Published in: Journal of Artificial Organs 4/2015

01-12-2015 | Original Article

Preclinical study of a novel hydrodynamically levitated centrifugal pump for long-term cardiopulmonary support

In vivo performance during percutaneous cardiopulmonary support

Authors: Tomonori Tsukiya, Toshihide Mizuno, Yoshiaki Takewa, Eisuke Tatsumi, Yoshiyuki Taenaka

Published in: Journal of Artificial Organs | Issue 4/2015

Login to get access

Abstract

An extracorporeal centrifugal blood pump with a hydrodynamically levitated impeller was developed for use in a durable extracorporeal membrane oxygenation (ECMO) system. The present study examined the biocompatibility of the blood pump during long-term use by conducting a series of 30-day chronic animal experiments. The ECMO system was used to produce a percutaneous venoarterial bypass between the venae cavae and carotid artery in adult goats. No anticoagulation or antiplatelet therapy was administered during the experiments. Three out of four animals survived for the scheduled 30-day period, and the blood pumps and membrane oxygenators both exhibited sufficient hydrodynamic performance and good antithrombogenicity, while one animal died of massive bleeding from the outflow cannulation site. The animals’ plasma free hemoglobin had returned to within the normal range by 1 week after the surgical intervention, and their hemodynamic and biochemistry parameters remained within their normal ranges throughout the experiment. The explanted centrifugal blood pumps did not display any trace of thrombus formation. Based on the biocompatibility demonstrated in this study, the examined centrifugal blood pump, which includes a hydrodynamically levitated impeller, is suitable for use in durable ECMO systems.
Literature
1.
go back to reference Paden ML, Conrad SA, Rycus PT, Thiagarajan RR; ELSO Registry. Extracorporeal life support organization registry report 2012. ASAIO J. 2013;59:202–10.CrossRefPubMed Paden ML, Conrad SA, Rycus PT, Thiagarajan RR; ELSO Registry. Extracorporeal life support organization registry report 2012. ASAIO J. 2013;59:202–10.CrossRefPubMed
2.
go back to reference Peek GJ, Mugford M, Tiruvoipati R, Wilson A, Allen E, Thalanany MM, Hibbert CL, Truesdale A, Clemens F, Cooper N, Firmin RK, Elbourne D. Efficacy and economic assessment of conventional ventilatory support versus extracorporeal membrane oxygenation for severe adult respiratory failure (CESAR): a multicentre randomised controlled trial. Lancet. 2009;374:1351–63.CrossRefPubMed Peek GJ, Mugford M, Tiruvoipati R, Wilson A, Allen E, Thalanany MM, Hibbert CL, Truesdale A, Clemens F, Cooper N, Firmin RK, Elbourne D. Efficacy and economic assessment of conventional ventilatory support versus extracorporeal membrane oxygenation for severe adult respiratory failure (CESAR): a multicentre randomised controlled trial. Lancet. 2009;374:1351–63.CrossRefPubMed
3.
go back to reference Fuehner T, Kuehn C, Hadem J, Wiesner O, Gottlieb J, Tudorache I, Olsson KM, Greer M, Sommer W, Welte T, Haverich A, Hoeper MM, Warnecke G. Extracorporeal membrane oxygenation in awake patients as bridge to lung transplantation. Am J Respir Crit Care Med. 2012;185:763–8.CrossRefPubMed Fuehner T, Kuehn C, Hadem J, Wiesner O, Gottlieb J, Tudorache I, Olsson KM, Greer M, Sommer W, Welte T, Haverich A, Hoeper MM, Warnecke G. Extracorporeal membrane oxygenation in awake patients as bridge to lung transplantation. Am J Respir Crit Care Med. 2012;185:763–8.CrossRefPubMed
4.
go back to reference Peek GJ, Killer HM, Reeves R, Sosnowski AW, Firmin RK. Early experience with a polymethyl pentene oxygenator for adult extracorporeal life support. ASAIO J. 2002;48:480–2.CrossRefPubMed Peek GJ, Killer HM, Reeves R, Sosnowski AW, Firmin RK. Early experience with a polymethyl pentene oxygenator for adult extracorporeal life support. ASAIO J. 2002;48:480–2.CrossRefPubMed
5.
go back to reference Kawakami H, Mori Y, Takagi J, Nagaoka S, Kanamori T, Shinbo T, Kubota S. Development of a novel polyimide hollow fiber for an intravascular oxygenator. ASAIO J. 1997;43:M490–4.CrossRefPubMed Kawakami H, Mori Y, Takagi J, Nagaoka S, Kanamori T, Shinbo T, Kubota S. Development of a novel polyimide hollow fiber for an intravascular oxygenator. ASAIO J. 1997;43:M490–4.CrossRefPubMed
