01-05-2021 | Laparoscopy

Monopolar stray energy in robotic surgery

Authors: Douglas M. Overbey, Heather Carmichael, Krzysztof J. Wikiel, Douglas A. Hirth, Brandon C. Chapman, John T. Moore, Carlton C. Barnett, Teresa S. Jones, Thomas N. Robinson, Edward L. Jones

Published in: Surgical Endoscopy | Issue 5/2021

Login to get access

Abstract

Introduction

Stray energy transfer from monopolar radiofrequency energy during laparoscopy can be potentially catastrophic. Robotic surgery is increasing in popularity; however, the risk of stray energy transfer during robotic surgery is unknown. The purpose of this study was to (1) quantify stray energy transfer using robotic instrumentation, (2) determine strategies to minimize the transfer of energy, and (3) compare robotic stray energy transfer to laparoscopy.

Methods

In a laparoscopic trainer, a monopolar instrument (L-hook) was activated with DaVinci Si (Intuitive, Sunnyvale, CA) robotic instruments. A camera and assistant grasper were inserted to mimic a minimally invasive cholecystectomy. During activation of the L-hook, the non-electric tips of the camera and grasper were placed adjacent to simulated tissue (saline-soaked sponge). The primary outcome was change in temperature from baseline (°C) measured nearest the tip of the non-electric instrument.

Results

Simulated tissue nearest the robotic grasper increased an average of 18.3 ± 5.8 °C; p < 0.001 from baseline. Tissue nearest the robotic camera tip increased (9.0 ± 2.1 °C; p < 0.001). Decreasing the power from 30 to 15 W (18.3 ± 5.8 vs. 2.6 ± 2.7 °C, p < 0.001) or using low-voltage cut mode (18.3 ± 5.8 vs. 3.1 ± 2.1 °C, p < 0.001) reduced stray energy transfer to the robotic grasper. Desiccating tissue, in contrast to open air activation, also significantly reduced stray energy transfer for the grasper (18.3 ± 5.8 vs. 0.15 ± 0.21 °C, p < 0.001) and camera (9.0 ± 2.1 vs. 0.24 ± 0.34 °C, p < 0.001).

