01-08-2017 | Original Article

Grasper having tactile sensing function using acoustic reflection for laparoscopic surgery

Authors: Hiep Hoang Ly, Yoshihiro Tanaka, Tomohiro Fukuda, Akihito Sano

Published in: International Journal of Computer Assisted Radiology and Surgery | Issue 8/2017

Login to get access

Abstract

Purpose:

In current minimally invasive surgery techniques, the tactile information available to the surgeon is limited. Improving tactile sensation could enhance the operability of surgical instruments. Considering surgical applications, requirements such as having electrical safety, a simple structure, and sterilization capability should be considered. The current study sought to develop a grasper that can measure grasping force at the tip, based on a previously proposed tactile sensing method using acoustic reflection. This method can satisfy the requirements for surgical applications because it has no electrical element within the part that is inserted into the patient’s body.

Methods:

We integrated our acoustic tactile sensing method into a conventional grasping forceps instrument. We designed the instrument so that acoustic cavities within a grasping arm and a fork sleeve were connected by a small cavity in a pivoting joint. In this design, when the angle between the two grasping arms changes during grasping, the total length and local curvature of the acoustic cavity remain unchanged. Thus, the grasping force can be measured regardless of the orientation of the grasping arm.

Results:

We developed a prototype sensorized grasper based on our proposed design. Fundamental tests revealed that sensor output increased with increasing contact force applied to the grasping arm, and the angle of the grasping arm did not significantly affect the sensor output. Moreover, the results of a grasping test, in which objects with different softness characteristics were held by the grasper, revealed that the grasping force could be appropriately adjusted to handle different objects on the basis of sensor output.

Conclusions:

