Vertical implantable collamer lens (ICL) rotation for the management of high vault due to lens oversizing

Authors: Francesco Matarazzo, Alexander Clifford Day, Luis Fernandez-Vega Cueto, Vincenzo Maurino

Published in: International Ophthalmology | Issue 6/2018

Login to get access

Abstract

Purpose

To describe the good outcome of implantable collamer lens (ICL) rotation to reduce post-operative vault.

Methods

Retrospective analysis of case report. A 43-year-old woman had V4c EVO + myopic non-toric ICL implantation and post-operatively she presented with anisocoria and high vault. She underwent surgery to rotate the ICL 90 degrees to a vertical orientation.

Results

We achieved a reduction in the vault from 1020 to 486 μm after vertical ICL rotation. Satisfactory refractive outcome and optimal vault were achieved and maintained during the period of follow-up.

Conclusion

Optimal ICL sizing is important as too high vault/clearance is associated with problems such as angle closure glaucoma, pupil dilatation and anisocoria and too low vault/clearance to increased risk of cataract formation. Non-toric ICL rotation can be a simple surgical technique to deal with oversized lenses thus avoiding ICL exchange.
Literature
4.
go back to reference Reinstein DZ, Archer TJ, Silverman RH, Rondeau MJ, Coleman DJ (2009) Correlation of anterior chamber angle and ciliary sulcus diameters with white-to-white corneal diameter in high myopes using Artemis VHF digital ultrasound. J Refract Surg 25(2):185–194PubMedPubMedCentral Reinstein DZ, Archer TJ, Silverman RH, Rondeau MJ, Coleman DJ (2009) Correlation of anterior chamber angle and ciliary sulcus diameters with white-to-white corneal diameter in high myopes using Artemis VHF digital ultrasound. J Refract Surg 25(2):185–194PubMedPubMedCentral
7.
go back to reference Sanders DR, Vukich JA, Doney K, Gaston M, Implantable Contact Lens in Treatment of Myopia Study Group (2003) U.S. Food and drug administration clinical trial of the implantable contact lens for moderate to high myopia. Ophthalmology 110(2):255–266CrossRef Sanders DR, Vukich JA, Doney K, Gaston M, Implantable Contact Lens in Treatment of Myopia Study Group (2003) U.S. Food and drug administration clinical trial of the implantable contact lens for moderate to high myopia. Ophthalmology 110(2):255–266CrossRef
Metadata
Title
Vertical implantable collamer lens (ICL) rotation for the management of high vault due to lens oversizing
Authors
Francesco Matarazzo
Alexander Clifford Day
Luis Fernandez-Vega Cueto
Vincenzo Maurino
Publication date
01-12-2018
Publisher
Springer Netherlands
Published in
International Ophthalmology / Issue 6/2018
Print ISSN: 0165-5701
Electronic ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-017-0757-2

Other articles of this Issue 6/2018

International Ophthalmology 6/2018 Go to the issue