01-07-2014 | Original Article

Canal shaping with WaveOne reciprocating files: influence of operator experience on instrument breakage and canal preparation time

Authors: Luigi Generali, Elena Righi, Maria Vittoria Todesca, Ugo Consolo

Published in: Odontology | Issue 2/2014

Login to get access

Abstract

The aim of this study was to determine the fracture incidence of WaveOne Primary reciprocating files and the time required for shaping of curved canals based on the experience of the operator. A total of 109 mesiobuccal canals of permanent molars extracted with an angle of curvature of >25°–45° according to Schneider were randomly assigned to four groups. An experienced operator (endodontist) and an inexperienced operator (student) each shaped one of two groups: one with the instrument WaveOne Primary to WL and the other after creation of a glide path with PathFile 1, 2 and 3 at the WL. Any fractures or visible deformations of the instruments during the shaping phase and the effective time required to prepare the canals for each instrument were recorded. No visible deformation or fracture was observed. The experienced operator tended to finish their shape faster than the inexperienced operator regardless of the technique applied. For the inexperienced operator, the usage time with only WaveOne Primary was significantly lower when the canals were preliminarily instrumented with the PathFile than when these instruments were not used (average time, 22.03 vs. 36.22 s, respectively; p < 0.001). The experience of the operator did not influence fracture of the WaveOne Primary instruments. The time required to prepare the canals was instead inversely proportional to the experience of the operator. However, the creation of a glide path with PathFile instruments reduced the time required by the inexperienced operator to prepare the canal.
Literature
1.
go back to reference Berutti E, Cantatore G, Castellucci A, Chiandussi G, Pera F, Migliaretti G, Pasqualini D. Use of nickel titanium rotary PathFile to create the glide path: comparison with manual preflaring in simulated root canals. J Endod. 2009;35:408–12.CrossRef Berutti E, Cantatore G, Castellucci A, Chiandussi G, Pera F, Migliaretti G, Pasqualini D. Use of nickel titanium rotary PathFile to create the glide path: comparison with manual preflaring in simulated root canals. J Endod. 2009;35:408–12.CrossRef
2.
go back to reference Pasqualini D, Bianchi CC, Paolino DS, Mancini L, Cemenasco A, Cantatore G, Castellucci A, Berutti E. Computed micro-tomographic evaluation of glide path with nickel-titanium rotary PathFile in maxillary first molars curved canals. J Endod. 2012;38:389–93.CrossRef Pasqualini D, Bianchi CC, Paolino DS, Mancini L, Cemenasco A, Cantatore G, Castellucci A, Berutti E. Computed micro-tomographic evaluation of glide path with nickel-titanium rotary PathFile in maxillary first molars curved canals. J Endod. 2012;38:389–93.CrossRef
3.
go back to reference Glossen CR, Haller RH, Dove SB, del Rio CE. A comparison of root canal preparations using NiTi hand, NiTi engine-driven, and K-Flex endodontic instruments. J Endod. 1995;25:434–40. Glossen CR, Haller RH, Dove SB, del Rio CE. A comparison of root canal preparations using NiTi hand, NiTi engine-driven, and K-Flex endodontic instruments. J Endod. 1995;25:434–40.
4.
go back to reference Setzer FC, Kwon TK, Karabucak B. Comparison of apical transportation between two rotary file systems and two hybrid rotary instrumentation sequences. J Endod. 2010;36:1991–4.CrossRef Setzer FC, Kwon TK, Karabucak B. Comparison of apical transportation between two rotary file systems and two hybrid rotary instrumentation sequences. J Endod. 2010;36:1991–4.CrossRef
5.
go back to reference Cheung GSP, Liu CSY. A retrospective study of endodontic treatment outcomes between nickel-titanium rotary and stainless steel hand filing techniques. J Endod. 2009;35:938–43.CrossRef Cheung GSP, Liu CSY. A retrospective study of endodontic treatment outcomes between nickel-titanium rotary and stainless steel hand filing techniques. J Endod. 2009;35:938–43.CrossRef
6.
go back to reference Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod. 2000;26:161–5.CrossRef Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod. 2000;26:161–5.CrossRef
