Rapid-onset plasma leakage of extracorporeal oxygenation membranes possibly due to hyperbilirubinemia

Authors: Yoshiko Kida, Shinichiro Ohshimo, Michihito Kyo, Yuko Tanabe, Kei Suzuki, Koji Hosokawa, Nobuaki Shime

Published in: Journal of Artificial Organs | Issue 4/2018

Login to get access

Abstract

Extracorporeal membrane oxygenation (ECMO) is an emerging tool for supporting cardiopulmonary function in patients with cardiorespiratory failure or arrest. The oxygenator of the ECMO circuit requires effective oxygenation and removal of carbon dioxide from the blood. Major problems that can occur with the oxygenator include plasma leakage, one of the late-onset serious complications necessitating device replacement. However, the rapid onset of plasma leakage is rare. We present a 1-year-old boy with acute respiratory failure due to Pneumocystis and Aspergillus pneumonia. He presented with tachypnea, tachycardia, and hypoxemia despite the ventilatory support, and was therefore placed on venoarterial ECMO with a drainage catheter from the right internal jugular vein (12 Fr) and a return catheter to the right internal carotid artery (10 Fr). Extracorporeal circulation was initiated at a blood flow of 1 L/min (145 mL/kg/min) and a sweep gas flow of 1 L/min with FiO2 of 0.7. Although he was successfully weaned from the venoarterial ECMO on day 15 with an improvement of cardiopulmonary function, he was later placed on venoarterial ECMO again because of the progression of pulmonary hypertension. Laboratory tests showed increased concentrations of hepatic enzymes and hyperbilirubinemia (total bilirubin 31.6 mg/dL). Six hours after starting ECMO circulation, plasma leakage from the oxygenator occurred. Although we replaced the oxygenator with a new one, the replacement showed plasma leakage after 6 h. Disassembly of the oxygenator revealed congestion from bilirubin in the membrane fibers. We described a case of repeated, rapid-onset plasma leakage after implementation of ECMO. Hyperbilirubinemia was likely associated with the plasma leakage of this patient.
Literature
1.
go back to reference Park TK, Yang JH, Jeon K, et al. Extracorporeal membrane oxygenation for refractory septic shock in adults. Eur J Cardiothorac Surg. 2015;47:e68–74.CrossRef Park TK, Yang JH, Jeon K, et al. Extracorporeal membrane oxygenation for refractory septic shock in adults. Eur J Cardiothorac Surg. 2015;47:e68–74.CrossRef
2.
go back to reference Australia and New Zealand Extracorporeal Membrane Oxygenation (ANZ ECMO) Influenza Investigators, Davies A, Jones D, et al. Extracorporeal membrane oxygenation for 2009 influenza A(H1N1) acute respiratory distress syndrome. JAMA 2009;302:1888–95.CrossRef Australia and New Zealand Extracorporeal Membrane Oxygenation (ANZ ECMO) Influenza Investigators, Davies A, Jones D, et al. Extracorporeal membrane oxygenation for 2009 influenza A(H1N1) acute respiratory distress syndrome. JAMA 2009;302:1888–95.CrossRef
3.
go back to reference Posluszny J, Rycus PT, Bartlett RH, et al. Outcome of adult respiratory failure patients receiving prolonged (≥ 14 days) ECMO. Ann Surg. 2016;263:573–81.CrossRef Posluszny J, Rycus PT, Bartlett RH, et al. Outcome of adult respiratory failure patients receiving prolonged (≥ 14 days) ECMO. Ann Surg. 2016;263:573–81.CrossRef
4.
go back to reference Kind K, Reiss N, Knobl HJ, et al. Introduction of a new oxygenator including a tight fiber for long-term ECMO in infants. ASAIO J. 2006;52:217–8.CrossRef Kind K, Reiss N, Knobl HJ, et al. Introduction of a new oxygenator including a tight fiber for long-term ECMO in infants. ASAIO J. 2006;52:217–8.CrossRef
5.
