Published in: Journal of Clinical Monitoring and Computing 5/2022

Open Access 05-10-2021 | Original Research

Non-invasive capnodynamic mixed venous oxygen saturation during major changes in oxygen delivery

Authors: Anders Svedmyr, Mark Konrad, Mats Wallin, Magnus Hallbäck, Per-Arne Lönnqvist, Jacob Karlsson

Published in: Journal of Clinical Monitoring and Computing | Issue 5/2022

Login to get access

Abstract

Mixed venous oxygen saturation (SvO2) is an important variable in anesthesia and intensive care but currently requires pulmonary artery catheterization. Recently, non-invasive determination of SvO2 (Capno-SvO2) using capnodynamics has shown good agreement against CO-oximetry in an animal model of modest hemodynamic changes. The purpose of the current study was to validate Capno-SvO2 against CO-oximetry during major alterations in oxygen delivery. Furthermore, evaluating fiberoptic SvO2 for its response to the same challenges. Eleven mechanically ventilated pigs were exposed to oxygen delivery changes: increased inhaled oxygen concentration, hemorrhage, crystalloid and blood transfusion, preload reduction and dobutamine infusion. Capno-SvO2 and fiberoptic SvO2 recordings were made in parallel with CO-oximetry. Respiratory quotient, needed for capnodynamic SvO2, was measured by analysis of mixed expired gases. Agreement of absolute values between CO-oximetry and Capno-SvO2 and fiberoptic SvO2 respectively, was assessed using Bland–Altman plots. Ability of Capno- SvO2 and fiberoptic SvO2 to detect change compared to CO-oximetry was assessed using concordance analysis. The interventions caused significant hemodynamic variations. Bias between Capno-SvO2 and CO-oximetry was + 3% points (95% limits of agreements – 7 to + 13). Bias between fiberoptic SvO2 and CO-oximetry was + 1% point, (95% limits of agreements − 7 to + 9). Concordance rate for Capno-SvO2 and fiberoptic SvO2 vs. CO-oximetry was 98% and 93%, respectively. Capno-SvO2 generates absolute values close to CO-oximetry. The performance of Capno-SvO2 vs. CO-oximetry was comparable to the performance of fiberoptic SvO2 vs. CO-oximetry. Capno-SvO2 appears to be a promising tool for non-invasive SvO2 monitoring.
Appendix
Available only for authorised users
Literature
1.
go back to reference Holm J, Håkanson E, Vánky F, Svedjeholm R. Mixed venous oxygen saturation predicts short- and long-term outcome after coronary artery bypass grafting surgery: a retrospective cohort analysis. Br J Anaesth. 2011;107:344–50.CrossRef Holm J, Håkanson E, Vánky F, Svedjeholm R. Mixed venous oxygen saturation predicts short- and long-term outcome after coronary artery bypass grafting surgery: a retrospective cohort analysis. Br J Anaesth. 2011;107:344–50.CrossRef
2.
go back to reference Kalra A, Heitner S, Topalian S. Iatrogenic pulmonary artery rupture during Swan-Ganz catheter placement: a novel therapeutic approach. Catheter Cardiovasc Interv. 2013;81:57–9.CrossRef Kalra A, Heitner S, Topalian S. Iatrogenic pulmonary artery rupture during Swan-Ganz catheter placement: a novel therapeutic approach. Catheter Cardiovasc Interv. 2013;81:57–9.CrossRef
3.
go back to reference Hadian M, Pinsky MR. Evidence-based review of the use of the pulmonary artery catheter: impact data and complications. Crit Care. 2006;10(Suppl 3):8.CrossRef Hadian M, Pinsky MR. Evidence-based review of the use of the pulmonary artery catheter: impact data and complications. Crit Care. 2006;10(Suppl 3):8.CrossRef
4.
go back to reference Karlsson J, Wallin M, Hallbäck M, Lönnqvist PA. Capnodynamic determination of cardiac output in hypoxia-induced pulmonary hypertension in pigs. Br J Anaesth. 2019;122:335–41.CrossRef Karlsson J, Wallin M, Hallbäck M, Lönnqvist PA. Capnodynamic determination of cardiac output in hypoxia-induced pulmonary hypertension in pigs. Br J Anaesth. 2019;122:335–41.CrossRef
5.
go back to reference Hällsjö Sander C, Hallbäck M, Wallin M, Emtell P, Oldner A, Björne H. Novel continuous capnodynamic method for cardiac output assessment during mechanical ventilation. Br J Anaesth. 2014;112:824–31.CrossRef Hällsjö Sander C, Hallbäck M, Wallin M, Emtell P, Oldner A, Björne H. Novel continuous capnodynamic method for cardiac output assessment during mechanical ventilation. Br J Anaesth. 2014;112:824–31.CrossRef
