Open Access 01-03-2011 | Experimental research

Minimally invasive percutaneous transpedicular screw fixation: increased accuracy and reduced radiation exposure by means of a novel electromagnetic navigation system

Authors: Ron von Jako, Michael A. Finn, Kenneth S. Yonemura, Ali Araghi, Larry T. Khoo, John A. Carrino, Mick Perez-Cruet

Published in: Acta Neurochirurgica | Issue 3/2011

Login to get access

Abstract

Background

Minimally invasive percutaneous pedicle screw instrumentation methods may increase the need for intraoperative fluoroscopy, resulting in excessive radiation exposure for the patient, surgeon, and support staff. Electromagnetic field (EMF)-based navigation may aid more accurate placement of percutaneous pedicle screws while reducing fluoroscopic exposure. We compared the accuracy, time of insertion, and radiation exposure of EMF with traditional fluoroscopic percutaneous pedicle screw placement.

Methods

Minimally invasive pedicle screw placement in T8 to S1 pedicles of eight fresh-frozen human cadaveric torsos was guided with EMF or standard fluoroscopy. Set-up, insertion, and fluoroscopic times and radiation exposure and accuracy (measured with post-procedural computed tomography) were analyzed in each group.

Results

Sixty-two pedicle screws were placed under fluoroscopic guidance and 60 under EMF guidance. Ideal trajectories were achieved more frequently with EMF over all segments (62.7% vs. 40%; p = 0.01). Greatest EMF accuracy was achieved in the lumbar spine, with significant improvements in both ideal trajectory and reduction of pedicle breaches over fluoroscopically guided placement (64.9% vs. 40%, p = 0.03, and 16.2% vs. 42.5%, p = 0.01, respectively). Fluoroscopy time was reduced 77% with the use of EMF (22 s vs. 5 s per level; p < 0.0001) over all spinal segments. Radiation exposure at the hand and body was reduced 60% (p = 0.058) and 32% (p = 0.073), respectively. Time for insertion did not vary between the two techniques.

