Cervical anterior transpedicular screw fixation (ATPS)—Part II. Accuracy of manual insertion and pull-out strength of ATPS

Authors: Heiko Koller, Frank Acosta, Mark Tauber, Michael Fox, Hudelmaier Martin, Rosmarie Forstner, Peter Augat, Rainer Penzkofer, Christian Pirich, H. Kässmann, Herbert Resch, Wolfgang Hitzl

Published in: European Spine Journal | Issue 4/2008

Login to get access

Abstract

Reconstruction after multilevel decompression of the cervical spine, especially in the weakened osteoporotic, neoplastic or infectious spine often requires circumferential stabilization and fusion. To avoid the additional posterior surgery in these cases while increasing rigidity of anterior-only screw-plate constructs, the authors introduce the concept of anterior transpedicular screw (ATPS) fixation. We demonstrated its morphological feasibility as well as its indications in a previous study in Part I of our project. Consequently, the objectives of the current study were to assess the ex vivo accuracy of placing ATPS into the cervical vertebra as well as the biomechanical performance of ATPS in comparison to traditional vertebral body screws (VBS) in terms of pull-out strength (POS). Twenty-three ATPS were inserted alternately to two screws into the pedicles and vertebral bodies, respectively, of six cadaveric specimens from C3–T1. For insertion of ATPS, a manual fluoroscopically assisted technique was used. Pre- and post insertional CT-scans were used to assess accuracy of ATPS insertion in the axial and sagittal planes. A newly designed grading system and accuracy score were used to delineate accuracy of ATPS insertion. Following insertion of screws, 23 ATPS and 22 VBS were subjected to pull-out testing (POT). The bone mineral density (BMD) of each specimen was assessed prior to POT. Statistical analysis showed that the incidence of correctly placed screws and non-critical pedicles breaches in axial plane was 78.3%, and 95.7% in sagittal plane. Hence, according to our definition of “critical” pedicle breach that exposes neurovascular structures at risk, 21.7% (n = 5) of all ATPS inserted showed a critical pedicle breach in axial plane. Notably, no critical pedicle perforation occurred at the C6 to T1 levels. Pull-out testing of ATPS and VBS revealed that pull-out resistance of ATPS was 2.5-fold that of VBS. Mean POS of 23 ATPS with a mean BMD of 0.566 g/cm2 and a mean osseus screw purchase of 27.2 mm was 467.8 N. In comparison, POS of 22 VBS screws with a mean BMD of 0.533 g/cm2 and a mean osseus screw purchase of 16.0 mm was 181.6 N. The difference in ultimate pull-out strength between the ATPS and VBS group was significant (p < 0.000001). Also, accuracy of ATPS placement in axial plane was shown to be significantly correlated with POS. In contrast, there was no correlation between screw-length, BMD, or level of insertion and the POS of ATPS or VBS. The study demonstrated that the use of ATPS might be a new technique worthy of further investigation. The use of ATPS shows the potential to increase construct rigidity in terms of screw-plate pull-out resistance. It might diminish construct failures during anterior-only reconstructions of the highly unstable decompressed cervical spine.
Appendix
Available only for authorised users
Literature
1.
go back to reference Abumi K, Shono Y, Ito M, Taneichi H, Kotani Y, Kaneda K (2000) Complications of pedicle screw fixation in reconstructive surgery of the cervical spine. Spine 25(8):962–969PubMedCrossRef Abumi K, Shono Y, Ito M, Taneichi H, Kotani Y, Kaneda K (2000) Complications of pedicle screw fixation in reconstructive surgery of the cervical spine. Spine 25(8):962–969PubMedCrossRef
2.
go back to reference Kotani Y, Abumi K, Ito M, Minami A (2005) Cervical spine injuries associated with lateral mass and facet joint fractures: new classification and surgical treatment with pedicle fixation. Eur Spine J 14:69–77PubMedCrossRef Kotani Y, Abumi K, Ito M, Minami A (2005) Cervical spine injuries associated with lateral mass and facet joint fractures: new classification and surgical treatment with pedicle fixation. Eur Spine J 14:69–77PubMedCrossRef
3.
go back to reference Holly LT, Foley KT (2006) Percutaneous placement of posterior cervical screws using three-dimensional fluoroscopy. Spine 31:536–540PubMedCrossRef Holly LT, Foley KT (2006) Percutaneous placement of posterior cervical screws using three-dimensional fluoroscopy. Spine 31:536–540PubMedCrossRef
4.
