Published in: Surgical Endoscopy 2/2019

01-02-2019 | New Technology

Efficacy and safety of an internal magnet traction device for endoscopic submucosal dissection: ex vivo study in a porcine model (with video)

Authors: Akira Dobashi, Andrew C. Storm, Louis M. Wong Kee Song, Christopher J. Gostout, Jodie L. Deters, Charles A. Miller, Mary A. Knipschield, Elizabeth Rajan

Published in: Surgical Endoscopy | Issue 2/2019

Login to get access

Abstract

Background

Appropriate traction allows for safer and easier endoscopic submucosal dissection (ESD). The aim of this study was to evaluate the efficacy and safety of an internal magnet traction device (MTD) for ESD in an ex vivo porcine model.

Methods

The MTD consisted of a small neodymium magnet and a suture attached to a through-the-scope clip. A circumferential mucosal incision was completed around a 30-mm diameter template that served as the target lesion. The first MTD was deployed at the proximal edge of the lesion. A second MTD was deployed on the wall opposite the lesion. With both magnets connected, this created traction or lifting of the target lesion towards the opposing wall during submucosal dissection. Primary endpoint was comparison of submucosal dissection times between conventional ESD (C-ESD) and MTD-assisted ESD (MTD-ESD).

Results

Twenty lesions along the anterior wall, posterior wall and greater curvature were resected using either C-ESD or MTD-ESD. The submucosal dissection time in MTD-ESD was significantly shorter than C-ESD (median: 6.4 [interquartile range {IQR} 4.6–8.7] min vs. 14.4 min [IQR 11.8–18.0], p < 0.05). There was a significant difference between MTD-ESD and C-ESD in total procedure times for lesions on the posterior gastric wall and greater curvature (median: 23.0 min [IQR 21.1–24.5] vs. 29.2 min [IQR 24.8–33.2], p < 0.05) with no difference for lesions on the anterior gastric wall (median: 18.8 min [IQR 15.5–20.5] vs. 17.1 min [IQR 13.1–20.0], p = 0.5). The number of muscularis propria injuries per lesion was significantly lower in MTD-ESD than C-ESD (median: 0 [IQR 0–0] vs. 1 [IQR 0–2], p < 0.05).

Conclusions

MTD for ESD is effective and safe when compared to C-ESD. This approach significantly reduced submucosal dissection times with less injury to the muscularis propria. Furthermore, MTD-ESD was particularly beneficial for more challenging gastric lesions located on the posterior wall and greater curvature.
Appendix
Available only for authorised users
Literature
1.
go back to reference Ono H, Kondo H, Gotoda T, Shirao K, Yamaguchi H, Saito D, Hosokawa K, Shimoda T, Yoshida S (2001) Endoscopic mucosal resection for treatment of early gastric cancer. Gut 48(2):225–229CrossRefPubMedPubMedCentral Ono H, Kondo H, Gotoda T, Shirao K, Yamaguchi H, Saito D, Hosokawa K, Shimoda T, Yoshida S (2001) Endoscopic mucosal resection for treatment of early gastric cancer. Gut 48(2):225–229CrossRefPubMedPubMedCentral
6.
go back to reference Fujiya M, Tanaka K, Dokoshi T, Tominaga M, Ueno N, Inaba Y, Ito T, Moriichi K, Kohgo Y (2015) Efficacy and adverse events of EMR and endoscopic submucosal dissection for the treatment of colon neoplasms: a meta-analysis of studies comparing EMR and endoscopic submucosal dissection. Gastrointest Endosc 81(3):583–595. https://doi.org/10.1016/j.gie.2014.07.034 CrossRefPubMed Fujiya M, Tanaka K, Dokoshi T, Tominaga M, Ueno N, Inaba Y, Ito T, Moriichi K, Kohgo Y (2015) Efficacy and adverse events of EMR and endoscopic submucosal dissection for the treatment of colon neoplasms: a meta-analysis of studies comparing EMR and endoscopic submucosal dissection. Gastrointest Endosc 81(3):583–595. https://​doi.​org/​10.​1016/​j.​gie.​2014.​07.​034 CrossRefPubMed
13.
go back to reference Sakamoto N, Osada T, Shibuya T, Beppu K, Matsumoto K, Shimada Y, Konno A, Kurosawa A, Nagahara A, Ohkusa T, Ogihara T, Watanabe S (2008) The facilitation of a new traction device (S-O clip) assisting endoscopic submucosal dissection for superficial colorectal neoplasms. Endoscopy 40(Suppl 2):E94–E95. https://doi.org/10.1055/s-2007-995603 CrossRefPubMed Sakamoto N, Osada T, Shibuya T, Beppu K, Matsumoto K, Shimada Y, Konno A, Kurosawa A, Nagahara A, Ohkusa T, Ogihara T, Watanabe S (2008) The facilitation of a new traction device (S-O clip) assisting endoscopic submucosal dissection for superficial colorectal neoplasms. Endoscopy 40(Suppl 2):E94–E95. https://​doi.​org/​10.​1055/​s-2007-995603 CrossRefPubMed
16.
go back to reference Imaeda H, Iwao Y, Ogata H, Ichikawa H, Mori M, Hosoe N, Masaoka T, Nakashita M, Suzuki H, Inoue N, Aiura K, Nagata H, Kumai K, Hibi T (2006) A new technique for endoscopic submucosal dissection for early gastric cancer using an external grasping forceps. Endoscopy 38(10):1007–1010. https://doi.org/10.1055/s-2006-925264 CrossRefPubMed Imaeda H, Iwao Y, Ogata H, Ichikawa H, Mori M, Hosoe N, Masaoka T, Nakashita M, Suzuki H, Inoue N, Aiura K, Nagata H, Kumai K, Hibi T (2006) A new technique for endoscopic submucosal dissection for early gastric cancer using an external grasping forceps. Endoscopy 38(10):1007–1010. https://​doi.​org/​10.​1055/​s-2006-925264 CrossRefPubMed
Metadata
Title
Efficacy and safety of an internal magnet traction device for endoscopic submucosal dissection: ex vivo study in a porcine model (with video)
Authors
Akira Dobashi
Andrew C. Storm
Louis M. Wong Kee Song
Christopher J. Gostout
Jodie L. Deters
Charles A. Miller
Mary A. Knipschield
Elizabeth Rajan
Publication date
01-02-2019
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 2/2019
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-018-6486-4

Other articles of this Issue 2/2019

Surgical Endoscopy 2/2019 Go to the issue