Published in: Journal of Clinical Monitoring and Computing 1/2014

01-02-2014 | Original Research

Evaluation of the estimated continuous cardiac output monitoring system in adults and children undergoing kidney transplant surgery: a pilot study

Authors: Takashi Terada, Yumi Maemura, Akiko Yoshida, Rika Muto, Ryoichi Ochiai

Published in: Journal of Clinical Monitoring and Computing | Issue 1/2014

Login to get access

Abstract

Evaluation of the estimated continuous cardiac output (esCCO) allows non-invasive and continuous assessment of cardiac output. However, the applicability of this approach in children has not been assessed thus far. We compared the correlation coefficient, bias, standard deviation (SD), and the lower and upper 95 % limits of agreement for esCCO and dye densitography-cardiac output (DDG-CO) measurements by pulse dye densitometry (PDD) in adults and children. On the basis of these assessments, we aimed to examine whether esCCO can be used in pediatric patients. DDG-CO was measured by pulse dye densitometry (PDD) using indocyanine green. Modified-pulse wave transit time, obtained using pulse oximetry and electrocardiography, was used to measure esCCO. Correlations between DDG-CO and esCCO in adults and children were analyzed using regression analysis with the least squares method. Differences between the two correlation coefficients were statistically analyzed using a correlation coefficient test. Bland–Altman plots were used to evaluate bias and SD for DDG-CO and esCCO in both adults and children, and 95 % limits of agreement (bias ± 1.96 SD) and percentage error (1.96 SD/mean DDG-CO) were calculated and compared. The average age of the adult patients (n = 10) was 39.3 ± 12.1 years, while the average age of the pediatric patients (n = 7) was 9.4 ± 3.1 years (p < 0.001). For adults, the correlation coefficient was 0.756; bias, −0.258 L/min; SD, 1.583 L/min; lower and upper 95 % limits of agreement for DDG-CO and esCCO, −3.360 and 2.844 L/min, respectively; and percentage error, 42.7 %. For children, the corresponding values were 0.904; −0.270; 0.908; −2.051 and 1.510 L/min, respectively; and 35.7 %. Due to the high percentage error values, we could not establish a correlation between esCCO and DDG-CO. However, the 95 % limits of agreement and percentage error were better in children than in adults. Due to the high percentage error, we could not confirm a correlation between esCCO and DDG-CO. However, the agreement between esCCO and DDG-CO seems to be higher in children than in adults. These results suggest that esCCO can also be used in children. Future studies with bigger study populations will be required to further investigate these conclusions.
Appendix
Available only for authorised users
Literature
1.
go back to reference Skowno JJ, Michael BF. Cardiac output measurement in pediatric anesthesia. Pediatr Anesth. 2008;18:1019–28.CrossRef Skowno JJ, Michael BF. Cardiac output measurement in pediatric anesthesia. Pediatr Anesth. 2008;18:1019–28.CrossRef
2.
go back to reference Kouchoukos NT, Sheppard LC, McDonald DA. Estimation of stroke volume in the dog by a pulse contour method. Circ Res. 1970;26:611–23.PubMedCrossRef Kouchoukos NT, Sheppard LC, McDonald DA. Estimation of stroke volume in the dog by a pulse contour method. Circ Res. 1970;26:611–23.PubMedCrossRef
3.
go back to reference Ochiai R, Takeda J, Hosaka H, et al. The relationship between modified pulse wave transit time and cardiovascular changes in isoflurane anesthetized dogs. J Clin Monit Comput. 1999;15:493–501.PubMedCrossRef Ochiai R, Takeda J, Hosaka H, et al. The relationship between modified pulse wave transit time and cardiovascular changes in isoflurane anesthetized dogs. J Clin Monit Comput. 1999;15:493–501.PubMedCrossRef
4.
go back to reference Sugo Y, Ukawa T, Takeda S, et al. A novel continuous cardiac output monitor based on pulse wave transit time. Conf Proc IEEE Eng Med Biol Soc. 2010;2010:2853–6.PubMed Sugo Y, Ukawa T, Takeda S, et al. A novel continuous cardiac output monitor based on pulse wave transit time. Conf Proc IEEE Eng Med Biol Soc. 2010;2010:2853–6.PubMed
5.