go back to reference Tatsumi E, Taenaka Y, Nakatani, et al. A VAD and novel high performance compact oxygenator for long-term ECMO with local anticoagulation. Trans Am Soc Artif Intern Organs. 1990;36:M480–3. Tatsumi E, Taenaka Y, Nakatani, et al. A VAD and novel high performance compact oxygenator for long-term ECMO with local anticoagulation. Trans Am Soc Artif Intern Organs. 1990;36:M480–3.
6.
go back to reference Thiara AP, Hoel TN, Kristiansen F, et al. Evaluation of oxygenators and centrifugal pumps for long-term pediatric extracorporeal membrane oxygenation. Perfusion. 2007;22:323–6.CrossRef Thiara AP, Hoel TN, Kristiansen F, et al. Evaluation of oxygenators and centrifugal pumps for long-term pediatric extracorporeal membrane oxygenation. Perfusion. 2007;22:323–6.CrossRef
7.
go back to reference Akasu H, Anazawa T. Development of a membrane oxygenator using novel polyolefin hollow fibers with blind-ended micropores. Jpn J Biomater. 1990;8:141–7. Akasu H, Anazawa T. Development of a membrane oxygenator using novel polyolefin hollow fibers with blind-ended micropores. Jpn J Biomater. 1990;8:141–7.
8.
go back to reference Meyns B, Vercaemst L, Vandezande E, et al. Plasma leakage of oxygenators in ECMO depends on the type of oxygenator and on patient variables. Int J Artif Organs. 2005;28:30–4.CrossRef Meyns B, Vercaemst L, Vandezande E, et al. Plasma leakage of oxygenators in ECMO depends on the type of oxygenator and on patient variables. Int J Artif Organs. 2005;28:30–4.CrossRef
9.
go back to reference Montoya JP, Shanley CJ, Merz SI, et al. Plasma leakage through microporous membranes. Role of phospholipids. ASAIO J. 1992;38:M399–405.CrossRef Montoya JP, Shanley CJ, Merz SI, et al. Plasma leakage through microporous membranes. Role of phospholipids. ASAIO J. 1992;38:M399–405.CrossRef
10.
go back to reference Eash HJ, Jones HM, Hattler BG, et al. Evaluation of plasma resistant hollow fiber membranes for artificial lungs. ASAIO J. 2004;50:491–7.CrossRef Eash HJ, Jones HM, Hattler BG, et al. Evaluation of plasma resistant hollow fiber membranes for artificial lungs. ASAIO J. 2004;50:491–7.CrossRef
11.
go back to reference Toomasian JM, Schreiner RJ, Meyer DE, et al. A polymethylpentene fiber gas exchanger for long-term extracorporeal life support. ASAIO J. 2005;51:390–7.CrossRef Toomasian JM, Schreiner RJ, Meyer DE, et al. A polymethylpentene fiber gas exchanger for long-term extracorporeal life support. ASAIO J. 2005;51:390–7.CrossRef
12.
go back to reference Jacobsen J. Binding of bilirubin to human serum albumin—determination of the dissociation constants. FEBS Lett. 1969;5:112–4.CrossRef Jacobsen J. Binding of bilirubin to human serum albumin—determination of the dissociation constants. FEBS Lett. 1969;5:112–4.CrossRef
13.
go back to reference Murrsy RK. Porphyrin and bile pigments. 24 ed Harper’s biochemistry. London: Appleton & Lange; 1996. Murrsy RK. Porphyrin and bile pigments. 24 ed Harper’s biochemistry. London: Appleton & Lange; 1996.
14.
go back to reference Wildschut ED, Ahsman MJ, Allegaert K, et al. Determinants of drug absorption in different ECMO circuits. Intensive Care Med. 2010;36:2109–16.CrossRef Wildschut ED, Ahsman MJ, Allegaert K, et al. Determinants of drug absorption in different ECMO circuits. Intensive Care Med. 2010;36:2109–16.CrossRef
Metadata
Title
Rapid-onset plasma leakage of extracorporeal oxygenation membranes possibly due to hyperbilirubinemia
Authors
Yoshiko Kida
Shinichiro Ohshimo
Michihito Kyo
Yuko Tanabe
Kei Suzuki
Koji Hosokawa
Nobuaki Shime
Publication date
01-12-2018
Publisher
Springer Japan
Published in
Journal of Artificial Organs / Issue 4/2018
Print ISSN: 1434-7229
Electronic ISSN: 1619-0904
DOI
https://doi.org/10.1007/s10047-018-1051-4

Other articles of this Issue 4/2018

Journal of Artificial Organs 4/2018 Go to the issue