go back to reference Jeanneret B, Gebhard JS, Magerl F (1994) Transpedicular screw fixation of articular mass fracture-separation: results of an anatomical study and operative technique. Spine 19:2529–2539CrossRef Jeanneret B, Gebhard JS, Magerl F (1994) Transpedicular screw fixation of articular mass fracture-separation: results of an anatomical study and operative technique. Spine 19:2529–2539CrossRef
5.
go back to reference Kamimura M, Ebara S, Itoh H, Tateiwa Y, Kinoshita T, Takaoka K (2003) Cervical pedicle screw insertion: assessment of safety and accuracy with computer-assisted image guidance. J Spinal Disord 13:218–224CrossRef Kamimura M, Ebara S, Itoh H, Tateiwa Y, Kinoshita T, Takaoka K (2003) Cervical pedicle screw insertion: assessment of safety and accuracy with computer-assisted image guidance. J Spinal Disord 13:218–224CrossRef
6.
go back to reference Karaikovic EE, Yingsakmongkol W, Gaines RW (2001) Accuracy of cervical pedicle screw placement using the funnel technique. Spine 26:2456–2462PubMedCrossRef Karaikovic EE, Yingsakmongkol W, Gaines RW (2001) Accuracy of cervical pedicle screw placement using the funnel technique. Spine 26:2456–2462PubMedCrossRef
7.
go back to reference Ludwig SC, Kowalski JM, Edwards CC 2nd, Heller JG (2000) Cervical pedicle screws: comparative accuracy of two insertion techniques. Spine 25:2675–2681PubMedCrossRef Ludwig SC, Kowalski JM, Edwards CC 2nd, Heller JG (2000) Cervical pedicle screws: comparative accuracy of two insertion techniques. Spine 25:2675–2681PubMedCrossRef
8.
go back to reference Ludwig SC, Kramer DL, Balderston RA, Vaccaro AR, Foley KF, Albert TJ (2000) Placement of pedicle screws in the human cadaveric cervical spine: comparative accuracy of three techniques. Spine 25:1655–1667PubMedCrossRef Ludwig SC, Kramer DL, Balderston RA, Vaccaro AR, Foley KF, Albert TJ (2000) Placement of pedicle screws in the human cadaveric cervical spine: comparative accuracy of three techniques. Spine 25:1655–1667PubMedCrossRef
9.
go back to reference Mao GP, Zhao JN, Wang YR, Li JS, chen YX, Wu SJ, Bao NR (2005) Design of cervical pedicle locator and three-dimensional location of cervical pedicle. Spine 30:1045–1050PubMedCrossRef Mao GP, Zhao JN, Wang YR, Li JS, chen YX, Wu SJ, Bao NR (2005) Design of cervical pedicle locator and three-dimensional location of cervical pedicle. Spine 30:1045–1050PubMedCrossRef
10.
go back to reference Miller RM, Ebraheim NA, Xu R, Yeasting RA (1996) Anatomic consideration of transpedicular screw placement in the cervical spine: an analysis of two approaches. Spine 21:2317–2322PubMedCrossRef Miller RM, Ebraheim NA, Xu R, Yeasting RA (1996) Anatomic consideration of transpedicular screw placement in the cervical spine: an analysis of two approaches. Spine 21:2317–2322PubMedCrossRef
11.
go back to reference Reinhold M, Magerl F, Rieger M, Blauth M (2006) Cervical pedicle screw placement: feasibility and accuracy of two new insertion techniques based on morphometric data. Eur Spine J 16:47–56PubMedCrossRef Reinhold M, Magerl F, Rieger M, Blauth M (2006) Cervical pedicle screw placement: feasibility and accuracy of two new insertion techniques based on morphometric data. Eur Spine J 16:47–56PubMedCrossRef
12.
go back to reference Richter M, Amiot LP, Puhl W (2002) Computer-assisted surgery in dorsal instrumentation of the cervical spine: an in vitro study. Orthopäde 31:372–377PubMedCrossRef Richter M, Amiot LP, Puhl W (2002) Computer-assisted surgery in dorsal instrumentation of the cervical spine: an in vitro study. Orthopäde 31:372–377PubMedCrossRef
13.
go back to reference Roh JS, Teng AL, Rice JA, Huang RC, Davis JA (2004) Accurate cervical pedicle screw placement using laser-guided fluoroscopy: the ‘perfect pedicle’ technique. 32nd annual meeting. Boston: Cervical Spine Research Society—A, Poster 16:69 Roh JS, Teng AL, Rice JA, Huang RC, Davis JA (2004) Accurate cervical pedicle screw placement using laser-guided fluoroscopy: the ‘perfect pedicle’ technique. 32nd annual meeting. Boston: Cervical Spine Research Society—A, Poster 16:69
14.
