High Cup Angle and Microseparation Increase the Wear of Hip Surface Replacements

Authors: Ian J. Leslie, PhD, Sophie Williams, PhD, Graham Isaac, PhD, Eileen Ingham, PhD, John Fisher, PhD

Published in: Clinical Orthopaedics and Related Research® | Issue 9/2009

Login to get access

Abstract

High wear rates and high patient ion levels have been associated with high (> 55°) cup inclination angles for metal-on-metal surface replacements. Wear rates and patterns have been simulated for ceramic-on-ceramic bearings by applying microseparation to replicate head offset deficiency. We tested 39-mm metal-on-metal surface replacements (n = 5) in a hip simulator with (A) an increased cup inclination angle of 60° and (B) an increased cup inclination angle and microseparation over 2 million cycles. (A) resulted in a ninefold increase in wear rate and (B) resulted in a 17-fold increase in wear rate compared to a standard gait condition study. Wear particles produced under microseparation conditions were larger than those produced under standard conditions but of similar shape (round to oval). The data suggest both head and cup position influence the wear of surface replacements; we believe it likely bearings with high wear either have a high cup inclination angle, an offset deficient head, or a combination of both.
Literature
1.
go back to reference Barbour PS, Stone MH, Fisher J. A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates. Proc Inst Mech Eng [H]. 1999;213:455–467. Barbour PS, Stone MH, Fisher J. A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates. Proc Inst Mech Eng [H]. 1999;213:455–467.
2.
go back to reference Brown C, Williams S, Tipper JL, Fisher J, Ingham E. Characterisation of wear particles produced by metal on metal and ceramic on metal hip prostheses under standard and microseparation simulation. J Mater Sci Mater Med. 2007;18:819–827.PubMedCrossRef Brown C, Williams S, Tipper JL, Fisher J, Ingham E. Characterisation of wear particles produced by metal on metal and ceramic on metal hip prostheses under standard and microseparation simulation. J Mater Sci Mater Med. 2007;18:819–827.PubMedCrossRef
3.
go back to reference Buergi ML, Walter WL. Hip resurfacing arthroplasty: the Australian experience. J Arthroplasty. 2007;22(7 Suppl 3):61–65.PubMedCrossRef Buergi ML, Walter WL. Hip resurfacing arthroplasty: the Australian experience. J Arthroplasty. 2007;22(7 Suppl 3):61–65.PubMedCrossRef
4.
go back to reference Buscher R, Tager G, Dudzinski W, Gleising B, Wimmer MA, Fischer A. Subsurface microstructure of metal-on-metal hip joints and its relationship to wear particle generation. J Biomed Mater Res B Appl Biomater. 2005;72:206–214.PubMedCrossRef Buscher R, Tager G, Dudzinski W, Gleising B, Wimmer MA, Fischer A. Subsurface microstructure of metal-on-metal hip joints and its relationship to wear particle generation. J Biomed Mater Res B Appl Biomater. 2005;72:206–214.PubMedCrossRef
5.
go back to reference Campbell P, Beaule PE, Ebramzadeh E, LeDuff M, De Smet K, Lu Z, Amstutz HC. The John Charnley Award: a study of implant failure in metal-on-metal surface arthroplasties. Clin Orthop Relat Res. 2006;453:35–46.PubMedCrossRef Campbell P, Beaule PE, Ebramzadeh E, LeDuff M, De Smet K, Lu Z, Amstutz HC. The John Charnley Award: a study of implant failure in metal-on-metal surface arthroplasties. Clin Orthop Relat Res. 2006;453:35–46.PubMedCrossRef
6.
go back to reference Catelas I, Campbell PA, Bobyn JD, Medley JB, Huk OL. Wear particles from metal-on-metal total hip replacements: effects of implant design and implantation time. Proc Inst Mech Eng [H]. 2006;220:195–208. Catelas I, Campbell PA, Bobyn JD, Medley JB, Huk OL. Wear particles from metal-on-metal total hip replacements: effects of implant design and implantation time. Proc Inst Mech Eng [H]. 2006;220:195–208.
7.
go back to reference De Haan R, Campbell PA, Su EP, De Smet KA. Revision of metal-on-metal resurfacing arthroplasty of the hip: the influence of malpositioning of the components. J Bone Joint Surg Br. 2008;90:1158–1163.PubMedCrossRef De Haan R, Campbell PA, Su EP, De Smet KA. Revision of metal-on-metal resurfacing arthroplasty of the hip: the influence of malpositioning of the components. J Bone Joint Surg Br. 2008;90:1158–1163.PubMedCrossRef
8.