Conclusions

Stray energy transfer occurs during robotic surgery. The assistant grasper carries the highest risk for thermal injury. Similar to laparoscopy, stray energy transfer can be reduced by lowering the power setting, utilizing a low-voltage cut mode instead of coagulation mode and avoiding open air activation. These practical findings can aid surgeons performing robotic surgery to reduce injuries from stray energy.
Literature
1.
go back to reference Tuncel U, Ozgenel GY (2005) Thermal injury due to electrosurgery. Ulus Travma Acil Cerrahi Derg 11(1):76–77PubMed Tuncel U, Ozgenel GY (2005) Thermal injury due to electrosurgery. Ulus Travma Acil Cerrahi Derg 11(1):76–77PubMed
2.
go back to reference Martin KE et al (2016) Quantifying inadvertent thermal bowel injury from the monopolar instrument. Surg Endosc 30(11):4776–4784CrossRef Martin KE et al (2016) Quantifying inadvertent thermal bowel injury from the monopolar instrument. Surg Endosc 30(11):4776–4784CrossRef
3.
go back to reference Morrison JE Jr, Jacobs VR (2007) Rupture of the spleen with the harmonic scalpel: case report of an unexplained complication. Jsls 11(2):268–271PubMedPubMedCentral Morrison JE Jr, Jacobs VR (2007) Rupture of the spleen with the harmonic scalpel: case report of an unexplained complication. Jsls 11(2):268–271PubMedPubMedCentral
4.
go back to reference Awwad JT, Isaacson K (1996) The harmonic scalpel: an intraoperative complication. Obstet Gynecol 88(4 Pt 2):718–720CrossRef Awwad JT, Isaacson K (1996) The harmonic scalpel: an intraoperative complication. Obstet Gynecol 88(4 Pt 2):718–720CrossRef
5.
go back to reference Harrell AG, Kercher KW, Heniford BT (2004) Energy sources in laparoscopy. Semin Laparosc Surg 11(3):201–209PubMed Harrell AG, Kercher KW, Heniford BT (2004) Energy sources in laparoscopy. Semin Laparosc Surg 11(3):201–209PubMed
6.
go back to reference Sankaranarayanan G et al (2013) Common uses and cited complications of energy in surgery. Surg Endosc 27(9):3056–3072CrossRef Sankaranarayanan G et al (2013) Common uses and cited complications of energy in surgery. Surg Endosc 27(9):3056–3072CrossRef
7.
go back to reference Willson PD, van der Walt JD, Rogers J (1995) Electrosurgical coupling to a metal cannula causing skin burns during laparoscopic surgery. Minim Invasive Ther 4(3):163–164CrossRef Willson PD, van der Walt JD, Rogers J (1995) Electrosurgical coupling to a metal cannula causing skin burns during laparoscopic surgery. Minim Invasive Ther 4(3):163–164CrossRef
8.
go back to reference Willson PD, McAnena OJ, Peters EE (1994) A fatal complication of diathermy in laparoscopic surgery. Minim Invasive Ther 3(1):19–20CrossRef Willson PD, McAnena OJ, Peters EE (1994) A fatal complication of diathermy in laparoscopic surgery. Minim Invasive Ther 3(1):19–20CrossRef
9.
go back to reference Tucker RD (1995) Laparoscopic electrosurgical injuries: survey results and their implications. Surg Laparosc Endosc 5(4):311–317PubMed Tucker RD (1995) Laparoscopic electrosurgical injuries: survey results and their implications. Surg Laparosc Endosc 5(4):311–317PubMed
10.
go back to reference Odell RC (2013) Surgical complications specific to monopolar electrosurgical energy: engineering changes that have made electrosurgery safer. J Minim Invasive Gynecol 20(3):288–298CrossRef Odell RC (2013) Surgical complications specific to monopolar electrosurgical energy: engineering changes that have made electrosurgery safer. J Minim Invasive Gynecol 20(3):288–298CrossRef
13.
go back to reference Guzman C et al (2019) Estimating the incidence of stray energy burns during laparoscopic surgery based on two statewide databases and retrospective rates: an opportunity to improve patient safety. Surg Technol Int 34:30–34PubMed Guzman C et al (2019) Estimating the incidence of stray energy burns during laparoscopic surgery based on two statewide databases and retrospective rates: an opportunity to improve patient safety. Surg Technol Int 34:30–34PubMed
18.
go back to reference Mendez-Probst CE et al (2011) Stray electrical currents in laparoscopic instruments used in da Vinci(R) robot-assisted surgery: an in vitro study. J Endourol 25(9):1513–1517CrossRef Mendez-Probst CE et al (2011) Stray electrical currents in laparoscopic instruments used in da Vinci(R) robot-assisted surgery: an in vitro study. J Endourol 25(9):1513–1517CrossRef
25.
go back to reference Wu MP et al (2000) Complications and recommended practices for electrosurgery in laparoscopy. Am J Surg 179(1):67–73CrossRef Wu MP et al (2000) Complications and recommended practices for electrosurgery in laparoscopy. Am J Surg 179(1):67–73CrossRef
