The cement leakage in cement-augmented pedicle screw instrumentation in degenerative lumbosacral diseases: a retrospective analysis of 202 cases and 950 augmented pedicle screws

Authors: Hui-zhi Guo, Yong-chao Tang, Dan-qing Guo, Shun-cong Zhang, Yong-xian Li, Guo-ye Mo, Pei-jie Luo, Ten-peng Zhou, Yan-huai Ma, De Liang, Xiao-bing Jiang

Published in: European Spine Journal | Issue 7/2019

Login to get access

Abstract

Purpose

To evaluate the incidence, type and risk factors of cement leakage (CL) with cement-augmented pedicle screw instrumentation (CAPSI) in degenerative lumbosacral disease.

Methods

Two hundred and two patients using a total of 950 cement-augmented screws were enrolled. CL was classified into three types: type S: leakage via segmental veins; type B: leakage via basivertebral veins; and type I: leakage via pedicle screw instrumentation to paravertebral soft tissue. The age, gender, operation stage (primary or later stage), body mass index, bone mineral density, the number and type of augmented screw, the position of the tip of screw (lateral or internal part of vertebral body), the position of screw (left or right side), the volume of bone cement, location of the augmented vertebra (lumbar or sacrum), the type of CL and complications were recorded. Binary logistic regression correlation was used to analyze risk factors of veins leakage (type S and type B).

Results

The CL was observed in 165 patients (81.68%) and 335 screws (35.26%), leakage types of S, B and I were seen in 255 (76.12%), 77 (22.99%), and 30 (8.96%) of screws, respectively. Besides, double or multiple routes of leakage were seen in 27 screws. Number of augmented screw was a risk factor for vein leakage (OR 0.58; 95% CI 0.44–0.77; P = 0.000). Furthermore, the doses of cement (OR 0.79; 95% CI 0.61–0.99; P = 0.038) and the position of screw (OR 0.39; 95% CI 0.29–0.53; P = 0.000) were identified as risk factors for type S, and the doses of bone cement (OR 0.37; 95% CI 0.25–0.54; P = 0.000) and the position of the tip of screw (OR 0.07; 95% CI 0.04–0.13; P = 0.000) were risk factors for type B.

Conclusions

CAPSI bears a high risk of asymptomatic CL, with a higher rate of leakage into segmental veins and basivertebral veins. As is known, more augmented screws and larger doses of cement are risk factors for veins leakage (type S and type B), while the tip of screw approaching to the midline of the vertebral body is another risk factor to type B. Thus, the CL could be reduced by the amelioration of operative techniques and procedures.