go back to reference Ito H, Neo M, Yoshida M, Fujibayashi S, Yoshitomi H, Nakamura T (2006) Efficacy of computer-assisted pedicle screw insertion for cervical instability in RA patients. Rheumatol Int 27:567–574PubMedCrossRef Ito H, Neo M, Yoshida M, Fujibayashi S, Yoshitomi H, Nakamura T (2006) Efficacy of computer-assisted pedicle screw insertion for cervical instability in RA patients. Rheumatol Int 27:567–574PubMedCrossRef
15.
go back to reference Kast E, Mohr K, Richter H-P, Börm W (2006) Complications of transpedicular screw fixation in the cervical spine. Eur Spine J 15:327–334PubMedCrossRef Kast E, Mohr K, Richter H-P, Börm W (2006) Complications of transpedicular screw fixation in the cervical spine. Eur Spine J 15:327–334PubMedCrossRef
16.
go back to reference Kotani Y, Abumi K, Ito M, Minami A (2003) Improved accuracy of computer-asisted cervical pedicle screw insertion. J Neurosurg 99(Spine 3):257–263PubMed Kotani Y, Abumi K, Ito M, Minami A (2003) Improved accuracy of computer-asisted cervical pedicle screw insertion. J Neurosurg 99(Spine 3):257–263PubMed
17.
go back to reference Neo M, Sakamoto T, Fujibayashi S, Nakamura T (2005) The clinical risk of vertebral artery injury from cervical pedicle screws in degenerative vertebrae. Spine 30:2800–2805PubMedCrossRef Neo M, Sakamoto T, Fujibayashi S, Nakamura T (2005) The clinical risk of vertebral artery injury from cervical pedicle screws in degenerative vertebrae. Spine 30:2800–2805PubMedCrossRef
18.
go back to reference Takahashi J, Shono Z, Nakamura I, Hirabayashi H, Kamimura M, Ebara S, Kato H (2006) Computer-assisted screw insertion of cervical disorders in rheumatoid arthritis. Eur Spine J 16:485–494PubMedCrossRef Takahashi J, Shono Z, Nakamura I, Hirabayashi H, Kamimura M, Ebara S, Kato H (2006) Computer-assisted screw insertion of cervical disorders in rheumatoid arthritis. Eur Spine J 16:485–494PubMedCrossRef
19.
go back to reference Yoshimoto H, Sato S, Hyakumachi T, Yanagibashi Y, Masuda T (2005) Spinal reconstruction using a cervical pedicle screw system. Clin Orthop Rel Res 431:111–119 Yoshimoto H, Sato S, Hyakumachi T, Yanagibashi Y, Masuda T (2005) Spinal reconstruction using a cervical pedicle screw system. Clin Orthop Rel Res 431:111–119
20.
go back to reference Yukawa Y, Kato F, Yoshirara H, Yanase M, Ito K (2006) Cervical pedicle screw fixation in 100 cases of unstable cervical injuries: pedicle axis views obtained using fluoroscopy. J Neurosurg 5:488–493 Yukawa Y, Kato F, Yoshirara H, Yanase M, Ito K (2006) Cervical pedicle screw fixation in 100 cases of unstable cervical injuries: pedicle axis views obtained using fluoroscopy. J Neurosurg 5:488–493
21.
go back to reference Koller H, Acosta F, Hempfing A, Scheiter A, Tauber A, Holz U, Resch H, Fox M, Hitzl M (2008). Cervical anterior transpedicular screw fixation. Part I: Study on morphological feasibility, indications, and technical prerequisites. Eur Spine J (in press) [E-pub ahead of print] Koller H, Acosta F, Hempfing A, Scheiter A, Tauber A, Holz U, Resch H, Fox M, Hitzl M (2008). Cervical anterior transpedicular screw fixation. Part I: Study on morphological feasibility, indications, and technical prerequisites. Eur Spine J (in press) [E-pub ahead of print]
22.
go back to reference Hirano T, Hasegawa K,Takahashi H, Uchiyama S, Hara T, Washio T, Sugiura T, Yokaichiya M, Ideka M (1997) Structural characteristics of the pedicle and its role in screw stability. Spine 22:2504–2510PubMedCrossRef Hirano T, Hasegawa K,Takahashi H, Uchiyama S, Hara T, Washio T, Sugiura T, Yokaichiya M, Ideka M (1997) Structural characteristics of the pedicle and its role in screw stability. Spine 22:2504–2510PubMedCrossRef
23.
go back to reference Weinstein JN, Rydevik BL, Rauschnig WR (1992) Anatomical and technical considerations of pedicle screw fixation. Clin Orthop 284:34–36PubMed Weinstein JN, Rydevik BL, Rauschnig WR (1992) Anatomical and technical considerations of pedicle screw fixation. Clin Orthop 284:34–36PubMed
24.