6.
go back to reference Wendel HP, Ziemer G. Coating-techniques to improve the hemocompatibility of artificial devices used for extracorporeal circulation. Eur J Cardiothorac Surg. 1999;16:342–50.CrossRefPubMed Wendel HP, Ziemer G. Coating-techniques to improve the hemocompatibility of artificial devices used for extracorporeal circulation. Eur J Cardiothorac Surg. 1999;16:342–50.CrossRefPubMed
7.
go back to reference Zimmermann AK, Aebert H, Freitag M, Husseini M, Ziemer G, Wendel HP. Hemocompatibility of PMEA coated oxygenators used for extracorporeal circulation procedures. ASAIO J. 2004;50:193–9.CrossRefPubMed Zimmermann AK, Aebert H, Freitag M, Husseini M, Ziemer G, Wendel HP. Hemocompatibility of PMEA coated oxygenators used for extracorporeal circulation procedures. ASAIO J. 2004;50:193–9.CrossRefPubMed
8.
go back to reference Eya K, Tatsumi E, Taenaka Y, Takewa Y, Wakisaka Y, Toda K, Nakatani T, Masuzawa T, Baba Y, Miyazaki K, Nishimura T, Ohno T, Ahn JM, Takano H. Development of a membrane oxygenator for long-term respiratory support and its experimental evaluation in prolonged ECMO. ASAIO J. 1996;42:M832–6.CrossRefPubMed Eya K, Tatsumi E, Taenaka Y, Takewa Y, Wakisaka Y, Toda K, Nakatani T, Masuzawa T, Baba Y, Miyazaki K, Nishimura T, Ohno T, Ahn JM, Takano H. Development of a membrane oxygenator for long-term respiratory support and its experimental evaluation in prolonged ECMO. ASAIO J. 1996;42:M832–6.CrossRefPubMed
9.
go back to reference Nishinaka T, Tatsumi E, Taenaka Y, Katagiri N, Ohnishi H, Shioya K, Fukuda T, Oshikawa M, Sato K, Tsukiya T, Homma A, Takewa Y, Takano H, Sato M, Kashiwabara S, Tanaka H, Sakai K, Matsuda T. At least 34 days of animal continuous perfusion by a newly developed extracorporeal membrane oxygenation system without systemic anticoagulants. Artif Organs. 2002;26:548–51.CrossRefPubMed Nishinaka T, Tatsumi E, Taenaka Y, Katagiri N, Ohnishi H, Shioya K, Fukuda T, Oshikawa M, Sato K, Tsukiya T, Homma A, Takewa Y, Takano H, Sato M, Kashiwabara S, Tanaka H, Sakai K, Matsuda T. At least 34 days of animal continuous perfusion by a newly developed extracorporeal membrane oxygenation system without systemic anticoagulants. Artif Organs. 2002;26:548–51.CrossRefPubMed
10.
go back to reference Thiara APS, Hoel TN, Kristiansen F, Karlsen HM, Flane EF, Svennevig JL. Evaluation of oxygenators and centrifugal pumps for long-term pediatric extracorporeal membrane oxygenation. Perfusion. 2007;22:323–6.CrossRefPubMed Thiara APS, Hoel TN, Kristiansen F, Karlsen HM, Flane EF, Svennevig JL. Evaluation of oxygenators and centrifugal pumps for long-term pediatric extracorporeal membrane oxygenation. Perfusion. 2007;22:323–6.CrossRefPubMed
11.
go back to reference Kashiwa K, Nishimura T, Saito A, Kubo H, Fukaya A, Tamai H, Yambe T, Kyo S, Ono M. Left heart bypass support with the Rotaflow Centrifugal Pump® as a bridge to decision and recovery in an adult. J Artif Organs. 2012;15:207–10.CrossRefPubMed Kashiwa K, Nishimura T, Saito A, Kubo H, Fukaya A, Tamai H, Yambe T, Kyo S, Ono M. Left heart bypass support with the Rotaflow Centrifugal Pump® as a bridge to decision and recovery in an adult. J Artif Organs. 2012;15:207–10.CrossRefPubMed
12.
go back to reference John R, Long JW, Massey HT, Griffith BP, Sun BC, Tector AJ, Frazier OH, Joyce LD. Outcomes of a multicenter trial of the Levitronix CentriMag ventricular assist system for short-term circulatory support. J Thorac Cardiovasc Surg. 2011;141:932–9.CrossRefPubMed John R, Long JW, Massey HT, Griffith BP, Sun BC, Tector AJ, Frazier OH, Joyce LD. Outcomes of a multicenter trial of the Levitronix CentriMag ventricular assist system for short-term circulatory support. J Thorac Cardiovasc Surg. 2011;141:932–9.CrossRefPubMed
13.
go back to reference Bhama JK, Kormos RL, Toyoda Y, Teuteberg JJ, Siegenthaler MP. Clinical experience using the Levitronix CentriMag system for temporary right ventricular mechanical circulatory support. J Heart Lung Transplant. 2009;28:971–6.CrossRefPubMed Bhama JK, Kormos RL, Toyoda Y, Teuteberg JJ, Siegenthaler MP. Clinical experience using the Levitronix CentriMag system for temporary right ventricular mechanical circulatory support. J Heart Lung Transplant. 2009;28:971–6.CrossRefPubMed
