Published in: Comparative Clinical Pathology 6/2015

01-11-2015 | Original Article

Monocytes in Icelandic horses: comparison between automatic gating by the ADVIA 2120i, a manual gating method and microscopic differentiation

Authors: E. Leidinger, G. Prechtl, J. Figl, J. Leidinger

Published in: Comparative Clinical Pathology | Issue 6/2015

Login to get access

Abstract

Inspection of the ADVIA 2120i PEROX cytograms from healthy Icelandic horses revealed a mismatch of the automatic gating (A) for monocytes, resulting in a lower percentage. The raw data of 40 Icelandic horses were transformed to FCS files. Data were imported to WinMDI 2.9 (Purdue Cytometry Laboratories) and gated manually (M). In addition, a manual 200-cell differential (D) was performed. ANOVA was used to test for significant differences. RefVal 2.1 (Toulouse Veterinary School) was used to calculate the reference values by different methods. Significance level was set at P < 0.05. Monocytes were significantly higher by M compared to A and D, neutrophils were significantly lower for A compared to M, but not for A compared to D. Data were used to calculate the RI as follows: neutrophils (A) 37.7–65.6 % (1.7–4.6 × 109/L); neutrophils (M): 33.8–61.8 % (1.6–4.3 × 109/L); lymphocytes (A): 26.0–52.7 % (1.1–4.6 × 109/L), lymphocytes (M): 28.1–55.5 % (1.2–4.0 × 109/L); monocytes (A): 1.5–4.6 % (0.1–0.4 × 109/L); monocytes (M): 3.4–6.9 % (0.2–0.5 × 109/L). The ADVIA 2120i built-in export function and the use of WinMDI allow a very precise recalculation of cell populations. Icelandic horses reportedly have the lowest monocyte number of all horse breeds. This may be possibly caused by higher MPO activity and/or an inaccuracy of the algorithm used.
Literature
go back to reference Bauer N, Nakagawa J, Dunker C, Failing K, Moritz A (2012) Evaluation of the automated hematology analyzer Sysmex XT-2000iVTM compared to the ADVIA® 2120 for its use in dogs, cats, and horses. Part II: accuracy of leukocyte differential and reticulocyte count, impact of anticoagulant and sample aging. J Vet Diag Invest 24:74–89CrossRef Bauer N, Nakagawa J, Dunker C, Failing K, Moritz A (2012) Evaluation of the automated hematology analyzer Sysmex XT-2000iVTM compared to the ADVIA® 2120 for its use in dogs, cats, and horses. Part II: accuracy of leukocyte differential and reticulocyte count, impact of anticoagulant and sample aging. J Vet Diag Invest 24:74–89CrossRef
go back to reference Björnsson GB, Sveinsson HJ (2007) The Icelandic horse. 2nd ed.: Edda Publishing, Reykjavik pp 20-33 Björnsson GB, Sveinsson HJ (2007) The Icelandic horse. 2nd ed.: Edda Publishing, Reykjavik pp 20-33
go back to reference Clark P, Mogg TD, Tvedten H, Korcal D (2002) Artifactual changes in equine blood following storage, dedected using the Advia 120 hematology analyser. Vet Clin Pathol 31:90–94CrossRefPubMed Clark P, Mogg TD, Tvedten H, Korcal D (2002) Artifactual changes in equine blood following storage, dedected using the Advia 120 hematology analyser. Vet Clin Pathol 31:90–94CrossRefPubMed
go back to reference Dura A (2005) Blutzellzählung und -differenzierung bei Pferd und Schwein mit dem Hämatologiesystem ADVIA 120. Gerätevalidierung und Softwareadaptation [Blood cell count and differentation in horses and swines. Instrument validation and software adaption]. Dissertation, Justus-Liebig-Universität, Giessen Dura A (2005) Blutzellzählung und -differenzierung bei Pferd und Schwein mit dem Hämatologiesystem ADVIA 120. Gerätevalidierung und Softwareadaptation [Blood cell count and differentation in horses and swines. Instrument validation and software adaption]. Dissertation, Justus-Liebig-Universität, Giessen