Graphical abstract

These slides can be retrieved under Electronic Supplementary Material.
Appendix
Available only for authorised users
Literature
1.
go back to reference Keorochana G, Pairuchvej S, Trathitephun W, Arirachakaran A, Predeeprompan P, Kongtharvonskul J (2017) Comparative outcomes of cortical screw trajectory fixation and pedicle screw fixation in lumbar spinal fusion: systematic review and meta-analysis. World Neurosurg 102:340–349CrossRefPubMed Keorochana G, Pairuchvej S, Trathitephun W, Arirachakaran A, Predeeprompan P, Kongtharvonskul J (2017) Comparative outcomes of cortical screw trajectory fixation and pedicle screw fixation in lumbar spinal fusion: systematic review and meta-analysis. World Neurosurg 102:340–349CrossRefPubMed
2.
go back to reference Lykissas MG, Jain VV, Nathan ST, Pawar V, Eismann EA, Sturm PF et al (2013) Mid- to long-term outcomes in adolescent idiopathic scoliosis after instrumented posterior spinal fusion: a meta-analysis. Spine (Phila Pa 1976) 38(2):E113–E119CrossRef Lykissas MG, Jain VV, Nathan ST, Pawar V, Eismann EA, Sturm PF et al (2013) Mid- to long-term outcomes in adolescent idiopathic scoliosis after instrumented posterior spinal fusion: a meta-analysis. Spine (Phila Pa 1976) 38(2):E113–E119CrossRef
3.
go back to reference Hoppe S, Keel MJB (2017) Pedicle screw augmentation in osteoporotic spine: indications, limitations and technical aspects. Eur J Trauma Emerg Surg 43(1):3–8CrossRefPubMed Hoppe S, Keel MJB (2017) Pedicle screw augmentation in osteoporotic spine: indications, limitations and technical aspects. Eur J Trauma Emerg Surg 43(1):3–8CrossRefPubMed
4.
go back to reference Halvorson TL, Kelley LA, Thomas KA, Whitecloud TS, Cook SD (1994) Effects of bone mineral density on pedicle screw fixation. Spine (Phila Pa 1976) 19(21):2415–2420CrossRef Halvorson TL, Kelley LA, Thomas KA, Whitecloud TS, Cook SD (1994) Effects of bone mineral density on pedicle screw fixation. Spine (Phila Pa 1976) 19(21):2415–2420CrossRef
5.
go back to reference Ramaswamy R, Evans S, Kosashvili Y (2010) Holding power of variable pitch screws in osteoporotic, osteopenic and normal bone: are all screws created equal. Injury 41(2):179–183CrossRefPubMed Ramaswamy R, Evans S, Kosashvili Y (2010) Holding power of variable pitch screws in osteoporotic, osteopenic and normal bone: are all screws created equal. Injury 41(2):179–183CrossRefPubMed
6.
go back to reference Choma TJ, Pfeiffer FM, Swope RW, Hirner JP (2012) Pedicle screw design and cement augmentation in osteoporotic vertebrae: effects of fenestrations and cement viscosity on fixation and extraction. Spine (Phila Pa 1976) 37(26):E1628–E1632CrossRef Choma TJ, Pfeiffer FM, Swope RW, Hirner JP (2012) Pedicle screw design and cement augmentation in osteoporotic vertebrae: effects of fenestrations and cement viscosity on fixation and extraction. Spine (Phila Pa 1976) 37(26):E1628–E1632CrossRef
7.
go back to reference Chang MC, Kao HC, Ying SH, Liu CL (2013) Polymethylmethacrylate augmentation of cannulated pedicle screws for fixation in osteoporotic spines and comparison of its clinical results and biomechanical characteristics with the needle injection method. J Spinal Disord Tech 26(6):305–315CrossRefPubMed Chang MC, Kao HC, Ying SH, Liu CL (2013) Polymethylmethacrylate augmentation of cannulated pedicle screws for fixation in osteoporotic spines and comparison of its clinical results and biomechanical characteristics with the needle injection method. J Spinal Disord Tech 26(6):305–315CrossRefPubMed
8.
go back to reference Moussazadeh N, Rubin DG, McLaughlin L, Lis E, Bilsky MH, Laufer I (2015) Short-segment percutaneous pedicle screw fixation with cement augmentation for tumor-induced spinal instability. Spine J 15(7):1609–1617CrossRefPubMed Moussazadeh N, Rubin DG, McLaughlin L, Lis E, Bilsky MH, Laufer I (2015) Short-segment percutaneous pedicle screw fixation with cement augmentation for tumor-induced spinal instability. Spine J 15(7):1609–1617CrossRefPubMed
9.
