01-10-2017 | Original Article

Comparison of Nellcor™ PM1000N and Masimo Radical-7® for detecting apnea in volunteers

Authors: Chanatthee Kitsiripant, Tomoko Fukada, Hiroko Iwakiri, Yuri Tsuchiya, Makoto Ozaki, Minoru Nomura

Published in: Journal of Anesthesia | Issue 5/2017

Login to get access

Abstract

Purpose

Although capnography is considered the gold standard for monitoring of ventilation, it may not work accurately in some situations. We compared the performance of two non-invasive continuous respiratory rate (RR) monitors that are alternatives for the detection of respiratory depression.

Methods

Fifty healthy volunteers ≥20 years old were enrolled in this study. After monitoring of the volunteers was started by the Masimo Radical-7® and Nellcor™ PM1000N, they breathed at the rate of 12 breaths/min for 3 min and then stopped breathing. As soon as the apnea alarm of either monitor went off, breathing was resumed at the same rate. This entire procedure was repeated three times. The data collected every 30 s included the RR, pulse rate (PR) and oxygen saturation (SpO2). The times of alarm on, alarm off and reappearance of RR on the monitor screens were also recorded.

Results

The biases of the RR, PR and SpO2 measurements from the two monitors were 0.5, 0.2 and −0.4, respectively. Of 143 procedures that stopped breathing for more than 30 s, 114 and 15 alarms were shown by the Masimo Radical-7® and Nellcor™ PM-1000N monitors, respectively. Most alarms of the Nellcor™ PM1000N followed from SpO2 <90%. Conversely, most alarms of the Masimo Radical-7® were caused by RR <10 breaths/min. Times to alarm on, alarm off and display of RR measured by the Masimo Radical-7® monitor were significantly shorter than the Nellcor™ PM-1000N monitor.