7.
go back to reference Parashos P, Messer HH. Rotary NiTi instruments fracture and its conseguences. J Endod. 2006;32:1031–43.CrossRef Parashos P, Messer HH. Rotary NiTi instruments fracture and its conseguences. J Endod. 2006;32:1031–43.CrossRef
8.
go back to reference Jamleh A, Kobayashi C, Yahata Y, Ebihara A, Suda H. Deflecting load of nickel titanium rotary instruments during cyclic fatigue. Dent Mater J. 2012;31:389–93.CrossRef Jamleh A, Kobayashi C, Yahata Y, Ebihara A, Suda H. Deflecting load of nickel titanium rotary instruments during cyclic fatigue. Dent Mater J. 2012;31:389–93.CrossRef
9.
go back to reference Martin B, Zelada G, Varela P, Bahillo JG, Magan F, Ahn S, Rodriguez C. Factors influencing the fracture of nickel-titanium rotary instruments. Int Endod J. 2003;36:262–6. Martin B, Zelada G, Varela P, Bahillo JG, Magan F, Ahn S, Rodriguez C. Factors influencing the fracture of nickel-titanium rotary instruments. Int Endod J. 2003;36:262–6.
10.
go back to reference Cheung GSP, Peng B, Bian Z, Shen Y, Darvell BW. Defects in ProTaper S1 instruments after clinical use: fractographic examination. Int Endod J. 2005;38:802–9. Cheung GSP, Peng B, Bian Z, Shen Y, Darvell BW. Defects in ProTaper S1 instruments after clinical use: fractographic examination. Int Endod J. 2005;38:802–9.
11.
go back to reference Kramkowski TR, Bahcall J. An in vitro comparison of torsional stress and cyclic fatigue resistance of ProFile GT and ProFile GT series X rotary nickel-titanium files. J Endod. 2009;35:404–7.CrossRef Kramkowski TR, Bahcall J. An in vitro comparison of torsional stress and cyclic fatigue resistance of ProFile GT and ProFile GT series X rotary nickel-titanium files. J Endod. 2009;35:404–7.CrossRef
12.
go back to reference Blum JY, Machtou P, Ruddle C, Micallef JP. Analysis of mechanical preparations in extracted teeth using ProTaper rotary instruments: value of the safety quotient. J Endod. 2003;29:567–75.CrossRef Blum JY, Machtou P, Ruddle C, Micallef JP. Analysis of mechanical preparations in extracted teeth using ProTaper rotary instruments: value of the safety quotient. J Endod. 2003;29:567–75.CrossRef
13.
go back to reference Patino PV, Biedma BM, Liebana CR, Cantatore G, Bahilo JG. The influence of a manual glide path on the separation rate of NiTi rotary instruments. J Endod. 2005;31:114–6.CrossRef Patino PV, Biedma BM, Liebana CR, Cantatore G, Bahilo JG. The influence of a manual glide path on the separation rate of NiTi rotary instruments. J Endod. 2005;31:114–6.CrossRef
14.
go back to reference Berutti E, Paolino DS, Chiandussi G, Alovisi M, Cantatore G, Castellucci A, Pasqualini D. Root canal anatomy preservation of WaveOne reciprocating files with or without glide path. J Endod. 2012;38:101–4.CrossRef Berutti E, Paolino DS, Chiandussi G, Alovisi M, Cantatore G, Castellucci A, Pasqualini D. Root canal anatomy preservation of WaveOne reciprocating files with or without glide path. J Endod. 2012;38:101–4.CrossRef
15.
go back to reference Berutti E, Negro AR, Lendini M, Pasqualini D. Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments. J Endod. 2004;30:228–30.CrossRef Berutti E, Negro AR, Lendini M, Pasqualini D. Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments. J Endod. 2004;30:228–30.CrossRef
16.
go back to reference Wan J, Rasimick BJ, Musikant BL, Deutsch AS. A comparison of cyclic fatigue resistance in reciprocating and rotary nickel-titanium instruments. Aust Endod J. 2011;37:122–7.CrossRef Wan J, Rasimick BJ, Musikant BL, Deutsch AS. A comparison of cyclic fatigue resistance in reciprocating and rotary nickel-titanium instruments. Aust Endod J. 2011;37:122–7.CrossRef
17.
go back to reference Gambarini G, Gergi R, Naaman A, Osta N, Al Sudani D. Cyclic fatigue analysis of twisted file rotary NiTi instruments used in reciprocating motion. Int Endod J. 2012;45:802–6. Gambarini G, Gergi R, Naaman A, Osta N, Al Sudani D. Cyclic fatigue analysis of twisted file rotary NiTi instruments used in reciprocating motion. Int Endod J. 2012;45:802–6.
18.
