Published in: Odontology 3/2016

01-09-2016 | Original Article

Performance of NiTi endodontic instrument under different temperatures

Authors: Ahmed Jamleh, Yoshio Yahata, Arata Ebihara, Amre R. Atmeh, Turki Bakhsh, Hideaki Suda

Published in: Odontology | Issue 3/2016

Login to get access

Abstract

The purpose of this study was to test nickel titanium (NiTi) instrument performance under different surrounding temperatures. Twenty-four superelastic NiTi instruments with a conical shape comprising a 0.30-mm-diameter tip and 0.06 taper were equally divided into 3 groups according to the temperature employed. Using a specially designed cyclic fatigue testing apparatus, each instrument was deflected to give a curvature 10 mm in radius and a 30° angle. This position was kept as the instrument was immersed in a continuous flow of water under a temperature of 10, 37, or 50 °C for 20 s to calculate the deflecting load (DL). In the same position, the instrument was then allowed to rotate at 300 rpm to fracture, and the working time was converted to the number of cycles to fracture (NCF). The statistical significance was set at p = 0.05. The mean DL (in N) and NCF (in cycles) of the groups at 10, 37, and 50 °C were 10.16 ± 1.36 and 135.50 ± 31.48, 13.50 ± 0.92 and 89.20 ± 16.44, and 14.70 ± 1.21 and 65.50 ± 15.90, respectively. The group at 10 °C had significantly the lowest DL that favorably resulted in the highest NCF. Within the limitations of this study, the surrounding temperature influences the cyclic fatigue resistance and DL of the superelastic NiTi instruments. Lower temperatures are found to favorably decrease the DL and extend the lifetime of the superelastic NiTi instrument. Further NiTi instrument failure studies should be performed under simulated body temperature.
Literature
1.
go back to reference Courtney TH. Mechanical behavior of materials. NY: McGraw Hill; 1990. p. 562. Courtney TH. Mechanical behavior of materials. NY: McGraw Hill; 1990. p. 562.
2.
go back to reference Parashos P, Gordon I, Messer HH. Factors influencing defects of rotary nickel-titanium endodontic instruments after clinical use. J Endod. 2004;30:722–5.CrossRefPubMed Parashos P, Gordon I, Messer HH. Factors influencing defects of rotary nickel-titanium endodontic instruments after clinical use. J Endod. 2004;30:722–5.CrossRefPubMed
3.
go back to reference Pruett JP, Clement DJ, Carnes DL Jr. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod. 1997;23:77–85.CrossRefPubMed Pruett JP, Clement DJ, Carnes DL Jr. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod. 1997;23:77–85.CrossRefPubMed
4.
go back to reference Turpin YL, Chagneau F, Vulcain JM. Impact of two theoretical cross-sections on torsional and bending stresses of nickel-titanium root canal instrument models. J Endod. 2000;26:414–7.CrossRefPubMed Turpin YL, Chagneau F, Vulcain JM. Impact of two theoretical cross-sections on torsional and bending stresses of nickel-titanium root canal instrument models. J Endod. 2000;26:414–7.CrossRefPubMed
5.
go back to reference Zinelis S, Eliades T, Eliades G. A metallurgical characterization of ten endodontic Ni-Ti instruments: assessing the clinical relevance of shape memory and superelastic properties of Ni-Ti endodontic instruments. Int Endod J. 2010;43:125–34.CrossRefPubMed Zinelis S, Eliades T, Eliades G. A metallurgical characterization of ten endodontic Ni-Ti instruments: assessing the clinical relevance of shape memory and superelastic properties of Ni-Ti endodontic instruments. Int Endod J. 2010;43:125–34.CrossRefPubMed
6.
go back to reference Gambarini G. Cyclic fatigue of ProFile rotary instruments after prolonged clinical use. Int Endod J. 2001;34:386–9.CrossRefPubMed Gambarini G. Cyclic fatigue of ProFile rotary instruments after prolonged clinical use. Int Endod J. 2001;34:386–9.CrossRefPubMed
7.
