two to five-year follow-up. J Bone Joint Surg Am. 1985;67:1074–1085.PubMed Callaghan JJ, Salvati EA, Pellicci PM, Wilson PD Jr, Ranawat CS. Results of revision for mechanical failure after cemented total hip replacement, 1979 to 1982. A two to five-year follow-up. J Bone Joint Surg Am. 1985;67:1074–1085.PubMed
5.
go back to reference Dastane MR, Long WT, Wan Z, Chao L, Dorr LD. Metal-on-metal hip arthroplasty does equally well in osteonecrosis and osteoarthritis. Clin Orthop Relat Res. 2008;466:1148–1153.CrossRefPubMed Dastane MR, Long WT, Wan Z, Chao L, Dorr LD. Metal-on-metal hip arthroplasty does equally well in osteonecrosis and osteoarthritis. Clin Orthop Relat Res. 2008;466:1148–1153.CrossRefPubMed
6.
go back to reference Delaunay CP. Metal-on-metal bearings in cementless primary total hip arthroplasty. J Arthroplasty. 2004;19:35–40.CrossRefPubMed Delaunay CP. Metal-on-metal bearings in cementless primary total hip arthroplasty. J Arthroplasty. 2004;19:35–40.CrossRefPubMed
7.
go back to reference DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed
8.
go back to reference Dorr LD, Hilton KR, Wan Z, Markovich GD, Bloebaum R. Modern metal on metal articulation for total hip replacements. Clin Orthop Relat Res. 1996;333:108–117.CrossRefPubMed Dorr LD, Hilton KR, Wan Z, Markovich GD, Bloebaum R. Modern metal on metal articulation for total hip replacements. Clin Orthop Relat Res. 1996;333:108–117.CrossRefPubMed
9.
go back to reference Dorr LD, Long WT, Sirianni L, Campana M, Wan Z. The argument for the use of Metasul as an articulation surface in total hip replacement. Clin Orthop Relat Res. 2004;429:80–85.CrossRefPubMed Dorr LD, Long WT, Sirianni L, Campana M, Wan Z. The argument for the use of Metasul as an articulation surface in total hip replacement. Clin Orthop Relat Res. 2004;429:80–85.CrossRefPubMed
10.
go back to reference Dorr LD, Malik A, Dastane M, Wan Z. Combined anteversion technique for total hip arthroplasty. Clin Orthop Relat Res. 2009;467:119–127.CrossRefPubMed Dorr LD, Malik A, Dastane M, Wan Z. Combined anteversion technique for total hip arthroplasty. Clin Orthop Relat Res. 2009;467:119–127.CrossRefPubMed
11.
go back to reference Dorr LD, Wan Z, Gruen T. Functional results in total hip replacement in patients 65 years and older. Clin Orthop Relat Res. 1997;336:143–151.CrossRefPubMed Dorr LD, Wan Z, Gruen T. Functional results in total hip replacement in patients 65 years and older. Clin Orthop Relat Res. 1997;336:143–151.CrossRefPubMed
12.
go back to reference Dorr LD, Wan Z, Longjohn DB, Dubois B, Murken R. Total hip arthroplasty with use of the Metasul metal-on-metal articulation. Four to seven-year results. J Bone Joint Surg Am. 2000;82:789–798.PubMed Dorr LD, Wan Z, Longjohn DB, Dubois B, Murken R. Total hip arthroplasty with use of the Metasul metal-on-metal articulation. Four to seven-year results. J Bone Joint Surg Am. 2000;82:789–798.PubMed
13.
go back to reference Emerson RH Jr, Sanders SB, Head WC, Higgins L. Effect of circumferential plasma-spray porous coating on the rate of femoral osteolysis after total hip arthroplasty. J Bone Joint Surg Am. 1999;81:1291–1298.PubMed Emerson RH Jr, Sanders SB, Head WC, Higgins L. Effect of circumferential plasma-spray porous coating on the rate of femoral osteolysis after total hip arthroplasty. J Bone Joint Surg Am. 1999;81:1291–1298.PubMed
14.
go back to reference Gruen TA, McNeice GM, Amstutz HC. ‘Modes of failure’ of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed Gruen TA, McNeice GM, Amstutz HC. ‘Modes of failure’ of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed
15.
go back to reference Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755.PubMed Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755.PubMed
16.
go back to reference Inaba Y, Dorr LD, Wan Z, Sirianni L, Boutary M. Operative and patient care techniques for posterior mini-incision total hip arthroplasty. Clin Orthop Relat Res. 2005;441:104–114.CrossRefPubMed Inaba Y, Dorr LD, Wan Z, Sirianni L, Boutary M. Operative and patient care techniques for posterior mini-incision total hip arthroplasty. Clin Orthop Relat Res. 2005;441:104–114.CrossRefPubMed
17.
go back to reference Johnston RC, Fitzgerald RH Jr, Harris WH, Poss R, Muller ME, Sledge CB. Clinical and radiographic evaluation of total hip replacement. A standard system of terminology for reporting results. J Bone Joint Surg Am. 1990;72:161–168.PubMed Johnston RC, Fitzgerald RH Jr, Harris WH, Poss R, Muller ME, Sledge CB. Clinical and radiographic evaluation of total hip replacement. A standard system of terminology for reporting results. J Bone Joint Surg Am. 1990;72:161–168.PubMed
18.
go back to reference Kim SY, Kyung HS, Ihn JC, Cho MR, Koo KH, Kim CY. Cementless Metasul metal-on-metal total hip arthroplasty in patients less than fifty years old. J Bone Joint Surg Am. 2004;86:2475–2481.CrossRefPubMed Kim SY, Kyung HS, Ihn JC, Cho MR, Koo KH, Kim CY. Cementless Metasul metal-on-metal total hip arthroplasty in patients less than fifty years old. J Bone Joint Surg Am. 2004;86:2475–2481.CrossRefPubMed
