Development of a simple device enabling percutaneous flow regulation for a small vascular graft for a Blalock–Taussig shunt capable of flow regulation: complete translation of a review article originally published in Pediatric Cardiology and Cardiac Surgery (154–159, 2016: vol. 32)

Authors: Yoshikazu Motohashi, Ryo Shimada, Tomoyasu Sasaki, Takahiro Katsumata, Kazunori Dan, Yasuhiro Tsutsui, Shintaro Nemoto

Published in: General Thoracic and Cardiovascular Surgery | Issue 3/2018

Login to get access

Abstract

Objectives

The Blalock–Taussig shunt (BTS) operation is a cornerstone as initial palliative surgery for congenital heart disease with severely reduced pulmonary blood flow (PBF). The ideal PBF provided by BTS is crucial for an uneventful postoperative course, since excess PBF results in acute distress of the systemic circulation and insufficient PBF requires another BTS surgery. Therefore, the goal of this study was to develop a simple device to control the shunt graft flow percutaneously using a constrictor balloon connected to a subcutaneous port.

Methods

The device consists of a cylindrical balloon and an anti-bending structure extension connected to the balloon center. A PTFE vascular graft wrapped by the device was connected to a simulated closed circuit to measure the relationship between pressure and blood flow while changing the inner volume of the balloon. In a beagle model of replacement of the right carotid artery, blood flow velocity was measured in the carotid artery after saline injection into the balloon. The blood flow velocity before and after balloon inflation was compared immediately after implantation of the device and at 3 months after implantation.

Results

The device provided good flow control by inflating and deflating the balloon ex vivo and in vivo for up to 3 months in a canine model with a small graft wrapped with the device.

Conclusions

The simple device developed in this study may enable regulation of PBF through a small vascular graft and help to prevent severe morbidity and mortality in the clinical setting of BTS.
Literature
1.
go back to reference Dirks V, Prêtre R, Knirsch W, Valsangiacomo-Buechel ER, Seifert B, Schweiger M, et al. Modified Blalock Taussig shunt: a not-so-simple palliative procedure. Eur J Cardiothorac Surg. 2013;44:1096–102.CrossRefPubMed Dirks V, Prêtre R, Knirsch W, Valsangiacomo-Buechel ER, Seifert B, Schweiger M, et al. Modified Blalock Taussig shunt: a not-so-simple palliative procedure. Eur J Cardiothorac Surg. 2013;44:1096–102.CrossRefPubMed
2.
go back to reference Petrucci O, O’Brien SM, Jacobs ML, Jacobs JP, Manning PB, Eghtesady P. Risk factors for mortality and morbidity after the neonatal Blalock–Taussig shunt procedure. Ann Thorac Surg 2011; 92: 642–52.CrossRefPubMed Petrucci O, O’Brien SM, Jacobs ML, Jacobs JP, Manning PB, Eghtesady P. Risk factors for mortality and morbidity after the neonatal Blalock–Taussig shunt procedure. Ann Thorac Surg 2011; 92: 642–52.CrossRefPubMed
3.
go back to reference McKenzie ED, Khan MS, Samayoa AX, Vener DS, Ishak YM, Santos AB, et al. The Blalock-Taussig shunt revisited: a contemporary experience. J Am Coll Surg 2013;216:699–704.CrossRefPubMed McKenzie ED, Khan MS, Samayoa AX, Vener DS, Ishak YM, Santos AB, et al. The Blalock-Taussig shunt revisited: a contemporary experience. J Am Coll Surg 2013;216:699–704.CrossRefPubMed
4.
go back to reference Higashidate M, Beppu T, Imai Y, Kurosawa H. Percutaneously adjustable pulmonary artery band. An experimental study. J Thorac Cardiovasc Surg. 1989;97:864–9.PubMed Higashidate M, Beppu T, Imai Y, Kurosawa H. Percutaneously adjustable pulmonary artery band. An experimental study. J Thorac Cardiovasc Surg. 1989;97:864–9.PubMed
5.
go back to reference Takeuchi T, Imai Y, Sawatari K, Ishihara K, Hoshino S, Higashidate M, et al. Experience of new adjustable vessel occluder. Jpn J Artif Organs 1993; 22: 24–6. Takeuchi T, Imai Y, Sawatari K, Ishihara K, Hoshino S, Higashidate M, et al. Experience of new adjustable vessel occluder. Jpn J Artif Organs 1993; 22: 24–6.
6.
go back to reference Douglas WI, Moore KB, Resig PP, Mohiuddin MW. The adjustable systemic-pulmonary artery shunt provides precise control of flow in vivo. ASAIO J 2010; 56: 73–6.CrossRefPubMed Douglas WI, Moore KB, Resig PP, Mohiuddin MW. The adjustable systemic-pulmonary artery shunt provides precise control of flow in vivo. ASAIO J 2010; 56: 73–6.CrossRefPubMed
7.
go back to reference Mohiuddin MW, Resig PP, Sexton KW, Douglas WI. Two-day control of pulmonary blood flow with an adjustable systemic-pulmonary artery shunt. ASAIO J. 2011;57:225–30.CrossRefPubMed Mohiuddin MW, Resig PP, Sexton KW, Douglas WI. Two-day control of pulmonary blood flow with an adjustable systemic-pulmonary artery shunt. ASAIO J. 2011;57:225–30.CrossRefPubMed
Metadata
Title
Development of a simple device enabling percutaneous flow regulation for a small vascular graft for a Blalock–Taussig shunt capable of flow regulation: complete translation of a review article originally published in Pediatric Cardiology and Cardiac Surgery (154–159, 2016: vol. 32)
Authors
Yoshikazu Motohashi
Ryo Shimada
Tomoyasu Sasaki
Takahiro Katsumata
Kazunori Dan
Yasuhiro Tsutsui
Shintaro Nemoto
Publication date
01-03-2018
Publisher
Springer Japan
Published in
General Thoracic and Cardiovascular Surgery / Issue 3/2018
Print ISSN: 1863-6705
Electronic ISSN: 1863-6713
DOI
https://doi.org/10.1007/s11748-017-0860-3

Other articles of this Issue 3/2018

General Thoracic and Cardiovascular Surgery 3/2018 Go to the issue