Published in: Clinical Oral Investigations 11/2019

01-11-2019 | Original Article

Osteotomy speed, heat development, and bone structure influence by various piezoelectric systems—an in vitro study

Authors: Nicolas Noetzel, Tim Fienitz, Matthias Kreppel, Matthias Zirk, Ali- Farid Safi, Daniel Rothamel

Published in: Clinical Oral Investigations | Issue 11/2019

Login to get access

Abstract

Objectives

The aim of this in vitro study was to evaluate osteotomy speed, heat development, and bone structure influence from osteotomies performed by various piezoelectric devices and insert tips. These devices and tips were compared among each other with regard to conventional rotatory and oscillating systems with special focus on the insert tip design and thickness.

Material and methods

The osteotomies were conducted on porcine ribs utilizing 12 different insert tips (straight and angulated) and three conventional systems. After time and temperature measurements, histological analysis was carried out. Light microscopy was used to evaluate the roughness of the osteotomic surface and to search for indications of thermal bone necrosis. A special software analyzing tool was employed to determine cutting width (mm) and debris (%).

Results

All piezoelectric tips created smooth cuts. Cutting widths in general were wider than the actual insert tip size with a tendency for narrow straight insert tips producing relatively wide osteotomies, whereas narrow angulated inserts produced relatively small osteotomies. None of the samples demonstrated distinct indication of necrosis. Overall, there was only a small amount of debris in all osteotomy gaps. Conventional rotatory saws were faster and created less heat compared to all tested piezoelectric systems. Straight tips proved faster osteotomy speed than angulated tips. Thin insert tips indicated to have a positive correlation to osteotomy time and performed faster than conventional microsaw. The average temperature rise was lower when using conventional systems, but critical exceeding temperatures were only observed in short-time exceptional cases. In general, temperature rise was less when using angulated inserts.

Conclusion

All tested tips are appropriate for bone surgery. Only small differences were found among the piezoelectric insert tips. Although conventional rotatory systems in general performed faster osteotomies, special designed and thin piezoelectric insert tips seem to have a positive influence on osteotomy speed. Ultimately, none of the tested devices or inserts combined all best features of speed, heat development, bone structure influence, and safety.

