Published in: International Journal of Computer Assisted Radiology and Surgery 11/2018

01-11-2018 | Original Article

Development of a shoulder-mounted robot for MRI-guided needle placement: phantom study

Authors: Reza Monfaredi, Iulian Iordachita, Emmanuel Wilson, Raymond Sze, Karun Sharma, Axel Krieger, Stanley Fricke, Kevin Cleary

Published in: International Journal of Computer Assisted Radiology and Surgery | Issue 11/2018

Login to get access

Abstract

Purpose

This paper presents new quantitative data on a signal-to-noise ratio (SNR) study, distortion study, and targeting accuracy phantom study for our patient-mounted robot (called Arthrobot). Arthrobot was developed as an MRI-guided needle placement device for diagnostic and interventional procedures such as arthrography.

Methods

We present the robot design and inverse kinematics. Quantitative assessment results for SNR and distortion study are also reported. A respiratory motion study was conducted to evaluate the shoulder mounting method. A phantom study was conducted to investigate end-to-end targeting accuracy. Combined error considering targeting accuracy, respiratory motion, and structure deformation is also reported.

Results

The SNR study showed that the SNR changes only 2% when the unpowered robot was placed on top of a standard water phantom. The distortion study showed that the maximum distortion from the ground truth was 2.57%. The average error associated with respiratory motion was 1.32 mm with standard deviation of 1.38 mm. Results of gel phantom targeting studies indicate average needle placement error of 1.64 mm, with a standard deviation of 0.90 mm.