go back to reference Flandrin G, Valensi F, Mazon C, Sassier, P. Ann Biol Clin (Paris) (1993) [The progresses in the determination of blood monocytosis and its perspextives of clinical application in pathology];51:787–791 Flandrin G, Valensi F, Mazon C, Sassier, P. Ann Biol Clin (Paris) (1993) [The progresses in the determination of blood monocytosis and its perspextives of clinical application in pathology];51:787–791
go back to reference Geffré A, Concordet D, Braun J-P, Trumel C (2011) Reference value advisor: a new freeware set of macroinstructions to calculate reference intervals with Microsoft Excel. Vet Clin Pathol 40:107–112CrossRefPubMed Geffré A, Concordet D, Braun J-P, Trumel C (2011) Reference value advisor: a new freeware set of macroinstructions to calculate reference intervals with Microsoft Excel. Vet Clin Pathol 40:107–112CrossRefPubMed
go back to reference Gieseler T (2012) Referenzwerte beim Pferd [Reference intervals in horses]. Dissertation, Vet Med Leipzig Gieseler T (2012) Referenzwerte beim Pferd [Reference intervals in horses]. Dissertation, Vet Med Leipzig
go back to reference Giordano A, Rossi G, Pieralisi C, Paltrinieri S (2008) Evaluation of equine hemograms using the Advia 120 as compared with an impedance counter and manual differential count. Vet Clin Pathol 37:21–30CrossRefPubMed Giordano A, Rossi G, Pieralisi C, Paltrinieri S (2008) Evaluation of equine hemograms using the Advia 120 as compared with an impedance counter and manual differential count. Vet Clin Pathol 37:21–30CrossRefPubMed
go back to reference Gorman ME, Villarroel A, Tornquist SJ, Flachsbart J, Warden A, Boeder L (2009) Comparison between manual and automated total nucleated cell counts using the ADVIA 120 for pleural and peritoneal fluid samples from dogs, cats, horses, and alpacas. Vet Clin Pathol 38:388–391CrossRefPubMed Gorman ME, Villarroel A, Tornquist SJ, Flachsbart J, Warden A, Boeder L (2009) Comparison between manual and automated total nucleated cell counts using the ADVIA 120 for pleural and peritoneal fluid samples from dogs, cats, horses, and alpacas. Vet Clin Pathol 38:388–391CrossRefPubMed
go back to reference Klenner S, Richartz J, Bauer N, Moritz A (2010) Myeloperoxidase deficiency in dogs observed with the ADVIA120. Tierärztl Praxis Kleint 3:139–46 Klenner S, Richartz J, Bauer N, Moritz A (2010) Myeloperoxidase deficiency in dogs observed with the ADVIA120. Tierärztl Praxis Kleint 3:139–46
go back to reference Lumsden JH, Rowe R, Mullen K (1980) Hematology and biochemistry reference values for the light horse. Can J Comp Med 44:32–42PubMedCentralPubMed Lumsden JH, Rowe R, Mullen K (1980) Hematology and biochemistry reference values for the light horse. Can J Comp Med 44:32–42PubMedCentralPubMed
go back to reference Moritz A (2000) Der Einsatz lasergestützter Multiparameter-Hämatologiesysteme in der Veterinärmedizin [The use of laser-based multiparameter hematology systems in veterinary medicine]. Dissertation, Justus-Liebig-Universität, Giessen Moritz A (2000) Der Einsatz lasergestützter Multiparameter-Hämatologiesysteme in der Veterinärmedizin [The use of laser-based multiparameter hematology systems in veterinary medicine]. Dissertation, Justus-Liebig-Universität, Giessen
