Determination of cardiac output from pulse pressure contour during intra-aortic balloon pumping in patients with low ejection fraction

Authors: Jos R. C. Jansen, Marcelo B Bastos, Pat Hanlon, Nicolas M. Van Mieghem, Ottavio Alfieri, Jan J. Schreuder

Published in: Journal of Clinical Monitoring and Computing | Issue 2/2020

Login to get access

Abstract

Evaluation of a new Windkessel model based pulse contour method (WKflow) to calculate stroke volume in patients undergoing intra-aortic balloon pumping (IABP). Preload changes were induced by vena cava occlusions (VCO) in twelve patients undergoing cardiac surgery to vary stroke volume (SV), which was measured by left ventricular conductance volume method (SVlv) and WKflow (SVwf). Twelve VCO series were carried out during IABP assist at a 1:2 ratio and seven VCO series were performed with IABP switched off. Additionally, SVwf was evaluated during nine episodes of severe arrhythmia. VCO’s produced marked changes in SV over 10–20 beats. 198 paired data sets of SVlv and SVwf were obtained. Bland–Altman analysis for the difference between SVlv and SVwf during IABP in 1:2 mode showed a bias (accuracy) of 1.04 ± 3.99 ml, precision 10.9% and limits of agreement (LOA) of − 6.94 to 9.02 ml. Without IABP bias was 0.48 ± 4.36 ml, precision 11.6% and LOA of − 8.24 to 9.20 ml. After one thermodilution calibration of SVwf per patient, during IABP the accuracy improved to 0.14 ± 3.07 ml, precision to 8.3% and LOA to − 6.00 to + 6.28 ml. Without IABP the accuracy improved to 0.01 ± 2.71 ml, precision to 7.5% and LOA to − 5.41 to + 5.43 ml. Changes in SVlv and SVwf were directionally concordant in response to VCO’s and during severe arrhythmia. (R2 = 0.868). The SVwf and SVlv methods are interchangeable with respect to measuring absolute stroke volume as well as tracking changes in stroke volume. The precision of the non-calibrated WKflow method is about 10% which improved to 7.5% after one calibration per patient.
Literature
1.
go back to reference Kantrowitz A. Experimental augmentation of coronary flow by retardation of arterial pressure pulse. Surgery. 1953;34:678.PubMed Kantrowitz A. Experimental augmentation of coronary flow by retardation of arterial pressure pulse. Surgery. 1953;34:678.PubMed
2.
go back to reference Kantrowitz A, Cardona RR, Freed PS. Percutaneous intraaortic balloon counterpulsation. Crit Care Clin. 1992;8:819–37.CrossRef Kantrowitz A, Cardona RR, Freed PS. Percutaneous intraaortic balloon counterpulsation. Crit Care Clin. 1992;8:819–37.CrossRef
3.
go back to reference Dunkman WB, Leinbach RC, Buckley MJ, et al. Clinical and hemodynamic results of intraaortic balloon pumping and surgery for cardiogenic shock. Circulation. 1972;46:465–77.CrossRef Dunkman WB, Leinbach RC, Buckley MJ, et al. Clinical and hemodynamic results of intraaortic balloon pumping and surgery for cardiogenic shock. Circulation. 1972;46:465–77.CrossRef
4.
go back to reference Nichols AB, Pohost GM, Gold HK, et al. Left ventricular function during intra-aortic balloon pumping assessed by multigated cardiac blood pool imaging. Circulation. 1978;58:176–83. Nichols AB, Pohost GM, Gold HK, et al. Left ventricular function during intra-aortic balloon pumping assessed by multigated cardiac blood pool imaging. Circulation. 1978;58:176–83.
5.
go back to reference Cheung AT, Savino JS, Weiss SJ. Beat-to-beat augmentation of left ventricular function by intraaortic counterpulsation. Anesthesiology. 1996;84:545–54.CrossRef Cheung AT, Savino JS, Weiss SJ. Beat-to-beat augmentation of left ventricular function by intraaortic counterpulsation. Anesthesiology. 1996;84:545–54.CrossRef
6.
go back to reference Kawaguchi O, Pae WE, Daily BB, et al. Ventriculoarterial coupling with intra-aortic balloon pump in ischemic heart failure. J Thorac Cardiovasc Surg. 1999;117:164–71.CrossRef Kawaguchi O, Pae WE, Daily BB, et al. Ventriculoarterial coupling with intra-aortic balloon pump in ischemic heart failure. J Thorac Cardiovasc Surg. 1999;117:164–71.CrossRef