Clinical relevance

Narrow and straight piezoelectric insert tips demonstrated reduced osteotomy times. Nevertheless, a combination of conventional and piezoelectric systems in clinical practice might be the best way to work time-efficient, patient-oriented, and safe. The choice of instrument should be based on clinical experience of the user and should be evaluated individually depending on the case.
Literature
3.
go back to reference Yaman Z, Suer BT (2013) Piezoelectric surgery in oral and maxillofacial surgery. Annals of Oral & Maxillofacial Surgery 1(1):5CrossRef Yaman Z, Suer BT (2013) Piezoelectric surgery in oral and maxillofacial surgery. Annals of Oral & Maxillofacial Surgery 1(1):5CrossRef
5.
go back to reference Vercellotti T (2000) Piezoelectric surgery in implantology: a case report—a new piezoelectric ridge expansion technique. Int J Periodontics Restorative Dent 20(4):358–365PubMed Vercellotti T (2000) Piezoelectric surgery in implantology: a case report—a new piezoelectric ridge expansion technique. Int J Periodontics Restorative Dent 20(4):358–365PubMed
6.
go back to reference Vercellotti T, De Paoli S, Nevins M (2001) The piezoelectric bony window osteotomy and sinus membrane elevation: introduction of a new technique for simplification of the sinus augmentation procedure. Int J Periodontics Restorative Dent 21(6):561–567PubMed Vercellotti T, De Paoli S, Nevins M (2001) The piezoelectric bony window osteotomy and sinus membrane elevation: introduction of a new technique for simplification of the sinus augmentation procedure. Int J Periodontics Restorative Dent 21(6):561–567PubMed
7.
go back to reference Torrella F, Pitarch J, Cabanes G, Anitua E (1998) Ultrasonic ostectomy for the surgical approach of the maxillary sinus: a technical note. Int J Oral Maxillofac Implants 13(5):697–700PubMed Torrella F, Pitarch J, Cabanes G, Anitua E (1998) Ultrasonic ostectomy for the surgical approach of the maxillary sinus: a technical note. Int J Oral Maxillofac Implants 13(5):697–700PubMed
9.
go back to reference Vercellotti T (2004) Technological characteristics and clinical indications of piezoelectric bone surgery. Minerva Stomatol 53(5):207–214PubMed Vercellotti T (2004) Technological characteristics and clinical indications of piezoelectric bone surgery. Minerva Stomatol 53(5):207–214PubMed
10.
go back to reference Vercellotti T, Nevins ML, Kim DM, Nevins M, Wada K, Schenk RK, Fiorellini JP (2005) Osseous response following resective therapy with piezosurgery. Int J Periodontics Restorative Dent 25(6):543–549PubMed Vercellotti T, Nevins ML, Kim DM, Nevins M, Wada K, Schenk RK, Fiorellini JP (2005) Osseous response following resective therapy with piezosurgery. Int J Periodontics Restorative Dent 25(6):543–549PubMed
11.
go back to reference Preti G, Martinasso G, Peirone B, Navone R, Manzella C, Muzio G, Russo C, Canuto RA, Schierano G (2007) Cytokines and growth factors involved in the osseointegration of oral titanium implants positioned using piezoelectric bone surgery versus a drill technique: a pilot study in minipigs. J Periodontol 78(4):716–722. https://doi.org/10.1902/jop.2007.060285 CrossRefPubMed Preti G, Martinasso G, Peirone B, Navone R, Manzella C, Muzio G, Russo C, Canuto RA, Schierano G (2007) Cytokines and growth factors involved in the osseointegration of oral titanium implants positioned using piezoelectric bone surgery versus a drill technique: a pilot study in minipigs. J Periodontol 78(4):716–722. https://​doi.​org/​10.​1902/​jop.​2007.​060285 CrossRefPubMed
13.
go back to reference Schlee M (2005) Ultraschallgestützte Chirurgie - Grundlagen und Möglichkeiten. Z Zahnärztl Impl:48–59 Schlee M (2005) Ultraschallgestützte Chirurgie - Grundlagen und Möglichkeiten. Z Zahnärztl Impl:48–59
14.
go back to reference Horton JE, Tarpley TM Jr, Jacoway JR (1981) Clinical applications of ultrasonic instrumentation in the surgical removal of bone. Oral Surg Oral Med Oral Pathol 51(3):236–242CrossRef Horton JE, Tarpley TM Jr, Jacoway JR (1981) Clinical applications of ultrasonic instrumentation in the surgical removal of bone. Oral Surg Oral Med Oral Pathol 51(3):236–242CrossRef
15.
go back to reference Sohn DS, Ahn MR, Lee WH, Yeo DS, Lim SY (2007) Piezoelectric osteotomy for intraoral harvesting of bone blocks. Int J Periodontics Restorative Dent 27(2):127–131PubMed Sohn DS, Ahn MR, Lee WH, Yeo DS, Lim SY (2007) Piezoelectric osteotomy for intraoral harvesting of bone blocks. Int J Periodontics Restorative Dent 27(2):127–131PubMed
18.
go back to reference Lundskog J (1972) Heat and bone tissue. An experimental investigation of the thermal properties of bone and threshold levels for thermal injury. Scand J Plast Reconstr Surg 9:1–80PubMed Lundskog J (1972) Heat and bone tissue. An experimental investigation of the thermal properties of bone and threshold levels for thermal injury. Scand J Plast Reconstr Surg 9:1–80PubMed
19.
go back to reference Eriksson AR, Albrektsson T (1983) Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent 50(1):101–107CrossRef Eriksson AR, Albrektsson T (1983) Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent 50(1):101–107CrossRef
24.
go back to reference Schweiberer L, Dambe LT, Eitel F, Klapp F (1974) [Revascularization of the tibia afterconservative and surgical fracture fixation]. Hefte Unfallheilkd 18–26 Schweiberer L, Dambe LT, Eitel F, Klapp F (1974) [Revascularization of the tibia afterconservative and surgical fracture fixation]. Hefte Unfallheilkd 18–26
25.
go back to reference Simonetti M, Facco G, Barberis F, Signorini G, Capurro M, Rebaudi A, Sammartino G (2014) Bone characteristics following osteotomy surgery: an in vitro SEM study comparing traditional Lindemann drill with sonic and ultrasonic instruments Simonetti M, Facco G, Barberis F, Signorini G, Capurro M, Rebaudi A, Sammartino G (2014) Bone characteristics following osteotomy surgery: an in vitro SEM study comparing traditional Lindemann drill with sonic and ultrasonic instruments
27.
go back to reference Khambay BS, Walmsley AD (2000) Investigations into the use of an ultrasonic chisel to cut bone. Part 2: cutting ability. J Dent 28(1):39–44CrossRef Khambay BS, Walmsley AD (2000) Investigations into the use of an ultrasonic chisel to cut bone. Part 2: cutting ability. J Dent 28(1):39–44CrossRef
28.
31.
go back to reference Khambay BS, Walmsley AD (2000) Investigations into the use of an ultrasonic chisel to cut bone. Part 1: forces applied by clinicians. J Dent 28(1):31–37CrossRef Khambay BS, Walmsley AD (2000) Investigations into the use of an ultrasonic chisel to cut bone. Part 1: forces applied by clinicians. J Dent 28(1):31–37CrossRef
34.
go back to reference Metzger MC, Bormann KH, Schoen R, Gellrich NC, Schmelzeisen R (2006) Inferior alveolar nerve transposition—an in vitro comparison between piezosurgery and conventional bur use. Journal of Oral Implantology 32(1):19–25. https://doi.org/10.1563/1548-1336(2006)32[19:IANTIV]2.0.CO;2 Metzger MC, Bormann KH, Schoen R, Gellrich NC, Schmelzeisen R (2006) Inferior alveolar nerve transposition—an in vitro comparison between piezosurgery and conventional bur use. Journal of Oral Implantology 32(1):19–25. https://​doi.​org/​10.​1563/​1548-1336(2006)32[19:IANTIV]2.0.CO;2
Metadata
Title
Osteotomy speed, heat development, and bone structure influence by various piezoelectric systems—an in vitro study
Authors
Nicolas Noetzel
Tim Fienitz
Matthias Kreppel
Matthias Zirk
Ali- Farid Safi
Daniel Rothamel
Publication date
01-11-2019
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 11/2019
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-019-02838-8

Other articles of this Issue 11/2019

Clinical Oral Investigations 11/2019 Go to the issue