Skip to main content
Top
Published in: International Ophthalmology 1/2020

01-01-2020 | Cataract | Original Paper

In vivo and in vitro results of an automated preloaded delivery system for IOL implantation in cataract surgery

Authors: Leonardo Mastropasqua, Lisa Toto, Erminia D’Ugo, Manuela Lanzini, Peter A. Mattei, Gennaro Falconio, Emanuele Doronzo, Francesca Diomede, Oriana Trubiani

Published in: International Ophthalmology | Issue 1/2020

Login to get access

Abstract

Purpose

To compare the corneal tissue trauma after the use of an automated preloaded injector and a manual injector and assess scanning electron microscope (SEM) and atomic force microscope (AFM) features of both injector cartridges.

Setting

Ophthalmology Clinic and Laboratory of Stem Cells and Regenerative Medicine University “G. d'Annunzio” of Chieti-Pescara, Chieti, Italy;

Design

Prospective randomized clinical study

Methods

Forty eyes of 40 patients for phacoemulsification were divided into two groups: implantation of intraocular lens was performed with AutonoMe automated delivery system (AutonoMe group: 20 eyes) and Monarch III injector system (Monarch group: 20 eyes). In vivo confocal microscopy (IVCM) and anterior segment optical coherence tomography (AS-OCT) were performed before surgery, at 1 h, 1 day and 1 month post-operatively. In addition, SEM and AFM were performed on cartridges of both injector systems after injection of the IOL.

Results

A greater increase in central corneal thickness and corneal thickness at the incision site were observed in Monarch group versus AutonoMe group 1 h and 1 day post-operatively (p < 0.05). Endothelial cell count loss was significantly higher in Monarch group compared with AutonoMe group (p < 0.05) at 1 and 24 h. AS-OCT showed less endothelial misalignment at 30 days (p < 0.05), and IVCM showed less tunnel inflammation at all time points (p < 0.05) in AutonoMe group compared with Monarch group; roughness analysis at AFM of the AutonoMe cartridge was significantly lower compared to Monarch D cartridge (p < 0.05).