go back to reference Elder BD, Lo SF, Holmes C, Goodwin CR et al (2015) The biomechanics of pedicle screw augmentation with cement. Spine J 15(6):1432–1445CrossRefPubMed Elder BD, Lo SF, Holmes C, Goodwin CR et al (2015) The biomechanics of pedicle screw augmentation with cement. Spine J 15(6):1432–1445CrossRefPubMed
10.
go back to reference Mueller JU, Baldauf J, Marx S, Kirsch M, Schroeder HW, Pillich DT (2016) Cement leakage in pedicle screw augmentation: a prospective analysis of 98 patients and 474 augmented pedicle screws. J Neurosurg Spine 25(1):103–109CrossRefPubMed Mueller JU, Baldauf J, Marx S, Kirsch M, Schroeder HW, Pillich DT (2016) Cement leakage in pedicle screw augmentation: a prospective analysis of 98 patients and 474 augmented pedicle screws. J Neurosurg Spine 25(1):103–109CrossRefPubMed
11.
go back to reference Martín-Fernández M, López-Herradón A, Piñera AR, Tomé-Bermejo F et al (2017) Potential risks of using cement-augmented screws for spinal fusion in patients with low bone quality. Spine J 17(8):1192–1199CrossRefPubMed Martín-Fernández M, López-Herradón A, Piñera AR, Tomé-Bermejo F et al (2017) Potential risks of using cement-augmented screws for spinal fusion in patients with low bone quality. Spine J 17(8):1192–1199CrossRefPubMed
12.
go back to reference Röllinghoff M, Siewe J, Eysel P, Delank KS (2010) Pulmonary cement embolism after augmentation of pedicle screws with bone cement. Acta Orthop Belg 76(2):269–273PubMed Röllinghoff M, Siewe J, Eysel P, Delank KS (2010) Pulmonary cement embolism after augmentation of pedicle screws with bone cement. Acta Orthop Belg 76(2):269–273PubMed
13.
go back to reference Yeom JS, Kim WJ, Choy WS, Lee CK, Chang BS, Kang JW (2003) Leakage of cement in percutaneous transpedicular vertebroplasty for painful osteoporotic compression fractures. J Bone Joint Surg Br 85(1):83–89CrossRefPubMed Yeom JS, Kim WJ, Choy WS, Lee CK, Chang BS, Kang JW (2003) Leakage of cement in percutaneous transpedicular vertebroplasty for painful osteoporotic compression fractures. J Bone Joint Surg Br 85(1):83–89CrossRefPubMed
14.
go back to reference Kerry G, Ruedinger C, Steiner HH (2013) Cement embolism into the venous system after pedicle screw fixation: case report, literature review, and prevention tips. Orthop Rev (Pavia) 5(3):e24CrossRef Kerry G, Ruedinger C, Steiner HH (2013) Cement embolism into the venous system after pedicle screw fixation: case report, literature review, and prevention tips. Orthop Rev (Pavia) 5(3):e24CrossRef
15.
go back to reference Ding J, Zhang Q, Zhu J, Tao W et al (2016) Risk factors for predicting cement leakage following percutaneous vertebroplasty for osteoporotic vertebral compression fractures. Eur Spine J 25(11):3411–3417CrossRefPubMed Ding J, Zhang Q, Zhu J, Tao W et al (2016) Risk factors for predicting cement leakage following percutaneous vertebroplasty for osteoporotic vertebral compression fractures. Eur Spine J 25(11):3411–3417CrossRefPubMed
16.
go back to reference Arabmotlagh M, Rickert M, Lukas A, Rauschmann M, Fleege C (2017) Small cavity creation in the vertebral body reduces the rate of cement leakage during vertebroplasty. J Orthop Res 35(1):154–159CrossRefPubMed Arabmotlagh M, Rickert M, Lukas A, Rauschmann M, Fleege C (2017) Small cavity creation in the vertebral body reduces the rate of cement leakage during vertebroplasty. J Orthop Res 35(1):154–159CrossRefPubMed
17.
go back to reference Zhan Y, Jiang J, Liao H, Tan H, Yang K (2017) Risk factors for cement leakage after vertebroplasty or kyphoplasty: a meta-analysis of published evidence. World Neurosurg 101:633–642CrossRefPubMed Zhan Y, Jiang J, Liao H, Tan H, Yang K (2017) Risk factors for cement leakage after vertebroplasty or kyphoplasty: a meta-analysis of published evidence. World Neurosurg 101:633–642CrossRefPubMed
18.
go back to reference Galbusera F, Volkheimer D, Reitmaier S, Berger-Roscher N, Kienle A, Wilke HJ (2015) Pedicle screw loosening: a clinically relevant complication. Eur Spine J 24(5):1005–1016CrossRefPubMed Galbusera F, Volkheimer D, Reitmaier S, Berger-Roscher N, Kienle A, Wilke HJ (2015) Pedicle screw loosening: a clinically relevant complication. Eur Spine J 24(5):1005–1016CrossRefPubMed
