Published in: Journal of Clinical Monitoring and Computing 3/2020

01-06-2020 | Original Research

Accuracy and trending ability of hemoglobin measurement by the Pulse CO-Oximeter during vascular surgery

Authors: Rosanna Carmela De Rosa, Giovanni Marco Romano, Roberta Abbate, Antonio Corcione, Edoardo De Robertis

Published in: Journal of Clinical Monitoring and Computing | Issue 3/2020

Login to get access

Abstract

Real time information of Hb concentration can guide a tailored patient blood management. The study investigates the accuracy, precision and trending ability of the Pulse CO-Oximeter (SpHb) and blood gas analyzer in measuring the Hb concentration, compared to hematological analysis, in surgery at high risk of hemorrhage. We performed an observational study, involving 48 patients undergoing abdominal aortic open surgery. The primary endpoints of the study were to compare the accuracy in measuring the Hb concentration using non-invasive method (Masimo rainbow SET® Radical 7 Pulse CO-Oximetry™) compared to the values provided by invasive conventional blood gas analyzer and hematological analysis. The secondary endpoint was to compare the differences between the baseline and the final value of the Hb after surgery (Δ-values), as well as the trending ability. Bias (precision) for the SpHb was 1.63 g/dL (± 0.05) with 95% limits of agreement between 0.85 and + 2.4 g/dL, while for the blood gas analyzer was 0.69 g/dL (± 0.04) with 95% limits of agreement between 0.07 and 1.3 g/dL. Δ-values values were not statistically different from the reference values of ΔHb obtained with the hematological analysis. Trending ability was good for both Pulse CO-Oximeter and blood gas analysis. Our results have shown that the SpHb is not precise enough to replace an invasive approach, but the trending ability of SpHb is accurate and may provide important information on the changes in hemoglobin concentration to guide blood management.
Literature
1.
go back to reference Cecconi M, Corredor C, Arulkumaran N, Abuella G, Ball J, Grounds RM, et al. Clinical review: goal-directed therapy—what is the evidence in surgical patients? The effect on different risk groups. Crit Care. 2013;17:209.PubMedPubMedCentral Cecconi M, Corredor C, Arulkumaran N, Abuella G, Ball J, Grounds RM, et al. Clinical review: goal-directed therapy—what is the evidence in surgical patients? The effect on different risk groups. Crit Care. 2013;17:209.PubMedPubMedCentral
2.
go back to reference Gupta PK, Sundaram A, Mactaggart JN, Johanning JM, Gupta H, Fang X, et al. Preoperative anemia is an independent predictor of postoperative mortality and adverse cardiac events in elderly patients undergoing elective vascular operations. Ann Surg. 2013;258:1096–102.PubMed Gupta PK, Sundaram A, Mactaggart JN, Johanning JM, Gupta H, Fang X, et al. Preoperative anemia is an independent predictor of postoperative mortality and adverse cardiac events in elderly patients undergoing elective vascular operations. Ann Surg. 2013;258:1096–102.PubMed
3.
go back to reference Berkow L. Factors affecting hemoglobin measurement. J Clin Monit Comput. 2013;27:499–508.PubMed Berkow L. Factors affecting hemoglobin measurement. J Clin Monit Comput. 2013;27:499–508.PubMed
4.
go back to reference Bourner G, Dhaliwal J, Sumner J. Performance evaluation of the latest fully automated hematology analyzers in a large, commercial laboratory setting: a 4-way, side-by-side study. Lab Hematol. 2005;11:285–97.PubMed Bourner G, Dhaliwal J, Sumner J. Performance evaluation of the latest fully automated hematology analyzers in a large, commercial laboratory setting: a 4-way, side-by-side study. Lab Hematol. 2005;11:285–97.PubMed