go back to reference You SY, Bae KS, Baek SH, Kum KY, Shon WJ, Lee W. Lifespan of one nickel-titanium rotary file with reciprocating motion in curved root canals. J Endod. 2010;36:1991–4.CrossRef You SY, Bae KS, Baek SH, Kum KY, Shon WJ, Lee W. Lifespan of one nickel-titanium rotary file with reciprocating motion in curved root canals. J Endod. 2010;36:1991–4.CrossRef
19.
go back to reference Schäfer E, Tepel J. Relationship between design features of endodontic instruments and their properties. Part 3 Resistance to bending and fracture. J Endod. 2001;27:299–303.CrossRef Schäfer E, Tepel J. Relationship between design features of endodontic instruments and their properties. Part 3 Resistance to bending and fracture. J Endod. 2001;27:299–303.CrossRef
20.
go back to reference Berutti E, Chiandussi G, Gaviglio I, Ibba A. Comparative analysis of torsional an bending stresses in two mathematical models of nickel-titanium rotary instruments: ProTaper versus ProFile. J Endod. 2003;29:15–9.CrossRef Berutti E, Chiandussi G, Gaviglio I, Ibba A. Comparative analysis of torsional an bending stresses in two mathematical models of nickel-titanium rotary instruments: ProTaper versus ProFile. J Endod. 2003;29:15–9.CrossRef
21.
go back to reference Mandel E, Adib-Yazdi M, Benhamou LM, Lachkar T, Mesgouez C, Sobel M. Rotary Ni-Ti profile systems for preparing curved canals in resin blocks: influence of operator on instrument breakage. Int Endod J. 1999;32:436–43. Mandel E, Adib-Yazdi M, Benhamou LM, Lachkar T, Mesgouez C, Sobel M. Rotary Ni-Ti profile systems for preparing curved canals in resin blocks: influence of operator on instrument breakage. Int Endod J. 1999;32:436–43.
22.
go back to reference Yared GM, Bou Dagher FE, Machtou P. Influence of rotational speed, torque and operator proficiency on ProFile failures. Int Endod J. 2001;34:47–53. Yared GM, Bou Dagher FE, Machtou P. Influence of rotational speed, torque and operator proficiency on ProFile failures. Int Endod J. 2001;34:47–53.
23.
go back to reference Yared GM, Bou Dagher FE, Machtou P, Kulkarni GK. Influence of rotational speed, torque and operator proficiency on failure of Greater Taper files. Int Endod J. 2002;35:7–12. Yared GM, Bou Dagher FE, Machtou P, Kulkarni GK. Influence of rotational speed, torque and operator proficiency on failure of Greater Taper files. Int Endod J. 2002;35:7–12.
24.
go back to reference Ye J, Gao Y. Metallurgical characterization of M-Wire nickel-titanium shape memory alloy used for endodontic rotary instruments during low-cycle fatigue. J Endod. 2012;38:105–7.CrossRef Ye J, Gao Y. Metallurgical characterization of M-Wire nickel-titanium shape memory alloy used for endodontic rotary instruments during low-cycle fatigue. J Endod. 2012;38:105–7.CrossRef
25.
go back to reference Webber J, Machtou P, Pertot W, Kuttler S, Ruddle C, West J. The WaveOne single-file reciprocating system. Roots. 2011;1:28–33. Webber J, Machtou P, Pertot W, Kuttler S, Ruddle C, West J. The WaveOne single-file reciprocating system. Roots. 2011;1:28–33.
26.
go back to reference Berutti E, Chiandussi G, Paolino DS, Scotti N, Cantatore G, Castellucci A, Pasqualini D. Effect of canal lenght and curvature on working lenght alteration with WaveOne reciprocating file. J Endod. 2011;37:1687–90.CrossRef Berutti E, Chiandussi G, Paolino DS, Scotti N, Cantatore G, Castellucci A, Pasqualini D. Effect of canal lenght and curvature on working lenght alteration with WaveOne reciprocating file. J Endod. 2011;37:1687–90.CrossRef
27.
go back to reference Schneider SW. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol. 1971;32:271–5.CrossRef Schneider SW. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol. 1971;32:271–5.CrossRef
28.
go back to reference Roane JB, Sabala CL, Duncanson MG Jr. The “balanced force” concept for instrumentation of curved canals. J Endod. 1985;11:203–11.CrossRef Roane JB, Sabala CL, Duncanson MG Jr. The “balanced force” concept for instrumentation of curved canals. J Endod. 1985;11:203–11.CrossRef
29.
go back to reference Varela-Patino P, Ibanez-Parraga A, Rivas-Mundina B, Cantatore G, Otero XL, Martin-Biedma B. Alternating versus continuous rotation: a comparative study of the effect on instrument life. J Endod. 2010;36:157–9.CrossRef Varela-Patino P, Ibanez-Parraga A, Rivas-Mundina B, Cantatore G, Otero XL, Martin-Biedma B. Alternating versus continuous rotation: a comparative study of the effect on instrument life. J Endod. 2010;36:157–9.CrossRef
