Accuracy of bedside glucometry in critically ill children with peripheral hypoperfusion

Authors: Anant Khositseth, Onsuthi Pharadornuwat, Rojjanee Lertbunrian, Nattachai Anantasit, Jarin Vaewpanich

Published in: Journal of Clinical Monitoring and Computing | Issue 1/2015

Login to get access

Abstract

The accuracy of glucose test strip in critically care has been questioned. We investigated the accuracy of glucose test strip in critically ill children. Patients, aged from 1 month to 18 years admitted in pediatric intensive care unit. Demographic data, hemodynamic parameters, and perfusion index (PI) were recorded. Glucose test strips were performed from finger stick blood [capillary blood glucose (CBG)] and from whole blood [whole blood glucose (WBG)] along with laboratory plasma blood glucose (PBG) from either arterial or venous blood samples. The accuracy of glucose test strips was defined according to ISO 15197 and Clarke error grid (CEG). One hundred and eighty one blood samplings including 117 arterial blood (CBG, WBGa, PBGa) and 64 venous blood (CBG, WBGv, PBGv) were obtained. The accuracy of WBGa was 98.3 and 95.2 % when compared to the accuracy of CBG (88.7 and 83.3 %. The accuracy of WBGv was 92.2 % and 87.0 when compared to the accuracy of CBG which was 79.7 and 72.9 % (ISO 15197: 2003 and 2013, respectively). Bland–Altman plot demonstrated bias and precision of 7.4 ± 17.7 mg/dL in acceptable PI group compared to 30.2 ± 23.4 mg/dL in low PI group (PI ≤ 0.3). The CBG test strip must be interpreted carefully in critically ill children. A low PI was associated with poor CBG strip accuracy. WBG test strip from arterial blood was more appropriate for glucose monitoring in children with peripheral hypoperfusion.
Literature
1.
go back to reference Mebis L, Gunst J, Langouche L, Vanhorebeek I, Van den Berghe G. Indication and practical use of intensive insulin therapy in the critically ill. Curr Opin Crit Care. 2007;13:392–8.PubMedCrossRef Mebis L, Gunst J, Langouche L, Vanhorebeek I, Van den Berghe G. Indication and practical use of intensive insulin therapy in the critically ill. Curr Opin Crit Care. 2007;13:392–8.PubMedCrossRef
2.
go back to reference Critchell CD, Savarese V, Callahan A, Aboud C, Jabbour S, Marik P. Accuracy of bedside capillary blood glucose measurements in critically ill patients. Intensive Care Med. 2007;33:2079–84.PubMedCrossRef Critchell CD, Savarese V, Callahan A, Aboud C, Jabbour S, Marik P. Accuracy of bedside capillary blood glucose measurements in critically ill patients. Intensive Care Med. 2007;33:2079–84.PubMedCrossRef
3.
go back to reference Cengiz E, Tamborlane WV. A tale of two compartments: interstitial versus blood glucose monitoring. Diabetes Technol Ther. 2009;11:S11–6.PubMedCrossRef Cengiz E, Tamborlane WV. A tale of two compartments: interstitial versus blood glucose monitoring. Diabetes Technol Ther. 2009;11:S11–6.PubMedCrossRef
4.
go back to reference Karon BS, Gandhi GY, Nuttall GA, Bryant SC, Schaff HV, McMahon MM, et al. Accuracy of roche accu-chek inform whole blood capillary, arterial, and venous glucose values in patients receiving intensive intravenous insulin therapy after cardiac surgery. Am J Clin Pathol. 2007;127:919–26.PubMedCrossRef Karon BS, Gandhi GY, Nuttall GA, Bryant SC, Schaff HV, McMahon MM, et al. Accuracy of roche accu-chek inform whole blood capillary, arterial, and venous glucose values in patients receiving intensive intravenous insulin therapy after cardiac surgery. Am J Clin Pathol. 2007;127:919–26.PubMedCrossRef
5.
go back to reference Desachy A, Vuagnat AC, Ghazali AD, Baudin OT, Longuet OH, Calvat SN, et al. Accuracy of bedside glucometry in critically ill patients: influence of clinical characteristics and perfusion index. Mayo Clin Proc. 2008;83:400–5.PubMedCrossRef Desachy A, Vuagnat AC, Ghazali AD, Baudin OT, Longuet OH, Calvat SN, et al. Accuracy of bedside glucometry in critically ill patients: influence of clinical characteristics and perfusion index. Mayo Clin Proc. 2008;83:400–5.PubMedCrossRef