Conclusions

Noise and distortion of the MR images were not significant, and image quality in the presence of the robot was satisfactory for MRI-guided targeting. Combined average total error, adding mounting stability errors and structure deformation errors to targeting error, is estimated to be 3.4 mm with a standard deviation of 1.65 mm. In clinical practice, needle placement accuracy under 5 mm is considered sufficient for successful joint injection during shoulder arthrography. Therefore, for the intended clinical procedure, these results indicate that Arthrobot has sufficient positioning accuracy.
Literature
1.
go back to reference Gopinathan A, Peh WCG (2011) Image-guided facet joint injection. Biomed Imaging Interv J 7(1):e4 Gopinathan A, Peh WCG (2011) Image-guided facet joint injection. Biomed Imaging Interv J 7(1):e4
2.
go back to reference Srinivasan S, Peh WCG (2011) Imaging-guided biopsy in musculoskeletal infections. Semin Musculoskelet Radiol 15(5):561–568CrossRefPubMed Srinivasan S, Peh WCG (2011) Imaging-guided biopsy in musculoskeletal infections. Semin Musculoskelet Radiol 15(5):561–568CrossRefPubMed
4.
go back to reference Magnusson A, Akerfeldt D (1991) CT-guided core biopsy using a new guidance device. Acta Radiol 32(1):83–85CrossRefPubMed Magnusson A, Akerfeldt D (1991) CT-guided core biopsy using a new guidance device. Acta Radiol 32(1):83–85CrossRefPubMed
5.
go back to reference Onik G, Cosman ER, Wells T, Goldberg H, Moss AA, Costello P, Kane RA, Hoddick W, Demas B (1988) CT-guided aspirations for the body: comparison of hand guidance with stereotaxis. Radiology 166(2):389–394CrossRefPubMed Onik G, Cosman ER, Wells T, Goldberg H, Moss AA, Costello P, Kane RA, Hoddick W, Demas B (1988) CT-guided aspirations for the body: comparison of hand guidance with stereotaxis. Radiology 166(2):389–394CrossRefPubMed
6.
go back to reference Wu HT, Chang CY, Chang H, Yen CC, Cheng H, Chen PC, Chiou HJ (2012) Magnetic resonance imaging guided biopsy of musculoskeletal lesions. J Chin Med Assoc 75(4):160–166CrossRefPubMed Wu HT, Chang CY, Chang H, Yen CC, Cheng H, Chen PC, Chiou HJ (2012) Magnetic resonance imaging guided biopsy of musculoskeletal lesions. J Chin Med Assoc 75(4):160–166CrossRefPubMed
7.
go back to reference Fritz J, Thomas C, Clasen S, Claussen CD, Lewin JS, Pereira PL (2009) Freehand real-time MRI-guided lumbar spinal injection procedures at 1.5 T: feasibility, accuracy, and safety. Vasc Interv Radiol 192(4):W161–W167CrossRef Fritz J, Thomas C, Clasen S, Claussen CD, Lewin JS, Pereira PL (2009) Freehand real-time MRI-guided lumbar spinal injection procedures at 1.5 T: feasibility, accuracy, and safety. Vasc Interv Radiol 192(4):W161–W167CrossRef
8.
go back to reference Mozer PC, Partin AW, Stoianovici D (2009) Robotic image-guided needle interventions of the prostate. Rev Urol 11(1):7–15PubMedPubMedCentral Mozer PC, Partin AW, Stoianovici D (2009) Robotic image-guided needle interventions of the prostate. Rev Urol 11(1):7–15PubMedPubMedCentral
9.
go back to reference Walsh C, Hanumara NC, Slocum AH, Shepard JA, Gupta R (2008) A patient-mounted, telerobotic tool for CT-guided percutaneous interventions. J Med Devices 2(1):011007.1–011007.10CrossRef Walsh C, Hanumara NC, Slocum AH, Shepard JA, Gupta R (2008) A patient-mounted, telerobotic tool for CT-guided percutaneous interventions. J Med Devices 2(1):011007.1–011007.10CrossRef
10.
go back to reference Wu FY, Torabi M, Yamadak A, Golden A, Fischer GS, Tuncali K, Frey DD, Walsh C (2013) An MRI coil-mounted multi-probe robotic positioner for cryoablation. In: Proceedings of the ASME International Design Engineering Technical Conferences & Computers and Information in Engineering Conference, Portland, Oregon, USA, pp 1–9 Wu FY, Torabi M, Yamadak A, Golden A, Fischer GS, Tuncali K, Frey DD, Walsh C (2013) An MRI coil-mounted multi-probe robotic positioner for cryoablation. In: Proceedings of the ASME International Design Engineering Technical Conferences & Computers and Information in Engineering Conference, Portland, Oregon, USA, pp 1–9
11.
go back to reference Maurin B, Bayle B, Piccin O, Gangloff J, de Mathelin M, Doignon C, Zanne P, Gangi A (2008) A patient-mounted robotic platform for CT-scan guided procedures. IEEE Trans Biomed Eng 55(10):2417–2425CrossRefPubMed Maurin B, Bayle B, Piccin O, Gangloff J, de Mathelin M, Doignon C, Zanne P, Gangi A (2008) A patient-mounted robotic platform for CT-scan guided procedures. IEEE Trans Biomed Eng 55(10):2417–2425CrossRefPubMed
12.
go back to reference Hungr N, Fouard C, Robert A, Bricault I, Cinquin P (2011) Interventional radiology robot for CT and MRI-guided percutaneous interventions. Med Image Comput Comput Assist Interv 14(1):137–144PubMed Hungr N, Fouard C, Robert A, Bricault I, Cinquin P (2011) Interventional radiology robot for CT and MRI-guided percutaneous interventions. Med Image Comput Comput Assist Interv 14(1):137–144PubMed
13.
go back to reference Bricault I, Jouniaux E, Zemiti N, Fouard C, Taillant E, Dorandeu F, Cinquin P (2008) A light puncture robot for CT and MRI interventions. IEEE Eng Med Biol Mag 27(3):42–50CrossRefPubMed Bricault I, Jouniaux E, Zemiti N, Fouard C, Taillant E, Dorandeu F, Cinquin P (2008) A light puncture robot for CT and MRI interventions. IEEE Eng Med Biol Mag 27(3):42–50CrossRefPubMed
14.
