Published in: Journal of Clinical Monitoring and Computing 5/2016

01-10-2016 | Original Research

The effect of variable arterial transducer level on the accuracy of pulse contour waveform-derived measurements in critically ill patients

Authors: Huai-wu He, Da-wei Liu, Yun Long, Xiao-ting Wang, Mei-ling Zhao, Xiao-li Lai

Published in: Journal of Clinical Monitoring and Computing | Issue 5/2016

Login to get access

Abstract

We know that a 10 cm departure from the reference level of pressure transducer position is equal to a 7.5 mmHg change of invasive hemodynamic pressure monitoring in a fluid-filled system. However, the relationship between the site level of a variable arterial pressure transducer and the pulse contour-derived parameters has yet to be established in critically ill patients. Moreover, the related quantitative analysis has never been investigated. Forty-two critically ill patients requiring PiCCO-Plus cardiac output monitoring were prospectively studied. The phlebostatic axis was defined as the zero reference level; the arterial pressure transducer was then vertically adjusted to different positions (+5, +10, +15, +20, −20, −15, −10, −5 cm) of departure from the zero reference site. The pulse contour waveform-derived parameters were recorded at each position. Elevation of the pressure transducer caused significantly positive changes in the continuous cardiac index (+CCI), stroke volume index (+SVI), and stroke volume variation (+SVV), and negative changes in the rate of left ventricular pressure rise during systole (−dP/dtmax), the systemic vascular resistance index (−SVRI), and vice versa. At the 5 cm position, the SVRI changes reached statistical significance with error. At the 10 cm position, the changes in CCI and dP/dtmax reached statistical significance with error, while the change in SVV reached statistical significance at 15 cm. The change rate of CCI was more than 5 % at the 15 cm position and approximately 10 % at the 20 cm position. On average, for every centimeter change of the transducer, there was a corresponding 0.014 L/min/m2 CCI change and 0.36 % change rate, a 1.41 mmHg/s dP/dtmax change and 0.13 % change rate, and a 25 dyne/s/cm5 SVRI change and 1.2 % change rate. The variation of arterial transducer position can result in inaccurate measurement of pulse contour waveform-derived parameters, especially when the transducer’s vertical distance is more than 10 cm from the phlebostatic axis. These findings have clinical implications for continuous hemodynamic monitoring.
Appendix
Available only for authorised users
Literature
1.
go back to reference Rauen CA, Makic MBF, Bridges E. Practice evidence-based practice habits: transforming research into bedside. Crit Care Nurse. 2009;29:46–59.CrossRefPubMed Rauen CA, Makic MBF, Bridges E. Practice evidence-based practice habits: transforming research into bedside. Crit Care Nurse. 2009;29:46–59.CrossRefPubMed
2.
go back to reference Winsor T, Burch G. Phlebostatic level: reference level for venous pressure measurement in man. Proc Soc Exp Biol Med. 1945;58:165–9.CrossRef Winsor T, Burch G. Phlebostatic level: reference level for venous pressure measurement in man. Proc Soc Exp Biol Med. 1945;58:165–9.CrossRef
3.
go back to reference McCann UG II, Schiller HJ, Carney DE, Kilpatrick J, Gatto LA, Paskanik AM, Nieman GF. Invasive arterial BP monitoring in trauma and critical care. Chest. 2001;120:1322–6.CrossRefPubMed McCann UG II, Schiller HJ, Carney DE, Kilpatrick J, Gatto LA, Paskanik AM, Nieman GF. Invasive arterial BP monitoring in trauma and critical care. Chest. 2001;120:1322–6.CrossRefPubMed
4.