go back to reference Rizzi TE, Meinkoth JH, Clinkenbeard KD (2010) Normal hematology of the dog. In: Weiss DJ, Wardrop KJ (eds) Schalm’s veterinary hematology, 6th edn. Blackwell, Ames, pp 799–810 Rizzi TE, Meinkoth JH, Clinkenbeard KD (2010) Normal hematology of the dog. In: Weiss DJ, Wardrop KJ (eds) Schalm’s veterinary hematology, 6th edn. Blackwell, Ames, pp 799–810
go back to reference Rümke CL (1959) Variability of results in differential cell counts on blood smears. Triangle 4:154–158 Rümke CL (1959) Variability of results in differential cell counts on blood smears. Triangle 4:154–158
go back to reference Ryseff JK (2012) Characterization of errors and predictability of disease using the ADVIA 120 hematology analyser leukocyte differential. Dissertation, Colorado State University Ryseff JK (2012) Characterization of errors and predictability of disease using the ADVIA 120 hematology analyser leukocyte differential. Dissertation, Colorado State University
go back to reference Sahraneshin Samani F, Moore JK, Khosravani P, Ebrahimi M (2010) Features of free software packages in flow cytometry: a comparison between four non-commercial software sources. Cytotechnology. doi:10.1007/s10616-013-9609-8 Sahraneshin Samani F, Moore JK, Khosravani P, Ebrahimi M (2010) Features of free software packages in flow cytometry: a comparison between four non-commercial software sources. Cytotechnology. doi:10.​1007/​s10616-013-9609-8
go back to reference Seiser M, Strasser A, Hofbauer B (2001) Der Einfluss von Alter und Geschlecht of diagnostisch wesentliche Blutparameter bei Islandpferden. [The influence of age and sex on significant haematological parameters in Icelandic horses]. Tierarztl Prax 29:324–331 Seiser M, Strasser A, Hofbauer B (2001) Der Einfluss von Alter und Geschlecht of diagnostisch wesentliche Blutparameter bei Islandpferden. [The influence of age and sex on significant haematological parameters in Icelandic horses]. Tierarztl Prax 29:324–331
go back to reference Takasu M, NAGATANI N, Tozaki T, Kakoi H, Maeda M, Murase T, Mukoyama H (2013) Hematological and biochemical reference values for the endangered Kiso horse. J Equ Sci 24:75–78CrossRef Takasu M, NAGATANI N, Tozaki T, Kakoi H, Maeda M, Murase T, Mukoyama H (2013) Hematological and biochemical reference values for the endangered Kiso horse. J Equ Sci 24:75–78CrossRef
go back to reference Tvedten H, Lilliehöok I, Nostell K, Edner A, Bröjer J (2012) Decrease in ADVIA 2120 MPXI and a monocyte identification error occur late after endotoxin treatment of horses. Seattle, WA: Proceedings 47th ASVCP Ann. Meeting. p10 Tvedten H, Lilliehöok I, Nostell K, Edner A, Bröjer J (2012) Decrease in ADVIA 2120 MPXI and a monocyte identification error occur late after endotoxin treatment of horses. Seattle, WA: Proceedings 47th ASVCP Ann. Meeting. p10
go back to reference van der Meer W (2006) Blood cell morphology: controversies and alternatives. Dissertation, Radboud University, Nijmegen, The Netherlands van der Meer W (2006) Blood cell morphology: controversies and alternatives. Dissertation, Radboud University, Nijmegen, The Netherlands
Metadata
Title
Monocytes in Icelandic horses: comparison between automatic gating by the ADVIA 2120i, a manual gating method and microscopic differentiation
Authors
E. Leidinger
G. Prechtl
J. Figl
J. Leidinger
Publication date
01-11-2015
Publisher
Springer London
Published in
Comparative Clinical Pathology / Issue 6/2015
Print ISSN: 1618-5641
Electronic ISSN: 1618-565X
DOI
https://doi.org/10.1007/s00580-015-2080-5

Other articles of this Issue 6/2015

Comparative Clinical Pathology 6/2015 Go to the issue