01-04-2020 | Stroke | Original Research

Comparison of cardiovascular parameter estimation methods using swine data

Authors: Tatsuya Arai, Kichang Lee, Richard J. Cohen

Published in: Journal of Clinical Monitoring and Computing | Issue 2/2020

Login to get access

Abstract

In this study, new and existing methods of estimating stroke volume, cardiac output and total peripheral resistance from analysis of the arterial blood pressure waveform were tested over a wide range of conditions. These pulse contour analysis methods (PCMs) were applied to data obtained in six swine during infusion of volume, phenylephrine, dobutamine, isoproterenol, esmolol and nitroglycerine as well as during progressive hemorrhage. Performance of PCMs was compared using true end-ejection pressures as well as estimated end-ejection pressures. There was considerable overlap in the accuracies of the PCMs when using true end-ejection measures. However, for perhaps the most clinically relevant condition, where radial artery pressure is the input, only Wesseling’s Corrected Impedance method and the Kouchoukos Correction method achieved statistically superior results. We introduced a method of estimating end-ejection by determining when the systolic pressure dropped to a value equal to the sum of the end-diastolic pressure plus a fraction of the pulse pressure. The most accurate estimation of end-ejection was obtained when that fraction was set to 60% for the central arterial pressure and to 50% for the femoral and radial arterial pressures. When the estimated end-ejection measures were used for the PCMs that depend on end-ejection measures and when radial artery pressure was used as the input, only Wesseling’s Corrected Impedance method and the modified Herd’s method achieved statistically superior results. This study provides a systematic comparison of multiple PCMs’ ability to estimate stroke volume, cardiac output, and total peripheral resistance and introduces a new method of estimating end-systole.
Literature
1.
go back to reference Barcroft H, Edholm OG, McMichael J, Sharpey-Schafer EP. Posthaemorrhagic fainting. Study by cardiac output and forearm flow. Lancet. 1944;1:489–91.CrossRef Barcroft H, Edholm OG, McMichael J, Sharpey-Schafer EP. Posthaemorrhagic fainting. Study by cardiac output and forearm flow. Lancet. 1944;1:489–91.CrossRef
2.
go back to reference Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ. A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol. 1971;27:392–6.CrossRef Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ. A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol. 1971;27:392–6.CrossRef
3.
go back to reference Ganz W, Swan HJ. Measurement of blood flow by thermodilution. Am J Cardiol. 1972;29:241–6.CrossRef Ganz W, Swan HJ. Measurement of blood flow by thermodilution. Am J Cardiol. 1972;29:241–6.CrossRef
4.
go back to reference Manecke GR Jr, Brown JC, Landau AA, Kapelanski DP, St Laurent CM, Auger WR. An unusual case of pulmonary artery catheter malfunction. Anesth Analg. 2002;95:302–4.PubMed Manecke GR Jr, Brown JC, Landau AA, Kapelanski DP, St Laurent CM, Auger WR. An unusual case of pulmonary artery catheter malfunction. Anesth Analg. 2002;95:302–4.PubMed
5.
go back to reference Vender JS, Gilbert HC. Monitoring the anesthetized patient. 3rd ed. Philadelphia: Lippincott-Raven Publishers; 1997. Vender JS, Gilbert HC. Monitoring the anesthetized patient. 3rd ed. Philadelphia: Lippincott-Raven Publishers; 1997.
6.
go back to reference Monnet X, Teboul JL. Transpulmonary thermodilution: advantages and limits. Crit Care. 2017;21:147.CrossRef Monnet X, Teboul JL. Transpulmonary thermodilution: advantages and limits. Crit Care. 2017;21:147.CrossRef
7.
go back to reference Huter L, Schwarzkopf KR, Schubert NP, Schreiber T. The level of cardiac output affects the relationship and agreement between pulmonary artery and transpulmonary aortic thermodilution measurements in an animal model. J Cardiotheorac Vasc Anesth. 2007;21:659–63.CrossRef Huter L, Schwarzkopf KR, Schubert NP, Schreiber T. The level of cardiac output affects the relationship and agreement between pulmonary artery and transpulmonary aortic thermodilution measurements in an animal model. J Cardiotheorac Vasc Anesth. 2007;21:659–63.CrossRef
8.
go back to reference Staier K, Wilhelm M, Wiesenack C, Thoma M, Keyl C. Pulmonary artery vs. transpulmonary thermodilution for the assessment of cardiac output in mitral regurgitation: a prospective observational study. Eur J Anaesthesiol. 2012;29:431–7.CrossRef Staier K, Wilhelm M, Wiesenack C, Thoma M, Keyl C. Pulmonary artery vs. transpulmonary thermodilution for the assessment of cardiac output in mitral regurgitation: a prospective observational study. Eur J Anaesthesiol. 2012;29:431–7.CrossRef