go back to reference Tagami T, Kushimoto S, Tosa R, et al. The precision of PiCCO® measurements in hypothermic post-cardiac arrest patients. Anaesthesia. 2012;67:236–43.CrossRefPubMed Tagami T, Kushimoto S, Tosa R, et al. The precision of PiCCO® measurements in hypothermic post-cardiac arrest patients. Anaesthesia. 2012;67:236–43.CrossRefPubMed
5.
go back to reference de Wilde RB, Schreuder JJ, van den Berg PC, Jansen JR. An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery. Anaesthesia. 2007;62:760–8.CrossRefPubMed de Wilde RB, Schreuder JJ, van den Berg PC, Jansen JR. An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery. Anaesthesia. 2007;62:760–8.CrossRefPubMed
6.
go back to reference Perner A, Haase N, Wiis J, White JO, Delaney A. Central venous oxygen saturation for the diagnosis of low cardiac output in septic shock patients. Acta Anaesthesiol Scand. 2010;54:98–102.CrossRefPubMed Perner A, Haase N, Wiis J, White JO, Delaney A. Central venous oxygen saturation for the diagnosis of low cardiac output in septic shock patients. Acta Anaesthesiol Scand. 2010;54:98–102.CrossRefPubMed
7.
go back to reference Scolletta Sabino, Bodson Laurent, Donadello Katia, Taccone Fabio S, Devigili Alessandro, Vincent Jean-Louis, De Backer Daniel. Assessment of left ventricular function by pulse wave analysis in critically ill patients. Intensive Care Med. 2013;39:1025–33.CrossRefPubMed Scolletta Sabino, Bodson Laurent, Donadello Katia, Taccone Fabio S, Devigili Alessandro, Vincent Jean-Louis, De Backer Daniel. Assessment of left ventricular function by pulse wave analysis in critically ill patients. Intensive Care Med. 2013;39:1025–33.CrossRefPubMed
8.
go back to reference He H, Liu D, Long Y, Wang X, Yuan Y, Li X, Guo H, Cai J, Fang N. The relationship between arterial transducer level and pulse contour waveform-derived measurements. Crit Care. 2015;19:31.CrossRefPubMedPubMedCentral He H, Liu D, Long Y, Wang X, Yuan Y, Li X, Guo H, Cai J, Fang N. The relationship between arterial transducer level and pulse contour waveform-derived measurements. Crit Care. 2015;19:31.CrossRefPubMedPubMedCentral
9.
go back to reference Knaus WA, Draper EA, Wagner DP, Zimmerman JE. APACHE II: a severity of disease classification system. Crit Care Med. 1985;13:818–8295.CrossRefPubMed Knaus WA, Draper EA, Wagner DP, Zimmerman JE. APACHE II: a severity of disease classification system. Crit Care Med. 1985;13:818–8295.CrossRefPubMed
10.
go back to reference Kleinman B, Powell S, Kumar P, Gardner RM. The fast flush test measures the dynamic response of the entire blood pressure monitoring system. Anesthesiology. 1992;77:1215–20.CrossRefPubMed Kleinman B, Powell S, Kumar P, Gardner RM. The fast flush test measures the dynamic response of the entire blood pressure monitoring system. Anesthesiology. 1992;77:1215–20.CrossRefPubMed
11.
go back to reference Watanabe H, Yagi S, Namiki A. Recommendation of a clinical impulse response analysis for catheter calibration-dumping coefficient and natural frequency are incomplete parameters for clinical evaluation. J Clin Monit Comput. 2006;20:37–42.CrossRefPubMed Watanabe H, Yagi S, Namiki A. Recommendation of a clinical impulse response analysis for catheter calibration-dumping coefficient and natural frequency are incomplete parameters for clinical evaluation. J Clin Monit Comput. 2006;20:37–42.CrossRefPubMed
12.
go back to reference Roy A. Estimating correlation coefficient between two variables with repeated observations using mixed effects model. Biom J. 2006;48:286–301.CrossRefPubMed Roy A. Estimating correlation coefficient between two variables with repeated observations using mixed effects model. Biom J. 2006;48:286–301.CrossRefPubMed
13.