Experimental results demonstrated that the prototype grasper can measure grasping force, enabling safe and stable grasping.
Literature
1.
go back to reference Semere W, Kitagawa M, Okamura AM (2004) Teleoperation with sensor/actuator asymmetry: task performance with partial force feedback. In: Proceedings of the IEEE international symposium on haptic interfaces for virtual environment and teleoperator systems. doi:10.1109/HAPTIC.2004.1287186 Semere W, Kitagawa M, Okamura AM (2004) Teleoperation with sensor/actuator asymmetry: task performance with partial force feedback. In: Proceedings of the IEEE international symposium on haptic interfaces for virtual environment and teleoperator systems. doi:10.​1109/​HAPTIC.​2004.​1287186
2.
go back to reference Wagner CR, Stylopoulos N, Howe RD (2002) The role of force feedback in surgery: analysis of blunt dissection. In: Proceedings of the IEEE symposium on haptic interfaces for virtual environment and teleoperator systems. doi:10.1109/HAPTIC.2002.998943 Wagner CR, Stylopoulos N, Howe RD (2002) The role of force feedback in surgery: analysis of blunt dissection. In: Proceedings of the IEEE symposium on haptic interfaces for virtual environment and teleoperator systems. doi:10.​1109/​HAPTIC.​2002.​998943
3.
go back to reference Braun EU, Mayer H, Knoll A, Lange R, Bauernschmitt R (2008) The must-have in robotic heart surgery: haptic feedback. In: Bozovic V (ed) Medical robotics. I-Tech Education and Publishing, Rijeka, pp 9–20 Braun EU, Mayer H, Knoll A, Lange R, Bauernschmitt R (2008) The must-have in robotic heart surgery: haptic feedback. In: Bozovic V (ed) Medical robotics. I-Tech Education and Publishing, Rijeka, pp 9–20
4.
go back to reference Puangmali P, Althoefer K, Seneviratne LD, Murphy D, Dasgupta P (2008) State-of-the-art in force and tactile sensing for minimally invasive surgery. IEEE Sens J 8(4):371–381CrossRef Puangmali P, Althoefer K, Seneviratne LD, Murphy D, Dasgupta P (2008) State-of-the-art in force and tactile sensing for minimally invasive surgery. IEEE Sens J 8(4):371–381CrossRef
5.
go back to reference King CH, Culjat MO, Franco ML, Bisley JW, Carman GP, Dutson EP, Grundfest WS (2009) A multielement tactile feedback system for robot-assisted minimally invasive surgery. IEEE Trans Haptics 2(1):52–56CrossRefPubMed King CH, Culjat MO, Franco ML, Bisley JW, Carman GP, Dutson EP, Grundfest WS (2009) A multielement tactile feedback system for robot-assisted minimally invasive surgery. IEEE Trans Haptics 2(1):52–56CrossRefPubMed
6.
go back to reference Kuwana K, Nakai A, Masamune K, Dohi T (2013): A grasping forceps with a triaxial MEMS tactile sensor for quantification of stresses on organs. In: Proceedings of the international conference of the IEEE engineering in medicine and biology society, pp 4490–4493 Kuwana K, Nakai A, Masamune K, Dohi T (2013): A grasping forceps with a triaxial MEMS tactile sensor for quantification of stresses on organs. In: Proceedings of the international conference of the IEEE engineering in medicine and biology society, pp 4490–4493
7.
go back to reference Valdastri P, Harada K, Menciassi A, Beccai L, Stefanini C, Fujie M, Dario P (2006) Integration of a miniaturised triaxial force sensor in a minimally invasive surgical tool. IEEE Trans Biomed Eng 53(11):2397–2400CrossRefPubMed Valdastri P, Harada K, Menciassi A, Beccai L, Stefanini C, Fujie M, Dario P (2006) Integration of a miniaturised triaxial force sensor in a minimally invasive surgical tool. IEEE Trans Biomed Eng 53(11):2397–2400CrossRefPubMed
8.
go back to reference Dargahi J, Parameswaran M, Payandeh S (2000) A micromachined piezoelectric tactile sensor for an endoscopic grasper—theory, fabrication and experiments. J Microelectromech Syst 9:329–335CrossRef Dargahi J, Parameswaran M, Payandeh S (2000) A micromachined piezoelectric tactile sensor for an endoscopic grasper—theory, fabrication and experiments. J Microelectromech Syst 9:329–335CrossRef
9.
go back to reference Qasaimeh MA, Sokhanvar S, Dargahi J, Kahrizi M (2009) PVDF-based microfabricated tactile sensor for minimally invasive surgery. J Microelectromech Syst 18(1):195–207CrossRef Qasaimeh MA, Sokhanvar S, Dargahi J, Kahrizi M (2009) PVDF-based microfabricated tactile sensor for minimally invasive surgery. J Microelectromech Syst 18(1):195–207CrossRef
10.
go back to reference Sokhanvar S, Packirisamy M, Dargahi J (2007) A multifunctional PVDF-based tactile sensor for minimally invasive surgery. Smart Mater Struct 16:989–998CrossRef Sokhanvar S, Packirisamy M, Dargahi J (2007) A multifunctional PVDF-based tactile sensor for minimally invasive surgery. Smart Mater Struct 16:989–998CrossRef