14.
go back to reference Whitson BA, D’Cunha J, Knutsen AC, Boyle AJ, Liao KK. Levitronix ventricular assist devices as a bridge to recovery after profound biventricular heart failure associated with pulmonary aspergillosis. J Heart Lung Transplant. 2007;26:345–9.CrossRefPubMed Whitson BA, D’Cunha J, Knutsen AC, Boyle AJ, Liao KK. Levitronix ventricular assist devices as a bridge to recovery after profound biventricular heart failure associated with pulmonary aspergillosis. J Heart Lung Transplant. 2007;26:345–9.CrossRefPubMed
15.
go back to reference Aziz TA, Singh G, Popjes E, Mulvey S, Pae W, El-Nanayosy A. Initial experience with CentriMag extracorporeal membrane oxygenation for support of critically ill patients with refractory cardiogenic shock. J Heart Lung Transplant. 2010;29:66–71.CrossRefPubMed Aziz TA, Singh G, Popjes E, Mulvey S, Pae W, El-Nanayosy A. Initial experience with CentriMag extracorporeal membrane oxygenation for support of critically ill patients with refractory cardiogenic shock. J Heart Lung Transplant. 2010;29:66–71.CrossRefPubMed
16.
go back to reference Van der Meer AL, James NL, Esmore DS, Rosenfeldt FL, Begg JD, Woodard J. Initial in vivo experience of the VentrAssist implantable rotary blood pump in sheep. Artif Organs. 2003;27:21–6.CrossRefPubMed Van der Meer AL, James NL, Esmore DS, Rosenfeldt FL, Begg JD, Woodard J. Initial in vivo experience of the VentrAssist implantable rotary blood pump in sheep. Artif Organs. 2003;27:21–6.CrossRefPubMed
17.
go back to reference Ochiai Y, Golding LA, Meaaiello AL, Medvedev AL, Gerhart RL, Chen JF, Takagaki M, Fukamachi K. In vivo hemodynamic performance of the Cleveland Clinic CorAide blood pump in calves. Ann Thorac Surg. 2001;72:747–52.CrossRefPubMed Ochiai Y, Golding LA, Meaaiello AL, Medvedev AL, Gerhart RL, Chen JF, Takagaki M, Fukamachi K. In vivo hemodynamic performance of the Cleveland Clinic CorAide blood pump in calves. Ann Thorac Surg. 2001;72:747–52.CrossRefPubMed
18.
go back to reference Tuzun E, Roberts K, Cohn WE, Sargin M, Gemmato CJ, Radovancevic B, Frazier OH. In vivo evaluation of the HeartWare centrifugal ventricular assist device. Tex Heart Inst J. 2007;34:406–11.PubMedCentralPubMed Tuzun E, Roberts K, Cohn WE, Sargin M, Gemmato CJ, Radovancevic B, Frazier OH. In vivo evaluation of the HeartWare centrifugal ventricular assist device. Tex Heart Inst J. 2007;34:406–11.PubMedCentralPubMed
19.
go back to reference Hoashi T, Kagisaki K, Yamashita K, Tatsumi E, Nishigaki T, Yoshida K, Hayashi T, Ichikawa H. Early clinical outcomes of new pediatric extracorporeal life support system (Endumo® 2000) in neonates and infants. J Artif Organs. 2013;16:267–72.CrossRefPubMed Hoashi T, Kagisaki K, Yamashita K, Tatsumi E, Nishigaki T, Yoshida K, Hayashi T, Ichikawa H. Early clinical outcomes of new pediatric extracorporeal life support system (Endumo® 2000) in neonates and infants. J Artif Organs. 2013;16:267–72.CrossRefPubMed
20.
go back to reference Kusajima K, Hoashi T, Kagisaki K, Yoshida K, Nishigaki T, Hayashi T, Ichikawa H. Clinical experience of more than 2 months usage of extracorporeal membrane oxygenation (Endumo® 4000) without circuit exchange. Artif Organs. 2014;17:99–102.CrossRef Kusajima K, Hoashi T, Kagisaki K, Yoshida K, Nishigaki T, Hayashi T, Ichikawa H. Clinical experience of more than 2 months usage of extracorporeal membrane oxygenation (Endumo® 4000) without circuit exchange. Artif Organs. 2014;17:99–102.CrossRef
Metadata
Title
Preclinical study of a novel hydrodynamically levitated centrifugal pump for long-term cardiopulmonary support
In vivo performance during percutaneous cardiopulmonary support
Authors
Tomonori Tsukiya
Toshihide Mizuno
Yoshiaki Takewa
Eisuke Tatsumi
Yoshiyuki Taenaka
Publication date
01-12-2015
Publisher
Springer Japan
Published in
Journal of Artificial Organs / Issue 4/2015
Print ISSN: 1434-7229
Electronic ISSN: 1619-0904
DOI
https://doi.org/10.1007/s10047-015-0842-0

Other articles of this Issue 4/2015

Journal of Artificial Organs 4/2015 Go to the issue