01-07-2017 | Clinical Investigation

Evaluation of kinetic programs in various automated perimeters

Authors: Shigeki Hashimoto, Chota Matsumoto, Mariko Eura, Sachiko Okuyama, Yoshikazu Shimomura

Published in: Japanese Journal of Ophthalmology | Issue 4/2017

Login to get access

Abstract

Purpose

Kinetic programs in four automated perimeters were evaluated and compared for their clinical usefulness using four simulated visual field (VF) patterns.

Methods

Using the results of conventional Goldmann manual kinetic perimetry (MKP), simulated fields with concentric contraction, a temporal residual island only, a small central island with a temporal island, and a ring scotoma were created. Four kinetic programs, Humphrey 750i Kinetic Test (Humphrey), OCULUS Twinfield 2 Kinetic Perimetry (OCULUS), OCTOPUS 900 Goldmann Kinetic Perimetry (OCTOPUS GKP), and Kowa AP-7000 Isopter (Kowa) were tested by the 4 simulated defect patterns using stimuli of V/4e, I/4e, I/3e, I/2e, and I/1e at speeds of 3 and 5°/s.

Results

Except Humphrey, OCULUS, OCTOPUS GKP, and Kowa could obtain isopters nearly comparable to those of Goldmann MKP. However, their results were considerably influenced by the examiner’s skill. Besides, Humphrey had restrictions on target presentation, and OCULUS and Kowa had problems in isopter drawing and in filling in the scotoma. OCTOPUS GKP was the only method that could correctly detect and depict all four defect patterns. It also had relatively shorter test durations among the three methods excluding Humphrey, which did not have a built-in function for test duration measurement. The perimeters’ test ranges for the periphery were 90° for Humphrey, OCULUS, and OCTOPUS GKP, and 80° for Kowa.