26.
go back to reference Abu-Rafea B et al (2011) Monopolar electrosurgery through single-port laparoscopy: a potential hidden hazard for bowel burns. J Minim Invasive Gynecol 18(6):734–740CrossRef Abu-Rafea B et al (2011) Monopolar electrosurgery through single-port laparoscopy: a potential hidden hazard for bowel burns. J Minim Invasive Gynecol 18(6):734–740CrossRef
27.
go back to reference Jones ELM, Dean J (2018) Surgical energy. In: Henke PK, Hochwald SN, Tiao GM, Fischer JEE, Christopher E (eds) Fischer's mastery of surgery. Wolters Kluwer, Philadelphia Jones ELM, Dean J (2018) Surgical energy. In: Henke PK, Hochwald SN, Tiao GM, Fischer JEE, Christopher E (eds) Fischer's mastery of surgery. Wolters Kluwer, Philadelphia
28.
go back to reference Gurtcheff SE (2008) Introduction to the MAUDE database. Clin Obstet Gynecol 51(1):120–123CrossRef Gurtcheff SE (2008) Introduction to the MAUDE database. Clin Obstet Gynecol 51(1):120–123CrossRef
30.
go back to reference Abraham NE, Simon R, Shah O (2011) Thermal injury causing delayed perforation of small bowel after transurethral resection of bladder tumor without evidence of bladder perforation. Can J Urol 18(4):5836–5838PubMed Abraham NE, Simon R, Shah O (2011) Thermal injury causing delayed perforation of small bowel after transurethral resection of bladder tumor without evidence of bladder perforation. Can J Urol 18(4):5836–5838PubMed
31.
go back to reference Cassaro S (2015) Delayed manifestations of laparoscopic bowel injury. Am Surg 81(5):478–482CrossRef Cassaro S (2015) Delayed manifestations of laparoscopic bowel injury. Am Surg 81(5):478–482CrossRef
32.
go back to reference Berry SM et al (1994) Thermal injury of the posterior duodenum during laparoscopic cholecystectomy. Surg Endosc 8(3):197–200CrossRef Berry SM et al (1994) Thermal injury of the posterior duodenum during laparoscopic cholecystectomy. Surg Endosc 8(3):197–200CrossRef
33.
go back to reference Campagnacci R et al (2007) Electrothermal bipolar vessel sealing device vs ultrasonic coagulating shears in laparoscopic colectomies: a comparative study. Surg Endosc 21(9):1526–1531CrossRef Campagnacci R et al (2007) Electrothermal bipolar vessel sealing device vs ultrasonic coagulating shears in laparoscopic colectomies: a comparative study. Surg Endosc 21(9):1526–1531CrossRef
34.
go back to reference Machado NO (2016) Duodenal injury post laparoscopic cholecystectomy: incidence, mechanism, management and outcome. World J Gastrointest Surg 8(4):335–344CrossRef Machado NO (2016) Duodenal injury post laparoscopic cholecystectomy: incidence, mechanism, management and outcome. World J Gastrointest Surg 8(4):335–344CrossRef
35.
go back to reference Madani A et al (2014) Fundamental use of surgical energy (FUSE): a curriculum on surgical energy-based devices. Surg Endosc 28(9):2509–2512CrossRef Madani A et al (2014) Fundamental use of surgical energy (FUSE): a curriculum on surgical energy-based devices. Surg Endosc 28(9):2509–2512CrossRef
36.
go back to reference Jones SB et al (2017) Fundamental use of surgical energy (FUSE): an essential educational program for operating room safety. Perm J 21:16–050PubMedPubMedCentral Jones SB et al (2017) Fundamental use of surgical energy (FUSE): an essential educational program for operating room safety. Perm J 21:16–050PubMedPubMedCentral
37.
go back to reference Reissis Y et al (2013) The effect of temperature on the viability of human mesenchymal stem cells. Stem Cell Res Ther 4(6):139CrossRef Reissis Y et al (2013) The effect of temperature on the viability of human mesenchymal stem cells. Stem Cell Res Ther 4(6):139CrossRef
38.
go back to reference Zhang Y et al (2018) Temperature-dependent cell death patterns induced by functionalized gold nanoparticle photothermal therapy in melanoma cells. Sci Rep 8(1):8720CrossRef Zhang Y et al (2018) Temperature-dependent cell death patterns induced by functionalized gold nanoparticle photothermal therapy in melanoma cells. Sci Rep 8(1):8720CrossRef
Metadata
Title
Monopolar stray energy in robotic surgery
Authors
Douglas M. Overbey
Heather Carmichael
Krzysztof J. Wikiel
Douglas A. Hirth
Brandon C. Chapman
John T. Moore
Carlton C. Barnett
Teresa S. Jones
Thomas N. Robinson
Edward L. Jones
Publication date
01-05-2021
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 5/2021
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-020-07605-5

Other articles of this Issue 5/2021

Surgical Endoscopy 5/2021 Go to the issue