11.
go back to reference Hong MB, Jo YH (2012) Design and evaluation of 2-DOF compliant forceps with force-sensing capability for minimally invasive robot surgery. IEEE Trans Robot 28(4):932–941CrossRef Hong MB, Jo YH (2012) Design and evaluation of 2-DOF compliant forceps with force-sensing capability for minimally invasive robot surgery. IEEE Trans Robot 28(4):932–941CrossRef
12.
go back to reference Seibold U, Kuebler B, Hirzinger G (2008) Prototypic force feedback instrument for minimally invasive robotic surgery. In: Bozovic V (ed) Medical robotics. I-Tech Education and Publishing, Rijeka, pp 377–400 Seibold U, Kuebler B, Hirzinger G (2008) Prototypic force feedback instrument for minimally invasive robotic surgery. In: Bozovic V (ed) Medical robotics. I-Tech Education and Publishing, Rijeka, pp 377–400
13.
go back to reference Hagn U, Konietschke R, Tobergte A, Nickl M, Jorg S, Kuebler B, Passig G, Groger M, Frohlich F, Seibold U, Le-Tien L, Albu- Schaffer A, Nothhelfer A, Hacker F, Grebenstein M, Hirzinger G (2010) DLR MiroSurge: a versatile system for research in endoscopic telesurgery. Int J CARS 5(2):183–193CrossRef Hagn U, Konietschke R, Tobergte A, Nickl M, Jorg S, Kuebler B, Passig G, Groger M, Frohlich F, Seibold U, Le-Tien L, Albu- Schaffer A, Nothhelfer A, Hacker F, Grebenstein M, Hirzinger G (2010) DLR MiroSurge: a versatile system for research in endoscopic telesurgery. Int J CARS 5(2):183–193CrossRef
14.
go back to reference Kim U, Lee DL, Yoon WJ, Hannaford B, Choi HR (2015) Force sensor integrated surgical forceps for minimally invasive robotic surgery. IEEE Trans Robot 31(5):1214–1224CrossRef Kim U, Lee DL, Yoon WJ, Hannaford B, Choi HR (2015) Force sensor integrated surgical forceps for minimally invasive robotic surgery. IEEE Trans Robot 31(5):1214–1224CrossRef
15.
go back to reference Trejos AL, Patel RV, Naish MD, Lyle AC, Schlachta CM (2009) A sensorized instrument for skills assessment and training in minimally invasive surgery. J Med Devices 3(4):041002CrossRef Trejos AL, Patel RV, Naish MD, Lyle AC, Schlachta CM (2009) A sensorized instrument for skills assessment and training in minimally invasive surgery. J Med Devices 3(4):041002CrossRef
16.
go back to reference Arata J, Mitsuishi M, Warisawa S, Tanaka K, Yoshizawa T, Hashizume M (2005) Development of a dexterous minimally invasive surgical system with augmented force feedback capability. In: Proceedings of the IEEE/RSJ international conference on intelligent robots and systems, pp 3738–3743 Arata J, Mitsuishi M, Warisawa S, Tanaka K, Yoshizawa T, Hashizume M (2005) Development of a dexterous minimally invasive surgical system with augmented force feedback capability. In: Proceedings of the IEEE/RSJ international conference on intelligent robots and systems, pp 3738–3743
17.
go back to reference Son HI, Bhattacharjee T, Lee DY (2010) Estimation of environmental force for the haptic interface of robotic surgery. Int J Med Robot Comput Assist Surg 6(2):221–230 Son HI, Bhattacharjee T, Lee DY (2010) Estimation of environmental force for the haptic interface of robotic surgery. Int J Med Robot Comput Assist Surg 6(2):221–230
18.
go back to reference Tadano K, Kawashima K (2009) Development of a pneumatically driven forceps manipulator IBIS IV. In: Proceedings of ICROS-SICE international joint conference, pp 3815–3818 Tadano K, Kawashima K (2009) Development of a pneumatically driven forceps manipulator IBIS IV. In: Proceedings of ICROS-SICE international joint conference, pp 3815–3818
19.
go back to reference Takaki T, Omasa Y, Ishii I, Kawahara T, Okajima M (2010) Force visualization mechanism using a moire fringe applied to endoscopic surgical instruments. In: Proceedings of the IEEE international conference on robotics and automation, pp 3648–3653 Takaki T, Omasa Y, Ishii I, Kawahara T, Okajima M (2010) Force visualization mechanism using a moire fringe applied to endoscopic surgical instruments. In: Proceedings of the IEEE international conference on robotics and automation, pp 3648–3653
20.
go back to reference Faragasso A, Stilli A, Bimbo J, Noh Y, Liu H, Nanayakkara T, Dasgupta P, Wurdemann HA, Althoefer K (2014) Endoscopic add-on stiffness probe for real-time soft surface characterisation in MIS. In: Proceedings of the international conference of the IEEE engineering in medicine and biology society, pp 6517–6520 Faragasso A, Stilli A, Bimbo J, Noh Y, Liu H, Nanayakkara T, Dasgupta P, Wurdemann HA, Althoefer K (2014) Endoscopic add-on stiffness probe for real-time soft surface characterisation in MIS. In: Proceedings of the international conference of the IEEE engineering in medicine and biology society, pp 6517–6520