Conclusion

To assess kinetic fields with various defect patterns, OCTOPUS GKP seems to be the most useful method.
Literature
1.
go back to reference Aulhorn E. Glaukoma-Gesichtsfield. Ophthalmologica. 1968;158:469–87 (in German).CrossRef Aulhorn E. Glaukoma-Gesichtsfield. Ophthalmologica. 1968;158:469–87 (in German).CrossRef
2.
go back to reference Grover S, Fishman GA, Brown J Jr. Patterns of visual field progression in patients with retinitis pigmentosa. Ophthalmology. 1988;105:1069–75.CrossRef Grover S, Fishman GA, Brown J Jr. Patterns of visual field progression in patients with retinitis pigmentosa. Ophthalmology. 1988;105:1069–75.CrossRef
3.
go back to reference Chauhan BC, Drance SM. The relationship between intraocular pressure and visual field progression in glaucoma. Graefes Arch Clin Expo Ophthalmol. 1992;230:521–6.CrossRef Chauhan BC, Drance SM. The relationship between intraocular pressure and visual field progression in glaucoma. Graefes Arch Clin Expo Ophthalmol. 1992;230:521–6.CrossRef
4.
go back to reference Goldmann H. Ein selbstregistrierendes Projektionskugelperimeter. Ophthalmologica. 1945;109:71–9 (in German).CrossRef Goldmann H. Ein selbstregistrierendes Projektionskugelperimeter. Ophthalmologica. 1945;109:71–9 (in German).CrossRef
5.
go back to reference Bittner AK, Iftikhar MH, Dagnelie G. Test-retest, within-visit variability of Goldmann visual fields in retinitis pigmentosa. Invest Ophthalmol Vis Sci. 2011;11:8042–6.CrossRef Bittner AK, Iftikhar MH, Dagnelie G. Test-retest, within-visit variability of Goldmann visual fields in retinitis pigmentosa. Invest Ophthalmol Vis Sci. 2011;11:8042–6.CrossRef
6.
go back to reference Schiefer U, Strasburger H, Becker ST, Vonthein R, Schiller J, Dietrich TJ, et al. Reaction time in automated kinetic perimetry: effects of stimulus luminance, eccentricity, and movement direction. Vision Res. 2001;41:2157–64.CrossRefPubMed Schiefer U, Strasburger H, Becker ST, Vonthein R, Schiller J, Dietrich TJ, et al. Reaction time in automated kinetic perimetry: effects of stimulus luminance, eccentricity, and movement direction. Vision Res. 2001;41:2157–64.CrossRefPubMed
7.
go back to reference Nowomiejska KE, Vonthein R, Paetzold J, Zagorski Z, Kardon R, Schiefer U. Comparison between semiautomated kinetic perimetry and conventional Goldmann manual kinetic perimetry in advanced visual field loss. Ophthalmoly. 2005;112:1343–54.CrossRef Nowomiejska KE, Vonthein R, Paetzold J, Zagorski Z, Kardon R, Schiefer U. Comparison between semiautomated kinetic perimetry and conventional Goldmann manual kinetic perimetry in advanced visual field loss. Ophthalmoly. 2005;112:1343–54.CrossRef
8.
go back to reference Schiefer U, Nowomiejska K, Krapp E, Paetzold J, Johnson CA. K-Train- a computer-based, interactive training propram with an incorporated certification system for practicing kinetic perimetry: evaluation of acceptance and success rate. Graefes Arch Clin Expo Ophthalmol. 2006;244:1300–9.CrossRef Schiefer U, Nowomiejska K, Krapp E, Paetzold J, Johnson CA. K-Train- a computer-based, interactive training propram with an incorporated certification system for practicing kinetic perimetry: evaluation of acceptance and success rate. Graefes Arch Clin Expo Ophthalmol. 2006;244:1300–9.CrossRef
9.
go back to reference Nevalainen J, Paetzold J, Krapp E, Vonthein R, Johnson CA, Schiefer U. The use of semi-automated kinetic perimetry (SPK) to monitor advanced glaucomatous visual field loss. Graefes Arch Clin Expo Ophthalmol. 2008;246:1331–9.CrossRef Nevalainen J, Paetzold J, Krapp E, Vonthein R, Johnson CA, Schiefer U. The use of semi-automated kinetic perimetry (SPK) to monitor advanced glaucomatous visual field loss. Graefes Arch Clin Expo Ophthalmol. 2008;246:1331–9.CrossRef
10.
go back to reference Nowomiejska K, Vonthein R, Paetzold J, Zagorski Z, Kardon R, Schiefer U. Reaction time during semi-automated kinetic perimetry (SPK) in patients with advanced visual field loss. Acta Ophthalmol. 2010;88:65–9.CrossRefPubMed Nowomiejska K, Vonthein R, Paetzold J, Zagorski Z, Kardon R, Schiefer U. Reaction time during semi-automated kinetic perimetry (SPK) in patients with advanced visual field loss. Acta Ophthalmol. 2010;88:65–9.CrossRefPubMed
