Published in: Japanese Journal of Radiology 2/2018

01-02-2018 | Original Article

CT brain artefact due to air bubbles in the oil cooling system: characteristic band-like configuration on sagittal reformatted image

Authors: Simon Sin Man Wong, James Siu Wai Cheung, Tom Wai Kin Lee, Chris Siu Chun Tsai, Alex Wing Hung Ng

Published in: Japanese Journal of Radiology | Issue 2/2018

Login to get access

Abstract

Purpose

To retrospectively review the imaging characteristics of CT artefacts due to air bubbles within the oil cooling system of the X-ray tube housing.

Materials and methods

Air bubbles were introduced into the oil cooling system of the X-ray tube housing during tube replacement in one of the CT scanners in the authors’ institution. All 126 CT brain studies performed in this period were retrospectively reviewed. One hundred and four studies were negative for artefacts. Artefacts were confirmed in 5 and considered probable in 17 studies, respectively. The imaging characteristics of artefacts in these 22 cases were analysed.

Results

All artefacts manifested as ill-defined hypoattenuations in the periventricular/subcortical white matter of bilateral cerebral hemispheres with/without involvement of the internal capsule and basal ganglia. The posterior fossa was also involved in two (40%) confirmed and four (24%) probable studies. A band-like configuration of hypoattenuations on sagittal images was observed in five (100%) confirmed and eight (47%) probable studies.

Conclusion

Air bubble artefacts manifested as hypoattenuations in the periventricular/subcortical white matter of the supratentorial brain. A characteristic band-like configuration was observed in the sagittal reformatted image, which is useful for differentiating it from periventricular small vessel disease.
Literature
1.
go back to reference McVey S, Kanodia AK, Dundas S, Main G, Pillai S, Prasad G, et al. “Air bubble artefact”: a new type of artefact on CT head examination. Clin Radiol. 2016;71(10):1059–65.CrossRefPubMed McVey S, Kanodia AK, Dundas S, Main G, Pillai S, Prasad G, et al. “Air bubble artefact”: a new type of artefact on CT head examination. Clin Radiol. 2016;71(10):1059–65.CrossRefPubMed
2.
go back to reference Trieu N, Xia R, Loneragan R, Ridley L, Trieu J. Artefact on CT brain images caused by the presence of air bubbles in the cooling oil of the X-ray tube. J Med Imaging Radiat Oncol. 2017;61(2):197–203.CrossRefPubMed Trieu N, Xia R, Loneragan R, Ridley L, Trieu J. Artefact on CT brain images caused by the presence of air bubbles in the cooling oil of the X-ray tube. J Med Imaging Radiat Oncol. 2017;61(2):197–203.CrossRefPubMed
3.
go back to reference Hedrick WR, Markovic MA, Short JH, Vera CD. Computed tomography artifact created by air in the X-ray tube oil. J Comput Assist Tomogr. 2016;40(1):67–70.CrossRefPubMed Hedrick WR, Markovic MA, Short JH, Vera CD. Computed tomography artifact created by air in the X-ray tube oil. J Comput Assist Tomogr. 2016;40(1):67–70.CrossRefPubMed
5.
go back to reference Törmänen J, Rautiainen J, Tahvonen P, Leinonen K, Nieminen MT, Tervonen O. The, “air in the CT X-ray tube oil” artifact—examples of the quality control images and the evaluation of four potential clinical patients’ head computed tomography cases. J Comput Assist Tomogr. 2017;41(3):489–93.CrossRefPubMed Törmänen J, Rautiainen J, Tahvonen P, Leinonen K, Nieminen MT, Tervonen O. The, “air in the CT X-ray tube oil” artifact—examples of the quality control images and the evaluation of four potential clinical patients’ head computed tomography cases. J Comput Assist Tomogr. 2017;41(3):489–93.CrossRefPubMed
6.
go back to reference Filippi M, Rocca MA, Ciccarelli O, De Stefano N, Evangelou N, Kappos L, et al. MRI criteria for the diagnosis of multiple sclerosis: MAGNIMS consensus guidelines. Lancet Neurol. 2016;15(3):292–303.CrossRefPubMedPubMedCentral Filippi M, Rocca MA, Ciccarelli O, De Stefano N, Evangelou N, Kappos L, et al. MRI criteria for the diagnosis of multiple sclerosis: MAGNIMS consensus guidelines. Lancet Neurol. 2016;15(3):292–303.CrossRefPubMedPubMedCentral
Metadata
Title
CT brain artefact due to air bubbles in the oil cooling system: characteristic band-like configuration on sagittal reformatted image
Authors
Simon Sin Man Wong
James Siu Wai Cheung
Tom Wai Kin Lee
Chris Siu Chun Tsai
Alex Wing Hung Ng
Publication date
01-02-2018
Publisher
Springer Japan
Published in
Japanese Journal of Radiology / Issue 2/2018
Print ISSN: 1867-1071
Electronic ISSN: 1867-108X
DOI
https://doi.org/10.1007/s11604-017-0702-3

Other articles of this Issue 2/2018

Japanese Journal of Radiology 2/2018 Go to the issue