5.
go back to reference Feiner JR, Bickler PE, Mannheimer PD. Accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia. Anesth Analg. 2010;111:143–8.PubMed Feiner JR, Bickler PE, Mannheimer PD. Accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia. Anesth Analg. 2010;111:143–8.PubMed
6.
go back to reference Butwick A, Hilton G, Carvalho B. Non-invasive haemoglobin measurement in patients undergoing elective Caesarean section. Br J Anaesth. 2012;108:271–7.PubMed Butwick A, Hilton G, Carvalho B. Non-invasive haemoglobin measurement in patients undergoing elective Caesarean section. Br J Anaesth. 2012;108:271–7.PubMed
7.
go back to reference Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011;39:2277–82.PubMed Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011;39:2277–82.PubMed
8.
go back to reference Giraud B, Frasca D, Debaene B, Mimoz O. Comparison of haemoglobin measurement methods in the operating theatre. Br J Anaesth. 2013;111:946–54.PubMed Giraud B, Frasca D, Debaene B, Mimoz O. Comparison of haemoglobin measurement methods in the operating theatre. Br J Anaesth. 2013;111:946–54.PubMed
9.
go back to reference Vos JJ, Kalmar AF, Struys MMRF, Porte RJ, Wietasch JKG, Scheeren TWL, et al. Accuracy of non-invasive measurement of haemoglobin concentration by pulse co-oximetry during steady-state and dynamic conditions in liver surgery. Br J Anaesth. 2012;109:522–8.PubMed Vos JJ, Kalmar AF, Struys MMRF, Porte RJ, Wietasch JKG, Scheeren TWL, et al. Accuracy of non-invasive measurement of haemoglobin concentration by pulse co-oximetry during steady-state and dynamic conditions in liver surgery. Br J Anaesth. 2012;109:522–8.PubMed
10.
go back to reference Patino M, Schultz L, Hossain M, Moeller J, Mahmoud M, Gunter J, et al. Trending and accuracy of noninvasive hemoglobin monitoring in pediatric perioperative patients. Anesth Analg. 2014;119:920–5.PubMed Patino M, Schultz L, Hossain M, Moeller J, Mahmoud M, Gunter J, et al. Trending and accuracy of noninvasive hemoglobin monitoring in pediatric perioperative patients. Anesth Analg. 2014;119:920–5.PubMed
11.
go back to reference Erdogan Kayhan G, Colak YZ, Sanli M, Ucar M, Toprak HI. Accuracy of non-invasive hemoglobin monitoring by pulse CO-oximeter during liver transplantation. Minerva Anestesiol. 2017;83:485–92.PubMed Erdogan Kayhan G, Colak YZ, Sanli M, Ucar M, Toprak HI. Accuracy of non-invasive hemoglobin monitoring by pulse CO-oximeter during liver transplantation. Minerva Anestesiol. 2017;83:485–92.PubMed
12.
go back to reference Zeng R, Svensen CH, Li H, Xu X, Skoog Svanberg A, Liu H, et al. Can noninvasive hemoglobin measurement reduce the need for preoperative venipuncture in pediatric outpatient surgery? Paediatr Anaesth. 2017;27:1131–5.PubMed Zeng R, Svensen CH, Li H, Xu X, Skoog Svanberg A, Liu H, et al. Can noninvasive hemoglobin measurement reduce the need for preoperative venipuncture in pediatric outpatient surgery? Paediatr Anaesth. 2017;27:1131–5.PubMed
13.
go back to reference Khalafallah AA, Chilvers CR, Thomas M, Chilvers CM, Sexton M, Vialle M, et al. Usefulness of non-invasive spectrophotometric haemoglobin estimation for detecting low haemoglobin levels when compared with a standard laboratory assay for preoperative assessment. Br J Anaesth. 2015;114:669–76.PubMed Khalafallah AA, Chilvers CR, Thomas M, Chilvers CM, Sexton M, Vialle M, et al. Usefulness of non-invasive spectrophotometric haemoglobin estimation for detecting low haemoglobin levels when compared with a standard laboratory assay for preoperative assessment. Br J Anaesth. 2015;114:669–76.PubMed