go back to reference Johnston LT, Karaikovic EE, Lautenschlager EP, Marcu D (2006) Cervical pedicle screws vs. lateral mass screws: uniplanar fatigue analysis and residual pullout strengths. Spine J 6:667–672PubMedCrossRef Johnston LT, Karaikovic EE, Lautenschlager EP, Marcu D (2006) Cervical pedicle screws vs. lateral mass screws: uniplanar fatigue analysis and residual pullout strengths. Spine J 6:667–672PubMedCrossRef
25.
go back to reference Jones EL, Heller JG, Silcox DH, Hutton WC (1997) Cervical pedicle screws versus lateral mass screws: anatomic feasibility and biomechanical comparison. Spine 22:977–982PubMedCrossRef Jones EL, Heller JG, Silcox DH, Hutton WC (1997) Cervical pedicle screws versus lateral mass screws: anatomic feasibility and biomechanical comparison. Spine 22:977–982PubMedCrossRef
26.
go back to reference Kotani Y, Cunningham BW, Abumi K, McAfee PC (1994) Biomechanical analysis of cervical stabilization systems. An assessment of transpedicular screw fixation in the cervical spine. Spine 19:2529–2539PubMedCrossRef Kotani Y, Cunningham BW, Abumi K, McAfee PC (1994) Biomechanical analysis of cervical stabilization systems. An assessment of transpedicular screw fixation in the cervical spine. Spine 19:2529–2539PubMedCrossRef
27.
go back to reference Kothe R, Rüter W, Schneider E, Linke B (2004) Biomechanical analysis of transpedicular screw fixation in the subaxial cervical spine. Spine 29:1869–1875PubMedCrossRef Kothe R, Rüter W, Schneider E, Linke B (2004) Biomechanical analysis of transpedicular screw fixation in the subaxial cervical spine. Spine 29:1869–1875PubMedCrossRef
28.
go back to reference KowalskiJ, Ludwig SC, Hutton WC, Heller JG (2000) Cervical spine pedicle screws: a biomechanical comparison of two insertion techniques. Spine 25:2865–2867CrossRef KowalskiJ, Ludwig SC, Hutton WC, Heller JG (2000) Cervical spine pedicle screws: a biomechanical comparison of two insertion techniques. Spine 25:2865–2867CrossRef
29.
go back to reference Krikpatrick JS, Levy JA, Carillo J, Moeini SR (1999) Reconstruction after multilevel corporectomy in the cervical spine. A sagittal plane biomechanical study. Spine 24:1186–1191; Discussion: Traynelis VC: 1191CrossRef Krikpatrick JS, Levy JA, Carillo J, Moeini SR (1999) Reconstruction after multilevel corporectomy in the cervical spine. A sagittal plane biomechanical study. Spine 24:1186–1191; Discussion: Traynelis VC: 1191CrossRef
30.
go back to reference Bozkus H, Ames CP, Chamberlain RH, Nottmeier EW, Sonntag VKH, Papadopoulos SM, Crawford NR (2005) Biomechanical analysis of rigid stabilization techniques for three-column injury in the lower cervical spine. Spine 30:915–922PubMedCrossRef Bozkus H, Ames CP, Chamberlain RH, Nottmeier EW, Sonntag VKH, Papadopoulos SM, Crawford NR (2005) Biomechanical analysis of rigid stabilization techniques for three-column injury in the lower cervical spine. Spine 30:915–922PubMedCrossRef
31.
go back to reference Do Koh Y, Lim TH, Won You J, Eck J, An HS (2001) A biomechanical comparison of modern anterior and posterior plate fixation of the cervical spine. Spine 26:15–21PubMedCrossRef Do Koh Y, Lim TH, Won You J, Eck J, An HS (2001) A biomechanical comparison of modern anterior and posterior plate fixation of the cervical spine. Spine 26:15–21PubMedCrossRef
32.
go back to reference Isomi T, Panjabi MM, Wang JL, Vaccaro AR, Garfin SR, Patel T (1999) Stabilizing potential of anterior cervical plates in multilevel corpectomies. Spine 24:2219–2223PubMedCrossRef Isomi T, Panjabi MM, Wang JL, Vaccaro AR, Garfin SR, Patel T (1999) Stabilizing potential of anterior cervical plates in multilevel corpectomies. Spine 24:2219–2223PubMedCrossRef
33.
go back to reference Isomi T Panjabi MM, Wang JL, et al (1999) Stabilizing potential of anterior cervical plates in multilevel corporectomies. Spine 24:2219–2223CrossRef Isomi T Panjabi MM, Wang JL, et al (1999) Stabilizing potential of anterior cervical plates in multilevel corporectomies. Spine 24:2219–2223CrossRef
34.