go back to reference Kazemi RB, Stenman E, Spångberg LSW. A comparison of stainless steel and nickel-titanium H-type instruments of identical design: torsional and bending tests. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2000;90:500–6.CrossRefPubMed Kazemi RB, Stenman E, Spångberg LSW. A comparison of stainless steel and nickel-titanium H-type instruments of identical design: torsional and bending tests. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2000;90:500–6.CrossRefPubMed
8.
go back to reference Alapati SB, Brantley WA, Nusstein JM, et al. Vickers hardness investigation of work-hardening in used NiTi rotary instruments. J Endod. 2006;32:1191–3.CrossRefPubMed Alapati SB, Brantley WA, Nusstein JM, et al. Vickers hardness investigation of work-hardening in used NiTi rotary instruments. J Endod. 2006;32:1191–3.CrossRefPubMed
9.
go back to reference Wagner MFX, Nayan N, Ramamurty U. Healing of fatigue damage in NiTi shape memory alloys. J Phys D. 2008;41:185408-1–4.CrossRef Wagner MFX, Nayan N, Ramamurty U. Healing of fatigue damage in NiTi shape memory alloys. J Phys D. 2008;41:185408-1–4.CrossRef
10.
go back to reference Plotino G, Grande NM, Cordaro M, et al. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod. 2009;35:1469–76.CrossRefPubMed Plotino G, Grande NM, Cordaro M, et al. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod. 2009;35:1469–76.CrossRefPubMed
11.
go back to reference Cheung GS, Peng B, Bian Z, et al. Defects in ProTaper S1 instruments after clinical use: fractographic examination. Int Endod J. 2005;38:802–9.CrossRefPubMed Cheung GS, Peng B, Bian Z, et al. Defects in ProTaper S1 instruments after clinical use: fractographic examination. Int Endod J. 2005;38:802–9.CrossRefPubMed
12.
go back to reference Gambarini G, Grande NM, Plotino G, et al. Fatigue resistance of engine-driven rotary nickel-titanium instruments produced by new manufacturing methods. J Endod. 2008;34:1003–5.CrossRefPubMed Gambarini G, Grande NM, Plotino G, et al. Fatigue resistance of engine-driven rotary nickel-titanium instruments produced by new manufacturing methods. J Endod. 2008;34:1003–5.CrossRefPubMed
13.
go back to reference Gao Y, Gutmann JL, Wilkinson K, et al. Evaluation of the impact of raw materials on the fatigue and mechanical properties of ProFile Vortex rotary instruments. J Endod. 2012;38:398–401.CrossRefPubMed Gao Y, Gutmann JL, Wilkinson K, et al. Evaluation of the impact of raw materials on the fatigue and mechanical properties of ProFile Vortex rotary instruments. J Endod. 2012;38:398–401.CrossRefPubMed
14.
go back to reference Shen Y, Coil JM, Zhou HM, et al. ProFile Vortex instruments after clinical use: a metallurgical properties study. J Endod. 2012;38:1613–7.CrossRefPubMed Shen Y, Coil JM, Zhou HM, et al. ProFile Vortex instruments after clinical use: a metallurgical properties study. J Endod. 2012;38:1613–7.CrossRefPubMed
15.
go back to reference Thompson SA. An overview of nickel-titanium alloys used in dentistry. Int Endod J. 2000;33:297–310.CrossRefPubMed Thompson SA. An overview of nickel-titanium alloys used in dentistry. Int Endod J. 2000;33:297–310.CrossRefPubMed
16.
go back to reference Cheung GS, Zhang EW, Zheng YF. A numerical method for predicting the bending fatigue life of NiTi and stainless steel root canal instruments. Int Endod J. 2011;44:357–61.CrossRefPubMed Cheung GS, Zhang EW, Zheng YF. A numerical method for predicting the bending fatigue life of NiTi and stainless steel root canal instruments. Int Endod J. 2011;44:357–61.CrossRefPubMed
17.
go back to reference Yahata Y, Yoneyama T, Hayashi Y, et al. Effect of heat treatment on transformation temperatures and bending properties of nickel-titanium endodontic instruments. Int Endod J. 2009;42:621–6.CrossRefPubMed Yahata Y, Yoneyama T, Hayashi Y, et al. Effect of heat treatment on transformation temperatures and bending properties of nickel-titanium endodontic instruments. Int Endod J. 2009;42:621–6.CrossRefPubMed
18.
go back to reference Miyai K, Ebihara A, Hayashi Y, et al. Influence of phase transformation on the torsional and bending properties of nickel-titanium rotary endodontic instruments. Int Endod J. 2006;39:119–26.CrossRefPubMed Miyai K, Ebihara A, Hayashi Y, et al. Influence of phase transformation on the torsional and bending properties of nickel-titanium rotary endodontic instruments. Int Endod J. 2006;39:119–26.CrossRefPubMed
