Skip to main content
Top
Published in: International Journal of Computer Assisted Radiology and Surgery 6/2017

01-06-2017 | Original Article

First clinical use of the EchoTrack guidance approach for radiofrequency ablation of thyroid gland nodules

Authors: Alfred Michael Franz, Alexander Seitel, Nasrin Bopp, Christian Erbelding, Dominique Cheray, Stefan Delorme, Frank Grünwald, Hüdayi Korkusuz, Lena Maier-Hein

Published in: International Journal of Computer Assisted Radiology and Surgery | Issue 6/2017

Login to get access

Abstract

Purpose

Percutaneous radiofrequency ablation (RFA) of thyroid nodules is an alternative to surgical resection that offers the benefits of minimal scars for the patient, lower complication rates, and shorter treatment times. Ultrasound (US) is the preferred modality for guiding these procedures. The needle is usually kept within the US scanning plane to ensure needle visibility. However, this restricts flexibility in both transducer and needle movement and renders the procedure difficult, especially for inexperienced users. Existing navigation solutions often involve electromagnetic (EM) tracking, which requires placement of an external field generator (FG) in close proximity of the intervention site in order to avoid distortion of the EM field. This complicates the clinical workflow as placing the FG while ensuring that it neither restricts the physician’s workspace nor affects tracking accuracy is awkward and time-consuming.

Methods

The EchoTrack concept overcomes these issues by combining the US probe and the EM FG in one modality, simultaneously providing both real-time US and tracking data without requiring the placement of an external FG for tracking. We propose a system and workflow to use EchoTrack for RFA of thyroid nodules.

Results

According to our results, the overall error of the EchoTrack system resulting from errors related to tracking and calibration is below 2 mm. Navigated thyroid RFA with the proposed concept is clinically feasible. Motion of internal critical structures relative to external markers can be up to several millimeters in extreme cases.