19.
go back to reference Chen YL, Chen WC, Chou CW, Chen JW et al (2014) Biomechanical study of expandable pedicle screw fixation in severe osteoporotic bone comparing with conventional and cement-augmented pedicle screws. Med Eng Phys 36(11):1416–1420CrossRefPubMed Chen YL, Chen WC, Chou CW, Chen JW et al (2014) Biomechanical study of expandable pedicle screw fixation in severe osteoporotic bone comparing with conventional and cement-augmented pedicle screws. Med Eng Phys 36(11):1416–1420CrossRefPubMed
20.
go back to reference Nakashima H, Kawakami N, Tsuji T, Ohara T et al (2015) Adjacent segment disease after posterior lumbar interbody fusion: based on cases with a minimum of 10 years of follow-up. Spine (Phila Pa 1976) 40(14):E831–E841CrossRef Nakashima H, Kawakami N, Tsuji T, Ohara T et al (2015) Adjacent segment disease after posterior lumbar interbody fusion: based on cases with a minimum of 10 years of follow-up. Spine (Phila Pa 1976) 40(14):E831–E841CrossRef
21.
go back to reference Gazzeri R, Roperto R, Fiore C (2016) Surgical treatment of degenerative and traumatic spinal diseases with expandable screws in patients with osteoporosis: 2-year follow-up clinical study. J Neurosurg Spine 25(5):610–619CrossRefPubMed Gazzeri R, Roperto R, Fiore C (2016) Surgical treatment of degenerative and traumatic spinal diseases with expandable screws in patients with osteoporosis: 2-year follow-up clinical study. J Neurosurg Spine 25(5):610–619CrossRefPubMed
22.
go back to reference Bullmann V, Liljenqvist UR, Rödl R, Schulte TL (2010) Pedicle screw augmentation from a biomechanical perspective. Orthopade 39(7):673–678CrossRefPubMed Bullmann V, Liljenqvist UR, Rödl R, Schulte TL (2010) Pedicle screw augmentation from a biomechanical perspective. Orthopade 39(7):673–678CrossRefPubMed
23.
go back to reference Tan JS, Bailey CS, Dvorak MF, Fisher CG, Cripton PA, Oxland TR (2007) Cement augmentation of vertebral screws enhances the interface strength between interbody device and vertebral body. Spine (Phila Pa 1976) 32(3):334–341CrossRef Tan JS, Bailey CS, Dvorak MF, Fisher CG, Cripton PA, Oxland TR (2007) Cement augmentation of vertebral screws enhances the interface strength between interbody device and vertebral body. Spine (Phila Pa 1976) 32(3):334–341CrossRef
24.
go back to reference Janssen I, Ryang YM, Gempt J, Bette S et al (2017) Risk of cement leakage and pulmonary embolism by bone cement-augmented pedicle screw fixation of the thoracolumbar spine. Spine J 17(6):837–844CrossRefPubMed Janssen I, Ryang YM, Gempt J, Bette S et al (2017) Risk of cement leakage and pulmonary embolism by bone cement-augmented pedicle screw fixation of the thoracolumbar spine. Spine J 17(6):837–844CrossRefPubMed
25.
go back to reference Hatzantonis C, Czyz M, Pyzik R, Boszczyk BM (2017) Intracardiac bone cement embolism as a complication of vertebroplasty: management strategy. Eur Spine J 26(12):3199–3205CrossRefPubMed Hatzantonis C, Czyz M, Pyzik R, Boszczyk BM (2017) Intracardiac bone cement embolism as a complication of vertebroplasty: management strategy. Eur Spine J 26(12):3199–3205CrossRefPubMed
26.
go back to reference Audat ZA, Alfawareh MD, Darwish FT, Alomari AA (2016) Intracardiac leakage of cement during kyphoplasty and vertebroplasty: a case report. Am J Case Rep 17:326–330CrossRefPubMedPubMedCentral Audat ZA, Alfawareh MD, Darwish FT, Alomari AA (2016) Intracardiac leakage of cement during kyphoplasty and vertebroplasty: a case report. Am J Case Rep 17:326–330CrossRefPubMedPubMedCentral
27.
go back to reference El SA, Kelm A, Meier S, Sander AL et al (2013) Intraoperative PEEP-ventilation during PMMA-injection for augmented pedicle screws: improvement of leakage rate in spinal surgery. Eur J Trauma Emerg Surg 39(5):461–468CrossRef El SA, Kelm A, Meier S, Sander AL et al (2013) Intraoperative PEEP-ventilation during PMMA-injection for augmented pedicle screws: improvement of leakage rate in spinal surgery. Eur J Trauma Emerg Surg 39(5):461–468CrossRef
28.