go back to reference McCulloch CE, Searle SR. Generalized, linear, and mixed models. London: Wiley; 2000.CrossRef McCulloch CE, Searle SR. Generalized, linear, and mixed models. London: Wiley; 2000.CrossRef
14.
go back to reference Wesseling KH, de Wit B, Weber JAP, Smith NT. A simple device for the continuous measurement of cardiac output. Adv Cardiovasc Phys. 1983;5:16–52. Wesseling KH, de Wit B, Weber JAP, Smith NT. A simple device for the continuous measurement of cardiac output. Adv Cardiovasc Phys. 1983;5:16–52.
15.
go back to reference Wesseling KH, Jansen JRC, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear three-element model. J Appl Physiol. 1984;74(5):2566–73. Wesseling KH, Jansen JRC, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear three-element model. J Appl Physiol. 1984;74(5):2566–73.
16.
go back to reference Wesseling, KH, Smith, NT, Nichols, WW et al. Beat to beat cardiac output from the arterial pressure pulse contour. In: Boerhave, SJ, Course on measurement in anaesthesia. Leiden: University of Leiden Press; 1974. pp. 150–64. Wesseling, KH, Smith, NT, Nichols, WW et al. Beat to beat cardiac output from the arterial pressure pulse contour. In: Boerhave, SJ, Course on measurement in anaesthesia. Leiden: University of Leiden Press; 1974. pp. 150–64.
17.
go back to reference Godje O, Hoke K, Goetz AE, Felbinger TW, Reuter DA, Reichart B, Friedl R, Hannekum A, Pfeiffer UJ. Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during hemodynamic instability. Crit Care Med. 2002;30:52–8.CrossRefPubMed Godje O, Hoke K, Goetz AE, Felbinger TW, Reuter DA, Reichart B, Friedl R, Hannekum A, Pfeiffer UJ. Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during hemodynamic instability. Crit Care Med. 2002;30:52–8.CrossRefPubMed
18.
go back to reference Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL. Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care. 2010;14(3):R109.CrossRefPubMedPubMedCentral Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL. Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care. 2010;14(3):R109.CrossRefPubMedPubMedCentral
19.
go back to reference Parry T, Hirsch N, Fauvel N. Comparison of direct blood pressure measurement at the radial and dorsalis pedis arteries during surgery in the horizontal and reverse Trendelenburg positions. Anaesthesia. 1995;50:553–5.CrossRefPubMed Parry T, Hirsch N, Fauvel N. Comparison of direct blood pressure measurement at the radial and dorsalis pedis arteries during surgery in the horizontal and reverse Trendelenburg positions. Anaesthesia. 1995;50:553–5.CrossRefPubMed
20.
go back to reference Chulay M, Holland S. Where should the transducer be leveled for radial or femoral arterial pressure monitoring? Crit Care Nurs. 1996;16:103–7. Chulay M, Holland S. Where should the transducer be leveled for radial or femoral arterial pressure monitoring? Crit Care Nurs. 1996;16:103–7.
21.
go back to reference Ahrens TS, Taylor LA. Technical considerations in obtaining hemodynamic waveform values. In: Darovic GO, editor. Hemodynamic waveform analysis. Philadelphia: W.B. Saunders press; 1992. p. 222. Ahrens TS, Taylor LA. Technical considerations in obtaining hemodynamic waveform values. In: Darovic GO, editor. Hemodynamic waveform analysis. Philadelphia: W.B. Saunders press; 1992. p. 222.
22.
go back to reference Courtois M, Fattal PG, Kovacs SJ, et al. Anatomically and physiologically based reference level for measurement of intracardiac pressures. Circulation. 1995;92:1994–2000.CrossRefPubMed Courtois M, Fattal PG, Kovacs SJ, et al. Anatomically and physiologically based reference level for measurement of intracardiac pressures. Circulation. 1995;92:1994–2000.CrossRefPubMed
