01-06-2020 | Cholecystectomy | 2019 EAES Oral

A robotic flexible endoscope with shared autonomy: a study of mockup cholecystectomy

Authors: Chengzhi Song, Xin Ma, Xianfeng Xia, Philip Wai Yan Chiu, Charing Ching Ning Chong, Zheng Li

Published in: Surgical Endoscopy | Issue 6/2020

Login to get access

Abstract

Background

Endoscope is the eye of surgeon in minimally invasive surgery (MIS). Prevailing handheld endoscopes are manually steered, which can cause endoscope-instrument fencing. Robotic endoscopes can reduce the fatigue but could not reduce collisions. Handheld endoscopes with a flexible bending tip can reduce the shaft pivoting and collisions. However, its steering is challenging. In this paper, we present a robotic flexible endoscope with auto-tracking function and compare it with the conventional rigid endoscopes.

Methods

A robotic flexible endoscope (RFE) with shared autonomy is developed. The RFE could either track the instruments automatically or be controlled by a foot pedal. A mockup cholecystectomy was designed to evaluate the performance. Five surgeons were invited to perform the mockup cholecystectomy in an abdominal cavity phantom with a manual rigid endoscope (MRE), a robotic rigid endoscope (RRE), and the RFE. Space occupation, time consumption, and questionnaires based on the NASA task load index were adopted to evaluate the performances and compare the three endoscope systems. An ex vivo experiment was conducted to demonstrate the feasibility of using the RFE in a biological tissue environment.

Results

All surgeons completed the mockup cholecystectomy with the RFE independently. Failure occurred in the cases involving the RRE and the MRE. Inside the body cavity, the space occupied when using the RFE is 17.28% and 23.95% (p < 0.05) of that when using the MRE and the RRE, respectively. Outside the body cavity, the space occupied when using the RFE is 14.60% and 15.53% (p < 0.05) of that by using MRE and RRE. Time consumed in the operations with MRE, RRE, and RFE are 28.3 s, 93.2 s and 34.8 s, respectively. Questionnaires reveal that the performance of the RFE is the best among the three endoscope systems.