6.
go back to reference Sigmundsson TS, Öhman T, Hallbäck M, Redondo E, Sipmann FS, Wallin M, Oldner A, Hällsjö Sander C, Björne H. Performance of a capnodynamic method estimating effective pulmonary blood flow during transient and sustained hypercapnia. J Clin Monit Comput. 2018;32:311–9.CrossRef Sigmundsson TS, Öhman T, Hallbäck M, Redondo E, Sipmann FS, Wallin M, Oldner A, Hällsjö Sander C, Björne H. Performance of a capnodynamic method estimating effective pulmonary blood flow during transient and sustained hypercapnia. J Clin Monit Comput. 2018;32:311–9.CrossRef
7.
go back to reference Peyton PJ, Wallin M, Hallbäck M. New generation continuous cardiac output monitoring from carbon dioxide elimination. BMC Anesthesiol. 2019;19:28.CrossRef Peyton PJ, Wallin M, Hallbäck M. New generation continuous cardiac output monitoring from carbon dioxide elimination. BMC Anesthesiol. 2019;19:28.CrossRef
9.
go back to reference Karlsson J, Winberg P, Scarr B, Lönnqvist PA, Neovius E, Wallin M, Hallbäck M. Validation of capnodynamic determination of cardiac output by measuring effective pulmonary blood flow: a study in anaesthetised children and piglets. Br J Anaesth. 2018;121:550–8.CrossRef Karlsson J, Winberg P, Scarr B, Lönnqvist PA, Neovius E, Wallin M, Hallbäck M. Validation of capnodynamic determination of cardiac output by measuring effective pulmonary blood flow: a study in anaesthetised children and piglets. Br J Anaesth. 2018;121:550–8.CrossRef
10.
go back to reference Ferrando C, Romero C, Tusman G, Suarez-Sipmann F, Canet J, Dosdá R, Valls P, Villena A, Serralta F, Jurado A, Carrizo J, Navarro J, Parrilla C, Romero JE, Pozo N, Soro M, Villar J, Belda FJ. The accuracy of postoperative, non-invasive Air-Test to diagnose atelectasis in healthy patients after surgery: a prospective, diagnostic pilot study. BMJ Open. 2017;7:e015560.CrossRef Ferrando C, Romero C, Tusman G, Suarez-Sipmann F, Canet J, Dosdá R, Valls P, Villena A, Serralta F, Jurado A, Carrizo J, Navarro J, Parrilla C, Romero JE, Pozo N, Soro M, Villar J, Belda FJ. The accuracy of postoperative, non-invasive Air-Test to diagnose atelectasis in healthy patients after surgery: a prospective, diagnostic pilot study. BMJ Open. 2017;7:e015560.CrossRef
11.
go back to reference Suarez-Sipmann F, Bohm SH, Tusman G. Volumetric capnography: the time has come. Curr Opin Crit Care. 2014;20:333–9.CrossRef Suarez-Sipmann F, Bohm SH, Tusman G. Volumetric capnography: the time has come. Curr Opin Crit Care. 2014;20:333–9.CrossRef
12.
go back to reference Fick A. Uber die messung des Blutquantums in den Hertzvent rikeln. Sitzber Physik Med Ges Wurzburg, 1870, pp 290–91. Fick A. Uber die messung des Blutquantums in den Hertzvent rikeln. Sitzber Physik Med Ges Wurzburg, 1870, pp 290–91.
13.
go back to reference Gedeon A, Krill P, Osterlund B. Pulmonary blood flow (cardiac output) and the effective lung volume determined from a short breath hold using the differential Fick method. J Clin Monit Comput. 2002;17:313–21.CrossRef Gedeon A, Krill P, Osterlund B. Pulmonary blood flow (cardiac output) and the effective lung volume determined from a short breath hold using the differential Fick method. J Clin Monit Comput. 2002;17:313–21.CrossRef
14.
go back to reference Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM. Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies–with specific reference to the measurement of cardiac output. Crit Care. 2009;13:201.CrossRef Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM. Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies–with specific reference to the measurement of cardiac output. Crit Care. 2009;13:201.CrossRef
15.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRef
16.
go back to reference Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17:571–82.CrossRef Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17:571–82.CrossRef
17.
go back to reference Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15:85–91.CrossRef Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15:85–91.CrossRef
18.