9.
go back to reference Mihm FG, Gettinger A, Hanson CW 3rd, Gilbert HC, Stover EP, Vender JS, Beerle B, Haddow G. A multicenter evaluation of a new continuous cardiac output pulmonary artery catheter system. Crit Care Med. 1998;26:1346–50.CrossRef Mihm FG, Gettinger A, Hanson CW 3rd, Gilbert HC, Stover EP, Vender JS, Beerle B, Haddow G. A multicenter evaluation of a new continuous cardiac output pulmonary artery catheter system. Crit Care Med. 1998;26:1346–50.CrossRef
10.
go back to reference Botero M, Kirby D, Lobato EB, Staples ED, Gravenstein N. Measurement of cardiac output before and after cardiopulmonary bypass: comparison among aortic transit-time ultrasound, thermodilution, and noninvasive partial CO2 rebreathing. J Cardiothorac Vasc Anesth. 2004;18:563–72.CrossRef Botero M, Kirby D, Lobato EB, Staples ED, Gravenstein N. Measurement of cardiac output before and after cardiopulmonary bypass: comparison among aortic transit-time ultrasound, thermodilution, and noninvasive partial CO2 rebreathing. J Cardiothorac Vasc Anesth. 2004;18:563–72.CrossRef
11.
go back to reference Fares WH, Blanchard SK, Stouffer GA, Chang PP, Rosamond WD, Ford HJ, Aris RM. Thermodilution and Fick cardiac outputs differ: impact on pulmonary hypertension evaluation. Can Respir J. 2012;19:261–6.CrossRef Fares WH, Blanchard SK, Stouffer GA, Chang PP, Rosamond WD, Ford HJ, Aris RM. Thermodilution and Fick cardiac outputs differ: impact on pulmonary hypertension evaluation. Can Respir J. 2012;19:261–6.CrossRef
12.
go back to reference Blohm ME, Obrecht D, Hartwich J, Mueller GC, Kersten JF, Weil J, Singer D. Impedance cardiography (electrical velocimetry) and transthoracic echocardiography for non-invasive cardiac output monitoring in pediatric intensive care patients: a prospective single-center observational study. Crit Care. 2014;18:603.CrossRef Blohm ME, Obrecht D, Hartwich J, Mueller GC, Kersten JF, Weil J, Singer D. Impedance cardiography (electrical velocimetry) and transthoracic echocardiography for non-invasive cardiac output monitoring in pediatric intensive care patients: a prospective single-center observational study. Crit Care. 2014;18:603.CrossRef
13.
go back to reference Schubert S, Schmitz T, Weiss M, Nagdyman N, Huebler M, Alexi-Meskishvili V, Berger F, Stiller B. Continuous, non-invasive techniques to determine cardiac output in children after cardiac surgery: evaluation of transesophageal Doppler and electric velocimetry. J Clin Monit Comput. 2008;22:299–307.CrossRef Schubert S, Schmitz T, Weiss M, Nagdyman N, Huebler M, Alexi-Meskishvili V, Berger F, Stiller B. Continuous, non-invasive techniques to determine cardiac output in children after cardiac surgery: evaluation of transesophageal Doppler and electric velocimetry. J Clin Monit Comput. 2008;22:299–307.CrossRef
14.
go back to reference Scolletta S, Franchi F, Romagnoli S, Carla R, Donati A, Fabbri LP, Forfori F, Alonso-Inigo JM, Laviola S, Mangani V, Maj G, Martinelli G, Mirabella L, Morelli A, Persona P, Payen D, PulseCOval Group. Comparison between doppler-echocardiography and uncalibrated pulse contour method for cardiac output measurement: a multicenter observational study. Crit Care Med. 2016;44:1370–9.CrossRef Scolletta S, Franchi F, Romagnoli S, Carla R, Donati A, Fabbri LP, Forfori F, Alonso-Inigo JM, Laviola S, Mangani V, Maj G, Martinelli G, Mirabella L, Morelli A, Persona P, Payen D, PulseCOval Group. Comparison between doppler-echocardiography and uncalibrated pulse contour method for cardiac output measurement: a multicenter observational study. Crit Care Med. 2016;44:1370–9.CrossRef
15.
go back to reference Patterson RP. Fundamentals of impedance cardiography. IEEE Eng Med Biol Mag. 1989;8:35–8.CrossRef Patterson RP. Fundamentals of impedance cardiography. IEEE Eng Med Biol Mag. 1989;8:35–8.CrossRef
16.
go back to reference Erlanger J, Hooker DR. An experimental study of blood-pressure and of pulse-pressure in man. Johns Hopkins Hosp Rep. 1904;12:145–378. Erlanger J, Hooker DR. An experimental study of blood-pressure and of pulse-pressure in man. Johns Hopkins Hosp Rep. 1904;12:145–378.
17.
go back to reference Herd JA, Leclair NR, Simon W. Arterial pressure pulse contours during hemorrhage in anesthetized dogs. J Appl Physiol. 1966;21:1864–8.CrossRef Herd JA, Leclair NR, Simon W. Arterial pressure pulse contours during hemorrhage in anesthetized dogs. J Appl Physiol. 1966;21:1864–8.CrossRef