go back to reference Tibby SM, Hatherill M, Murdoch IA. Use of transesophageal Doppler ultrasonography in ventilated pediatric patients: derivation of cardiac output. Crit Care Med. 2000;6:2045–50.CrossRef Tibby SM, Hatherill M, Murdoch IA. Use of transesophageal Doppler ultrasonography in ventilated pediatric patients: derivation of cardiac output. Crit Care Med. 2000;6:2045–50.CrossRef
6.
go back to reference Chew MS, Poelaert J. Accuracy and repeatability of pediatric cardiac output measurement using Doppler: 20-year review of the literature. Intensive Care Med. 2003;29:1889–94.PubMedCrossRef Chew MS, Poelaert J. Accuracy and repeatability of pediatric cardiac output measurement using Doppler: 20-year review of the literature. Intensive Care Med. 2003;29:1889–94.PubMedCrossRef
7.
go back to reference Peyton PJ, Chong SW. Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision. Anesthesiology. 2010;113:1220–35.PubMedCrossRef Peyton PJ, Chong SW. Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision. Anesthesiology. 2010;113:1220–35.PubMedCrossRef
8.
go back to reference Funk DJ, Moretti EW, Gan TJ. Minimally invasive cardiac output monitoring in the perioperative setting. Anesth Analg. 2009;108:887–97.PubMedCrossRef Funk DJ, Moretti EW, Gan TJ. Minimally invasive cardiac output monitoring in the perioperative setting. Anesth Analg. 2009;108:887–97.PubMedCrossRef
9.
go back to reference Feldman JM. Is it a bird? Is it a plane? The role of patient monitors in medical decision making. Anesth Analg. 2009;108:707–10.PubMedCrossRef Feldman JM. Is it a bird? Is it a plane? The role of patient monitors in medical decision making. Anesth Analg. 2009;108:707–10.PubMedCrossRef
10.
go back to reference Nagano K, Kusaka T, Okubo K, et al. Estimation of circulating blood volume in infants using the pulse dye densitometry method. Paediatr Anesth. 2005;15:125–30.CrossRef Nagano K, Kusaka T, Okubo K, et al. Estimation of circulating blood volume in infants using the pulse dye densitometry method. Paediatr Anesth. 2005;15:125–30.CrossRef
11.
go back to reference Taguchi N, Nakagawa S, Miyasaka K, et al. Cardiac output measurement by pulse dye densitometry using three wavelengths. Pediatr Crit Care Med. 2004;5:343–50.PubMedCrossRef Taguchi N, Nakagawa S, Miyasaka K, et al. Cardiac output measurement by pulse dye densitometry using three wavelengths. Pediatr Crit Care Med. 2004;5:343–50.PubMedCrossRef
12.
go back to reference Ishihara H, Okawa H, Tanabe K, et al. A new non-invasive continuous cardiac output trend solely utilizing routine cardiovascular monitors. J Clin Monit Comput. 2004;18:313–20.PubMedCrossRef Ishihara H, Okawa H, Tanabe K, et al. A new non-invasive continuous cardiac output trend solely utilizing routine cardiovascular monitors. J Clin Monit Comput. 2004;18:313–20.PubMedCrossRef
13.
go back to reference Yamada T, Tsutsui M, Sugo Y, et al. Multicenter study verifying a method of non-invasive continuous cardiac output measurement using pulse wave transit time: a comparison with the intermittent bolus thermodilution cardiac output. Anesth Analg. 2012;115:82–7.PubMedCrossRef Yamada T, Tsutsui M, Sugo Y, et al. Multicenter study verifying a method of non-invasive continuous cardiac output measurement using pulse wave transit time: a comparison with the intermittent bolus thermodilution cardiac output. Anesth Analg. 2012;115:82–7.PubMedCrossRef
14.
go back to reference Ishihara H, Sugo Y, Tsutsui M, et al. The ability of a new continuous cardiac output monitor to measure trends in cardiac output following implementation of a patient information calibration and an automated exclusion algorithm. J Clin Monit Comput. 2012;26:465–71.PubMedCentralPubMedCrossRef Ishihara H, Sugo Y, Tsutsui M, et al. The ability of a new continuous cardiac output monitor to measure trends in cardiac output following implementation of a patient information calibration and an automated exclusion algorithm. J Clin Monit Comput. 2012;26:465–71.PubMedCentralPubMedCrossRef
Metadata
Title
Evaluation of the estimated continuous cardiac output monitoring system in adults and children undergoing kidney transplant surgery: a pilot study
Authors
Takashi Terada
Yumi Maemura
Akiko Yoshida
Rika Muto
Ryoichi Ochiai
Publication date
01-02-2014
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 1/2014
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-013-9501-2

Other articles of this Issue 1/2014

Journal of Clinical Monitoring and Computing 1/2014 Go to the issue