Conclusions

The EchoTrack concept with its simple setup, flexibility, improved needle visualization, and additional guidance information has high potential to be clinically used for thyroid RFA.
Literature
1.
go back to reference Baek JH, Lee JH, Valcavi R, Pacella CM, Rhim H, Na DG (2011) Thermal ablation for benign thyroid nodules: radiofrequency and laser. Korean J Radiol 12(5):525–540CrossRefPubMedPubMedCentral Baek JH, Lee JH, Valcavi R, Pacella CM, Rhim H, Na DG (2011) Thermal ablation for benign thyroid nodules: radiofrequency and laser. Korean J Radiol 12(5):525–540CrossRefPubMedPubMedCentral
2.
go back to reference Erbelding C, Franz A, Seitel A, Bopp N, Kohlhase K, Grunwald F, Maier-Hein L (2017) Ultrasound-navigated radiofrequency ablation of thyroid nodules with integrated electromagnetic tracking: comparison with conventional ultrasound guidance in gelatin models. Int J Comput Assist Radiol Surg. doi:10.1007/s11548-017-1544-2 Erbelding C, Franz A, Seitel A, Bopp N, Kohlhase K, Grunwald F, Maier-Hein L (2017) Ultrasound-navigated radiofrequency ablation of thyroid nodules with integrated electromagnetic tracking: comparison with conventional ultrasound guidance in gelatin models. Int J Comput Assist Radiol Surg. doi:10.​1007/​s11548-017-1544-2
3.
go back to reference Franz AM, Haidegger T, Birkfellner W, Cleary K, Peters TM, Maier-Hein L (2014) Electromagnetic tracking in medicine-a review of technology, validation, and applications. IEEE Trans Med Imaging 33(8):1702–1725CrossRefPubMed Franz AM, Haidegger T, Birkfellner W, Cleary K, Peters TM, Maier-Hein L (2014) Electromagnetic tracking in medicine-a review of technology, validation, and applications. IEEE Trans Med Imaging 33(8):1702–1725CrossRefPubMed
4.
go back to reference Franz AM, März K, Hummel J, Birkfellner W, Bendl R, Delorme S, Schlemmer HP, Meinzer HP, Maier-Hein L (2012) Electromagnetic tracking for US-guided interventions: standardized assessment of a new compact field generator. Int J Comput Assist Radiol Surg 7(6):813–818CrossRefPubMed Franz AM, März K, Hummel J, Birkfellner W, Bendl R, Delorme S, Schlemmer HP, Meinzer HP, Maier-Hein L (2012) Electromagnetic tracking for US-guided interventions: standardized assessment of a new compact field generator. Int J Comput Assist Radiol Surg 7(6):813–818CrossRefPubMed
5.
go back to reference Franz AM, Simpfendörfer T, Garoussi C, Majlesara A, März K, Meinzer HP, Mehrabi A, Teber D, Maier-Hein L (2015) EchoTrack for simultaneous EM tracking and US imaging: initial experience in ventilated swine cadaver. Int J Comput Assist Radiol Surg 10(Suppl 1):52 Franz AM, Simpfendörfer T, Garoussi C, Majlesara A, März K, Meinzer HP, Mehrabi A, Teber D, Maier-Hein L (2015) EchoTrack for simultaneous EM tracking and US imaging: initial experience in ventilated swine cadaver. Int J Comput Assist Radiol Surg 10(Suppl 1):52
6.
go back to reference Horn BKP (1987) Closed-form solution of absolute orientation using unit quaternions. J Opt Soc Am A 4(4):629–642CrossRef Horn BKP (1987) Closed-form solution of absolute orientation using unit quaternions. J Opt Soc Am A 4(4):629–642CrossRef
7.
go back to reference Hummel JB, Bax MR, Figl ML, Kang Y, Maurer C, Birkfellner WW, Bergmann H, Shahidi R (2005) Design and application of an assessment protocol for electromagnetic tracking systems. Med Phys 32(7):2371–2379CrossRef Hummel JB, Bax MR, Figl ML, Kang Y, Maurer C, Birkfellner WW, Bergmann H, Shahidi R (2005) Design and application of an assessment protocol for electromagnetic tracking systems. Med Phys 32(7):2371–2379CrossRef
8.
go back to reference Jeong WK, Baek JH, Rhim H, Kim YS, Kwak MS, Jeong HJ, Lee D (2008) Radiofrequency ablation of benign thyroid nodules: safety and imaging follow-up in 236 patients. Eur Radiol 18(6):1244–1250CrossRefPubMed Jeong WK, Baek JH, Rhim H, Kim YS, Kwak MS, Jeong HJ, Lee D (2008) Radiofrequency ablation of benign thyroid nodules: safety and imaging follow-up in 236 patients. Eur Radiol 18(6):1244–1250CrossRefPubMed
9.
go back to reference Korkusuz Y, Erbelding C, Kohlhase K, Luboldt W, Happel C, Grunwald F (2016) Bipolar radiofrequency ablation of benign symptomatic thyroid nodules: initial experience. Rofo 188(7):671–675PubMed Korkusuz Y, Erbelding C, Kohlhase K, Luboldt W, Happel C, Grunwald F (2016) Bipolar radiofrequency ablation of benign symptomatic thyroid nodules: initial experience. Rofo 188(7):671–675PubMed
10.
go back to reference Maier-Hein L, Franz AM, Birkfellner W, Hummel J, Gergel I, Wegner I, Meinzer HP (2012) Standardized assessment of new electromagnetic field generators in an interventional radiology setting. Med Phys 39(6):3424–3434CrossRefPubMed Maier-Hein L, Franz AM, Birkfellner W, Hummel J, Gergel I, Wegner I, Meinzer HP (2012) Standardized assessment of new electromagnetic field generators in an interventional radiology setting. Med Phys 39(6):3424–3434CrossRefPubMed
11.