23.
go back to reference Whitman G, Howaniak D, Verga T. Comparison of cardiac output measurements in 20-degree supine and 20-degree right and left lateral recumbent positions. Heart Lung. 1982;11:256–7. Whitman G, Howaniak D, Verga T. Comparison of cardiac output measurements in 20-degree supine and 20-degree right and left lateral recumbent positions. Heart Lung. 1982;11:256–7.
24.
go back to reference Dobbin K, Wallace S, Ahlberg J, Chulay M. Pulmonary artery pressure measurement in patients with elevated pressures: effect of backrest elevation and method of measurement. Am J Crit Care. 1992;1:61–9.PubMed Dobbin K, Wallace S, Ahlberg J, Chulay M. Pulmonary artery pressure measurement in patients with elevated pressures: effect of backrest elevation and method of measurement. Am J Crit Care. 1992;1:61–9.PubMed
25.
go back to reference Wilson AE, Bermingham-Mitchell K, Wells N, Zachary K. Effect of backrest position on hemodynamic and right ventricular measurements in critically ill adults. Am J Crit Care. 1996;5:264–70.PubMed Wilson AE, Bermingham-Mitchell K, Wells N, Zachary K. Effect of backrest position on hemodynamic and right ventricular measurements in critically ill adults. Am J Crit Care. 1996;5:264–70.PubMed
26.
go back to reference Giuliano KK, Scott SS, Brown V, Olson M. Backrest angle and cardiac output measurement in critically ill patients. Nurs Res. 2003;52:242–8.CrossRefPubMed Giuliano KK, Scott SS, Brown V, Olson M. Backrest angle and cardiac output measurement in critically ill patients. Nurs Res. 2003;52:242–8.CrossRefPubMed
27.
go back to reference Grose BL, Woods SL, Laurent DJ. Effect of backrest position on cardiac output measured by the thermodilution method in acutely ill patients. Heart Lung. 1981;10:661–5.PubMed Grose BL, Woods SL, Laurent DJ. Effect of backrest position on cardiac output measured by the thermodilution method in acutely ill patients. Heart Lung. 1981;10:661–5.PubMed
28.
go back to reference Monnet X, Bataille A, Magalhaes E, Barrois J, Le Corre M, Gosset C, Guerin L, Richard C, Teboul JL. End-tidal carbon dioxide is better than arterial pressure for predicting volume responsiveness by the passive leg raising test. Intensive Care Med. 2013;39:93–100.CrossRefPubMed Monnet X, Bataille A, Magalhaes E, Barrois J, Le Corre M, Gosset C, Guerin L, Richard C, Teboul JL. End-tidal carbon dioxide is better than arterial pressure for predicting volume responsiveness by the passive leg raising test. Intensive Care Med. 2013;39:93–100.CrossRefPubMed
29.
go back to reference Jabot J, Teboul JL, Richard C, Monnet X. Passive leg raising for predicting fluid responsiveness: importance of the postural change. Intensive Care Med. 2009;35:85–90.CrossRefPubMed Jabot J, Teboul JL, Richard C, Monnet X. Passive leg raising for predicting fluid responsiveness: importance of the postural change. Intensive Care Med. 2009;35:85–90.CrossRefPubMed
30.
go back to reference He HW, Liu DW. Passive leg raising: influence of blood pressure transducer site. Intensive Care Med. 2013;39(9):1668.CrossRefPubMed He HW, Liu DW. Passive leg raising: influence of blood pressure transducer site. Intensive Care Med. 2013;39(9):1668.CrossRefPubMed
Metadata
Title
The effect of variable arterial transducer level on the accuracy of pulse contour waveform-derived measurements in critically ill patients
Authors
Huai-wu He
Da-wei Liu
Yun Long
Xiao-ting Wang
Mei-ling Zhao
Xiao-li Lai
Publication date
01-10-2016
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 5/2016
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-015-9756-x

Other articles of this Issue 5/2016

Journal of Clinical Monitoring and Computing 5/2016 Go to the issue