Abstract

A compact ubiquitous-health monitor operated by single 8-bit microcontroller was made. An integer signal processing algorithm for this microcontroller was developed and digital filtering of ECG (electrocardiogram) and PPG (photoplethysmogram) was performed. Rounding-off errors due to integer operation was solved by increasing the number of effective integer digits during CPU operation; digital filter coefficients and data expressed in decimal points were multiplied by a certain number and converted into integers. After filter operation, the actual values were retrieved by dividing with the same number and selecting available highest bits. Our results showed comparable accuracies to those computed by a commercial software. Compared with a floating-point calculation by the same microcontroller, the computation speed became faster by 1.45 ∼ 2.0 times depending on various digital filtering cases. Our algorithm was successfully tested for remote health monitoring with multiple users. If our algorithm were not used, our health monitor should have used additional microcontrollers or DSP chip. The proposed algorithm reduced the size and cost of our health monitor substantially.
Literature
1.
go back to reference Otto, C., Milenkovic, A., Sanders, C., and Jovanov, E., System architecture of a wireless body area sensor network for ubiquitous health monitoring. J. Mobile Multimedia. 1(4):307–326, 2006. Otto, C., Milenkovic, A., Sanders, C., and Jovanov, E., System architecture of a wireless body area sensor network for ubiquitous health monitoring. J. Mobile Multimedia. 1(4):307–326, 2006.
2.
go back to reference Kim, S., Nakamura, H., Yoshida, T., Kishimoto, M., Imai, Y., Matsuki, N., Ishikawa, T., and Yamaguchi, T., Development of a wearable system module for monitoring physical and mental workload. Telemedicine and e-Health. 14(9):939–945, 2008.CrossRef Kim, S., Nakamura, H., Yoshida, T., Kishimoto, M., Imai, Y., Matsuki, N., Ishikawa, T., and Yamaguchi, T., Development of a wearable system module for monitoring physical and mental workload. Telemedicine and e-Health. 14(9):939–945, 2008.CrossRef
3.
5.
go back to reference Herrera-Bendezu, L. G., Garcia, J., Denys, B. G., Real-time digital filters for ECG signals: evaluation and new designs, Computers in Cardiology. Proceedings 23–26(Sept), 133–136, 1991 Herrera-Bendezu, L. G., Garcia, J., Denys, B. G., Real-time digital filters for ECG signals: evaluation and new designs, Computers in Cardiology. Proceedings 23–26(Sept), 133–136, 1991
9.
go back to reference Huh, Y., and Yoon, G., Accuracy improvement of respiration rate based on photo-plethysmography by enhancing motion artifact. J Korean Sensors Soc. 17(6)447–453, 2008.CrossRef Huh, Y., and Yoon, G., Accuracy improvement of respiration rate based on photo-plethysmography by enhancing motion artifact. J Korean Sensors Soc. 17(6)447–453, 2008.CrossRef
10.
go back to reference Yoon, Y., Cho, J. H., Yoon, G., Non-constrained blood pressure monitoring using ECG and PPG for personal healthcare. J. Med. Syst. in press. Yoon, Y., Cho, J. H., Yoon, G., Non-constrained blood pressure monitoring using ECG and PPG for personal healthcare. J. Med. Syst. in press.
11.
go back to reference Helleputte, N., Tomasik, J. M., Galjan, W., Mora-Sanchez, A., Schroeder, D., Krautschneider, W. H., and Puer, R., A flexible system-on-chip (SoC) for biomedical signal acquisition and processing. Sensors Actuators A. 142:361–368, 2008. doi:10.1016/j.sna.2007.04.026.CrossRef Helleputte, N., Tomasik, J. M., Galjan, W., Mora-Sanchez, A., Schroeder, D., Krautschneider, W. H., and Puer, R., A flexible system-on-chip (SoC) for biomedical signal acquisition and processing. Sensors Actuators A. 142:361–368, 2008. doi:10.​1016/​j.​sna.​2007.​04.​026.CrossRef
12.
go back to reference Yang, Y., Huang, X., Yu, X., Real-time monitoring system based on FPGA, Industrial Electronics Society, 2007. IECON 2007. 33rd Annual Conference of the IEEE, 2136–2140, 5–8 Nov., 2007. Yang, Y., Huang, X., Yu, X., Real-time monitoring system based on FPGA, Industrial Electronics Society, 2007. IECON 2007. 33rd Annual Conference of the IEEE, 2136–2140, 5–8 Nov., 2007.
Metadata
Title
ECG/PPG Integer Signal Processing for a Ubiquitous Health Monitoring System
Authors
Woosik Shin
Yong Dae Cha
Gilwon Yoon
Publication date
01-10-2010
Publisher
Springer US
Published in
Journal of Medical Systems / Issue 5/2010
Print ISSN: 0148-5598
Electronic ISSN: 1573-689X
DOI
https://doi.org/10.1007/s10916-009-9304-7

Other articles of this Issue 5/2010

Journal of Medical Systems 5/2010 Go to the issue