go back to reference März K, Franz AM, Seitel A, Winterstein A, Hafezi M, Saffari A, Bendl R, Stieltjes B, Meinzer HP, Mehrabi A, Maier-Hein L (2014) Interventional real-time ultrasound imaging with an integrated electromagnetic field generator. Int J Comput Assist Radiol Surg 9(5):759–768CrossRefPubMed März K, Franz AM, Seitel A, Winterstein A, Hafezi M, Saffari A, Bendl R, Stieltjes B, Meinzer HP, Mehrabi A, Maier-Hein L (2014) Interventional real-time ultrasound imaging with an integrated electromagnetic field generator. Int J Comput Assist Radiol Surg 9(5):759–768CrossRefPubMed
12.
go back to reference Mercier L, Lang T, Lindseth F, Collins D (2005) A review of calibration techniques for freehand 3-d ultrasound systems. Ultrasound Med Biol 31(4):449–471CrossRefPubMed Mercier L, Lang T, Lindseth F, Collins D (2005) A review of calibration techniques for freehand 3-d ultrasound systems. Ultrasound Med Biol 31(4):449–471CrossRefPubMed
13.
go back to reference Nolden M, Zelzer S, Seitel A, Wald D, Müller M, Franz AM, Maleike D, Fangerau M, Baumhauer M, Maier-Hein L, Maier-Hein KH, Meinzer HP, Wolf I (2013) The medical imaging interaction toolkit: challenges and advances : 10 years of open-source development. Int J Comput Assist Radiol Surg 8(4):607–620 Nolden M, Zelzer S, Seitel A, Wald D, Müller M, Franz AM, Maleike D, Fangerau M, Baumhauer M, Maier-Hein L, Maier-Hein KH, Meinzer HP, Wolf I (2013) The medical imaging interaction toolkit: challenges and advances : 10 years of open-source development. Int J Comput Assist Radiol Surg 8(4):607–620
14.
go back to reference Reiners C, Wegscheider K, Schicha H, Theissen P, Vaupel R, Wrbitzky R, Schumm-Draeger PM (2004) Prevalence of thyroid disorders in the working population of Germany: ultrasonography screening in 96,278 unselected employees. Thyroid 14(11):926–932CrossRefPubMed Reiners C, Wegscheider K, Schicha H, Theissen P, Vaupel R, Wrbitzky R, Schumm-Draeger PM (2004) Prevalence of thyroid disorders in the working population of Germany: ultrasonography screening in 96,278 unselected employees. Thyroid 14(11):926–932CrossRefPubMed
15.
go back to reference Shin JH, Baek JH, Ha EJ, Lee JH (2012) Radiofrequency ablation of thyroid nodules: basic princicples and clinical applications. Int J Endocrinol 2012:1–7CrossRef Shin JH, Baek JH, Ha EJ, Lee JH (2012) Radiofrequency ablation of thyroid nodules: basic princicples and clinical applications. Int J Endocrinol 2012:1–7CrossRef
16.
go back to reference Stolka P, Foroughi P, Rendina M, Weiss C, Hager G, Boctor E (2014) Needle guidance using handheld stereo vision and projection for ultrasound-based interventions. Med Image Comput Comput Assist Interv 8674:684–691 Stolka P, Foroughi P, Rendina M, Weiss C, Hager G, Boctor E (2014) Needle guidance using handheld stereo vision and projection for ultrasound-based interventions. Med Image Comput Comput Assist Interv 8674:684–691
17.
go back to reference Turtulici G, Orlandi D, Corazza A, Sartoris R, Derchi LE, Silvestri E, Baek JH (2014) Percutaneous radiofrequency ablation of benign thyroid nodules assisted by a virtual needle tracking system. Ultrasound Med Biol 40(7):1447–1452CrossRefPubMed Turtulici G, Orlandi D, Corazza A, Sartoris R, Derchi LE, Silvestri E, Baek JH (2014) Percutaneous radiofrequency ablation of benign thyroid nodules assisted by a virtual needle tracking system. Ultrasound Med Biol 40(7):1447–1452CrossRefPubMed
18.
go back to reference Wilson E, Yaniv Z, Zhang H, Nafis C, Shen E, Shechter G, Wiles A D, Peters T, Lindisch D, Cleary K (2007) A hardware and software protocol for the evaluation of electromagnetic tracker accuracy in the clinical environment: a multi-center study. In: Proceedings of SPIE medical imaging: visualization and image-guided procedures 6509: 1–11 Wilson E, Yaniv Z, Zhang H, Nafis C, Shen E, Shechter G, Wiles A D, Peters T, Lindisch D, Cleary K (2007) A hardware and software protocol for the evaluation of electromagnetic tracker accuracy in the clinical environment: a multi-center study. In: Proceedings of SPIE medical imaging: visualization and image-guided procedures 6509: 1–11
19.
go back to reference Wood BJ, Krücker J, Abi-Jaoudeh N, Locklin JK, Levy E, Xu S, Solbiati L, Kapoor A, Amalou H, Venkatesan AM (2010) Navigation systems for ablation. J Vasc Interv Radiol 21(8):S257–263CrossRefPubMedPubMedCentral Wood BJ, Krücker J, Abi-Jaoudeh N, Locklin JK, Levy E, Xu S, Solbiati L, Kapoor A, Amalou H, Venkatesan AM (2010) Navigation systems for ablation. J Vasc Interv Radiol 21(8):S257–263CrossRefPubMedPubMedCentral
Metadata
Title
First clinical use of the EchoTrack guidance approach for radiofrequency ablation of thyroid gland nodules
Authors
Alfred Michael Franz
Alexander Seitel
Nasrin Bopp
Christian Erbelding
Dominique Cheray
Stefan Delorme
Frank Grünwald
Hüdayi Korkusuz
Lena Maier-Hein
Publication date
01-06-2017
Publisher
Springer International Publishing
Published in
International Journal of Computer Assisted Radiology and Surgery / Issue 6/2017
Print ISSN: 1861-6410
Electronic ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-017-1560-2

Other articles of this Issue 6/2017

International Journal of Computer Assisted Radiology and Surgery 6/2017 Go to the issue