Conclusions

Minimally invasive pedicle screw placement with the aid of EMF image guidance reduces fluoroscopy time and increases placement accuracy when compared with traditional fluoroscopic guidance while adding no additional time to the procedure.
Literature
1.
go back to reference Acosta FL Jr, Thompson TL, Campbell S, Weinstein PR, Ames CP (2005) Use of intraoperative isocentric C-arm 3D fluoroscopy for sextant percutaneous pedicle screw placement: case report and review of the literature. Spine J 5:339–343CrossRefPubMed Acosta FL Jr, Thompson TL, Campbell S, Weinstein PR, Ames CP (2005) Use of intraoperative isocentric C-arm 3D fluoroscopy for sextant percutaneous pedicle screw placement: case report and review of the literature. Spine J 5:339–343CrossRefPubMed
2.
go back to reference Assaker R, Reyns N, Vinchon M, Demondion X, Louis E (2001) Transpedicular screw placement: image-guided versus lateral-view fluoroscopy: in vitro simulation. Spine 26:2160–2164CrossRefPubMed Assaker R, Reyns N, Vinchon M, Demondion X, Louis E (2001) Transpedicular screw placement: image-guided versus lateral-view fluoroscopy: in vitro simulation. Spine 26:2160–2164CrossRefPubMed
3.
go back to reference Austin MS, Vaccaro AR, Brislin B, Nachwalter R, Hilibrand AS, Albert TJ (2002) Image-guided spine surgery: a cadaver study comparing conventional open laminoforaminotomy and two image-guided techniques for pedicle screw placement in posterolateral fusion and nonfusion models. Spine 27:2503–2508CrossRefPubMed Austin MS, Vaccaro AR, Brislin B, Nachwalter R, Hilibrand AS, Albert TJ (2002) Image-guided spine surgery: a cadaver study comparing conventional open laminoforaminotomy and two image-guided techniques for pedicle screw placement in posterolateral fusion and nonfusion models. Spine 27:2503–2508CrossRefPubMed
4.
go back to reference Biswas D, Bible JE, Whang PG, Simpson AK, Grauer JN (2008) Sterility of C-arm fluoroscopy during spinal surgery. Spine 33:1913–1917CrossRefPubMed Biswas D, Bible JE, Whang PG, Simpson AK, Grauer JN (2008) Sterility of C-arm fluoroscopy during spinal surgery. Spine 33:1913–1917CrossRefPubMed
5.
go back to reference Brodke DS, Dick JC, Kunz DN, McCabe R, Zdeblick TA (1997) Posterior lumbar interbody fusion. A biomechanical comparison, including a new threaded cage. Spine 22:26–31CrossRefPubMed Brodke DS, Dick JC, Kunz DN, McCabe R, Zdeblick TA (1997) Posterior lumbar interbody fusion. A biomechanical comparison, including a new threaded cage. Spine 22:26–31CrossRefPubMed
6.
go back to reference Choi WW, Green BA, Levi AD (2000) Computer-assisted fluoroscopic targeting system for pedicle screw insertion. Neurosurgery 47:872–878CrossRefPubMed Choi WW, Green BA, Levi AD (2000) Computer-assisted fluoroscopic targeting system for pedicle screw insertion. Neurosurgery 47:872–878CrossRefPubMed
7.
go back to reference Foley KT, Gupta SK (2002) Percutaneous pedicle screw fixation of the lumbar spine: preliminary clinical results. J Neurosurg 97:7–12PubMed Foley KT, Gupta SK (2002) Percutaneous pedicle screw fixation of the lumbar spine: preliminary clinical results. J Neurosurg 97:7–12PubMed
8.
9.
go back to reference Foley KT, Simon DA, Rampersaud YR (2001) Virtual fluoroscopy: computer-assisted fluoroscopic navigation. Spine 26:347–351CrossRefPubMed Foley KT, Simon DA, Rampersaud YR (2001) Virtual fluoroscopy: computer-assisted fluoroscopic navigation. Spine 26:347–351CrossRefPubMed
10.
go back to reference Foley KT, Smith MM (1996) Image-guided spine surgery. Neurosurg Clin N Am 7:171–186PubMed Foley KT, Smith MM (1996) Image-guided spine surgery. Neurosurg Clin N Am 7:171–186PubMed
11.
go back to reference Fraser J, Von Jako R, Carrino J, Härtl R (2008) Electromagnetic navigation in minimally invasive spine surgery: results of a cadaveric study to evaluate percutaneous pedicle screw insertion. SAS J 2:43–47CrossRef Fraser J, Von Jako R, Carrino J, Härtl R (2008) Electromagnetic navigation in minimally invasive spine surgery: results of a cadaveric study to evaluate percutaneous pedicle screw insertion. SAS J 2:43–47CrossRef
12.
go back to reference Gaines RW Jr (2000) The use of pedicle-screw internal fixation for the operative treatment of spinal disorders. J Bone Joint Surg Am 82-A:1458–1476PubMed Gaines RW Jr (2000) The use of pedicle-screw internal fixation for the operative treatment of spinal disorders. J Bone Joint Surg Am 82-A:1458–1476PubMed
13.
go back to reference Hart RA, Hansen BL, Shea M, Hsu F, Anderson GJ (2005) Pedicle screw placement in the thoracic spine: a comparison of image-guided and manual techniques in cadavers. Spine 30:E326–E331CrossRefPubMed Hart RA, Hansen BL, Shea M, Hsu F, Anderson GJ (2005) Pedicle screw placement in the thoracic spine: a comparison of image-guided and manual techniques in cadavers. Spine 30:E326–E331CrossRefPubMed