11.
go back to reference Wilscher S, Wabbels B, Lorenz B. Feasibility and outocome of automated kinetic perimetry in children. Graefes Arch Clin Exp Ophthalmolo. 2010;248:1493–500.CrossRef Wilscher S, Wabbels B, Lorenz B. Feasibility and outocome of automated kinetic perimetry in children. Graefes Arch Clin Exp Ophthalmolo. 2010;248:1493–500.CrossRef
12.
go back to reference Damgaard-Jenson L. Vertical steps in isopters at the hemianopic border in normal and glacomatous eye. Acta Ophthalmol. 1977;55:111–21.CrossRef Damgaard-Jenson L. Vertical steps in isopters at the hemianopic border in normal and glacomatous eye. Acta Ophthalmol. 1977;55:111–21.CrossRef
13.
go back to reference Stewart WC, Shields MB, Ollie AR. Peripheral visual field testing by automated kinetic perimetry in glaucoma. Arch Ophthalmol. 1988;106:202–6.CrossRefPubMed Stewart WC, Shields MB, Ollie AR. Peripheral visual field testing by automated kinetic perimetry in glaucoma. Arch Ophthalmol. 1988;106:202–6.CrossRefPubMed
14.
go back to reference Miller KN, Shields MB, Ollie AR. Automated kinetic perimetry with two peripheral isopters in glaucoma. Arch Ophthalmol. 1989;107:1316–20.CrossRefPubMed Miller KN, Shields MB, Ollie AR. Automated kinetic perimetry with two peripheral isopters in glaucoma. Arch Ophthalmol. 1989;107:1316–20.CrossRefPubMed
15.
go back to reference Gilpin LB, Stewart WC, Shields MB, Miller KN. Hemianopic offsets in the visual field of patients with glaucoma. Graefes Arch Clin Expo Ophthalmol. 1990;228:450–3.CrossRef Gilpin LB, Stewart WC, Shields MB, Miller KN. Hemianopic offsets in the visual field of patients with glaucoma. Graefes Arch Clin Expo Ophthalmol. 1990;228:450–3.CrossRef
16.
go back to reference Portney GL, Krohn MA. Automated perimetry, background, insutruments and methods. Sury Ophthalmol. 1978;22:271–8.CrossRef Portney GL, Krohn MA. Automated perimetry, background, insutruments and methods. Sury Ophthalmol. 1978;22:271–8.CrossRef
17.
go back to reference Heijl A, Drance SM. A clinical comparison of three computerized automatic perimeters in the detection of glaucoma defects. Arch Ophthalmol. 1981;99:832–6.CrossRefPubMed Heijl A, Drance SM. A clinical comparison of three computerized automatic perimeters in the detection of glaucoma defects. Arch Ophthalmol. 1981;99:832–6.CrossRefPubMed
18.
go back to reference Lynn JR, Swanson WH, Fellmann RL. Evaluation of automated kinetic perimetry (AKP) with the Humphrey Field Analyser. Perimetry Update. 1991;1990(1991):433–52. Lynn JR, Swanson WH, Fellmann RL. Evaluation of automated kinetic perimetry (AKP) with the Humphrey Field Analyser. Perimetry Update. 1991;1990(1991):433–52.
19.
go back to reference Ballon BJ, Echelman DA, Shields MB, Ollie AR. Peripheral visual field testing in glaucoma by automated kinetic perimetry with the Humphrey Field Analyzer. Arch Ophthalmol. 1992;110:1730–2.CrossRefPubMed Ballon BJ, Echelman DA, Shields MB, Ollie AR. Peripheral visual field testing in glaucoma by automated kinetic perimetry with the Humphrey Field Analyzer. Arch Ophthalmol. 1992;110:1730–2.CrossRefPubMed
20.
go back to reference Omodaka K, Kunimatsu-Sanuki S, Morin R, Tsuda S, Yokoyama Y, Takahashi H, et al. Development of a new strategy of visual field testing for macular dysfunction in patients with open angle glaucoma. Jpn J Ophthalmol. 2013;57:457–62.CrossRefPubMed Omodaka K, Kunimatsu-Sanuki S, Morin R, Tsuda S, Yokoyama Y, Takahashi H, et al. Development of a new strategy of visual field testing for macular dysfunction in patients with open angle glaucoma. Jpn J Ophthalmol. 2013;57:457–62.CrossRefPubMed
Metadata
Title
Evaluation of kinetic programs in various automated perimeters
Authors
Shigeki Hashimoto
Chota Matsumoto
Mariko Eura
Sachiko Okuyama
Yoshikazu Shimomura
Publication date
01-07-2017
Publisher
Springer Japan
Published in
Japanese Journal of Ophthalmology / Issue 4/2017
Print ISSN: 0021-5155
Electronic ISSN: 1613-2246
DOI
https://doi.org/10.1007/s10384-017-0516-y

Other articles of this Issue 4/2017

Japanese Journal of Ophthalmology 4/2017 Go to the issue