14.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.PubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.PubMed
16.
go back to reference Myles PS, Cui J. Using the Bland-Altman method to measure agreement with repeated measures. Br J Anaesth. 2007;99:309–11.PubMed Myles PS, Cui J. Using the Bland-Altman method to measure agreement with repeated measures. Br J Anaesth. 2007;99:309–11.PubMed
17.
go back to reference Morey TE, Gravenstein N, Rice MJ. Let’s think clinically instead of mathematically about device accuracy. Anesth Analg. 2011;113:89–91.PubMed Morey TE, Gravenstein N, Rice MJ. Let’s think clinically instead of mathematically about device accuracy. Anesth Analg. 2011;113:89–91.PubMed
18.
go back to reference Critchley LA, Lee A, Ho AMH. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.PubMed Critchley LA, Lee A, Ho AMH. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.PubMed
19.
go back to reference Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth. 2011;25:536–46.PubMed Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth. 2011;25:536–46.PubMed
20.
go back to reference Skelton VA, Wijayasinghe N, Sharafudeen S, Sange A, Parry NS, Junghans C. Evaluation of point-of-care haemoglobin measuring devices: a comparison of Radical-7™ pulse co-oximetry, HemoCue(®) and laboratory haemoglobin measurements in obstetric patients*. Anaesthesia. 2013;68:40–5.PubMed Skelton VA, Wijayasinghe N, Sharafudeen S, Sange A, Parry NS, Junghans C. Evaluation of point-of-care haemoglobin measuring devices: a comparison of Radical-7™ pulse co-oximetry, HemoCue(®) and laboratory haemoglobin measurements in obstetric patients*. Anaesthesia. 2013;68:40–5.PubMed
21.
go back to reference Miller RD, Ward TA, McCulloch CE, Cohen NH. Does a digital regional nerve block improve the accuracy of noninvasive hemoglobin monitoring? J Anesth. 2012;26:845–50.PubMed Miller RD, Ward TA, McCulloch CE, Cohen NH. Does a digital regional nerve block improve the accuracy of noninvasive hemoglobin monitoring? J Anesth. 2012;26:845–50.PubMed
22.
go back to reference Mannheimer PD. The light-tissue interaction of pulse oximetry. Anesth Analg. 2007;105:S10–7.PubMed Mannheimer PD. The light-tissue interaction of pulse oximetry. Anesth Analg. 2007;105:S10–7.PubMed
23.
go back to reference Severinghaus JW, Astrup PB. History of blood gas analysis. VI. Oximetry. J Clin Monit. 1986;2:270–88.PubMed Severinghaus JW, Astrup PB. History of blood gas analysis. VI. Oximetry. J Clin Monit. 1986;2:270–88.PubMed
24.
go back to reference Shamir MY, Avramovich A, Smaka T. The current status of continuous noninvasive measurement of total, carboxy, and methemoglobin concentration. Anesth Analg. 2012;114:972–8.PubMed Shamir MY, Avramovich A, Smaka T. The current status of continuous noninvasive measurement of total, carboxy, and methemoglobin concentration. Anesth Analg. 2012;114:972–8.PubMed
25.
go back to reference Macknet MR, Allard M, Applegate RL, Rook J. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010;111:1424–6.PubMed Macknet MR, Allard M, Applegate RL, Rook J. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010;111:1424–6.PubMed
26.
go back to reference Berkow L, Rotolo S, Mirski E. Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011;113:1396–402.PubMed Berkow L, Rotolo S, Mirski E. Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011;113:1396–402.PubMed
27.