21.
go back to reference Li J, Liu H, Althoefer K, Seneviratne LD (2012) A stiffness probe based on force and vision sensing for soft tissue diagnosis. In: Proceedings of the international conference of the IEEE engineering in medicine and biology society, pp 944–947 Li J, Liu H, Althoefer K, Seneviratne LD (2012) A stiffness probe based on force and vision sensing for soft tissue diagnosis. In: Proceedings of the international conference of the IEEE engineering in medicine and biology society, pp 944–947
22.
go back to reference Watanabe T, Iwai T, Fujihira Y, Koyama T, Yoneyama T (2016) Stiffness measurement system using endoscopes with a visualization method. IEEE Sens J 16(15):5889–5897CrossRef Watanabe T, Iwai T, Fujihira Y, Koyama T, Yoneyama T (2016) Stiffness measurement system using endoscopes with a visualization method. IEEE Sens J 16(15):5889–5897CrossRef
23.
go back to reference Kawahara T, Kaneko M (2005) Non-contact stiffness imager for medical application. In: Proceedings of the IEEE international conference on information acquisition, pp 350–355 Kawahara T, Kaneko M (2005) Non-contact stiffness imager for medical application. In: Proceedings of the IEEE international conference on information acquisition, pp 350–355
24.
go back to reference Puangmali P, Liu H, Seneviratne LD, Dasgupta P, Althoefer K (2012) Miniature 3-axis distal force sensor for minimally invasive surgical palpation. IEEE ASME Trans Mechatron 17(4):646–656CrossRef Puangmali P, Liu H, Seneviratne LD, Dasgupta P, Althoefer K (2012) Miniature 3-axis distal force sensor for minimally invasive surgical palpation. IEEE ASME Trans Mechatron 17(4):646–656CrossRef
25.
go back to reference Tanaka Y, Fukuda T, Fujiwara M, Sano A (2015) Tactile sensor using acoustic reflection for lump detection in laparoscopic surgery. Int J CARS 10(2):183–193CrossRef Tanaka Y, Fukuda T, Fujiwara M, Sano A (2015) Tactile sensor using acoustic reflection for lump detection in laparoscopic surgery. Int J CARS 10(2):183–193CrossRef
26.
go back to reference Fukuda T, Tanaka Y, Fujiwara M, Sano A (2015) Softness measurement by forceps-type tactile sensor using acoustic reflection. In: Proceedings of the IEEE/RSJ international conference on intelligent robots and systems, pp 3791–3796 Fukuda T, Tanaka Y, Fujiwara M, Sano A (2015) Softness measurement by forceps-type tactile sensor using acoustic reflection. In: Proceedings of the IEEE/RSJ international conference on intelligent robots and systems, pp 3791–3796
27.
go back to reference Figliola RS, Beasley DE (2011) Theory and design for mechanical measurements, 5th edn. Wiley, New York Figliola RS, Beasley DE (2011) Theory and design for mechanical measurements, 5th edn. Wiley, New York
28.
go back to reference Rosen J, Brown JD, De S, Sinanan M, Hannaford B (2008) Biomechanical properties of abdominal organs in vivo and postmortem under compression loads. J Biomech Eng 130(2):021020CrossRefPubMed Rosen J, Brown JD, De S, Sinanan M, Hannaford B (2008) Biomechanical properties of abdominal organs in vivo and postmortem under compression loads. J Biomech Eng 130(2):021020CrossRefPubMed
29.
go back to reference Tanaka Y, Aragaki S, Fukuda T, Fujiwara M, Sano A (2014) A study on tactile display for haptic sensing system with sensory feedback for laparoscopic surgery. In: Proceedings of the international symposium on micro-nanomechatronics and human science. doi:10.1109/MHS.2014.7006169 Tanaka Y, Aragaki S, Fukuda T, Fujiwara M, Sano A (2014) A study on tactile display for haptic sensing system with sensory feedback for laparoscopic surgery. In: Proceedings of the international symposium on micro-nanomechatronics and human science. doi:10.​1109/​MHS.​2014.​7006169
Metadata
Title
Grasper having tactile sensing function using acoustic reflection for laparoscopic surgery
Authors
Hiep Hoang Ly
Yoshihiro Tanaka
Tomohiro Fukuda
Akihito Sano
Publication date
01-08-2017
Publisher
Springer International Publishing
Published in
International Journal of Computer Assisted Radiology and Surgery / Issue 8/2017
Print ISSN: 1861-6410
Electronic ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-017-1592-7

Other articles of this Issue 8/2017

International Journal of Computer Assisted Radiology and Surgery 8/2017 Go to the issue