go back to reference Kim S-M, Lim TJ, Paterno J, Park BSJ, Kim DH (2004) A biomechanical comparison of three surgical approaches in bilateral cervical facet dislocation. J Neurosurg 1(Spine 1):108–115 Kim S-M, Lim TJ, Paterno J, Park BSJ, Kim DH (2004) A biomechanical comparison of three surgical approaches in bilateral cervical facet dislocation. J Neurosurg 1(Spine 1):108–115
35.
go back to reference Mummaneni PV, Haid RW, Traynelis VC, Sasso RC, Subach BR, Fiore AJ, Rodts GE (2002) Posterior cervical fixation using a new polyaxial screw and rod system: technique and surgical results. Neurosurg Focus 12:Article 8 Mummaneni PV, Haid RW, Traynelis VC, Sasso RC, Subach BR, Fiore AJ, Rodts GE (2002) Posterior cervical fixation using a new polyaxial screw and rod system: technique and surgical results. Neurosurg Focus 12:Article 8
36.
go back to reference Panjabi MM, Isomi T, Wang JL (1999) Loosening at the screw-vertebra junction in multilevel anterior cervical plate constructs. Spine 24:2383–2388PubMedCrossRef Panjabi MM, Isomi T, Wang JL (1999) Loosening at the screw-vertebra junction in multilevel anterior cervical plate constructs. Spine 24:2383–2388PubMedCrossRef
37.
go back to reference Sasso RC, Ruggiero RA Jr, Reilly TM, Hall PV (2003) Early reconstruction failures after multilevel cervical corpectomy. Spine 28:140–142PubMedCrossRef Sasso RC, Ruggiero RA Jr, Reilly TM, Hall PV (2003) Early reconstruction failures after multilevel cervical corpectomy. Spine 28:140–142PubMedCrossRef
38.
go back to reference Schmidt R, Wilke HJ, Claes L, Puhl W, Richter M (2005) Effect of constrained posterior screw and rod systems for primary stability: biomechanical in vitro comparison of various instrumentations in a single-level corporectomy model. Eur Spine J 14:372–380PubMedCrossRef Schmidt R, Wilke HJ, Claes L, Puhl W, Richter M (2005) Effect of constrained posterior screw and rod systems for primary stability: biomechanical in vitro comparison of various instrumentations in a single-level corporectomy model. Eur Spine J 14:372–380PubMedCrossRef
39.
go back to reference Schmidt R, Wilke HJ, Claes L, Puhl W, Richter M (2003) Pedicle screws enhance primary stability in multilevel cervical corporectomies: biomechanical in vitro comparison of different implants including constrained and nonconstrained posterior instruments. Spine 16:1821–1828CrossRef Schmidt R, Wilke HJ, Claes L, Puhl W, Richter M (2003) Pedicle screws enhance primary stability in multilevel cervical corporectomies: biomechanical in vitro comparison of different implants including constrained and nonconstrained posterior instruments. Spine 16:1821–1828CrossRef
40.
go back to reference Schulz KD Jr, McLaughlin MR, Haud RW Jr, et al (2000) Single-stage anterior-posterior decompression and stabilization for complex cervical spine disorders. J Neursurg 93(Spine 2):214–221 Schulz KD Jr, McLaughlin MR, Haud RW Jr, et al (2000) Single-stage anterior-posterior decompression and stabilization for complex cervical spine disorders. J Neursurg 93(Spine 2):214–221
41.
go back to reference Singh K, Vaccaro AR, Kim J, Lorenz EP, Lim TH, An HS (2003) Biomechanical comparison of cervical spine reconstructive techniques after a multilevel corporectomy of the cervical spine. Spine 28:2352–2357PubMedCrossRef Singh K, Vaccaro AR, Kim J, Lorenz EP, Lim TH, An HS (2003) Biomechanical comparison of cervical spine reconstructive techniques after a multilevel corporectomy of the cervical spine. Spine 28:2352–2357PubMedCrossRef
42.
go back to reference Ito Y, Kai N, Haesgawa Y, Nakago K, Toda K, Yagata Y, Shibara M (2005) Cervical or thoracic instrumentation surgery with registration-free 3D-navigation. Eur Spine J 14(Suppl 1):Abstract Ito Y, Kai N, Haesgawa Y, Nakago K, Toda K, Yagata Y, Shibara M (2005) Cervical or thoracic instrumentation surgery with registration-free 3D-navigation. Eur Spine J 14(Suppl 1):Abstract
43.
