Neck fracture of femoral stems with a sharp slot at the neck: biomechanical analysis

Authors: Kensei Yoshimoto, Yasuharu Nakashima, Akihiro Nakamura, Taro Mawatari, Mitsugu Todo, Daisuke Hara, Yukihide Iwamoto

Published in: Journal of Orthopaedic Science | Issue 5/2015

Login to get access

Abstract

Background

Fracture of the femoral stem in total hip arthroplasty (THA) is a rare complication. We have encountered 2 cases of neck fractures of the femoral stem occurring 9 and 12 years after THA. Morphological and biomechanical analysis were performed to investigate the mechanism of these fractures.

Method

A titanium alloy femoral stem having a slot with sharp corners (R = 0.2 mm) at the neck had been implanted in both cases. Fracture surfaces were examined by use of scanning electron microscopy (SEM). Stress concentration was simulated by using a finite element method (FEM) to compare slots with sharp (R = 0.2 mm) and smooth (R = 2 mm) corners.

Results

Study of the retrieved stems revealed that neck fractures had occurred at the distal end of the slot in both cases. SEM revealed numerous fine fissures extending from the anterolateral edge, striations on the middle of the fracture surface, and dimples on the posteromedial surface, suggesting that the fractures had occurred from the anterolateral aspect toward the posteromedial aspect because of metallic fatigue. FEM analysis showed that mechanical stress was concentrated at the distal and anterolateral corners of the slot. Under 3500-N loading force, the stress at the sharp corner was 556 MPa, which was approximately twofold that at the smooth corner and exceeded the fatigue strength of titanium alloy.