Conclusions

The Masimo Radical-7® monitor provides better detection of apnea in volunteers than the Nellcor™ PM-1000N.
Literature
1.
go back to reference Bittner EA, George E. Respiratory problems in the post-anesthesia care unit (PACU). In: UpToDate [Internet]. Waltham (MA): UpToDate, Inc; c2016 [updated 2016 May 31; cited 2016 Nov 10]. http://www.uptodate.com. Accessed 10 Nov 2016 Bittner EA, George E. Respiratory problems in the post-anesthesia care unit (PACU). In: UpToDate [Internet]. Waltham (MA): UpToDate, Inc; c2016 [updated 2016 May 31; cited 2016 Nov 10]. http://​www.​uptodate.​com. Accessed 10 Nov 2016
2.
go back to reference Karcz M, Papadakos PJ. Respiratory complications in the postanesthesia care unit: a review of pathophysiological mechanisms. Can J Respir Ther. 2013;49:21–9.PubMedPubMedCentral Karcz M, Papadakos PJ. Respiratory complications in the postanesthesia care unit: a review of pathophysiological mechanisms. Can J Respir Ther. 2013;49:21–9.PubMedPubMedCentral
3.
go back to reference Apfelbaum JL, Silverstein JH, Chung FF, Connis RT, Fillmore RB, Hunt SE, Nickinovich DG, Schreiner MS. Practice guidelines for postanesthetic care: an updated report by the American Society of Anesthesiologists Task Force on postanesthetic care. Anesthesiology. 2013;118:291–307.CrossRefPubMed Apfelbaum JL, Silverstein JH, Chung FF, Connis RT, Fillmore RB, Hunt SE, Nickinovich DG, Schreiner MS. Practice guidelines for postanesthetic care: an updated report by the American Society of Anesthesiologists Task Force on postanesthetic care. Anesthesiology. 2013;118:291–307.CrossRefPubMed
4.
go back to reference Keidan I, Gravenstein D, Berkenstadt H, Ziv A, Shavit I, Sidi A. Supplemental oxygen compromises the use of pulse oximetry for detection of apnea and hypoventilation during sedation in simulated pediatric patients. Pediatrics. 2008;122:293–8.CrossRefPubMed Keidan I, Gravenstein D, Berkenstadt H, Ziv A, Shavit I, Sidi A. Supplemental oxygen compromises the use of pulse oximetry for detection of apnea and hypoventilation during sedation in simulated pediatric patients. Pediatrics. 2008;122:293–8.CrossRefPubMed
5.
go back to reference Overdyk FJ, Carter R, Maddox RR, Callura J, Herrin AE, Henriquez C. Continuous oximetry/capnometry monitoring reveals frequent desaturation and bradypnea during patient-controlled analgesia. Anesth Analg. 2007;105:412–8.CrossRefPubMed Overdyk FJ, Carter R, Maddox RR, Callura J, Herrin AE, Henriquez C. Continuous oximetry/capnometry monitoring reveals frequent desaturation and bradypnea during patient-controlled analgesia. Anesth Analg. 2007;105:412–8.CrossRefPubMed
6.
go back to reference Mimoz O, Benard T, Gaucher A, Frasca D, Debaene B. Accuracy of respiratory rate monitoring using a non-invasive acoustic method after general anaesthesia. Br J Anaesth. 2012;108:872–5.CrossRefPubMed Mimoz O, Benard T, Gaucher A, Frasca D, Debaene B. Accuracy of respiratory rate monitoring using a non-invasive acoustic method after general anaesthesia. Br J Anaesth. 2012;108:872–5.CrossRefPubMed
7.
go back to reference Ouchi K, Fujiwara S, Sugiyama K. Acoustic method respiratory rate monitoring is useful in patients under intravenous anesthesia. J Clin Monit Comput [Internet]. 2016. doi:10.1007/s10877-015-9822-4. Ouchi K, Fujiwara S, Sugiyama K. Acoustic method respiratory rate monitoring is useful in patients under intravenous anesthesia. J Clin Monit Comput [Internet]. 2016. doi:10.​1007/​s10877-015-9822-4.
8.
go back to reference Gottlieb KT, Banerjee S, Barth BA, Bhat YM, Desilets DJ, Maple JT, Pfau PR, Pleskow DK, Siddiqui UD, Tokar JL, Wang A, Wong Kee Song LM, Rodriguez SA. Monitoring equipment for endoscopy. Gastrointest Endosc. 2013;77:175–80. doi:10.1016/j.gie.2012.09.02.CrossRefPubMed Gottlieb KT, Banerjee S, Barth BA, Bhat YM, Desilets DJ, Maple JT, Pfau PR, Pleskow DK, Siddiqui UD, Tokar JL, Wang A, Wong Kee Song LM, Rodriguez SA. Monitoring equipment for endoscopy. Gastrointest Endosc. 2013;77:175–80. doi:10.​1016/​j.​gie.​2012.​09.​02.CrossRefPubMed
9.
go back to reference Patino M, Redford DT, Quigley TW, Mahmoud M, Kurth CD, Szmuk P. Accuracy of acoustic respiration rate monitoring in pediatric patients. Pediatr Anaesth. 2013;23:1166–73.CrossRef Patino M, Redford DT, Quigley TW, Mahmoud M, Kurth CD, Szmuk P. Accuracy of acoustic respiration rate monitoring in pediatric patients. Pediatr Anaesth. 2013;23:1166–73.CrossRef
10.
go back to reference Fukada T, Iwakiri H, Nomura M, Ozaki M. Clinical evaluation of acoustic respiration rate monitoring compared with conventional systems in the postanaesthesia care unit. Eur J Anaesthesiol. 2015;32:58–68.CrossRef Fukada T, Iwakiri H, Nomura M, Ozaki M. Clinical evaluation of acoustic respiration rate monitoring compared with conventional systems in the postanaesthesia care unit. Eur J Anaesthesiol. 2015;32:58–68.CrossRef
11.
go back to reference Ramsay MAE, Usman M, Lagow E, Mendoza M, Untalan E, De Vol E. The accuracy, precision and reliability of measuring ventilatory rate and detecting ventilatory pause by rainbow acoustic monitoring and capnometry. Anesth Analg. 2013;117:69–75.CrossRefPubMed Ramsay MAE, Usman M, Lagow E, Mendoza M, Untalan E, De Vol E. The accuracy, precision and reliability of measuring ventilatory rate and detecting ventilatory pause by rainbow acoustic monitoring and capnometry. Anesth Analg. 2013;117:69–75.CrossRefPubMed
12.
go back to reference Addison PS, Watson JN, Mestek ML, Mecca RS. Developing an algorithm for pulse oximetry derived respiratory rate (RRoxi): a healthy volunteer study. J Clin Monit Comput. 2012;26:45–51.CrossRefPubMedPubMedCentral Addison PS, Watson JN, Mestek ML, Mecca RS. Developing an algorithm for pulse oximetry derived respiratory rate (RRoxi): a healthy volunteer study. J Clin Monit Comput. 2012;26:45–51.CrossRefPubMedPubMedCentral
13.
go back to reference Addison PS, Watson JN, Mestek ML, Ochs JP, Uribe AA, Bergese SD. Pulse oximetry-derived respiratory rate in general care floor patients. J Clin Monit Comput. 2015;29:113–20.CrossRefPubMed Addison PS, Watson JN, Mestek ML, Ochs JP, Uribe AA, Bergese SD. Pulse oximetry-derived respiratory rate in general care floor patients. J Clin Monit Comput. 2015;29:113–20.CrossRefPubMed
14.
go back to reference Touw HRW, Verheul MH, Tuinman PR, Smit J, Thöne D, Schober P, Boer C. Photoplethysmography respiratory rate monitoring in patients receiving procedural sedation and analgesia for upper gastrointestinal endoscopy. J Clin Monit Comput [Internet]. 2016. doi:10.1007/s10877-016-9890-0.PubMedCentral Touw HRW, Verheul MH, Tuinman PR, Smit J, Thöne D, Schober P, Boer C. Photoplethysmography respiratory rate monitoring in patients receiving procedural sedation and analgesia for upper gastrointestinal endoscopy. J Clin Monit Comput [Internet]. 2016. doi:10.​1007/​s10877-016-9890-0.PubMedCentral
15.
go back to reference Eisenberg ME, Levin R, Givony D, Overdyk FJ. Comparison study between nellcor respiratory rate technology (RRoxi) and masimo acoustic respiratory rate (RRa). In: Poster session presented at: ANESTHESIOLOGY® 2016 Annual Meeting; 2016 October 23; Chicago, IL, USA. (A2031). Eisenberg ME, Levin R, Givony D, Overdyk FJ. Comparison study between nellcor respiratory rate technology (RRoxi) and masimo acoustic respiratory rate (RRa). In: Poster session presented at: ANESTHESIOLOGY® 2016 Annual Meeting; 2016 October 23; Chicago, IL, USA. (A2031).
Metadata
Title
Comparison of Nellcor™ PM1000N and Masimo Radical-7® for detecting apnea in volunteers
Authors
Chanatthee Kitsiripant
Tomoko Fukada
Hiroko Iwakiri
Yuri Tsuchiya
Makoto Ozaki
Minoru Nomura
Publication date
01-10-2017
Publisher
Springer Japan
Published in
Journal of Anesthesia / Issue 5/2017
Print ISSN: 0913-8668
Electronic ISSN: 1438-8359
DOI
https://doi.org/10.1007/s00540-017-2385-4

Other articles of this Issue 5/2017

Journal of Anesthesia 5/2017 Go to the issue