6.
go back to reference De Felice C, Goldstein MR, Parrini S, Verrotti A, Criscuolo M, Latini G. Early dynamic changes in pulse oximetry signals in preterm newborns with histologic chorioamnionitis. Pediatr Crit Care Med. 2006;7:138–42.PubMedCrossRef De Felice C, Goldstein MR, Parrini S, Verrotti A, Criscuolo M, Latini G. Early dynamic changes in pulse oximetry signals in preterm newborns with histologic chorioamnionitis. Pediatr Crit Care Med. 2006;7:138–42.PubMedCrossRef
7.
go back to reference Lima AP, Beelen P, Bakker J. Use of a peripheral perfusion index derived from the pulse oximetry signal as a noninvasive indicator of perfusion. Crit Care Med. 2002;30:1210–3.PubMedCrossRef Lima AP, Beelen P, Bakker J. Use of a peripheral perfusion index derived from the pulse oximetry signal as a noninvasive indicator of perfusion. Crit Care Med. 2002;30:1210–3.PubMedCrossRef
8.
go back to reference Bland JM, Altman DG. Statistical method for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.PubMedCrossRef Bland JM, Altman DG. Statistical method for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.PubMedCrossRef
9.
go back to reference ISO 15197. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Geneva, Switzerland: International Organization for Standardization; 2003. ISO 15197. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Geneva, Switzerland: International Organization for Standardization; 2003.
10.
go back to reference ISO 15197. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Geneva, Switzerland: International Organization for Standardization; 2013. ISO 15197. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Geneva, Switzerland: International Organization for Standardization; 2013.
11.
go back to reference Clark WL, Cox D, Gonder-Frederick LA, Carter WL, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care. 1987;10:662–8.CrossRef Clark WL, Cox D, Gonder-Frederick LA, Carter WL, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care. 1987;10:662–8.CrossRef
12.
go back to reference Colagiuri S, Sandbaek A, Carstensen B, Christensen J, Glumer C, Lauritzen T, et al. Comparability of venous and capillary glucose measurements in blood. Diabet Med. 2003;20:953–6.PubMedCrossRef Colagiuri S, Sandbaek A, Carstensen B, Christensen J, Glumer C, Lauritzen T, et al. Comparability of venous and capillary glucose measurements in blood. Diabet Med. 2003;20:953–6.PubMedCrossRef
13.
go back to reference Meex C, Poncin J, Chapelle JP, Cavalier E. Analytical validation of the new plasma calibrated accu-chek test strips (roche diagnostics). Clin Chem Lab Med. 2006;44:1376–8.PubMedCrossRef Meex C, Poncin J, Chapelle JP, Cavalier E. Analytical validation of the new plasma calibrated accu-chek test strips (roche diagnostics). Clin Chem Lab Med. 2006;44:1376–8.PubMedCrossRef
14.
go back to reference Meynaar IA, van Spreuwel M, Tangkau PL, Dawson L, Sleeswijk Visser S, Rijks L, et al. Accuracy of accuchek glucose measurement in intensive care patients. Crit Care Med. 2009;37:2691–6.PubMedCrossRef Meynaar IA, van Spreuwel M, Tangkau PL, Dawson L, Sleeswijk Visser S, Rijks L, et al. Accuracy of accuchek glucose measurement in intensive care patients. Crit Care Med. 2009;37:2691–6.PubMedCrossRef
15.
go back to reference Kanji S, Buffie J, Hutton B, Bunting PS, Singh A, McDonald K, et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med. 2005;33:2778–85.PubMedCrossRef Kanji S, Buffie J, Hutton B, Bunting PS, Singh A, McDonald K, et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med. 2005;33:2778–85.PubMedCrossRef
Metadata
Title
Accuracy of bedside glucometry in critically ill children with peripheral hypoperfusion
Authors
Anant Khositseth
Onsuthi Pharadornuwat
Rojjanee Lertbunrian
Nattachai Anantasit
Jarin Vaewpanich
Publication date
01-02-2015
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 1/2015
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-014-9579-1

Other articles of this Issue 1/2015

Journal of Clinical Monitoring and Computing 1/2015 Go to the issue