30.
go back to reference Gavini G, Caldeira CL, Akisue E, Candeiro GT, Kawakami DA. Resistance to flexural fatigue of Reciproc R25 files under continuous rotation and reciprocating motion. J Endod. 2012;38:684–7.CrossRef Gavini G, Caldeira CL, Akisue E, Candeiro GT, Kawakami DA. Resistance to flexural fatigue of Reciproc R25 files under continuous rotation and reciprocating motion. J Endod. 2012;38:684–7.CrossRef
31.
go back to reference Johnson E, Lloyd A, Kuttler S, Namerow K. Comparison between a novel nickel-titanium alloy and 508 nitinol on the cyclic fatigue life of ProFile 25/.04 rotary instruments. J Endod. 2008;34:1406–9.CrossRef Johnson E, Lloyd A, Kuttler S, Namerow K. Comparison between a novel nickel-titanium alloy and 508 nitinol on the cyclic fatigue life of ProFile 25/.04 rotary instruments. J Endod. 2008;34:1406–9.CrossRef
32.
go back to reference Bouska J, Justman B, Williamson A, DeLong C, Qian F. Resistance to cyclic fatigue of a new endodontic rotary file. J Endod. 2012;38:667–9.CrossRef Bouska J, Justman B, Williamson A, DeLong C, Qian F. Resistance to cyclic fatigue of a new endodontic rotary file. J Endod. 2012;38:667–9.CrossRef
33.
go back to reference Yared G. Canal preparation using only one Ni-Ti rotary instruments: preliminary observations. Int Endod J. 2008;41:339–44. Yared G. Canal preparation using only one Ni-Ti rotary instruments: preliminary observations. Int Endod J. 2008;41:339–44.
34.
go back to reference Bürklein S, Hinschitza K, Dammaschke T, Schäfer E. Shaping ability and cleaning effectiveness of two single-file systems in severely curved root canals of extracted teeth: reciproc and WaveOne versus Mtwo and ProTaper. Int Endod J. 2012;45:449–61. Bürklein S, Hinschitza K, Dammaschke T, Schäfer E. Shaping ability and cleaning effectiveness of two single-file systems in severely curved root canals of extracted teeth: reciproc and WaveOne versus Mtwo and ProTaper. Int Endod J. 2012;45:449–61.
35.
go back to reference Baumann MA. WaveOne—First experiences of third-year students. Roots. 2012;2:20–3. Baumann MA. WaveOne—First experiences of third-year students. Roots. 2012;2:20–3.
36.
go back to reference Plotino G, Grande NM, Testarelli L, Gambarini G. Cyclic fatigue of Reciproc and WaveOne reciprocating instruments. Int Endod J. 2012;45:614–8. Plotino G, Grande NM, Testarelli L, Gambarini G. Cyclic fatigue of Reciproc and WaveOne reciprocating instruments. Int Endod J. 2012;45:614–8.
37.
go back to reference Mesgouez C, Rilliard F, Matossian L, Nassiri K, Mandel E. Influence of operator experience on canal preparation time when using the rotary Ni-Ti ProFile system in simulated curved canals. Int Endod J. 2003;36:161–5. Mesgouez C, Rilliard F, Matossian L, Nassiri K, Mandel E. Influence of operator experience on canal preparation time when using the rotary Ni-Ti ProFile system in simulated curved canals. Int Endod J. 2003;36:161–5.
38.
go back to reference Bürklein S, Schäfer E. Apically extruded debris with reciprocating single-file and full-sequence rotary instrumentation systems. J Endod. 2012;38:850–2.CrossRef Bürklein S, Schäfer E. Apically extruded debris with reciprocating single-file and full-sequence rotary instrumentation systems. J Endod. 2012;38:850–2.CrossRef
39.
go back to reference Berutti E, Chiandussi G, Paolino DS, Scotti N, Cantatore G, Castellucci A, Pasqualini D. Canal shaping with WaveOne primary reciprocating files and ProTaper system: a comparative study. J Endod. 2012;38:505–9.CrossRef Berutti E, Chiandussi G, Paolino DS, Scotti N, Cantatore G, Castellucci A, Pasqualini D. Canal shaping with WaveOne primary reciprocating files and ProTaper system: a comparative study. J Endod. 2012;38:505–9.CrossRef
Metadata
Title
Canal shaping with WaveOne reciprocating files: influence of operator experience on instrument breakage and canal preparation time
Authors
Luigi Generali
Elena Righi
Maria Vittoria Todesca
Ugo Consolo
Publication date
01-07-2014
Publisher
Springer Japan
Published in
Odontology / Issue 2/2014
Print ISSN: 1618-1247
Electronic ISSN: 1618-1255
DOI
https://doi.org/10.1007/s10266-013-0118-1

Other articles of this Issue 2/2014

Odontology 2/2014 Go to the issue