go back to reference Hiscock R, Kumar D, Simmons SW. Systematic review and meta-analysis of method comparison studies of Masimo pulse co-oximeters (Radical-7™ or Pronto-7™) and HemoCue® absorption spectrometers (B-Hemoglobin or 201+) with laboratory haemoglobin estimation. Anaesth Intensive Care. 2015;43:341–50.PubMed Hiscock R, Kumar D, Simmons SW. Systematic review and meta-analysis of method comparison studies of Masimo pulse co-oximeters (Radical-7™ or Pronto-7™) and HemoCue® absorption spectrometers (B-Hemoglobin or 201+) with laboratory haemoglobin estimation. Anaesth Intensive Care. 2015;43:341–50.PubMed
28.
go back to reference Kim S-H, Lilot M, Murphy LS, Sidhu KS, Yu Z, Rinehart J, et al. Accuracy of continuous noninvasive hemoglobin monitoring: a systematic review and meta-analysis. Anesth Analg. 2014;119:332–46.PubMed Kim S-H, Lilot M, Murphy LS, Sidhu KS, Yu Z, Rinehart J, et al. Accuracy of continuous noninvasive hemoglobin monitoring: a systematic review and meta-analysis. Anesth Analg. 2014;119:332–46.PubMed
29.
go back to reference Yang ZW, Yang SH, Chen L, Qu J, Zhu J, Tang Z. Comparison of blood counts in venous, fingertip and arterial blood and their measurement variation. Clin Lab Haematol. 2001;23:155–9.PubMed Yang ZW, Yang SH, Chen L, Qu J, Zhu J, Tang Z. Comparison of blood counts in venous, fingertip and arterial blood and their measurement variation. Clin Lab Haematol. 2001;23:155–9.PubMed
30.
go back to reference Gehring H, Duembgen L, Peterlein M, Hagelberg S, Dibbelt L. Hemoximetry as the “gold standard”? Error assessment based on differences among identical blood gas analyzer devices of five manufacturers. Anesth Analg. 2007;105:S24–30.PubMed Gehring H, Duembgen L, Peterlein M, Hagelberg S, Dibbelt L. Hemoximetry as the “gold standard”? Error assessment based on differences among identical blood gas analyzer devices of five manufacturers. Anesth Analg. 2007;105:S24–30.PubMed
31.
go back to reference Patel KP, Hay GW, Cheteri MK, Holt DW. Hemoglobin test result variability and cost analysis of eight different analyzers during open heart surgery. J Extra Corpor Technol. 2007;39:10–7.PubMedPubMedCentral Patel KP, Hay GW, Cheteri MK, Holt DW. Hemoglobin test result variability and cost analysis of eight different analyzers during open heart surgery. J Extra Corpor Technol. 2007;39:10–7.PubMedPubMedCentral
32.
go back to reference Barker SJ, Shander A, Ramsay MA. Continuous noninvasive hemoglobin monitoring: a measured response to a critical review. Anesth Analg. 2016;122:565–72.PubMed Barker SJ, Shander A, Ramsay MA. Continuous noninvasive hemoglobin monitoring: a measured response to a critical review. Anesth Analg. 2016;122:565–72.PubMed
33.
go back to reference Adel A, Awada W, Abdelhamid B, Omar H, Abd-El-Dayem O, Hasanin A, et al. Accuracy and trending of non-invasive hemoglobin measurement during different volume and perfusion statuses. J Clin Monit Comput. 2018;32(6):1025–31.PubMed Adel A, Awada W, Abdelhamid B, Omar H, Abd-El-Dayem O, Hasanin A, et al. Accuracy and trending of non-invasive hemoglobin measurement during different volume and perfusion statuses. J Clin Monit Comput. 2018;32(6):1025–31.PubMed
34.
go back to reference Colquhoun DA, Forkin KT, Durieux ME, Thiele RH. Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin. J Clin Monit Comput. 2012;26:69–73.PubMed Colquhoun DA, Forkin KT, Durieux ME, Thiele RH. Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin. J Clin Monit Comput. 2012;26:69–73.PubMed
35.
go back to reference Marques NR, Kramer GC, Voigt RB, Salter MG, Kinsky MP. Trending, accuracy, and precision of noninvasive hemoglobin monitoring during human hemorrhage and fixed crystalloid bolus. Shock. 2015;44(Suppl 1):45–9.PubMed Marques NR, Kramer GC, Voigt RB, Salter MG, Kinsky MP. Trending, accuracy, and precision of noninvasive hemoglobin monitoring during human hemorrhage and fixed crystalloid bolus. Shock. 2015;44(Suppl 1):45–9.PubMed