Conclusion

These findings showed that the sharp corner of slot increased stress concentrations at the anterolateral aspect and led to the neck fractures.
Literature
1.
go back to reference Charnley J. Fracture of femoral prostheses in total hip replacement. A clinical study. Clin Orthop Relat Res. 1975;111:105–20.CrossRefPubMed Charnley J. Fracture of femoral prostheses in total hip replacement. A clinical study. Clin Orthop Relat Res. 1975;111:105–20.CrossRefPubMed
2.
go back to reference Martens M, Aernoudt E, de Meester P, Ducheyne P, Mulier JC, de Langh R, Kestelijn P. Factors in the mechanical failure of the femoral component in total hip prosthesis: report of six fatigue fractures of the femoral stem and results of experimental loading tests. Acta Orthop Scand. 1974;45(5):693–710.CrossRefPubMed Martens M, Aernoudt E, de Meester P, Ducheyne P, Mulier JC, de Langh R, Kestelijn P. Factors in the mechanical failure of the femoral component in total hip prosthesis: report of six fatigue fractures of the femoral stem and results of experimental loading tests. Acta Orthop Scand. 1974;45(5):693–710.CrossRefPubMed
3.
go back to reference Vatani N, Comando D, Acuña J, Prieto D, Caviglia H. Faulty design increases the risk of neck fracture in hip prosthesis. Acta Orthop Scand. 2002;73(5):513–7.CrossRefPubMed Vatani N, Comando D, Acuña J, Prieto D, Caviglia H. Faulty design increases the risk of neck fracture in hip prosthesis. Acta Orthop Scand. 2002;73(5):513–7.CrossRefPubMed
4.
go back to reference Grivas TB, Savvidou OD, Psarakis SA, Bernard PF, Triantafyllopoulos G, Kovanis I, Alexandropoulos P. Neck fracture of a cementless forged titanium alloy femoral stem following total hip arthroplasty: a case report and review of the literature. J Med Case Rep. 2007;6(1):174–80.CrossRef Grivas TB, Savvidou OD, Psarakis SA, Bernard PF, Triantafyllopoulos G, Kovanis I, Alexandropoulos P. Neck fracture of a cementless forged titanium alloy femoral stem following total hip arthroplasty: a case report and review of the literature. J Med Case Rep. 2007;6(1):174–80.CrossRef
5.
go back to reference Dangles CJ, Altstetter CJ. Failure of the modular neck in a total hip arthroplasty. J Arthroplasty. 2010 Oct; 25(7):1169.e5-7. Dangles CJ, Altstetter CJ. Failure of the modular neck in a total hip arthroplasty. J Arthroplasty. 2010 Oct; 25(7):1169.e5-7.
6.
go back to reference Wright G, Sporer S, Urban R, Jacobs J. Fracture of a modular femoral neck after total hip arthroplasty: a case report. J Bone Joint Surg Am. 2010;92(6):1518–21.PubMedCentralCrossRefPubMed Wright G, Sporer S, Urban R, Jacobs J. Fracture of a modular femoral neck after total hip arthroplasty: a case report. J Bone Joint Surg Am. 2010;92(6):1518–21.PubMedCentralCrossRefPubMed
7.
go back to reference Nakashima Y, Sato T, Yamamoto T, Motomura G, Ohishi M, Hamai S, Akiyama M, Hirata M, Hara D, Iwamoto Y. Results at a minimum of 10 years of follow-up for AMS and PerFix HA-coated cementless total hip arthroplasty: impact of cross-linked polyethylene on implant longevity. J Orthop Sci. 2013;18(6):962–8.CrossRefPubMed Nakashima Y, Sato T, Yamamoto T, Motomura G, Ohishi M, Hamai S, Akiyama M, Hirata M, Hara D, Iwamoto Y. Results at a minimum of 10 years of follow-up for AMS and PerFix HA-coated cementless total hip arthroplasty: impact of cross-linked polyethylene on implant longevity. J Orthop Sci. 2013;18(6):962–8.CrossRefPubMed
8.
go back to reference Nakashima Y, Hayashi K, Inadome T, Uenoyama K, Hara T, Kanemaru T, Sugioka Y, Noda I. Hydroxyapatite-coating on titanium arc sprayed titanium implants. J Biomed Mater Res. 1997;35(3):287–98.CrossRefPubMed Nakashima Y, Hayashi K, Inadome T, Uenoyama K, Hara T, Kanemaru T, Sugioka Y, Noda I. Hydroxyapatite-coating on titanium arc sprayed titanium implants. J Biomed Mater Res. 1997;35(3):287–98.CrossRefPubMed
9.