go back to reference Taneichi H, Suda K, Kajino T, Kaneda K (2005) Traumatically induced vertebral artery occlusion associated with cervical spine injuries: prospective study using magnetic resonance angiography. Spine 30:1955–1962PubMedCrossRef Taneichi H, Suda K, Kajino T, Kaneda K (2005) Traumatically induced vertebral artery occlusion associated with cervical spine injuries: prospective study using magnetic resonance angiography. Spine 30:1955–1962PubMedCrossRef
44.
go back to reference Ludwig SC (2004) Complications of cervical spine pedicle screw placement. In: Clark CR (ed). The cervical spine, 4th edn. Lippincott Williams & Wilkins, Philadelphia Ludwig SC (2004) Complications of cervical spine pedicle screw placement. In: Clark CR (ed). The cervical spine, 4th edn. Lippincott Williams & Wilkins, Philadelphia
45.
go back to reference Zhang QH, Tan SH, Chou SM (2004) Investigation of fixation screw pull-out strength on human spine. J Biomech 37:479–485PubMedCrossRef Zhang QH, Tan SH, Chou SM (2004) Investigation of fixation screw pull-out strength on human spine. J Biomech 37:479–485PubMedCrossRef
46.
go back to reference Liu J, Shafiq Q, Ebrahim NA, Karkare N, Asaad M, Woldenberg L, Yeasting RA (2005) Value of intraoperative true latera lradiograph of C2 pedicle for C1-2 transarticular screw insertion. Spine J 5:434–440PubMedCrossRef Liu J, Shafiq Q, Ebrahim NA, Karkare N, Asaad M, Woldenberg L, Yeasting RA (2005) Value of intraoperative true latera lradiograph of C2 pedicle for C1-2 transarticular screw insertion. Spine J 5:434–440PubMedCrossRef
47.
go back to reference Richter M (2005) Posterior instrumentation of the cervical spine using the Neon Occipito-Cervical System. Part 2: cervical and cervicothoracic instrumentation. Oper Orthop Traumatol 6:579–600CrossRef Richter M (2005) Posterior instrumentation of the cervical spine using the Neon Occipito-Cervical System. Part 2: cervical and cervicothoracic instrumentation. Oper Orthop Traumatol 6:579–600CrossRef
48.
go back to reference Oda I, Abumi K, Ito M, Kotani Y, Oya T, Hasegawa K, Minami A (2006) Palliative spinal reconstruction using cervical pedicle screws for metastatic lesions of the spine. A retrospective analysis of 32 cases. Spine 31:1439–1444PubMedCrossRef Oda I, Abumi K, Ito M, Kotani Y, Oya T, Hasegawa K, Minami A (2006) Palliative spinal reconstruction using cervical pedicle screws for metastatic lesions of the spine. A retrospective analysis of 32 cases. Spine 31:1439–1444PubMedCrossRef
49.
go back to reference Koller H Kammermeier V, Ulbricht D, Assuncao A, Karolus S, v-d Berg B, Holz U (2006) Anterior retropharyngeal fixation C1-2 for stabilization of atlantoaxial instabilities. Study of feasibility, technical description and preliminary results. Eur Spine J 15:1326–1338CrossRef Koller H Kammermeier V, Ulbricht D, Assuncao A, Karolus S, v-d Berg B, Holz U (2006) Anterior retropharyngeal fixation C1-2 for stabilization of atlantoaxial instabilities. Study of feasibility, technical description and preliminary results. Eur Spine J 15:1326–1338CrossRef
51.
go back to reference Kim HK, Heller JG, Hudgins PA, Fountain JA (2003) The accuracy of computed tomography in assessing pedicle screw placement. Spine 28:2441–2446PubMedCrossRef Kim HK, Heller JG, Hudgins PA, Fountain JA (2003) The accuracy of computed tomography in assessing pedicle screw placement. Spine 28:2441–2446PubMedCrossRef
52.
go back to reference Rao G, Brodke D, Rondina M, et al (2002) Comparison of computerized tomography and direct visualization in thoracic pedicle screw placement. J Neurosurgery 97:223–226 Rao G, Brodke D, Rondina M, et al (2002) Comparison of computerized tomography and direct visualization in thoracic pedicle screw placement. J Neurosurgery 97:223–226
53.
go back to reference Langston H, Foley KT (2006) Percutaneous placement of posterior cervical screws using three-dimensional fluoroscopy. Spine 31:536–540CrossRef Langston H, Foley KT (2006) Percutaneous placement of posterior cervical screws using three-dimensional fluoroscopy. Spine 31:536–540CrossRef
54.