Conclusions

The AutonoMe injector provided less corneal tissue trauma compared with Monarch III injector. The AutonoMe cartridge showed lower roughness at AFM compared to the Monarch D cartridge.
Literature
1.
go back to reference Ong HS, Subash M, Sandhu A, Wilkins MR (2013) Intraocular lens delivery characteristics of the preloaded AcrySof IQ SN60WS/AcrySert injectable lens system. Am J Ophthalmol 156(1):77–81CrossRef Ong HS, Subash M, Sandhu A, Wilkins MR (2013) Intraocular lens delivery characteristics of the preloaded AcrySof IQ SN60WS/AcrySert injectable lens system. Am J Ophthalmol 156(1):77–81CrossRef
2.
go back to reference Oshika T, Nagahara K, Yaguchi S et al (1998) Three year prospective, randomized evaluation of intraocular lens implantation through 3.2 and 5.5 mm incisions. J Cataract Refract Surg 24(4):509–514CrossRef Oshika T, Nagahara K, Yaguchi S et al (1998) Three year prospective, randomized evaluation of intraocular lens implantation through 3.2 and 5.5 mm incisions. J Cataract Refract Surg 24(4):509–514CrossRef
3.
go back to reference Zheng L, Merriam JC, Zaider M (1997) Astigmatism and visual recovery after ‘large incision’ extracapsular cataract surgery and ‘small’ incisions for phakoemulsification. Trans Am Ophthalmol Soc 95:387–415PubMedPubMedCentral Zheng L, Merriam JC, Zaider M (1997) Astigmatism and visual recovery after ‘large incision’ extracapsular cataract surgery and ‘small’ incisions for phakoemulsification. Trans Am Ophthalmol Soc 95:387–415PubMedPubMedCentral
4.
go back to reference Steinert RF, Brint SF, White SM, Fine IH (1991) Astigmatism after small incision cataract surgery. A prospective, randomized multicenter comparison of 4- and 6.5-mm incisions. Ophthalmology 98(4):417–424CrossRef Steinert RF, Brint SF, White SM, Fine IH (1991) Astigmatism after small incision cataract surgery. A prospective, randomized multicenter comparison of 4- and 6.5-mm incisions. Ophthalmology 98(4):417–424CrossRef
5.
go back to reference Belazzougui R, Monod SD, Baudouin C, Labbé A (2010) Architectural analysis of clear corneal incisions using Visante OCT in acute postoperative endophthalmitis. J Fr Ophtalmol 33(1):10–15CrossRef Belazzougui R, Monod SD, Baudouin C, Labbé A (2010) Architectural analysis of clear corneal incisions using Visante OCT in acute postoperative endophthalmitis. J Fr Ophtalmol 33(1):10–15CrossRef
6.
go back to reference Palanker DV et al (2010) Femtosecond laser-assisted cataract surgery with integrated optical coherence tomography. Sci Transl Med 17(2):58 Palanker DV et al (2010) Femtosecond laser-assisted cataract surgery with integrated optical coherence tomography. Sci Transl Med 17(2):58
7.
go back to reference Taban M et al (2005) Ingress of India ink into the anterior chamber through sutureless clear corneal cataract wounds. Arch Ophthalmol 123:643–648CrossRef Taban M et al (2005) Ingress of India ink into the anterior chamber through sutureless clear corneal cataract wounds. Arch Ophthalmol 123:643–648CrossRef
8.
go back to reference Fine IH, Hoffman RS, Packer M (2007) Profile of clear corneal cataract incisions demonstrated by ocular coherence tomography. J Cataract Refract Surg 33(1):94–97CrossRef Fine IH, Hoffman RS, Packer M (2007) Profile of clear corneal cataract incisions demonstrated by ocular coherence tomography. J Cataract Refract Surg 33(1):94–97CrossRef
9.
go back to reference Mencucci R, Favuzza E, Salvatici MC, Spadea L, Allen D (2018) Corneal incision architecture after IOL implantation with three different injectors: an environmental scanning electron microscopy study. Int Ophthalmol 1 (Epub ahead of print) Mencucci R, Favuzza E, Salvatici MC, Spadea L, Allen D (2018) Corneal incision architecture after IOL implantation with three different injectors: an environmental scanning electron microscopy study. Int Ophthalmol 1 (Epub ahead of print)
10.
go back to reference Chung B, Lee H, Choi M, Seo KY, Kim EK, Kim TI (2018) Preloaded and non-preloaded intraocular lens delivery system and characteristics: human and porcine eyes trial. Int J Ophthalmol 18(11):6–11 Chung B, Lee H, Choi M, Seo KY, Kim EK, Kim TI (2018) Preloaded and non-preloaded intraocular lens delivery system and characteristics: human and porcine eyes trial. Int J Ophthalmol 18(11):6–11
11.
