Validation of continuous and noninvasive hemoglobin monitoring by pulse CO-oximetry in Japanese surgical patients

Authors: Tsuyoshi Isosu, Shinju Obara, Atsuyuki Hosono, Satoshi Ohashi, Yuko Nakano, Tsuyoshi Imaizumi, Midori Mogami, Masahiro Murakawa

Published in: Journal of Clinical Monitoring and Computing | Issue 1/2013

Login to get access

Abstract

We evaluated the accuracy of noninvasive and continuous total hemoglobin (SpHb) monitoring with the Radical-7® Pulse CO-Oximeter in Japanese surgical patients before and after an in vivo adjustment of the first SpHb value to match the first reference value from a satellite laboratory CO-Oximeter. Twenty patients undergoing surgical procedures with general anesthesia were monitored with Pulse CO-Oximetry for SpHb. Laboratory CO-Oximeter values (tHb) were compared to SpHb at the time of the blood draws. Bias, precision, limits of agreement and correlation coefficient of SpHb compared to tHb were calculated before and after SpHb values were adjusted by subtracting the difference between the first SpHb and tHb value from all subsequent SpHb values. Trending of SpHb to tHb and the effect of perfusion index (PI) on the agreement of SpHb to tHb were also analyzed. Ninety-two tHb values were compared to the SpHb. Bias ± 1SD was 0.2 ± 1.5 g/dL before in vivo adjustment and −0.7 ± 1.0 g/dL after in vivo adjustment. Bland–Altman analysis showed limits of agreement of −2.8 to 3.1 g/dL before in vivo adjustment and −2.8 to 1.4 g/dL after in vivo adjustment. The correlation coefficient was 0.76 prior to in vivo adjustment and 0.87 after in vivo adjustment. In patients with adequate perfusion (PI ≥1.4) the correlation coefficient was 0.89. In vivo adjustment of SpHb significantly improved the accuracy in our cohort of Japanese surgical patients. The strongest correlation between SpHb and tHb values was observed in patients with adequate peripheral perfusion suggesting that low perfusion may affect the accuracy of SpHb monitoring.
Literature
5.
go back to reference Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571–82.PubMedCrossRef Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571–82.PubMedCrossRef
7.
go back to reference Feiner JR, Severinghaus JW, Bickler PE. Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: the effects of oximeter probe type and gender. Anesth Analg. 2007; 105(6 Suppl):S18–23, tables of contents. doi:10.1213/01.ane.0000285988.35174.d9. Feiner JR, Severinghaus JW, Bickler PE. Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: the effects of oximeter probe type and gender. Anesth Analg. 2007; 105(6 Suppl):S18–23, tables of contents. doi:10.​1213/​01.​ane.​0000285988.​35174.​d9.
8.
go back to reference Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue(R) with automated laboratory hemoglobin measurement. Anesthesiology. 2011;115(3):548–54. doi:10.1097/ALN.0b013e3182270c22.PubMedCrossRef Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue(R) with automated laboratory hemoglobin measurement. Anesthesiology. 2011;115(3):548–54. doi:10.​1097/​ALN.​0b013e3182270c22​.PubMedCrossRef
9.
go back to reference International Committee for Standardization in Haematology, Expert Panel on Haemoglobinometry. Recommendations for reference method for haemoglobinometry in human blood (ICSH standard 1986) and specifications for international haemiglobincyanide reference preparation (3rd edn). Clin Lab Haematol. 1987; 9(1):73–9. International Committee for Standardization in Haematology, Expert Panel on Haemoglobinometry. Recommendations for reference method for haemoglobinometry in human blood (ICSH standard 1986) and specifications for international haemiglobincyanide reference preparation (3rd edn). Clin Lab Haematol. 1987; 9(1):73–9.
Metadata
Title
Validation of continuous and noninvasive hemoglobin monitoring by pulse CO-oximetry in Japanese surgical patients
Authors
Tsuyoshi Isosu
Shinju Obara
Atsuyuki Hosono
Satoshi Ohashi
Yuko Nakano
Tsuyoshi Imaizumi
Midori Mogami
Masahiro Murakawa
Publication date
01-02-2013
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 1/2013
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-012-9397-2

Other articles of this Issue 1/2013

Journal of Clinical Monitoring and Computing 1/2013 Go to the issue