go back to reference Datir SPD, Mitra SR (2004) Morphometric study of the thoracic vertebral pedicle in an Indian population. Spine 29:1174–1181PubMedCrossRef Datir SPD, Mitra SR (2004) Morphometric study of the thoracic vertebral pedicle in an Indian population. Spine 29:1174–1181PubMedCrossRef
55.
go back to reference Ugur HC, Attar A, Uz A, Tekdemir I, Egemen N, Genc Y (2001) Thoracic pedicle: surgical anatomic evaluation and relations. J Spinal Disord 14:39–45PubMedCrossRef Ugur HC, Attar A, Uz A, Tekdemir I, Egemen N, Genc Y (2001) Thoracic pedicle: surgical anatomic evaluation and relations. J Spinal Disord 14:39–45PubMedCrossRef
56.
go back to reference Richter M, Amiot LP, Neller S, Kluger P, Puhl W (2000) Computer-assisted surgery in posterior instrumentation of the cervical spine: an in-vitro feasibility study. Eur Spine J 9:S65–70PubMedCrossRef Richter M, Amiot LP, Neller S, Kluger P, Puhl W (2000) Computer-assisted surgery in posterior instrumentation of the cervical spine: an in-vitro feasibility study. Eur Spine J 9:S65–70PubMedCrossRef
57.
go back to reference Sanelli PC, Tong S, Gonzalez G, et al (2002) Normal variation of vertebra artery on CT angiography and its implications for diagnosis of acquired pathology. J Comput Assist Tomogr 26:462–470PubMedCrossRef Sanelli PC, Tong S, Gonzalez G, et al (2002) Normal variation of vertebra artery on CT angiography and its implications for diagnosis of acquired pathology. J Comput Assist Tomogr 26:462–470PubMedCrossRef
58.
go back to reference Ugur HC, Attar A, Uz A, Tekdemir I, Egemen N, Caglar S, Genc Y (2000) Surgical anatomic evaluation of the cervical pedicle and adjacent neural structures. Neurosurgery 47:1162–1168; comments 1168–1169PubMedCrossRef Ugur HC, Attar A, Uz A, Tekdemir I, Egemen N, Caglar S, Genc Y (2000) Surgical anatomic evaluation of the cervical pedicle and adjacent neural structures. Neurosurgery 47:1162–1168; comments 1168–1169PubMedCrossRef
59.
go back to reference Xu R, Kang A, Ebraheim NA, Xoung RA (1999) Anatomic relation between the cervical pedicle and the adjacent neural structures. Spine 24:451–454PubMedCrossRef Xu R, Kang A, Ebraheim NA, Xoung RA (1999) Anatomic relation between the cervical pedicle and the adjacent neural structures. Spine 24:451–454PubMedCrossRef
60.
go back to reference Heller JG, Silcox DH, Sutterlin CE (1995) Complications of posterior cervical plating. Spine 20:2442–2448PubMedCrossRef Heller JG, Silcox DH, Sutterlin CE (1995) Complications of posterior cervical plating. Spine 20:2442–2448PubMedCrossRef
61.
go back to reference Conrad B, Cordista A, Horodyski MB, Rechtine G (2005) Biomechanical evaluation of the pullout strength of cervical screws. J Spinal Disord Tech 18:506–510PubMedCrossRef Conrad B, Cordista A, Horodyski MB, Rechtine G (2005) Biomechanical evaluation of the pullout strength of cervical screws. J Spinal Disord Tech 18:506–510PubMedCrossRef
62.
go back to reference Hitchon PW, Brenton MD, Coppes JK, From AM, Torner JC (2003) Factors affecting pullout strength of self-drilling and self-tapping anterior cervical screws. Spine 28:9–13PubMedCrossRef Hitchon PW, Brenton MD, Coppes JK, From AM, Torner JC (2003) Factors affecting pullout strength of self-drilling and self-tapping anterior cervical screws. Spine 28:9–13PubMedCrossRef
63.
go back to reference Maiman DJ, Pintar FA, Yoganandan N, Reinartz J, Toselli R, Woodward E, Haid R (1992) Pull-out strength of Caspar cervical screw. Neurosurg 31:1097–1101CrossRef Maiman DJ, Pintar FA, Yoganandan N, Reinartz J, Toselli R, Woodward E, Haid R (1992) Pull-out strength of Caspar cervical screw. Neurosurg 31:1097–1101CrossRef
64.
go back to reference Pitzen T, Barbier D, Tintinger F, Steudel WI, Strowitzki M (2002) Screw fixation to the posterior cortical shell does not influence peak torque and pullout in anterior cervical plating. Eur Spine J 11:494–499PubMedCrossRef Pitzen T, Barbier D, Tintinger F, Steudel WI, Strowitzki M (2002) Screw fixation to the posterior cortical shell does not influence peak torque and pullout in anterior cervical plating. Eur Spine J 11:494–499PubMedCrossRef
65.