19.
go back to reference Hayashi Y, Yoneyama T, Yahata Y, et al. Phase transformation behaviour and bending properties of hybrid nickel-titanium rotary endodontic instruments. Int Endod J. 2007;40:247–53.CrossRefPubMed Hayashi Y, Yoneyama T, Yahata Y, et al. Phase transformation behaviour and bending properties of hybrid nickel-titanium rotary endodontic instruments. Int Endod J. 2007;40:247–53.CrossRefPubMed
20.
go back to reference Huang X, Ackland GJ, Rabe KM. Crystal structures and shape-memory behaviour of NiTi. Nat Mater. 2003;2:307–11.CrossRefPubMed Huang X, Ackland GJ, Rabe KM. Crystal structures and shape-memory behaviour of NiTi. Nat Mater. 2003;2:307–11.CrossRefPubMed
21.
go back to reference Parashos P, Messer HH. Rotary NiTi instrument fracture and its consequences. J Endod. 2006;32:1031–43.CrossRefPubMed Parashos P, Messer HH. Rotary NiTi instrument fracture and its consequences. J Endod. 2006;32:1031–43.CrossRefPubMed
22.
go back to reference Kuhn G, Tavernier B, Jordan L. Influence of structure on nickel-titanium endodontic instruments failure. J Endod. 2001;27:516–20.CrossRefPubMed Kuhn G, Tavernier B, Jordan L. Influence of structure on nickel-titanium endodontic instruments failure. J Endod. 2001;27:516–20.CrossRefPubMed
23.
go back to reference Pérez-Higueras J, Arias A, de la Macorra J, et al. Differences in cyclic fatigue resistance between ProTaper Next and ProTaper universal instruments at different levels. J Endod. 2014;40:1477–81.CrossRefPubMed Pérez-Higueras J, Arias A, de la Macorra J, et al. Differences in cyclic fatigue resistance between ProTaper Next and ProTaper universal instruments at different levels. J Endod. 2014;40:1477–81.CrossRefPubMed
24.
go back to reference Elnaghy A. Cyclic fatigue resistance of ProTaper next nickel-titanium rotary files. Int Endod J. 2014;47:1034–9.CrossRefPubMed Elnaghy A. Cyclic fatigue resistance of ProTaper next nickel-titanium rotary files. Int Endod J. 2014;47:1034–9.CrossRefPubMed
25.
go back to reference Plotino G, Grande N, Cotti E, et al. Blue treatment enhances cyclic fatigue resistance of vortex nickel-titanium rotary files. J Endod. 2014;40:1451–3.CrossRefPubMed Plotino G, Grande N, Cotti E, et al. Blue treatment enhances cyclic fatigue resistance of vortex nickel-titanium rotary files. J Endod. 2014;40:1451–3.CrossRefPubMed
26.
go back to reference de Hemptinne F, Slaus G, Vandendael M, Jacquet W, De Moor RJ, Bottenberg P. In vivo intracanal temperature evolution during endodontic treatment after the injection of room temperature or preheated sodium hypochlorite. J Endod. 2015;. doi:10.1016/j.joen.2015.02.011.PubMed de Hemptinne F, Slaus G, Vandendael M, Jacquet W, De Moor RJ, Bottenberg P. In vivo intracanal temperature evolution during endodontic treatment after the injection of room temperature or preheated sodium hypochlorite. J Endod. 2015;. doi:10.​1016/​j.​joen.​2015.​02.​011.PubMed
27.
go back to reference Sirtes G, Waltimo T, Schaetzle M, et al. The effects of temperature on sodium hypochlorite short-term stability, pulp dissolution capacity, and antimicrobial efficacy. J Endod. 2005;31:669–71.CrossRefPubMed Sirtes G, Waltimo T, Schaetzle M, et al. The effects of temperature on sodium hypochlorite short-term stability, pulp dissolution capacity, and antimicrobial efficacy. J Endod. 2005;31:669–71.CrossRefPubMed
28.
go back to reference Jamleh A, Sadr A, Nomura N, et al. Nano-indentation testing of new and fractured nickel-titanium endodontic instruments. Int Endod J. 2012;45:462–8.CrossRefPubMed Jamleh A, Sadr A, Nomura N, et al. Nano-indentation testing of new and fractured nickel-titanium endodontic instruments. Int Endod J. 2012;45:462–8.CrossRefPubMed
29.
go back to reference Jamleh A, Kobayashi C, Yahata Y, et al. Deflecting load of nickel titanium rotary instruments during cyclic fatigue. Dent Mat J. 2012;31:389–93.CrossRef Jamleh A, Kobayashi C, Yahata Y, et al. Deflecting load of nickel titanium rotary instruments during cyclic fatigue. Dent Mat J. 2012;31:389–93.CrossRef