go back to reference Usui M, Tanaka T (2015) Resistance force for intraocular lens insertion through lens cartridges and syringe-type injectors. J Cataract Refract Surg 41(8):1745–1751CrossRef Usui M, Tanaka T (2015) Resistance force for intraocular lens insertion through lens cartridges and syringe-type injectors. J Cataract Refract Surg 41(8):1745–1751CrossRef
12.
go back to reference Kohnen T, Klaproth OK (2008) Incision sizes before and after implantation of SN60WF intraocular lenses using the Monarch injector system with C and D cartridges. J Cataract Refract Surg 34:1748–1753CrossRef Kohnen T, Klaproth OK (2008) Incision sizes before and after implantation of SN60WF intraocular lenses using the Monarch injector system with C and D cartridges. J Cataract Refract Surg 34:1748–1753CrossRef
13.
go back to reference Kleinmann G, Kleinmann I (2014) Intraocular lens injector-induced stress on the corneal incisions during lens implantation. Am J Ophthalmol 158(1):185–191CrossRef Kleinmann G, Kleinmann I (2014) Intraocular lens injector-induced stress on the corneal incisions during lens implantation. Am J Ophthalmol 158(1):185–191CrossRef
14.
go back to reference Mastropasqua R, Agnifili L, Fasanella V, Curcio C, Brescia L, Lanzini M, Fresina M, Mastropasqua L, Marchini G (2015) Corneoscleral limbus in glaucoma patients. In vivo confocal microscopy and immunocytological study. Invest Ophthalmol Vis Sci. 56:2050–2058CrossRef Mastropasqua R, Agnifili L, Fasanella V, Curcio C, Brescia L, Lanzini M, Fresina M, Mastropasqua L, Marchini G (2015) Corneoscleral limbus in glaucoma patients. In vivo confocal microscopy and immunocytological study. Invest Ophthalmol Vis Sci. 56:2050–2058CrossRef
15.
go back to reference Mastropasqua L, Toto L, Vecchiarino L, Di Nicola M, Mastropasqua R (2011) Microcoaxial torsional cataract surgery 18 mm versus 22 mm: functional and morphological assessment. Ophthalmic Surg Lasers Imaging. 42(2):114–124PubMed Mastropasqua L, Toto L, Vecchiarino L, Di Nicola M, Mastropasqua R (2011) Microcoaxial torsional cataract surgery 18 mm versus 22 mm: functional and morphological assessment. Ophthalmic Surg Lasers Imaging. 42(2):114–124PubMed
16.
go back to reference Khokhar S, Sharma R, Patil B, Aron N, Gupta S (2014) Comparison of new motorized injector vs manual injector for implantation of foldable intraocular lenses on wound integrity: an ASOCT study. Eye (Lond) 28(10):1174–1178CrossRef Khokhar S, Sharma R, Patil B, Aron N, Gupta S (2014) Comparison of new motorized injector vs manual injector for implantation of foldable intraocular lenses on wound integrity: an ASOCT study. Eye (Lond) 28(10):1174–1178CrossRef
17.
go back to reference Kohnen T, Klaproth OK (2008) Incision sizes before and after implantation of SN60WF intraocular lenses using the Monarch injector system with C and D cartridges. J Cataract Refract Surg. 34(10):1748–1753CrossRef Kohnen T, Klaproth OK (2008) Incision sizes before and after implantation of SN60WF intraocular lenses using the Monarch injector system with C and D cartridges. J Cataract Refract Surg. 34(10):1748–1753CrossRef
17.
go back to reference Imai H, Fujii A, Tani E, Azumi A (2016) Wound integrity of 2.0 mm transconjunctival single-plane sclerocorneal incision: a comparison between a motorized injector with/without pause time and a manual injector. J Ophthalmol. 2016, 8129298.PubMedPubMedCentral Imai H, Fujii A, Tani E, Azumi A (2016) Wound integrity of 2.0 mm transconjunctival single-plane sclerocorneal incision: a comparison between a motorized injector with/without pause time and a manual injector. J Ophthalmol. 2016, 8129298.PubMedPubMedCentral
19.
go back to reference Ho JW, Afshari NA (2015) Advances in cataract surgery: preserving the corneal endothelium. Curr Opin Ophthalmol 26(1):22–27CrossRef Ho JW, Afshari NA (2015) Advances in cataract surgery: preserving the corneal endothelium. Curr Opin Ophthalmol 26(1):22–27CrossRef
Metadata
Title
In vivo and in vitro results of an automated preloaded delivery system for IOL implantation in cataract surgery
Authors
Leonardo Mastropasqua
Lisa Toto
Erminia D’Ugo
Manuela Lanzini
Peter A. Mattei
Gennaro Falconio
Emanuele Doronzo
Francesca Diomede
Oriana Trubiani
Publication date
01-01-2020
Publisher
Springer Netherlands
Keywords
Cataract
Incision
Published in
International Ophthalmology / Issue 1/2020
Print ISSN: 0165-5701
Electronic ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-019-01154-0

Other articles of this Issue 1/2020

International Ophthalmology 1/2020 Go to the issue