14.
go back to reference Holly LT, Foley KT (2003) Three-dimensional fluoroscopy-guided percutaneous thoracolumbar pedicle screw placement. Technical note. J Neurosurg 99:324–329PubMed Holly LT, Foley KT (2003) Three-dimensional fluoroscopy-guided percutaneous thoracolumbar pedicle screw placement. Technical note. J Neurosurg 99:324–329PubMed
15.
go back to reference Isaacs RE, Podichetty VK, Santiago P, Sandhu FA, Spears J, Kelly K, Rice L, Fessler RG (2005) Minimally invasive microendoscopy-assisted transforaminal lumbar interbody fusion with instrumentation. J Neurosurg Spine 3:98–105CrossRefPubMed Isaacs RE, Podichetty VK, Santiago P, Sandhu FA, Spears J, Kelly K, Rice L, Fessler RG (2005) Minimally invasive microendoscopy-assisted transforaminal lumbar interbody fusion with instrumentation. J Neurosurg Spine 3:98–105CrossRefPubMed
16.
go back to reference Jang JS, Lee SH (2005) Minimally invasive transforaminal lumbar interbody fusion with ipsilateral pedicle screw and contralateral facet screw fixation. J Neurosurg Spine 3:218–223CrossRefPubMed Jang JS, Lee SH (2005) Minimally invasive transforaminal lumbar interbody fusion with ipsilateral pedicle screw and contralateral facet screw fixation. J Neurosurg Spine 3:218–223CrossRefPubMed
17.
go back to reference Johnson RG, Murphy M, Miller M (1989) Fusions and transfusions. An analysis of blood loss and autologous replacement during lumbar fusions. Spine 14:358–362CrossRefPubMed Johnson RG, Murphy M, Miller M (1989) Fusions and transfusions. An analysis of blood loss and autologous replacement during lumbar fusions. Spine 14:358–362CrossRefPubMed
18.
go back to reference Kawaguchi Y, Matsui H, Tsuji H (1994) Back muscle injury after posterior lumbar spine surgery. Part 2: histologic and histochemical analyses in humans. Spine 19:2598–2602CrossRefPubMed Kawaguchi Y, Matsui H, Tsuji H (1994) Back muscle injury after posterior lumbar spine surgery. Part 2: histologic and histochemical analyses in humans. Spine 19:2598–2602CrossRefPubMed
19.
go back to reference Kawaguchi Y, Matsui H, Tsuji H (1996) Back muscle injury after posterior lumbar spine surgery. A histologic and enzymatic analysis. Spine 21:941–944CrossRefPubMed Kawaguchi Y, Matsui H, Tsuji H (1996) Back muscle injury after posterior lumbar spine surgery. A histologic and enzymatic analysis. Spine 21:941–944CrossRefPubMed
20.
go back to reference Kim DY, Lee SH, Chung SK, Lee HY (2005) Comparison of multifidus muscle atrophy and trunk extension muscle strength: percutaneous versus open pedicle screw fixation. Spine 30:123–129CrossRefPubMed Kim DY, Lee SH, Chung SK, Lee HY (2005) Comparison of multifidus muscle atrophy and trunk extension muscle strength: percutaneous versus open pedicle screw fixation. Spine 30:123–129CrossRefPubMed
21.
go back to reference Kosmopoulos V, Schizas C (2007) Pedicle screw placement accuracy: a meta-analysis. Spine 32:E111–E120CrossRefPubMed Kosmopoulos V, Schizas C (2007) Pedicle screw placement accuracy: a meta-analysis. Spine 32:E111–E120CrossRefPubMed
22.
go back to reference Krag MH, Beynnon BD, Pope MH, Frymoyer JW, Haugh LD, Weaver DL (1986) An internal fixator for posterior application to short segments of the thoracic, lumbar, or lumbosacral spine. Design and testing. Clin Orthop Relat Res 203:75–98PubMed Krag MH, Beynnon BD, Pope MH, Frymoyer JW, Haugh LD, Weaver DL (1986) An internal fixator for posterior application to short segments of the thoracic, lumbar, or lumbosacral spine. Design and testing. Clin Orthop Relat Res 203:75–98PubMed
23.
go back to reference Laine T, Schlenzka D, Makitalo K, Tallroth K, Nolte LP, Visarius H (1997) Improved accuracy of pedicle screw insertion with computer-assisted surgery. A prospective clinical trial of 30 patients. Spine 22:1254–1258CrossRefPubMed Laine T, Schlenzka D, Makitalo K, Tallroth K, Nolte LP, Visarius H (1997) Improved accuracy of pedicle screw insertion with computer-assisted surgery. A prospective clinical trial of 30 patients. Spine 22:1254–1258CrossRefPubMed
24.
go back to reference Little J, Taveras J (1992) Radiology: diagnosis, imaging intervention. JB Lippincott, Philadelphia Little J, Taveras J (1992) Radiology: diagnosis, imaging intervention. JB Lippincott, Philadelphia
25.
go back to reference Maciunas R (1997) Interactive image guided neurosurgery. In: Tindall G, Cooper P, Barrow D (eds) The practice of neurosurgery. Williams & Wilkins, Philadelphia Maciunas R (1997) Interactive image guided neurosurgery. In: Tindall G, Cooper P, Barrow D (eds) The practice of neurosurgery. Williams & Wilkins, Philadelphia
26.
go back to reference Mehlman CT, DiPasquale TG (1997) Radiation exposure to the orthopaedic surgical team during fluoroscopy: “how far away is far enough?” J Orthop Trauma 11:392–398CrossRefPubMed Mehlman CT, DiPasquale TG (1997) Radiation exposure to the orthopaedic surgical team during fluoroscopy: “how far away is far enough?” J Orthop Trauma 11:392–398CrossRefPubMed