36.
go back to reference Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011;112:858–63.PubMed Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011;112:858–63.PubMed
37.
go back to reference Applegate RL, Barr SJ, Collier CE, Rook JL, Mangus DB, Allard MW. Evaluation of pulse cooximetry in patients undergoing abdominal or pelvic surgery. Anesthesiology. 2012;116:65–72.PubMed Applegate RL, Barr SJ, Collier CE, Rook JL, Mangus DB, Allard MW. Evaluation of pulse cooximetry in patients undergoing abdominal or pelvic surgery. Anesthesiology. 2012;116:65–72.PubMed
38.
go back to reference Hinkelbein J, Genzwuerker HV, Fiedler F. Detection of a systolic pressure threshold for reliable readings in pulse oximetry. Resuscitation. 2005;64:315–9.PubMed Hinkelbein J, Genzwuerker HV, Fiedler F. Detection of a systolic pressure threshold for reliable readings in pulse oximetry. Resuscitation. 2005;64:315–9.PubMed
39.
go back to reference Hinkelbein J, Genzwuerker HV, Sogl R, Fiedler F. Effect of nail polish on oxygen saturation determined by pulse oximetry in critically ill patients. Resuscitation. 2007;72:82–91.PubMed Hinkelbein J, Genzwuerker HV, Sogl R, Fiedler F. Effect of nail polish on oxygen saturation determined by pulse oximetry in critically ill patients. Resuscitation. 2007;72:82–91.PubMed
40.
go back to reference Hinkelbein J, Koehler H, Genzwuerker HV, Fiedler F. Artificial acrylic finger nails may alter pulse oximetry measurement. Resuscitation. 2007;74:75–82.PubMed Hinkelbein J, Koehler H, Genzwuerker HV, Fiedler F. Artificial acrylic finger nails may alter pulse oximetry measurement. Resuscitation. 2007;74:75–82.PubMed
41.
go back to reference Bergek C, Zdolsek JH, Hahn RG. Accuracy of noninvasive haemoglobin measurement by pulse oximetry depends on the type of infusion fluid. Eur J Anaesthesiol. 2012;29:586–92.PubMed Bergek C, Zdolsek JH, Hahn RG. Accuracy of noninvasive haemoglobin measurement by pulse oximetry depends on the type of infusion fluid. Eur J Anaesthesiol. 2012;29:586–92.PubMed
42.
go back to reference Park Y-H, Lee J-H, Song H-G, Byon H-J, Kim H-S, Kim J-T. The accuracy of noninvasive hemoglobin monitoring using the radical-7 pulse CO-Oximeter in children undergoing neurosurgery. Anesth Analg. 2012;115:1302–7.PubMed Park Y-H, Lee J-H, Song H-G, Byon H-J, Kim H-S, Kim J-T. The accuracy of noninvasive hemoglobin monitoring using the radical-7 pulse CO-Oximeter in children undergoing neurosurgery. Anesth Analg. 2012;115:1302–7.PubMed
43.
go back to reference Awada WN, Mohmoued MF, Radwan TM, Hussien GZ, Elkady HW. Continuous and noninvasive hemoglobin monitoring reduces red blood cell transfusion during neurosurgery: a prospective cohort study. J Clin Monit Comput. 2015;29:733–40.PubMedPubMedCentral Awada WN, Mohmoued MF, Radwan TM, Hussien GZ, Elkady HW. Continuous and noninvasive hemoglobin monitoring reduces red blood cell transfusion during neurosurgery: a prospective cohort study. J Clin Monit Comput. 2015;29:733–40.PubMedPubMedCentral
Metadata
Title
Accuracy and trending ability of hemoglobin measurement by the Pulse CO-Oximeter during vascular surgery
Authors
Rosanna Carmela De Rosa
Giovanni Marco Romano
Roberta Abbate
Antonio Corcione
Edoardo De Robertis
Publication date
01-06-2020
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 3/2020
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-019-00337-5

Other articles of this Issue 3/2020

Journal of Clinical Monitoring and Computing 3/2020 Go to the issue