Abstract

Purpose

To describe recent changes in the dimensions of the ERG-jet contact lens electrode, and to suggest corresponding changes in the dimensions of cylindrical stents to facilitate the use of ERG-jet contact lens electrodes in infants and small children.

Methods

Upon the observation that existing cylindrical stents no longer fit newly purchased ERG-jet contact lenses, the spacings between the “bumps” on the front surface of old and new electrodes were measured with calipers, and prototype stents were newly machined from acrylic plastic.

Results

Cylindrical stents with an outside diameter of 0.277 inches (about 7.04 mm) were found to fit snugly between the front surface “bumps” of new ERG-jet contact lens electrodes.

Conclusions

Recent modifications in the injection molds for ERG-jet contact lens electrodes have resulted in an increase in the space between the “bumps” on the front surface of the electrodes, necessitating a corresponding increase in the outside diameter of cylindrical stents intended to fit snugly between the “bumps” for ease of insertion and prevention of lid closure.
Literature
1.
go back to reference Brodie SE, Breton ME (1996) Modified ERG-jet contact lens electrodes for use in infants and toddlers. Doc Ophthalmol 91:141–146CrossRef Brodie SE, Breton ME (1996) Modified ERG-jet contact lens electrodes for use in infants and toddlers. Doc Ophthalmol 91:141–146CrossRef
Metadata
Title
Modified ERG-Jet™ contact lens electrodes for use in infants and toddlers—update
Author
Scott E. Brodie
Publication date
01-12-2018
Publisher
Springer Berlin Heidelberg
Published in
Documenta Ophthalmologica / Issue 3/2018
Print ISSN: 0012-4486
Electronic ISSN: 1573-2622
DOI
https://doi.org/10.1007/s10633-018-9662-x

Other articles of this Issue 3/2018

Documenta Ophthalmologica 3/2018 Go to the issue