27.
go back to reference Quinones-Hinojosa A, Robert Kolen E, Jun P, Rosenberg WS, Weinstein PR (2006) Accuracy over space and time of computer-assisted fluoroscopic navigation in the lumbar spine in vivo. J Spinal Disord Tech 19:109–113CrossRefPubMed Quinones-Hinojosa A, Robert Kolen E, Jun P, Rosenberg WS, Weinstein PR (2006) Accuracy over space and time of computer-assisted fluoroscopic navigation in the lumbar spine in vivo. J Spinal Disord Tech 19:109–113CrossRefPubMed
28.
go back to reference Rampersaud YR, Foley KT, Shen AC, Williams S, Solomito M (2000) Radiation exposure to the spine surgeon during fluoroscopically assisted pedicle screw insertion. Spine 25:2637–2645CrossRefPubMed Rampersaud YR, Foley KT, Shen AC, Williams S, Solomito M (2000) Radiation exposure to the spine surgeon during fluoroscopically assisted pedicle screw insertion. Spine 25:2637–2645CrossRefPubMed
29.
go back to reference Resnick DK (2003) Prospective comparison of virtual fluoroscopy to fluoroscopy and plain radiographs for placement of lumbar pedicle screws. J Spinal Disord Tech 16:254–260PubMed Resnick DK (2003) Prospective comparison of virtual fluoroscopy to fluoroscopy and plain radiographs for placement of lumbar pedicle screws. J Spinal Disord Tech 16:254–260PubMed
30.
go back to reference Ringel F, Stoffel M, Stuer C, Meyer B (2006) Minimally invasive transmuscular pedicle screw fixation of the thoracic and lumbar spine. Neurosurgery 59:ONS361–ONS367CrossRefPubMed Ringel F, Stoffel M, Stuer C, Meyer B (2006) Minimally invasive transmuscular pedicle screw fixation of the thoracic and lumbar spine. Neurosurgery 59:ONS361–ONS367CrossRefPubMed
31.
go back to reference Sasso RC, Cotler HB (1993) Posterior instrumentation and fusion for unstable fractures and fracture-dislocations of the thoracic and lumbar spine. A comparative study of three fixation devices in 70 patients. Spine 18:450–460PubMed Sasso RC, Cotler HB (1993) Posterior instrumentation and fusion for unstable fractures and fracture-dislocations of the thoracic and lumbar spine. A comparative study of three fixation devices in 70 patients. Spine 18:450–460PubMed
32.
go back to reference Schulze CJ, Munzinger E, Weber U (1998) Clinical relevance of accuracy of pedicle screw placement. A computed tomographic-supported analysis. Spine 23:2215–2221CrossRefPubMed Schulze CJ, Munzinger E, Weber U (1998) Clinical relevance of accuracy of pedicle screw placement. A computed tomographic-supported analysis. Spine 23:2215–2221CrossRefPubMed
33.
go back to reference von Jako R (2006) Lumbosacral spine navigation and visualization of an access needle system. US patent no. 20060063998 von Jako R (2006) Lumbosacral spine navigation and visualization of an access needle system. US patent no. 20060063998
34.
go back to reference von Jako R, Cselik Z (2009) Percutaneous laser discectomy with stereotactic computer-assisted surgical navigation. J Lasers Surg Med 41:42–51CrossRef von Jako R, Cselik Z (2009) Percutaneous laser discectomy with stereotactic computer-assisted surgical navigation. J Lasers Surg Med 41:42–51CrossRef
35.
go back to reference von Jako RA, Carrino JA, Yonemura KS, Noda GA, Zhue W, Blaskiewicz D, Rajue M, Groszmann DE, Weber G (2009) Electromagnetic navigation for percutaneous guide-wire insertion: accuracy and efficiency compared to conventional fluoroscopic guidance. Neuroimage 47(Suppl 2):T127–T132CrossRef von Jako RA, Carrino JA, Yonemura KS, Noda GA, Zhue W, Blaskiewicz D, Rajue M, Groszmann DE, Weber G (2009) Electromagnetic navigation for percutaneous guide-wire insertion: accuracy and efficiency compared to conventional fluoroscopic guidance. Neuroimage 47(Suppl 2):T127–T132CrossRef
36.
go back to reference Zindrick MR, Wiltse LL, Widell EH, Thomas JC, Holland WR, Field BT, Spencer CW (1986) A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine. Clin Orthop Relat Res 203:99–112PubMed Zindrick MR, Wiltse LL, Widell EH, Thomas JC, Holland WR, Field BT, Spencer CW (1986) A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine. Clin Orthop Relat Res 203:99–112PubMed
Metadata
Title
Minimally invasive percutaneous transpedicular screw fixation: increased accuracy and reduced radiation exposure by means of a novel electromagnetic navigation system
Authors
Ron von Jako
Michael A. Finn
Kenneth S. Yonemura
Ali Araghi
Larry T. Khoo
John A. Carrino
Mick Perez-Cruet
Publication date
01-03-2011
Publisher
Springer Vienna
Published in
Acta Neurochirurgica / Issue 3/2011
Print ISSN: 0001-6268
Electronic ISSN: 0942-0940
DOI
https://doi.org/10.1007/s00701-010-0882-4

Other articles of this Issue 3/2011

Acta Neurochirurgica 3/2011 Go to the issue