go back to reference Herner A, Haller B, Mayr U, Rasch S, Offman L, Schmid R, Huber W. Accuracy and precision of ScvO2 measured with the CeVOX-device: A prospective study in patients with a wide variation of ScvO2-values. PLoS One. 2018;13:e0192073.CrossRef Herner A, Haller B, Mayr U, Rasch S, Offman L, Schmid R, Huber W. Accuracy and precision of ScvO2 measured with the CeVOX-device: A prospective study in patients with a wide variation of ScvO2-values. PLoS One. 2018;13:e0192073.CrossRef
19.
go back to reference Kissoon N, Spenceley N, Krahn G, Milner R. Continuous central venous oxygen saturation monitoring under varying physiological conditions in an animal model. Anaesth Intensive Care. 2010;38:883–9.CrossRef Kissoon N, Spenceley N, Krahn G, Milner R. Continuous central venous oxygen saturation monitoring under varying physiological conditions in an animal model. Anaesth Intensive Care. 2010;38:883–9.CrossRef
20.
go back to reference Iodice FG, Ricci Z, Haiberger R, Favia I, Cogo P. Fiberoptic monitoring of central venous oxygen saturation (PediaSat) in small children undergoing cardiac surgery: continuous is not continuous. F1000Res. 2014;3:23.CrossRef Iodice FG, Ricci Z, Haiberger R, Favia I, Cogo P. Fiberoptic monitoring of central venous oxygen saturation (PediaSat) in small children undergoing cardiac surgery: continuous is not continuous. F1000Res. 2014;3:23.CrossRef
22.
go back to reference Critchley LA, Lee A, Ho AM. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.CrossRef Critchley LA, Lee A, Ho AM. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.CrossRef
23.
go back to reference Porath M, Sinha P, Dudenhausen JW, Luttkus AK. Systematic instrumental errors between oxygen saturation analysers in fetal blood during deep hypoxemia. Clin Chim Acta. 2001;307:151–7.CrossRef Porath M, Sinha P, Dudenhausen JW, Luttkus AK. Systematic instrumental errors between oxygen saturation analysers in fetal blood during deep hypoxemia. Clin Chim Acta. 2001;307:151–7.CrossRef
24.
go back to reference Beards SC, Edwards JD, Nightingale P. The need for quality control in measurement of mixed venous oxygen saturation. Anaesthesia. 1994;49:886–8.CrossRef Beards SC, Edwards JD, Nightingale P. The need for quality control in measurement of mixed venous oxygen saturation. Anaesthesia. 1994;49:886–8.CrossRef
25.
go back to reference Shepherd SJ, Pearse RM. Role of central and mixed venous oxygen saturation measurement in perioperative care. Anesthesiology. 2009;111:649–56.CrossRef Shepherd SJ, Pearse RM. Role of central and mixed venous oxygen saturation measurement in perioperative care. Anesthesiology. 2009;111:649–56.CrossRef
27.
go back to reference Frankenfield DC. Impact of feeding on resting metabolic rate and gas exchange in critically ill patients. J Parenter Enteral Nutr. 2019;43:226–33.CrossRef Frankenfield DC. Impact of feeding on resting metabolic rate and gas exchange in critically ill patients. J Parenter Enteral Nutr. 2019;43:226–33.CrossRef
28.
go back to reference Liu H, Chen Y, Ming D, Wang J, Li Z, Ma X, van Milgen J, Wang F. Integrative analysis of indirect calorimetry and metabolomics profiling reveals alterations in energy metabolism between fed and fasted pigs. J Anim Sci Biotechnol. 2018;9:41.CrossRef Liu H, Chen Y, Ming D, Wang J, Li Z, Ma X, van Milgen J, Wang F. Integrative analysis of indirect calorimetry and metabolomics profiling reveals alterations in energy metabolism between fed and fasted pigs. J Anim Sci Biotechnol. 2018;9:41.CrossRef
29.
go back to reference Sander CH, Sigmundsson T, Hallbäck M, Sipmann FS, Wallin M, Oldner A, Björne H. A modified breathing pattern improves the performance of a continuous capnodynamic method for estimation of effective pulmonary blood flow. J Clin Monit Comput. 2017;31:717–25.CrossRef Sander CH, Sigmundsson T, Hallbäck M, Sipmann FS, Wallin M, Oldner A, Björne H. A modified breathing pattern improves the performance of a continuous capnodynamic method for estimation of effective pulmonary blood flow. J Clin Monit Comput. 2017;31:717–25.CrossRef
Metadata
Title
Non-invasive capnodynamic mixed venous oxygen saturation during major changes in oxygen delivery
Authors
Anders Svedmyr
Mark Konrad
Mats Wallin
Magnus Hallbäck
Per-Arne Lönnqvist
Jacob Karlsson
Publication date
05-10-2021
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 5/2022
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-021-00762-5

Other articles of this Issue 5/2022

Journal of Clinical Monitoring and Computing 5/2022 Go to the issue