go back to reference Koller H, Hempfing A, Ferraris L, Meier O, Metz-Stavenhagen P (2006) 4- and 5-level anterior fusions of the cervical spine: review of literature and clinical results. Eur Spine J 16(12):2055–2071CrossRef Koller H, Hempfing A, Ferraris L, Meier O, Metz-Stavenhagen P (2006) 4- and 5-level anterior fusions of the cervical spine: review of literature and clinical results. Eur Spine J 16(12):2055–2071CrossRef
66.
go back to reference Ryken TC, Clausen JD, Traynelis VC, Goel VK (1995) Biomechanical analysis of bone mineral density, inseriton technique, screw torque, and holding strenght of anterior cervical plates. Neurosurgery 83:325–329 Ryken TC, Clausen JD, Traynelis VC, Goel VK (1995) Biomechanical analysis of bone mineral density, inseriton technique, screw torque, and holding strenght of anterior cervical plates. Neurosurgery 83:325–329
67.
go back to reference Spivak JM, Chen D, Kummer FJ (1999) The effect of locking fixation screws on the stability of anterior cervical plating. Spine 24:334–338PubMedCrossRef Spivak JM, Chen D, Kummer FJ (1999) The effect of locking fixation screws on the stability of anterior cervical plating. Spine 24:334–338PubMedCrossRef
68.
go back to reference Heller JG, Estes BT, M Zaouali M, Diop A (1996) Biomechanical study of screws in the lateral masses: variables affecting pull-out resistance. J Bone Joint Surg Am 78(9):1315–1321PubMed Heller JG, Estes BT, M Zaouali M, Diop A (1996) Biomechanical study of screws in the lateral masses: variables affecting pull-out resistance. J Bone Joint Surg Am 78(9):1315–1321PubMed
69.
go back to reference Schlenk RP, Stewart T, Benzel EC (2003) The biomechanics of iatrogenic spinal destabilization and implant failure. Neurosurg Focus 15(3):Article 2 Schlenk RP, Stewart T, Benzel EC (2003) The biomechanics of iatrogenic spinal destabilization and implant failure. Neurosurg Focus 15(3):Article 2
70.
go back to reference Burval DJ, McLain RF, Milks R, Inceoglu S (2007) Primary pedicle screw augmentation in osteoporotic lumbar vertebrae—biomechanical analysis of pedicle fixation strength. Spine 32:1077–1083PubMedCrossRef Burval DJ, McLain RF, Milks R, Inceoglu S (2007) Primary pedicle screw augmentation in osteoporotic lumbar vertebrae—biomechanical analysis of pedicle fixation strength. Spine 32:1077–1083PubMedCrossRef
71.
go back to reference Barrey C, Mertens P, Rumelhart C, Cotton F, Jund J, Perrin G (2004) Biomechanical evaluation of cervical lateral mass fixation: a comparison of the Roy-Camille and Magerl screw techniques. J Neurosurg 100(Spine 3):268–276PubMed Barrey C, Mertens P, Rumelhart C, Cotton F, Jund J, Perrin G (2004) Biomechanical evaluation of cervical lateral mass fixation: a comparison of the Roy-Camille and Magerl screw techniques. J Neurosurg 100(Spine 3):268–276PubMed
72.
go back to reference Seybold EA, Baker JA, Criscitiello AA, Ordway BR, Park CK, Connolly PJ (1999) Characteristics of unicortical and bicortical bilateral mass screws in the cervical spine. Spine 24:2397–2403PubMedCrossRef Seybold EA, Baker JA, Criscitiello AA, Ordway BR, Park CK, Connolly PJ (1999) Characteristics of unicortical and bicortical bilateral mass screws in the cervical spine. Spine 24:2397–2403PubMedCrossRef
Metadata
Title
Cervical anterior transpedicular screw fixation (ATPS)—Part II. Accuracy of manual insertion and pull-out strength of ATPS
Authors
Heiko Koller
Frank Acosta
Mark Tauber
Michael Fox
Hudelmaier Martin
Rosmarie Forstner
Peter Augat
Rainer Penzkofer
Christian Pirich
H. Kässmann
Herbert Resch
Wolfgang Hitzl
Publication date
01-04-2008
Publisher
Springer-Verlag
Published in
European Spine Journal / Issue 4/2008
Print ISSN: 0940-6719
Electronic ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-007-0573-x

Other articles of this Issue 4/2008

European Spine Journal 4/2008 Go to the issue

Announcements

Announcements