go back to reference De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 2008;90:1291–1297.PubMedCrossRef De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 2008;90:1291–1297.PubMedCrossRef
9.
go back to reference Glaser D, Komistek RD, Cates HE, Mahfouz MR. Clicking and squeaking: in vivo correlation of sound and separation for different bearing surfaces. J Bone Joint Surg Am. 2008;90 (4):112–120.PubMedCrossRef Glaser D, Komistek RD, Cates HE, Mahfouz MR. Clicking and squeaking: in vivo correlation of sound and separation for different bearing surfaces. J Bone Joint Surg Am. 2008;90 (4):112–120.PubMedCrossRef
10.
go back to reference Kang L, Galvin AL, Brown TD, Fisher J, Jin ZM. Wear simulation of ultra-high molecular weight polyethylene hip implants by incorporating the effects of cross-shear and contact pressure. Proc Inst Mech Eng [H]. 2008;222:1049–1064. Kang L, Galvin AL, Brown TD, Fisher J, Jin ZM. Wear simulation of ultra-high molecular weight polyethylene hip implants by incorporating the effects of cross-shear and contact pressure. Proc Inst Mech Eng [H]. 2008;222:1049–1064.
11.
go back to reference Komistek RD, Dennis DA, Ochoa JA, Haas BD, Hammill C. In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty. J Bone Joint Surg Am. 2002;84:1836–1841.PubMed Komistek RD, Dennis DA, Ochoa JA, Haas BD, Hammill C. In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty. J Bone Joint Surg Am. 2002;84:1836–1841.PubMed
12.
go back to reference Langton DJ, Jameson SS, Joyce TJ, Webb J, Nargol AV. The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip. J Bone Joint Surg Br. 2008;90:1143–1151.PubMed Langton DJ, Jameson SS, Joyce TJ, Webb J, Nargol AV. The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip. J Bone Joint Surg Br. 2008;90:1143–1151.PubMed
13.
go back to reference Leslie I, Williams S, Brown C, Isaac G, Jin Z, Ingham E, Fisher J. Effect of bearing size on the long-term wear, wear debris, and ion levels of large diameter metal-on-metal hip replacements-An in vitro study. J Biomed Mater Res B Appl Biomater. 2008;87:163–172.PubMed Leslie I, Williams S, Brown C, Isaac G, Jin Z, Ingham E, Fisher J. Effect of bearing size on the long-term wear, wear debris, and ion levels of large diameter metal-on-metal hip replacements-An in vitro study. J Biomed Mater Res B Appl Biomater. 2008;87:163–172.PubMed
14.
go back to reference Liu F, Jin ZM, Hirt F, Rieker C, Roberts P, Grigoris P. Effect of wear of bearing surfaces on elastohydrodynamic lubrication of metal-on-metal hip implants. Proc Inst Mech Eng [H]. 2005;219:319–328. Liu F, Jin ZM, Hirt F, Rieker C, Roberts P, Grigoris P. Effect of wear of bearing surfaces on elastohydrodynamic lubrication of metal-on-metal hip implants. Proc Inst Mech Eng [H]. 2005;219:319–328.
15.
go back to reference Morlock MM, Bishop N, Zustin J, Hahn M, Ruther W, Amling M. Modes of implant failure after hip resurfacing: morphological and wear analysis of 267 retrieval specimens. J Bone Joint Surg Am. 2008;90(3):89–95.PubMedCrossRef Morlock MM, Bishop N, Zustin J, Hahn M, Ruther W, Amling M. Modes of implant failure after hip resurfacing: morphological and wear analysis of 267 retrieval specimens. J Bone Joint Surg Am. 2008;90(3):89–95.PubMedCrossRef
16.
go back to reference Nevelos J, Ingham E, Doyle C, Streicher R, Nevelos A, Walter W, Fisher J. Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. J Arthroplasty. 2000;15:793–795.PubMedCrossRef Nevelos J, Ingham E, Doyle C, Streicher R, Nevelos A, Walter W, Fisher J. Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. J Arthroplasty. 2000;15:793–795.PubMedCrossRef
17.
go back to reference Nevelos JE, Ingham E, Doyle C, Fisher J, Nevelos AB. Analysis of retrieved alumina ceramic components from Mittelmeier total hip prostheses. Biomaterials. 1999;20:1833–1840.PubMedCrossRef Nevelos JE, Ingham E, Doyle C, Fisher J, Nevelos AB. Analysis of retrieved alumina ceramic components from Mittelmeier total hip prostheses. Biomaterials. 1999;20:1833–1840.PubMedCrossRef
18.
go back to reference Nevelos JE, Ingham E, Doyle C, Nevelos AB, Fisher J. The influence of acetabular cup angle on the wear of “BIOLOX Forte” alumina ceramic bearing couples in a hip joint simulator. J Mater Sci Mater Med. 2001;12:141–144.PubMedCrossRef Nevelos JE, Ingham E, Doyle C, Nevelos AB, Fisher J. The influence of acetabular cup angle on the wear of “BIOLOX Forte” alumina ceramic bearing couples in a hip joint simulator. J Mater Sci Mater Med. 2001;12:141–144.PubMedCrossRef