18.
go back to reference Jones WB, Hefner LL, Bancroft WH Jr, Klip W. Velocity of blood flow and stroke volume obtained from the pressure pulse. J Clin Investig. 1959;38:2087–90.CrossRef Jones WB, Hefner LL, Bancroft WH Jr, Klip W. Velocity of blood flow and stroke volume obtained from the pressure pulse. J Clin Investig. 1959;38:2087–90.CrossRef
19.
go back to reference Kouchoukos NT, Sheppard LC, McDonald DA. Estimation of stroke volume in the dog by a pulse contour method. Circ Res. 1970;26:611–23.CrossRef Kouchoukos NT, Sheppard LC, McDonald DA. Estimation of stroke volume in the dog by a pulse contour method. Circ Res. 1970;26:611–23.CrossRef
20.
go back to reference Liljestrand G, Zander E. Vergleichende Bestimmung des Minutenvolumens des Herzens beim Menschen mittels der Stickoxydulmethode und durch Blutdruckmessung. Zeitschrift fur die gesamte experimentelle Medizin. 1928;59:105–22.CrossRef Liljestrand G, Zander E. Vergleichende Bestimmung des Minutenvolumens des Herzens beim Menschen mittels der Stickoxydulmethode und durch Blutdruckmessung. Zeitschrift fur die gesamte experimentelle Medizin. 1928;59:105–22.CrossRef
21.
go back to reference Mukkamala R, Reisner AT, Hojman HM, Mark RG, Cohen RJ. Continuous cardiac output monitoring by peripheral blood pressure waveform analysis. IEEE Trans Biomed Eng. 2006;53:459–67.CrossRef Mukkamala R, Reisner AT, Hojman HM, Mark RG, Cohen RJ. Continuous cardiac output monitoring by peripheral blood pressure waveform analysis. IEEE Trans Biomed Eng. 2006;53:459–67.CrossRef
22.
go back to reference Parlikar T, Heldt T, Ranade GV, Verghese G. Model-based estimation of cardiac output and total peripheral resistance. Comput Cardiol. 2007;34:379–82. Parlikar T, Heldt T, Ranade GV, Verghese G. Model-based estimation of cardiac output and total peripheral resistance. Comput Cardiol. 2007;34:379–82.
23.
go back to reference Sun JX, Reisner AT, Saeed M, Heldt T, Mark RG. The cardiac output from blood pressure algorithms trial. Crit Care Med. 2009;37:72–80.CrossRef Sun JX, Reisner AT, Saeed M, Heldt T, Mark RG. The cardiac output from blood pressure algorithms trial. Crit Care Med. 2009;37:72–80.CrossRef
24.
go back to reference Verdouw PD, Beaune J, Roelandt J, Hugenholtz PG. Stroke volume from central aortic pressure? A critical assessment of the various formulae as to their clinical value. Basic Res Cardiol. 1975;70:377–89.CrossRef Verdouw PD, Beaune J, Roelandt J, Hugenholtz PG. Stroke volume from central aortic pressure? A critical assessment of the various formulae as to their clinical value. Basic Res Cardiol. 1975;70:377–89.CrossRef
25.
go back to reference Wesseling KH, De Werr B, Weber JAP, Smith NT. A simple device for the continuous measurement of cardiac output. Its model basis and experimental verification. Adv Cardiovasc Phys. 1983;5:16–52. Wesseling KH, De Werr B, Weber JAP, Smith NT. A simple device for the continuous measurement of cardiac output. Its model basis and experimental verification. Adv Cardiovasc Phys. 1983;5:16–52.
26.
go back to reference Arai T, Lee K, Marini RP, Cohen RJ. Estimation of changes in instantaneous aortic blood flow by the analysis of arterial blood pressure. J Appl Physiol. 2012;112:1832–8.CrossRef Arai T, Lee K, Marini RP, Cohen RJ. Estimation of changes in instantaneous aortic blood flow by the analysis of arterial blood pressure. J Appl Physiol. 2012;112:1832–8.CrossRef
27.
go back to reference Parlikar T. Modeling and monitoring of cardiovascular dynamics for patients in critical care. Cambridge: Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology; 2007. Parlikar T. Modeling and monitoring of cardiovascular dynamics for patients in critical care. Cambridge: Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology; 2007.
28.
go back to reference Bourgeois MJ, Gilbert BK, Von Bernuth G, Wood EH. Continuous determination of beat to beat stroke volume from aortic pulse pressures in the dog. Circ Res. 1976;39:15–24.CrossRef Bourgeois MJ, Gilbert BK, Von Bernuth G, Wood EH. Continuous determination of beat to beat stroke volume from aortic pulse pressures in the dog. Circ Res. 1976;39:15–24.CrossRef
Metadata
Title
Comparison of cardiovascular parameter estimation methods using swine data
Authors
Tatsuya Arai
Kichang Lee
Richard J. Cohen
Publication date
01-04-2020
Publisher
Springer Netherlands
Keyword
Stroke
Published in
Journal of Clinical Monitoring and Computing / Issue 2/2020
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-019-00322-y

Other articles of this Issue 2/2020

Journal of Clinical Monitoring and Computing 2/2020 Go to the issue