7.
go back to reference Kern MJ, Aguirre F, Tatineni S, et al. Enhanced coronary blood flow velocity during intra-aortic balloon counterpulsation in critically ill patients. J Am Coll Cardiol. 1993;21:359–68.CrossRef Kern MJ, Aguirre F, Tatineni S, et al. Enhanced coronary blood flow velocity during intra-aortic balloon counterpulsation in critically ill patients. J Am Coll Cardiol. 1993;21:359–68.CrossRef
8.
go back to reference Schreuder JJ, Maisano F, Donelli A, et al. Beat-to-beat effects of intraaortic balloon pump timing on left ventricular performance in patients with low ejection fraction. Ann Thorac Surg. 2005;79:872–80.CrossRef Schreuder JJ, Maisano F, Donelli A, et al. Beat-to-beat effects of intraaortic balloon pump timing on left ventricular performance in patients with low ejection fraction. Ann Thorac Surg. 2005;79:872–80.CrossRef
9.
go back to reference Schreuder JJ, Castiglioni A, Donelli A, et al. Automatic intraaortic balloon pump timing using an intrabeat dicrotic notch prediction algorithm. Ann Thorac Surg. 2005;79:1017–22.CrossRef Schreuder JJ, Castiglioni A, Donelli A, et al. Automatic intraaortic balloon pump timing using an intrabeat dicrotic notch prediction algorithm. Ann Thorac Surg. 2005;79:1017–22.CrossRef
10.
go back to reference Pantalos GM, Koenig SC, Gillars KJ, et al. Intraaortic balloon pump timing discrepancies in adult patients. Artif Org. 2011;35:857–66.CrossRef Pantalos GM, Koenig SC, Gillars KJ, et al. Intraaortic balloon pump timing discrepancies in adult patients. Artif Org. 2011;35:857–66.CrossRef
11.
go back to reference Høie J. Determination of cardiac output from pulse pressure contour during intra-aortic balloon pumping. Scand J Clin Lab Investig. 1980;40:445–9.CrossRef Høie J. Determination of cardiac output from pulse pressure contour during intra-aortic balloon pumping. Scand J Clin Lab Investig. 1980;40:445–9.CrossRef
12.
go back to reference Herzlinger GA. Absolute determination of cardiac output in intra-aortic balloon pumped patients using the radial arterial pressure trace. Circulation. 1976;53:417–21.CrossRef Herzlinger GA. Absolute determination of cardiac output in intra-aortic balloon pumped patients using the radial arterial pressure trace. Circulation. 1976;53:417–21.CrossRef
13.
go back to reference Lorsomradee S, Lorsomradee S, Cromheecke S, De Hert SG. Uncalibrated arterial pulse contour analysis versus continuous thermodilution technique: effects of alterations in arterial waveform. J Cardiothorac Vasc Anesth. 2007;21:636–43.CrossRef Lorsomradee S, Lorsomradee S, Cromheecke S, De Hert SG. Uncalibrated arterial pulse contour analysis versus continuous thermodilution technique: effects of alterations in arterial waveform. J Cardiothorac Vasc Anesth. 2007;21:636–43.CrossRef
14.
go back to reference Schreuder JJ, van der Veen FH, van der Velde ET, et al. Beat-to-beat analysis of left ventricular pressure-volume relation and stroke volume by conductance catheter and aortic modelflow in cardiomyoplasty patients. Circulation. 1995;91:2010–7.CrossRef Schreuder JJ, van der Veen FH, van der Velde ET, et al. Beat-to-beat analysis of left ventricular pressure-volume relation and stroke volume by conductance catheter and aortic modelflow in cardiomyoplasty patients. Circulation. 1995;91:2010–7.CrossRef
15.
go back to reference Wesseling KH, Jansen JRC, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol. 1993;74:2566–73.CrossRef Wesseling KH, Jansen JRC, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol. 1993;74:2566–73.CrossRef
16.
go back to reference Langewouters GJ, Wesseling KH, Groedhard WJA. The static elastic properties of 45 human thoracic and 20 abdominal aortas in vitro and the parameters of a new model. J Biomech. 1984;17:425–35.CrossRef Langewouters GJ, Wesseling KH, Groedhard WJA. The static elastic properties of 45 human thoracic and 20 abdominal aortas in vitro and the parameters of a new model. J Biomech. 1984;17:425–35.CrossRef
17.