go back to reference Duda GN, Schneider E, Chao EY. Internal forces and moments in the femur during walking. J Biomech. 1997;30(9):933–41.CrossRefPubMed Duda GN, Schneider E, Chao EY. Internal forces and moments in the femur during walking. J Biomech. 1997;30(9):933–41.CrossRefPubMed
10.
go back to reference Edwards WB, Gillette JC, Thomas JM, Derrick TR. Internal femoral forces and moments during running: implications for stress fracture development. Clin Biomech (Bristol, Avon). 2008 Dec; 23(10):1269-1278. Edwards WB, Gillette JC, Thomas JM, Derrick TR. Internal femoral forces and moments during running: implications for stress fracture development. Clin Biomech (Bristol, Avon). 2008 Dec; 23(10):1269-1278.
11.
go back to reference Magnissalis EA, Zinelis S, Karachalios T, Hartofilakidis G. Failure analysis of two Ti-alloy hip arthroplasty femoral stems fractured in vivo. J Biomed Mater Res B Appl Biomater. 2003;66(1):299–305.CrossRefPubMed Magnissalis EA, Zinelis S, Karachalios T, Hartofilakidis G. Failure analysis of two Ti-alloy hip arthroplasty femoral stems fractured in vivo. J Biomed Mater Res B Appl Biomater. 2003;66(1):299–305.CrossRefPubMed
12.
go back to reference Gilbert JL, Buckley CA, Jacobs JJ, Bertin KC, Zernich MR. Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems. A failure analysis of two implants. J Bone Joint Surg Am. 1994;76(1):110–5.PubMed Gilbert JL, Buckley CA, Jacobs JJ, Bertin KC, Zernich MR. Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems. A failure analysis of two implants. J Bone Joint Surg Am. 1994;76(1):110–5.PubMed
13.
go back to reference Lee EW, Kim HT. Early fatigue failures of cemented, forged, cobalt chromium femoral stems at the neck shoulder junction: a case report. J Arthroplasty. 2001;16(2):236–8.CrossRefPubMed Lee EW, Kim HT. Early fatigue failures of cemented, forged, cobalt chromium femoral stems at the neck shoulder junction: a case report. J Arthroplasty. 2001;16(2):236–8.CrossRefPubMed
14.
go back to reference Heck DA, Partridge CM, Reuben JD, Lanzer WL, Lewis CG, Keating EM. Prosthetic component failures in hip arthroplasty surgery. J Arthroplasty. 1995;10(5):575–80.CrossRefPubMed Heck DA, Partridge CM, Reuben JD, Lanzer WL, Lewis CG, Keating EM. Prosthetic component failures in hip arthroplasty surgery. J Arthroplasty. 1995;10(5):575–80.CrossRefPubMed
15.
go back to reference Burstein AH, Wright TM. Neck fractures of femoral prostheses. A report of two cases. J Bone Joint Surg Am. 1985;67(3):497–9.PubMed Burstein AH, Wright TM. Neck fractures of femoral prostheses. A report of two cases. J Bone Joint Surg Am. 1985;67(3):497–9.PubMed
16.
go back to reference Rand JA, Chao EY. Femoral implant neck fracture following total hip arthroplasty. A report of three cases. Clin Orthop Relat Res. 1987;221:255–9.PubMed Rand JA, Chao EY. Femoral implant neck fracture following total hip arthroplasty. A report of three cases. Clin Orthop Relat Res. 1987;221:255–9.PubMed
17.
go back to reference Aspenberg P, Kolmert L, Person L, Onnerfalt R. Fracture of hip prostheses due to inadequate welding. Acta Orhop Scand. 1987;58(5):479–82.CrossRef Aspenberg P, Kolmert L, Person L, Onnerfalt R. Fracture of hip prostheses due to inadequate welding. Acta Orhop Scand. 1987;58(5):479–82.CrossRef
18.
go back to reference Jang B, Kanawati A, Brazil D, Bruce W. Laser etching causing fatigue fracture at the neck-shoulder junction of an uncemented femoral stem: A case report. J Orthop. 2013;10(2):95–8.PubMedCentralCrossRefPubMed Jang B, Kanawati A, Brazil D, Bruce W. Laser etching causing fatigue fracture at the neck-shoulder junction of an uncemented femoral stem: A case report. J Orthop. 2013;10(2):95–8.PubMedCentralCrossRefPubMed
19.
go back to reference Collier JP, Mayor MB, Jensen RE, Surprenant VA, Surprenant HP, McNamar JL, Belec L. Mechanisms of failure of modular prostheses. Clin Orthop Relat Res. 1992;285:129–39.PubMed Collier JP, Mayor MB, Jensen RE, Surprenant VA, Surprenant HP, McNamar JL, Belec L. Mechanisms of failure of modular prostheses. Clin Orthop Relat Res. 1992;285:129–39.PubMed