go back to reference Zhu SY, Zhong ZM, Wu Q, Chen JT (2016) Risk factors for bone cement leakage in percutaneous vertebroplasty: a retrospective study of four hundred and eighty five patients. Int Orthop 40(6):1205–1210CrossRefPubMed Zhu SY, Zhong ZM, Wu Q, Chen JT (2016) Risk factors for bone cement leakage in percutaneous vertebroplasty: a retrospective study of four hundred and eighty five patients. Int Orthop 40(6):1205–1210CrossRefPubMed
29.
go back to reference Fölsch C, Goost H, Figiel J, Paletta JR, Schultz W, Lakemeier S (2012) Correlation of pull-out strength of cement-augmented pedicle screws with CT-volumetric measurement of cement. Biomed Tech (Berl) 57(6):473–480CrossRef Fölsch C, Goost H, Figiel J, Paletta JR, Schultz W, Lakemeier S (2012) Correlation of pull-out strength of cement-augmented pedicle screws with CT-volumetric measurement of cement. Biomed Tech (Berl) 57(6):473–480CrossRef
30.
go back to reference Bokov A, Mlyavykh S, Aleynik A, Kutlaeva M, Anderson G (2016) The potential impact of venobasillar system morphology and applied technique on epidural cement leakage with percutaneous vertebroplasty. Pain Physician 19(6):357–362PubMed Bokov A, Mlyavykh S, Aleynik A, Kutlaeva M, Anderson G (2016) The potential impact of venobasillar system morphology and applied technique on epidural cement leakage with percutaneous vertebroplasty. Pain Physician 19(6):357–362PubMed
31.
go back to reference Kasó G, Horváth Z, Szenohradszky K, Sándor J, Dóczi T (2008) Comparison of CT characteristics of extravertebral cement leakages after vertebroplasty performed by different navigation and injection techniques. Acta Neurochir (Wien) 150(7):677–683 (discussion 683) CrossRef Kasó G, Horváth Z, Szenohradszky K, Sándor J, Dóczi T (2008) Comparison of CT characteristics of extravertebral cement leakages after vertebroplasty performed by different navigation and injection techniques. Acta Neurochir (Wien) 150(7):677–683 (discussion 683) CrossRef
32.
go back to reference Tajsic T, Patel K, Farmer R, Mannion RJ, Trivedi RA (2018) Spinal navigation for minimally invasive thoracic and lumbosacral spine fixation: implications for radiation exposure, operative time, and accuracy of pedicle screw placement. Eur Spine J 27(8):1918–1924CrossRefPubMed Tajsic T, Patel K, Farmer R, Mannion RJ, Trivedi RA (2018) Spinal navigation for minimally invasive thoracic and lumbosacral spine fixation: implications for radiation exposure, operative time, and accuracy of pedicle screw placement. Eur Spine J 27(8):1918–1924CrossRefPubMed
33.
go back to reference Kafchitsas K, Geiger F, Rauschmann M, Schmidt S (2010) Cement distribution in vertebroplasty pedicle screws with different designs. Orthopade 39(7):679–686CrossRefPubMed Kafchitsas K, Geiger F, Rauschmann M, Schmidt S (2010) Cement distribution in vertebroplasty pedicle screws with different designs. Orthopade 39(7):679–686CrossRefPubMed
34.
go back to reference Loeffel M, Ferguson SJ, Nolte LP, Kowal JH (2008) Vertebroplasty: experimental characterization of polymethylmethacrylate bone cement spreading as a function of viscosity, bone porosity, and flow rate. Spine (Phila Pa 1976) 33(12):1352–1359CrossRef Loeffel M, Ferguson SJ, Nolte LP, Kowal JH (2008) Vertebroplasty: experimental characterization of polymethylmethacrylate bone cement spreading as a function of viscosity, bone porosity, and flow rate. Spine (Phila Pa 1976) 33(12):1352–1359CrossRef
Metadata
Title
The cement leakage in cement-augmented pedicle screw instrumentation in degenerative lumbosacral diseases: a retrospective analysis of 202 cases and 950 augmented pedicle screws
Authors
Hui-zhi Guo
Yong-chao Tang
Dan-qing Guo
Shun-cong Zhang
Yong-xian Li
Guo-ye Mo
Pei-jie Luo
Ten-peng Zhou
Yan-huai Ma
De Liang
Xiao-bing Jiang
Publication date
01-07-2019
Publisher
Springer Berlin Heidelberg
Published in
European Spine Journal / Issue 7/2019
Print ISSN: 0940-6719
Electronic ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-019-05985-4

Other articles of this Issue 7/2019

European Spine Journal 7/2019 Go to the issue

Announcements

Announcement

Editorial

Editorial