Conclusions

The RFE provides a wider field of view (FOV) and occupies less space than rigid endoscopes.
Literature
1.
2.
go back to reference Vitiello V, Lee SL, Cundy TP, Yang GZ (2012) Emerging robotic platforms for minimally invasive surgery. IEEE Rev Biomed Eng 6:111–126CrossRef Vitiello V, Lee SL, Cundy TP, Yang GZ (2012) Emerging robotic platforms for minimally invasive surgery. IEEE Rev Biomed Eng 6:111–126CrossRef
3.
go back to reference Dakin GF, Gagner M (2003) Comparison of laparoscopic skills performance between standard instruments and two surgical robotic systems. Surg Endosc 17:574–579CrossRef Dakin GF, Gagner M (2003) Comparison of laparoscopic skills performance between standard instruments and two surgical robotic systems. Surg Endosc 17:574–579CrossRef
4.
go back to reference Freschi C, Ferrari V, Melfi F, Ferrari M, Mosca F, Cuschieri A (2013) Technical review of the da Vinci surgical telemanipulator. Med Robot Comput Assist Surg 9:396–406CrossRef Freschi C, Ferrari V, Melfi F, Ferrari M, Mosca F, Cuschieri A (2013) Technical review of the da Vinci surgical telemanipulator. Med Robot Comput Assist Surg 9:396–406CrossRef
5.
go back to reference Wang Y (1996) Automated endoscope system for optimal positioning. CIM-INT MANUF 12(4):372CrossRef Wang Y (1996) Automated endoscope system for optimal positioning. CIM-INT MANUF 12(4):372CrossRef
6.
go back to reference Taylor RH, Funda J, Eldridge B, Gomory S, Gruben K, LaRose D, Talamini M, Kavoussi L, Anderson J (1995) A telerobotic assistant for laparoscopic surgery. IEEE Eng Med Biol Mag 14(3):279–288CrossRef Taylor RH, Funda J, Eldridge B, Gomory S, Gruben K, LaRose D, Talamini M, Kavoussi L, Anderson J (1995) A telerobotic assistant for laparoscopic surgery. IEEE Eng Med Biol Mag 14(3):279–288CrossRef
7.
go back to reference Gilbert JM (2009) The EndoAssist™ robotic camera holder as an aid to the introduction of laparoscopic colorectal surgery. Ann R Coll Surg Engl 91(5):389–393CrossRef Gilbert JM (2009) The EndoAssist™ robotic camera holder as an aid to the introduction of laparoscopic colorectal surgery. Ann R Coll Surg Engl 91(5):389–393CrossRef
8.
go back to reference Voros S, Haber GP, Menudet JF, Long JA, Cinquin P (2010) ViKY robotic scope holder: initial clinical experience and preliminary results using instrument tracking. IEEE-ASME T MECH 15(6):879–886 Voros S, Haber GP, Menudet JF, Long JA, Cinquin P (2010) ViKY robotic scope holder: initial clinical experience and preliminary results using instrument tracking. IEEE-ASME T MECH 15(6):879–886
9.
go back to reference Taniguchi K, Nishikawa A, Sekimoto M, Yasui M, Takiguchi S, Seki Y, Monden M, Miyazaki F (2006) Design of a novel wearable laparoscope manipulator: SMART (Synthetic Muscle Actuator based Robotic Technology). Int J Comput Assist Radiol Surg 1:213CrossRef Taniguchi K, Nishikawa A, Sekimoto M, Yasui M, Takiguchi S, Seki Y, Monden M, Miyazaki F (2006) Design of a novel wearable laparoscope manipulator: SMART (Synthetic Muscle Actuator based Robotic Technology). Int J Comput Assist Radiol Surg 1:213CrossRef
10.
go back to reference Mirbagheri A, Farahmand F, Meghdari A, Karimian F (2011) Design and development of an effective low-cost robotic cameraman for laparoscopic surgery: RoboLens. Sci Iran 18(1):105–114CrossRef Mirbagheri A, Farahmand F, Meghdari A, Karimian F (2011) Design and development of an effective low-cost robotic cameraman for laparoscopic surgery: RoboLens. Sci Iran 18(1):105–114CrossRef
11.
go back to reference Kristin J, Geiger R, Kraus P, Klenzner T (2015) Assessment of the endoscopic range of motion for head and neck surgery using the SOLOASSIST endoscope holder. Int J Med Robot 11(4):418–423CrossRef Kristin J, Geiger R, Kraus P, Klenzner T (2015) Assessment of the endoscopic range of motion for head and neck surgery using the SOLOASSIST endoscope holder. Int J Med Robot 11(4):418–423CrossRef
12.
go back to reference Wijsman PJ, Broeders IA, Brenkman HJ, Szold A, Forgione A, Schreuder HW, Consten EC, Draaisma WA, Verheijen PM, Ruurda JP, Kaufman Y (2018) First experience with THE AUTOLAP™ SYSTEM: an image-based robotic camera steering device. Surg Endosc 32(5):2560–2566CrossRef Wijsman PJ, Broeders IA, Brenkman HJ, Szold A, Forgione A, Schreuder HW, Consten EC, Draaisma WA, Verheijen PM, Ruurda JP, Kaufman Y (2018) First experience with THE AUTOLAP™ SYSTEM: an image-based robotic camera steering device. Surg Endosc 32(5):2560–2566CrossRef
13.
go back to reference Mettler L, Ibrahim M, Jonat W (1998) One year of experience working with the aid of a robotic assistant (the voice-controlled optic holder AESOP) in gynecological endoscopic surgery. Hum Reprod (Oxford, England) 13(10):2748–2750CrossRef Mettler L, Ibrahim M, Jonat W (1998) One year of experience working with the aid of a robotic assistant (the voice-controlled optic holder AESOP) in gynecological endoscopic surgery. Hum Reprod (Oxford, England) 13(10):2748–2750CrossRef
14.
go back to reference Arezzo A, Ulmer F, Weiss O, Schurr M, Hamad M, Buess GF (2000) Experimental trial on solo surgery for minimally invasive therapy. Surg Endosc 14(10):955–959CrossRef Arezzo A, Ulmer F, Weiss O, Schurr M, Hamad M, Buess GF (2000) Experimental trial on solo surgery for minimally invasive therapy. Surg Endosc 14(10):955–959CrossRef
15.
go back to reference Holländer SW, Klingen HJ, Fritz M, Djalali P, Birk D (2014) Robotic camera assistance and its benefit in 1033 traditional laparoscopic procedures: prospective clinical trial using a joystick-guided camera holder. Surg Technol Int 25:19–23PubMed Holländer SW, Klingen HJ, Fritz M, Djalali P, Birk D (2014) Robotic camera assistance and its benefit in 1033 traditional laparoscopic procedures: prospective clinical trial using a joystick-guided camera holder. Surg Technol Int 25:19–23PubMed
