Skip to main content
Top
Published in: International Ophthalmology 4/2018

01-08-2018 | Case Report

Opacification of the C-flex 570C intraocular lens after sulcus fixation

Authors: Patrick Daigle, Marjorie Carbonneau

Published in: International Ophthalmology | Issue 4/2018

Login to get access

Abstract

Purpose

To report a case of diffuse intraocular lens (IOL) opacification in a patient who started complaining of blurred vision in his left eye over the course of three years after having phacoemulsification surgery combined with capsular bag fixation of a C-flex 570C IOL. The IOL had been repositioned in the ciliary sulcus following its subluxation.

Method

An IOL exchange was performed, and the explanted IOL was sent for histopathological analysis.

Results

Scanning electron microscopy identified multiple crystalline-like deposits on both sides of the IOL. The optic was more extensively involved than the haptics. Energy-dispersive X-ray spectroscopy confirmed the predominance of calcium within the deposits.

Conclusions

The breakdown of the blood-aqueous barrier seems to be a key component of this calcification process. Ciliary sulcus fixation is not a suitable option for C-flex 570C IOLs.
Literature
1.
go back to reference Werner L (2007) Causes of intraocular lens opacification or discoloration. J Cataract Refract Surg 33:713–726CrossRefPubMed Werner L (2007) Causes of intraocular lens opacification or discoloration. J Cataract Refract Surg 33:713–726CrossRefPubMed
2.
go back to reference Michelson J, Werner L, Ollerton A et al (2012) Light scattering and light transmittance in intraocular lenses explanted because of optic opacification. J Cataract Refract Surg 38:1476–1485CrossRefPubMed Michelson J, Werner L, Ollerton A et al (2012) Light scattering and light transmittance in intraocular lenses explanted because of optic opacification. J Cataract Refract Surg 38:1476–1485CrossRefPubMed
3.
go back to reference Lee SJ, Choi JH, Sun HJ et al (2010) Surface calcification of hydrophilic acrylic intraocular lens related to inflammatory membrane formation after combined vitrectomy and cataract surgery. J Cataract Refract Surg 36:676–681CrossRefPubMed Lee SJ, Choi JH, Sun HJ et al (2010) Surface calcification of hydrophilic acrylic intraocular lens related to inflammatory membrane formation after combined vitrectomy and cataract surgery. J Cataract Refract Surg 36:676–681CrossRefPubMed
4.
go back to reference Werner L, Wilbanks G, Nieuwendaal CP et al (2015) Localized opacification of hydrophilic acrylic intraocular lenses after procedures using intracameral injection of air or gas. J Cataract Refract Surg 41:199–207CrossRefPubMed Werner L, Wilbanks G, Nieuwendaal CP et al (2015) Localized opacification of hydrophilic acrylic intraocular lenses after procedures using intracameral injection of air or gas. J Cataract Refract Surg 41:199–207CrossRefPubMed
5.
go back to reference Lee J, Lee D, Choi SK, Kim JH (2011) Surface calcification of hydrophilic acrylic intraocular lens related to neovacular glaucoma: 9 cases with rayner C-flex 570c. Invest Ophthalmol Vis Sci 52:6210 Lee J, Lee D, Choi SK, Kim JH (2011) Surface calcification of hydrophilic acrylic intraocular lens related to neovacular glaucoma: 9 cases with rayner C-flex 570c. Invest Ophthalmol Vis Sci 52:6210
6.
go back to reference Fung SSM, Sykakis E, Islam NM et al (2015) Intraocular lens opacification following intracameral injection of recombinant tissue plasminogen activator to treat inflammatory membranes after cataract surgery. J Ophthalmol 2015:975075CrossRefPubMedPubMedCentral Fung SSM, Sykakis E, Islam NM et al (2015) Intraocular lens opacification following intracameral injection of recombinant tissue plasminogen activator to treat inflammatory membranes after cataract surgery. J Ophthalmol 2015:975075CrossRefPubMedPubMedCentral
7.
go back to reference Morgan-Warren PJ, Andreatta W, Patel AK (2015) Opacification of hydrophilic intraocular lenses after descemet stripping automated endothelial keratoplasty. Clin Ophthalmol Auckl NZ 9:277–283CrossRef Morgan-Warren PJ, Andreatta W, Patel AK (2015) Opacification of hydrophilic intraocular lenses after descemet stripping automated endothelial keratoplasty. Clin Ophthalmol Auckl NZ 9:277–283CrossRef
8.
go back to reference Lim AKE, Goh PP, Azura R, Mariam I (2011) Opacification of AcriFlex 50CSE hydrophilic acrylic intraocular lenses. J Cataract Refract Surg 37:655–659CrossRefPubMed Lim AKE, Goh PP, Azura R, Mariam I (2011) Opacification of AcriFlex 50CSE hydrophilic acrylic intraocular lenses. J Cataract Refract Surg 37:655–659CrossRefPubMed
9.
go back to reference Moriarty AP, Spalton DJ, Moriarty BJ et al (1994) Studies of the blood-aqueous barrier in diabetes mellitus. Am J Ophthalmol 117:768–771CrossRefPubMed Moriarty AP, Spalton DJ, Moriarty BJ et al (1994) Studies of the blood-aqueous barrier in diabetes mellitus. Am J Ophthalmol 117:768–771CrossRefPubMed
11.
go back to reference Taravati P, Lam DL, Leveque T, Van Gelder RN (2012) Postcataract surgical inflammation. Curr Opin Ophthalmol 23:12–18CrossRefPubMed Taravati P, Lam DL, Leveque T, Van Gelder RN (2012) Postcataract surgical inflammation. Curr Opin Ophthalmol 23:12–18CrossRefPubMed
12.
go back to reference Skip Nichamin LD (2009) Effect of repositioning on IOL opacification. J Cataract Refract Surg 35:965CrossRefPubMed Skip Nichamin LD (2009) Effect of repositioning on IOL opacification. J Cataract Refract Surg 35:965CrossRefPubMed
13.
go back to reference Miyake K, Asakura M, Kobayashi H (1984) Effect of intraocular lens fixation on the blood-aqueous barrier. Am J Ophthalmol 98:451–455CrossRefPubMed Miyake K, Asakura M, Kobayashi H (1984) Effect of intraocular lens fixation on the blood-aqueous barrier. Am J Ophthalmol 98:451–455CrossRefPubMed
14.
go back to reference Chang DF, Masket S, Miller KM et al (2009) Complications of sulcus placement of single-piece acrylic intraocular lenses: recommendations for backup IOL implantation following posterior capsule rupture. J Cataract Refract Surg 35:1445–1458CrossRefPubMed Chang DF, Masket S, Miller KM et al (2009) Complications of sulcus placement of single-piece acrylic intraocular lenses: recommendations for backup IOL implantation following posterior capsule rupture. J Cataract Refract Surg 35:1445–1458CrossRefPubMed
Metadata
Title
Opacification of the C-flex 570C intraocular lens after sulcus fixation
Authors
Patrick Daigle
Marjorie Carbonneau
Publication date
01-08-2018
Publisher
Springer Netherlands
Published in
International Ophthalmology / Issue 4/2018
Print ISSN: 0165-5701
Electronic ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-017-0643-y

Other articles of this Issue 4/2018

International Ophthalmology 4/2018 Go to the issue