20.
go back to reference Botti TP, Gent J, Martell JM, Manning DW. Trunion Fracture of a Fully Porous-Coated Femoral Stem. Case report. J Arthroplasty. 2005;20(7):943–5.CrossRefPubMed Botti TP, Gent J, Martell JM, Manning DW. Trunion Fracture of a Fully Porous-Coated Femoral Stem. Case report. J Arthroplasty. 2005;20(7):943–5.CrossRefPubMed
21.
go back to reference Morgan-Hough CV, Tavakkolizadeh A, Purkayastha S. Fatigue failure of the femoral component of a cementless total hip arthroplasty. J Arthoplasty. 2004;19(5):658–60.CrossRef Morgan-Hough CV, Tavakkolizadeh A, Purkayastha S. Fatigue failure of the femoral component of a cementless total hip arthroplasty. J Arthoplasty. 2004;19(5):658–60.CrossRef
22.
go back to reference Unnanuntana A, Chen DX, Unnanuntana A, Wright TM. Trunnion fracture of the Anatomic Medullary Locking A Plus femoral component. J Arthroplasty 2011;26(3):504.e13–16. Unnanuntana A, Chen DX, Unnanuntana A, Wright TM. Trunnion fracture of the Anatomic Medullary Locking A Plus femoral component. J Arthroplasty 2011;26(3):504.e13–16.
23.
go back to reference Skendzel JG, Blaha JD, Urquhart AG. Total hip arthroplasty modular neck failure. J Arthroplasty. 2011; 26(2):338.e1–4. Skendzel JG, Blaha JD, Urquhart AG. Total hip arthroplasty modular neck failure. J Arthroplasty. 2011; 26(2):338.e1–4.
24.
go back to reference Sotereanos NG, Sauber TJ, Tupis TT. Modular femoral neck fracture after primary total hip arthroplasty. J Arthroplasty. 2013; 28(1):196.e7–9. Sotereanos NG, Sauber TJ, Tupis TT. Modular femoral neck fracture after primary total hip arthroplasty. J Arthroplasty. 2013; 28(1):196.e7–9.
25.
go back to reference Ellman MB, Levine BR. Fracture of the modular femoral neck component in total hip arthroplasty. J Arthroplasty. 2013 Jan; 28(1):196.e1-5. Ellman MB, Levine BR. Fracture of the modular femoral neck component in total hip arthroplasty. J Arthroplasty. 2013 Jan; 28(1):196.e1-5.
26.
go back to reference Grupp TM, Weik T, Bloemer W, Knaebel HP. Modular titanium alloy neck adapter failures in hip replacement–failure mode analysis and influence of implant material. BMC Musculoskelet Disord. 2010;4(11):3–14.CrossRef Grupp TM, Weik T, Bloemer W, Knaebel HP. Modular titanium alloy neck adapter failures in hip replacement–failure mode analysis and influence of implant material. BMC Musculoskelet Disord. 2010;4(11):3–14.CrossRef
27.
go back to reference Takahashi K, Sato E. Influence of Surface Treatments on Fatigue Strength of Ti6Al4 V Alloy. Material Transactions. 2010;51:694–8.CrossRef Takahashi K, Sato E. Influence of Surface Treatments on Fatigue Strength of Ti6Al4 V Alloy. Material Transactions. 2010;51:694–8.CrossRef
28.
go back to reference Welsch G, Boyer R, Collings EW. Materials Properties Handbook: Titanium Alloys. ASM International: Almere; 1993. p. 553. Welsch G, Boyer R, Collings EW. Materials Properties Handbook: Titanium Alloys. ASM International: Almere; 1993. p. 553.
29.
go back to reference Oshkour AA, Abu Osman NA, Yau YH, Tarlochan F, Abas WA. Design of new generation femoral prostheses using functionally graded materials: a finite element analysis. Proc Inst Mech Eng H. 2013;227(1):3–17.CrossRefPubMed Oshkour AA, Abu Osman NA, Yau YH, Tarlochan F, Abas WA. Design of new generation femoral prostheses using functionally graded materials: a finite element analysis. Proc Inst Mech Eng H. 2013;227(1):3–17.CrossRefPubMed
Metadata
Title
Neck fracture of femoral stems with a sharp slot at the neck: biomechanical analysis
Authors
Kensei Yoshimoto
Yasuharu Nakashima
Akihiro Nakamura
Taro Mawatari
Mitsugu Todo
Daisuke Hara
Yukihide Iwamoto
Publication date
01-09-2015
Publisher
Springer Japan
Published in
Journal of Orthopaedic Science / Issue 5/2015
Print ISSN: 0949-2658
Electronic ISSN: 1436-2023
DOI
https://doi.org/10.1007/s00776-015-0745-1

Other articles of this Issue 5/2015

Journal of Orthopaedic Science 5/2015 Go to the issue