19.
go back to reference Long WT, Dorr LD, Gendelman V. An American experience with metal-on-metal total hip arthroplasties: a 7-year follow-up study. J Arthroplasty. 2004;19:29–34.PubMed Long WT, Dorr LD, Gendelman V. An American experience with metal-on-metal total hip arthroplasties: a 7-year follow-up study. J Arthroplasty. 2004;19:29–34.PubMed
20.
go back to reference Maheshwari AV, Boutary M, Yun AG, Sirianni LE, Dorr LD. Multimodal analgesia without routine parenteral narcotics for total hip arthroplasty. Clin Orthop Relat Res. 2006;453:231–238.CrossRefPubMed Maheshwari AV, Boutary M, Yun AG, Sirianni LE, Dorr LD. Multimodal analgesia without routine parenteral narcotics for total hip arthroplasty. Clin Orthop Relat Res. 2006;453:231–238.CrossRefPubMed
21.
go back to reference Manning DW, Ponce BA, Chiang PP, Harris WH, Burke DW. Isolated acetabular revision through the posterior approach: short-term results after revision of a recalled acetabular component. J Arthroplasty. 2005;20:723–729.CrossRefPubMed Manning DW, Ponce BA, Chiang PP, Harris WH, Burke DW. Isolated acetabular revision through the posterior approach: short-term results after revision of a recalled acetabular component. J Arthroplasty. 2005;20:723–729.CrossRefPubMed
22.
go back to reference Migaud H, Jobin A, Chantelot C, Giraud F, Laffargue P, Duquennoy A. Cementless metal-on-metal hip arthroplasty in patients less than 50 years of age: comparison with a matched control group using ceramic-on-polyethylene after a minimum 5-year follow-up. J Arthroplasty. 2004;19:23–28.PubMed Migaud H, Jobin A, Chantelot C, Giraud F, Laffargue P, Duquennoy A. Cementless metal-on-metal hip arthroplasty in patients less than 50 years of age: comparison with a matched control group using ceramic-on-polyethylene after a minimum 5-year follow-up. J Arthroplasty. 2004;19:23–28.PubMed
23.
go back to reference Shahrdar C, Campbell P, Mirra J, Dorr LD. Painful metal-on-metal total hip arthroplasty. J Arthroplasty. 2006;21:289–293.CrossRefPubMed Shahrdar C, Campbell P, Mirra J, Dorr LD. Painful metal-on-metal total hip arthroplasty. J Arthroplasty. 2006;21:289–293.CrossRefPubMed
24.
go back to reference Southwell DG, Bechtold JE, Lew WD, Schmidt AH. Improving the detection of acetabular osteolysis using oblique radiographs. J Bone Joint Surg Br. 1999;81:289–295.CrossRefPubMed Southwell DG, Bechtold JE, Lew WD, Schmidt AH. Improving the detection of acetabular osteolysis using oblique radiographs. J Bone Joint Surg Br. 1999;81:289–295.CrossRefPubMed
25.
go back to reference Udomkiat P, Wan Z, Dorr LD. Comparison of preoperative radiographs and intraoperative findings of fixation of hemispheric porous-coated sockets. J Bone Joint Surg Am. 2001;83:1865–1870.PubMed Udomkiat P, Wan Z, Dorr LD. Comparison of preoperative radiographs and intraoperative findings of fixation of hemispheric porous-coated sockets. J Bone Joint Surg Am. 2001;83:1865–1870.PubMed
26.
go back to reference Vendittoli PA, Mottard S, Roy AG, Dupont C, Lavigne M. Chromium and cobalt ion release following the Durom high carbon content, forged metal-on-metal surface replacement of the hip. J Bone Joint Surg Br. 2007;89:441–448.CrossRefPubMed Vendittoli PA, Mottard S, Roy AG, Dupont C, Lavigne M. Chromium and cobalt ion release following the Durom high carbon content, forged metal-on-metal surface replacement of the hip. J Bone Joint Surg Br. 2007;89:441–448.CrossRefPubMed
27.
go back to reference Wan Z, Boutary M, Dorr LD. Precision and limitation of measuring two-dimensional wear on clinical radiographs. Clin Orthop Relat Res. 2006;449:267–274.PubMed Wan Z, Boutary M, Dorr LD. Precision and limitation of measuring two-dimensional wear on clinical radiographs. Clin Orthop Relat Res. 2006;449:267–274.PubMed
Metadata
Title
Failure of the Durom Metasul® Acetabular Component
Authors
William T. Long, MD
Manish Dastane, MD
Michael J. Harris, MD
Zhinian Wan, MD
Lawrence D. Dorr, MD
Publication date
01-02-2010
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 2/2010
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-009-1071-8

Other articles of this Issue 2/2010

Clinical Orthopaedics and Related Research® 2/2010 Go to the issue

Symposium: Papers Presented at the Hip Society Meetings 2009

Biographical Sketch: Virgil Pendelton Gibney, MD, 1847–1927

Symposium: Papers Presented at the Hip Society Meetings 2009

Femoral Revision Hip Arthroplasty: A Comparison of Two Stem Designs

Symposium: Papers Presented at the Hip Society Meetings 2009

Incidence of ‘Squeaking’ After Ceramic-on-Ceramic Total Hip Arthroplasty

Letter to the Editor

Reply to Letter to the Editor

Symposium: Papers Presented at the Hip Society Meetings 2009

Femoral Anteversion in THA and its Lack of Correlation with Native Acetabular Anteversion