30.
go back to reference Shen Y, Zhou H, Wang Z, et al. Phase transformation behavior and mechanical properties of thermomechanically treated K3XF nickel-titanium instruments. J Endod. 2013;39:919–23.CrossRefPubMed Shen Y, Zhou H, Wang Z, et al. Phase transformation behavior and mechanical properties of thermomechanically treated K3XF nickel-titanium instruments. J Endod. 2013;39:919–23.CrossRefPubMed
31.
go back to reference Miyara K, Yahata Y, Hayashi Y, et al. The influence of heat treatment on the mechanical properties of Ni-Ti file materials. Dent Mat J. 2014;33:27–31.CrossRef Miyara K, Yahata Y, Hayashi Y, et al. The influence of heat treatment on the mechanical properties of Ni-Ti file materials. Dent Mat J. 2014;33:27–31.CrossRef
32.
go back to reference De Azevedo Bahia MG, Fonseca Dias R, Lopes Buono VT. The influence of high amplitude cyclic straining on the behaviour of superelastic Ni-Ti. Int J Fatigue. 2006;28:1087–91.CrossRef De Azevedo Bahia MG, Fonseca Dias R, Lopes Buono VT. The influence of high amplitude cyclic straining on the behaviour of superelastic Ni-Ti. Int J Fatigue. 2006;28:1087–91.CrossRef
33.
go back to reference VilaverdeCorreia S, Nogueira MT, Silva RJC, et al. Phase transformations in NiTi endodontic files and fatigue resistance. ESOMAT. 2009;1:1046–51. VilaverdeCorreia S, Nogueira MT, Silva RJC, et al. Phase transformations in NiTi endodontic files and fatigue resistance. ESOMAT. 2009;1:1046–51.
34.
go back to reference Young JM, Van Vliet KJ. Predicting in vivo failure of pseudoelastic NiTi devices under low cycle, high amplitude fatigue. J Biomed Mater Res B Appl Biomater. 2005;72:17–26.CrossRefPubMed Young JM, Van Vliet KJ. Predicting in vivo failure of pseudoelastic NiTi devices under low cycle, high amplitude fatigue. J Biomed Mater Res B Appl Biomater. 2005;72:17–26.CrossRefPubMed
35.
go back to reference Peters O, Boessler C, Zehnder M. Effect of liquid and paste-type lubricants on torque values during simulated rotary root canal instrumentation. Int Endod J. 2005;38:223–9.CrossRefPubMed Peters O, Boessler C, Zehnder M. Effect of liquid and paste-type lubricants on torque values during simulated rotary root canal instrumentation. Int Endod J. 2005;38:223–9.CrossRefPubMed
36.
go back to reference Tobushi H, Yamada S, Hachisuka T, et al. Thermomechanical properties due to martensitic and R-phase transformations of TiNi shape memory alloy subjected to cyclic loadings. Smart Mater Struct. 1996;5:788–95.CrossRef Tobushi H, Yamada S, Hachisuka T, et al. Thermomechanical properties due to martensitic and R-phase transformations of TiNi shape memory alloy subjected to cyclic loadings. Smart Mater Struct. 1996;5:788–95.CrossRef
37.
go back to reference McKelvey AL, Ritchie RO. On the temperature dependence of the superelastic strength and the prediction of the theoretical uniaxial transformation strain in Nitinol. Philo Mag. 2000;80:1759–68.CrossRef McKelvey AL, Ritchie RO. On the temperature dependence of the superelastic strength and the prediction of the theoretical uniaxial transformation strain in Nitinol. Philo Mag. 2000;80:1759–68.CrossRef
38.
go back to reference Siqueira JF Jr, Alves FR, Almeida BM, de Oliveira JC, Rôças IN. Ability of chemomechanical preparation with either rotary instruments or self-adjusting file to disinfect oval-shaped root canals. J Endod. 2010;36:1860–5.CrossRefPubMed Siqueira JF Jr, Alves FR, Almeida BM, de Oliveira JC, Rôças IN. Ability of chemomechanical preparation with either rotary instruments or self-adjusting file to disinfect oval-shaped root canals. J Endod. 2010;36:1860–5.CrossRefPubMed
Metadata
Title
Performance of NiTi endodontic instrument under different temperatures
Authors
Ahmed Jamleh
Yoshio Yahata
Arata Ebihara
Amre R. Atmeh
Turki Bakhsh
Hideaki Suda
Publication date
01-09-2016
Publisher
Springer Japan
Published in
Odontology / Issue 3/2016
Print ISSN: 1618-1247
Electronic ISSN: 1618-1255
DOI
https://doi.org/10.1007/s10266-015-0214-5

Other articles of this Issue 3/2016

Odontology 3/2016 Go to the issue