19.
go back to reference Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons CL, Ostlere S, Athanasou N, Gill HS, Murray DW. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90:847–851.PubMedCrossRef Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons CL, Ostlere S, Athanasou N, Gill HS, Murray DW. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90:847–851.PubMedCrossRef
20.
go back to reference Silva M, Lee KH, Heisel C, Dela Rosa MA, Schmalzried TP. The biomechanical results of total hip resurfacing arthroplasty. J. Bone Joint Surg Am. 2004;86:40–46.PubMed Silva M, Lee KH, Heisel C, Dela Rosa MA, Schmalzried TP. The biomechanical results of total hip resurfacing arthroplasty. J. Bone Joint Surg Am. 2004;86:40–46.PubMed
21.
go back to reference Silva M, Shepherd EF, Jackson WO, Dorey FJ, Schmalzried TP. Average patient walking activity approaches 2 million cycles per year: pedometers under-record walking activity. J Arthroplasty. 2002;17:693–697.PubMedCrossRef Silva M, Shepherd EF, Jackson WO, Dorey FJ, Schmalzried TP. Average patient walking activity approaches 2 million cycles per year: pedometers under-record walking activity. J Arthroplasty. 2002;17:693–697.PubMedCrossRef
22.
go back to reference Tipper JL, Hatton A, Nevelos JE, Ingham E, Doyle C, Streicher R, Nevelos AB, Fisher J. Alumina-alumina artificial hip joints. Part II: characterisation of the wear debris from in vitro hip joint simulations. Biomaterials. 2002;23:3441–3448.PubMedCrossRef Tipper JL, Hatton A, Nevelos JE, Ingham E, Doyle C, Streicher R, Nevelos AB, Fisher J. Alumina-alumina artificial hip joints. Part II: characterisation of the wear debris from in vitro hip joint simulations. Biomaterials. 2002;23:3441–3448.PubMedCrossRef
23.
go back to reference Williams S, Leslie I, Isaac G, Jin Z, Ingham E, Fisher J. Tribology and wear of metal-on-metal hip prostheses: influence of cup angle and head position. J Bone Joint Surg Am. 2008;90 (3):111–117.PubMedCrossRef Williams S, Leslie I, Isaac G, Jin Z, Ingham E, Fisher J. Tribology and wear of metal-on-metal hip prostheses: influence of cup angle and head position. J Bone Joint Surg Am. 2008;90 (3):111–117.PubMedCrossRef
24.
go back to reference Yan Y, Neville A, Dowson D, Williams S, Fisher J. Tribo-corrosion analysis of wear and metal ion release interactions from metal-on-metal and ceramic-on-metal contacts for the application in artificial hip prostheses. P I Mech Eng J-J Eng. 2008;222(J3):483–492. Yan Y, Neville A, Dowson D, Williams S, Fisher J. Tribo-corrosion analysis of wear and metal ion release interactions from metal-on-metal and ceramic-on-metal contacts for the application in artificial hip prostheses. P I Mech Eng J-J Eng. 2008;222(J3):483–492.
Metadata
Title
High Cup Angle and Microseparation Increase the Wear of Hip Surface Replacements
Authors
Ian J. Leslie, PhD
Sophie Williams, PhD
Graham Isaac, PhD
Eileen Ingham, PhD
John Fisher, PhD
Publication date
01-09-2009
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 9/2009
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-009-0830-x

Other articles of this Issue 9/2009

Clinical Orthopaedics and Related Research® 9/2009 Go to the issue

Symposium: Papers Presented at the Annual Closed Meeting of the International Hip Society

Trabecular Metal™ Cups for Acetabular Defects With 50% or Less Host Bone Contact

Symposium: Papers Presented at the Annual Closed Meeting of the International Hip Society

Enhanced Cell Integration to Titanium Alloy by Surface Treatment with Microarc Oxidation: A Pilot Study

Symposium: Papers Presented at the Annual Closed Meeting of the International Hip Society

Intra- and Intersurgeon Variability in Image-free Navigation System for THA

Symposium: Papers Presented at the Annual Closed Meeting of the International Hip Society

Factors Influencing the Stability of Stems Fixed with Impaction Graft in Vitro

Symposium: Papers Presented at the Annual Closed Meeting of the International Hip Society

Periacetabular Osteotomy for Acetabular Dysplasia in Patients Older than 40 Years: A Preliminary Study