go back to reference Hoeksel SA, Jansen JR, Blom JA, Schreuder JJ. Detection of dicrotic notch in arterial pressure signals. J Clin Monit. 1997;13:309–16.CrossRef Hoeksel SA, Jansen JR, Blom JA, Schreuder JJ. Detection of dicrotic notch in arterial pressure signals. J Clin Monit. 1997;13:309–16.CrossRef
18.
go back to reference Donelli A, Jansen JR, Hoeksel B, et al. Performance of a real-time dicrotic notch detection and prediction algorithm in arrhythmic human aortic pressure signals. J Clin Monit Comput. 2002;17:181–5.CrossRef Donelli A, Jansen JR, Hoeksel B, et al. Performance of a real-time dicrotic notch detection and prediction algorithm in arrhythmic human aortic pressure signals. J Clin Monit Comput. 2002;17:181–5.CrossRef
19.
go back to reference Jansen JR, Schreuder JJ, Settels JJ, et al. An adequate strategy for the thermodilution technique in patients during mechanical ventilation. Intensive Care Med. 1990;16:422–5.CrossRef Jansen JR, Schreuder JJ, Settels JJ, et al. An adequate strategy for the thermodilution technique in patients during mechanical ventilation. Intensive Care Med. 1990;16:422–5.CrossRef
20.
go back to reference Jansen JR, Schreuder JJ, Mulier JP, et al. A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients. Br J Anaesth. 2001;87:212–22.CrossRef Jansen JR, Schreuder JJ, Mulier JP, et al. A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients. Br J Anaesth. 2001;87:212–22.CrossRef
21.
go back to reference Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15:85–91.CrossRef Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15:85–91.CrossRef
22.
go back to reference Kornet L, Jansen JR, Gussenhoven EJ, et al. Conductance method for the measurement of cross-sectional areas of the aorta. Ann Biomed Eng. 1999;27:141–50.CrossRef Kornet L, Jansen JR, Gussenhoven EJ, et al. Conductance method for the measurement of cross-sectional areas of the aorta. Ann Biomed Eng. 1999;27:141–50.CrossRef
23.
go back to reference Pauca AL, Wallenhaupt SL, Kon ND. Reliability of the radial arterial pressure during anesthesia. Is wrist compression a possible diagnostic test? Chest. 1994;105:69–75.CrossRef Pauca AL, Wallenhaupt SL, Kon ND. Reliability of the radial arterial pressure during anesthesia. Is wrist compression a possible diagnostic test? Chest. 1994;105:69–75.CrossRef
24.
go back to reference Manecke GR Jr, Parimucha M, Stratmann G, et al. Deep hypothermic circulatory arrest and the femoral-to-radial arterial pressure gradient. J Cardiothorac Vasc Anesth. 2004;18:175–9.CrossRef Manecke GR Jr, Parimucha M, Stratmann G, et al. Deep hypothermic circulatory arrest and the femoral-to-radial arterial pressure gradient. J Cardiothorac Vasc Anesth. 2004;18:175–9.CrossRef
25.
go back to reference Hatib F, Jansen JR, Pinsky MR. Peripheral vascular decoupling in porcine endotoxic shock. J Appl Physiol. 2011;111:853–60.CrossRef Hatib F, Jansen JR, Pinsky MR. Peripheral vascular decoupling in porcine endotoxic shock. J Appl Physiol. 2011;111:853–60.CrossRef
26.
go back to reference Schreuder JJ, Castiglioni A, Maisano F, et al. Acute decrease of left ventricular mechanical dyssynchrony and improvement of contractile state and energy efficiency after left ventricular restoration. J Thorac Cardiovasc Surg. 2005;129:138–45.CrossRef Schreuder JJ, Castiglioni A, Maisano F, et al. Acute decrease of left ventricular mechanical dyssynchrony and improvement of contractile state and energy efficiency after left ventricular restoration. J Thorac Cardiovasc Surg. 2005;129:138–45.CrossRef
Metadata
Title
Determination of cardiac output from pulse pressure contour during intra-aortic balloon pumping in patients with low ejection fraction
Authors
Jos R. C. Jansen
Marcelo B Bastos
Pat Hanlon
Nicolas M. Van Mieghem
Ottavio Alfieri
Jan J. Schreuder
Publication date
01-04-2020
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 2/2020
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-019-00320-0

Other articles of this Issue 2/2020

Journal of Clinical Monitoring and Computing 2/2020 Go to the issue