16.
go back to reference Ozaki R, Kumakiri J, Kikuchi I, Kitade M, Matsuoka S, Jinushi M, Kono A, Takeda S (2011) Beneficial use of ENDOCAMELEON a novel optical devise for single port laparoscopic surgery for large ovarian cyst. J Minim Invasive Gynecol 18(6):S113CrossRef Ozaki R, Kumakiri J, Kikuchi I, Kitade M, Matsuoka S, Jinushi M, Kono A, Takeda S (2011) Beneficial use of ENDOCAMELEON a novel optical devise for single port laparoscopic surgery for large ovarian cyst. J Minim Invasive Gynecol 18(6):S113CrossRef
18.
go back to reference Li Z, Oo MZ, Nalam V, Thang VD, Ren H, Kofidis T, Yu H (2016) Design of a novel flexible endoscope—cardioscope. J Mech Robot 8(5):051014CrossRef Li Z, Oo MZ, Nalam V, Thang VD, Ren H, Kofidis T, Yu H (2016) Design of a novel flexible endoscope—cardioscope. J Mech Robot 8(5):051014CrossRef
19.
go back to reference Ko SY, Kim J, Lee WJ, Kwon DS (2007) Compact laparoscopic assistant robot using a bending mechanism. ADV Robot 21(5–6):689–709CrossRef Ko SY, Kim J, Lee WJ, Kwon DS (2007) Compact laparoscopic assistant robot using a bending mechanism. ADV Robot 21(5–6):689–709CrossRef
20.
go back to reference Luo RC, Wang J, Chang CK, Perng YW (2014, August) Surgeon’s third hand: an assistive robot endoscopic system with intuitive maneuverability for laparoscopic surgery. In: 5th Proceedings of the IEEE RAS EMBS International Conference Biomedical Robotics and Biomechatronics, pp 138–143 Luo RC, Wang J, Chang CK, Perng YW (2014, August) Surgeon’s third hand: an assistive robot endoscopic system with intuitive maneuverability for laparoscopic surgery. In: 5th Proceedings of the IEEE RAS EMBS International Conference Biomedical Robotics and Biomechatronics, pp 138–143
21.
go back to reference Kazanzides P, Chen Z, Deguet A, Fischer GS, Taylor RH, DiMaio SP (2014, May) An open-source research kit for the da Vinci® Surgical System. In: 2014 IEEE International Conference on Robotics and Automation, pp 6434–6439 Kazanzides P, Chen Z, Deguet A, Fischer GS, Taylor RH, DiMaio SP (2014, May) An open-source research kit for the da Vinci® Surgical System. In: 2014 IEEE International Conference on Robotics and Automation, pp 6434–6439
22.
go back to reference Li Z, Du R (2013) Design and analysis of a bio-inspired wire-driven multi-section flexible robot. IJARS 10(4):209 Li Z, Du R (2013) Design and analysis of a bio-inspired wire-driven multi-section flexible robot. IJARS 10(4):209
23.
go back to reference Li Z, Ren H, Chiu PWY, Du R, Yu H (2016) A novel constrained wire-driven flexible mechanism and its kinematic analysis. MMT 95:59–75 Li Z, Ren H, Chiu PWY, Du R, Yu H (2016) A novel constrained wire-driven flexible mechanism and its kinematic analysis. MMT 95:59–75
24.
go back to reference Ma X, Song C, Chiu WYP, Li Z (2019) Autonomous flexible endoscope for minimally invasive surgery with enhanced safety. IEEE Robot Autom Lett (RA-L) 4(3):2607–2613CrossRef Ma X, Song C, Chiu WYP, Li Z (2019) Autonomous flexible endoscope for minimally invasive surgery with enhanced safety. IEEE Robot Autom Lett (RA-L) 4(3):2607–2613CrossRef
25.
go back to reference Wang Z, Liu Z, Ma Q, Cheng A, Liu YH, Kim S, Deguet A, Reiter A, Kazanzides P, Taylor RH (2017) Vision-based calibration of dual RCM-based robot arms in human-robot collaborative minimally invasive surgery. IEEE Robot Autom Lett (RA-L) 3(2):672–679CrossRef Wang Z, Liu Z, Ma Q, Cheng A, Liu YH, Kim S, Deguet A, Reiter A, Kazanzides P, Taylor RH (2017) Vision-based calibration of dual RCM-based robot arms in human-robot collaborative minimally invasive surgery. IEEE Robot Autom Lett (RA-L) 3(2):672–679CrossRef
26.
go back to reference O'Donnell CRD, Thomas Eggemeier F (1986) Workload assessment methodology. In: Boff KR, Kaufman L, Thomas JP (eds) Handbook of perception and human performance. Cognitive processes and performance, vol II. Wiley-interscience, New York, p 42-1–42-43 O'Donnell CRD, Thomas Eggemeier F (1986) Workload assessment methodology. In: Boff KR, Kaufman L, Thomas JP (eds) Handbook of perception and human performance. Cognitive processes and performance, vol II. Wiley-interscience, New York, p 42-1–42-43
27.
go back to reference Hart SG, Staveland LE (1988) Development of NASA-TLX (Task Load Index): results of empirical and theoretical research. Adv Psychol 52:139–183CrossRef Hart SG, Staveland LE (1988) Development of NASA-TLX (Task Load Index): results of empirical and theoretical research. Adv Psychol 52:139–183CrossRef
28.
go back to reference Yurko YY, Scerbo MW, Prabhu AS, Acker CE, Stefanidis D (2010) Higher mental workload is associated with poorer laparoscopic performance as measured by the NASA-TLX tool. Simul Healthc 5(5):267–271CrossRef Yurko YY, Scerbo MW, Prabhu AS, Acker CE, Stefanidis D (2010) Higher mental workload is associated with poorer laparoscopic performance as measured by the NASA-TLX tool. Simul Healthc 5(5):267–271CrossRef
Metadata
Title
A robotic flexible endoscope with shared autonomy: a study of mockup cholecystectomy
Authors
Chengzhi Song
Xin Ma
Xianfeng Xia
Philip Wai Yan Chiu
Charing Ching Ning Chong
Zheng Li
Publication date
01-06-2020
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 6/2020
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-019-07241-8

Other articles of this Issue 6/2020

Surgical Endoscopy 6/2020 Go to the issue