go back to reference Song SE, Tokuda J, Tuncali K, Yamada A, Torabi M, Hata N (2013) Design evaluation of a double ring RCM mechanism for robotic needle guidance in MRI-guided liver interventions. In: IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), Tokyo, Japan, pp 4078–4083 Song SE, Tokuda J, Tuncali K, Yamada A, Torabi M, Hata N (2013) Design evaluation of a double ring RCM mechanism for robotic needle guidance in MRI-guided liver interventions. In: IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), Tokyo, Japan, pp 4078–4083
15.
go back to reference Li G, Hao S, Cole GA, Shang W, Harrington K, Camilo A, Pilitsis JG, Fischer GS (2015) Robotic system for MRI-guided stereotactic neurosurgery. IEEE Trans Biomed Eng 62(4):1077–1088CrossRefPubMedPubMedCentral Li G, Hao S, Cole GA, Shang W, Harrington K, Camilo A, Pilitsis JG, Fischer GS (2015) Robotic system for MRI-guided stereotactic neurosurgery. IEEE Trans Biomed Eng 62(4):1077–1088CrossRefPubMedPubMedCentral
16.
go back to reference Franco E, Brujic D, Rea M, Gedroyc WM, Ristic M (2015) Needle-guiding robot for laser ablation of liver tumors under MRI guidance. IEEE ASME Trans Mechatron PP(99):1–14 Franco E, Brujic D, Rea M, Gedroyc WM, Ristic M (2015) Needle-guiding robot for laser ablation of liver tumors under MRI guidance. IEEE ASME Trans Mechatron PP(99):1–14
17.
go back to reference Elhawary H, Tse ZT, Hamed A, Rea M, Davies BL, Lamperth MU (2008) The case for MR-compatible robotics: a review of the state of the art. Int J Med Robot 4(2):105–113CrossRefPubMed Elhawary H, Tse ZT, Hamed A, Rea M, Davies BL, Lamperth MU (2008) The case for MR-compatible robotics: a review of the state of the art. Int J Med Robot 4(2):105–113CrossRefPubMed
18.
go back to reference Cleary K, Melzer A, Watson V, Kronreif G, Stoianovici D (2006) Interventional robotic systems: applications and technology state-of-the-art. Minim Invasive Ther Allied Technol 15(2):101–113CrossRefPubMedPubMedCentral Cleary K, Melzer A, Watson V, Kronreif G, Stoianovici D (2006) Interventional robotic systems: applications and technology state-of-the-art. Minim Invasive Ther Allied Technol 15(2):101–113CrossRefPubMedPubMedCentral
19.
go back to reference Monfaredi R, Sze R, Safdar N, Sharma K, Cleary K (2013) Patient mounted CT and MRI-compatible shoulder arthrography robot for needle guidance in pediatric interventional procedures. The Hamlyn Symposium on Medical Robotics, London, pp 117–118 Monfaredi R, Sze R, Safdar N, Sharma K, Cleary K (2013) Patient mounted CT and MRI-compatible shoulder arthrography robot for needle guidance in pediatric interventional procedures. The Hamlyn Symposium on Medical Robotics, London, pp 117–118
20.
go back to reference Monfaredi R, Seifabadi R, Iordachita I, Sze R, Safdar NM, Sharma K, Fricke S, Krieger A, Cleary K (2014) A prototype body-mounted MRI-compatible robot for needle guidance in shoulder arthrography. In: Proceedings of IEEE RAS & EMBS Int. Conf. Biomedical Robotics and Biomechatronics (BioRob), Sao Paulo, Brazil, pp 40–45 Monfaredi R, Seifabadi R, Iordachita I, Sze R, Safdar NM, Sharma K, Fricke S, Krieger A, Cleary K (2014) A prototype body-mounted MRI-compatible robot for needle guidance in shoulder arthrography. In: Proceedings of IEEE RAS & EMBS Int. Conf. Biomedical Robotics and Biomechatronics (BioRob), Sao Paulo, Brazil, pp 40–45
21.
go back to reference Monfaredi R, Wilson E, Sze R, Sharma K, Azizi B, Iordachita I, Cleary K (2015) Shoulder-mounted robot for MRI-guided arthrography: accuracy and mounting study. In: International Conference of the IEEE Engineering in Medicine and Biology Society, Italy, pp 3643–3646 Monfaredi R, Wilson E, Sze R, Sharma K, Azizi B, Iordachita I, Cleary K (2015) Shoulder-mounted robot for MRI-guided arthrography: accuracy and mounting study. In: International Conference of the IEEE Engineering in Medicine and Biology Society, Italy, pp 3643–3646
22.
go back to reference Churchill E, McConville JE (1976) Sampling and data gathering strategies for future USAF Anthropometry. Webb Associates, Inc. A/F Aerospace Medical Res 2-76, OHIO, AMRL-TR-74-102 1976 Churchill E, McConville JE (1976) Sampling and data gathering strategies for future USAF Anthropometry. Webb Associates, Inc. A/F Aerospace Medical Res 2-76, OHIO, AMRL-TR-74-102 1976
25.
go back to reference Tokuda J, Song S, Tuncali K, Tempany C, Hata N (2013) Configurable automatic detection and registration of fiducial frames for device to image registration in MRI guided prostate interventions. Med Image Comput Comput Assist Interv 16(03):355–362PubMedPubMedCentral Tokuda J, Song S, Tuncali K, Tempany C, Hata N (2013) Configurable automatic detection and registration of fiducial frames for device to image registration in MRI guided prostate interventions. Med Image Comput Comput Assist Interv 16(03):355–362PubMedPubMedCentral
26.
go back to reference Jussila J, Leppaniemi A, Paronen M, Kulomaki E (2005) Ballistic skin simulant. Forensic Sci Int 1(150):63–71CrossRef Jussila J, Leppaniemi A, Paronen M, Kulomaki E (2005) Ballistic skin simulant. Forensic Sci Int 1(150):63–71CrossRef
Metadata
Title
Development of a shoulder-mounted robot for MRI-guided needle placement: phantom study
Authors
Reza Monfaredi
Iulian Iordachita
Emmanuel Wilson
Raymond Sze
Karun Sharma
Axel Krieger
Stanley Fricke
Kevin Cleary
Publication date
01-11-2018
Publisher
Springer International Publishing
Published in
International Journal of Computer Assisted Radiology and Surgery / Issue 11/2018
Print ISSN: 1861-6410
Electronic ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-018-1839-y

Other articles of this Issue 11/2018

International Journal of Computer Assisted Radiology and Surgery 11/2018 Go to the issue