Skip to main content
Top
Published in: Surgical Endoscopy 10/2018

Open Access 01-10-2018 | Review Article

Safety of orogastric tubes in foregut and bariatric surgery

Authors: Kulvir Nandra, Richard Ing

Published in: Surgical Endoscopy | Issue 10/2018

Login to get access

Abstract

Background

Orogastric tubes have traditionally aided foregut procedures with sizing and organ protection. The rise of bariatric surgery has led to the creation of novel medical devices aimed at facilitating the laparoscopic sleeve gastrectomy. While approved by the FDA, the long-term safety profile of these devices in the general population is often unknown. This review looks at complications associated with novel Orogastric Tubes compared to the traditional bougie.

Methods

We performed a review of the Food and Drug Administration’s (FDA) Manufacturer and User Facility Device Experience (MAUDE) database for complications associated with the traditional bougie, Boehringer Labs ViSiGi 3D® and the Medtronic GastriSail™ since 2011. In addition, we looked for reported cases in the literature of complications with these devices.

Results

Overall complication rates reported in the MAUDE database varied in number and severity. The bougie had seven reported complications, one of which was an organ perforation. The ViSiGi 3D® had zero reported complications. The GastriSail™ had 36 total reported complications with 17 perforations. A literature review shows that rates of bougie complications are extremely rare with no case reports or reviews of complications from the novel orogastric tubes.

Conclusions

The complication rates between the traditional bougie and novel devices vary in number and severity, with the GastriSail™ having the highest reported complication rate. Despite rigorous testing for FDA approval, ongoing research into performance of new medical devices in the general population remains important.
Literature
1.
go back to reference Gagner M, Deitel M, Erickson AL, Crosby RD (2013). Survey on laparoscopic sleeve gastrectomy (LSG) at the Fourth International Consensus Summit on Sleeve Gastrectomy. Obes Surg 23(12):2013CrossRef Gagner M, Deitel M, Erickson AL, Crosby RD (2013). Survey on laparoscopic sleeve gastrectomy (LSG) at the Fourth International Consensus Summit on Sleeve Gastrectomy. Obes Surg 23(12):2013CrossRef
2.
go back to reference Gagner M, Huang RY (2016) Comparison between orogastric tube/bougie and a suction calibration system for effects on operative duration, staple-line corkscrewing, and esophageal perforation during laparoscopic sleeve gastrectomy. Surg Endosc 30(4):1648–1655CrossRef Gagner M, Huang RY (2016) Comparison between orogastric tube/bougie and a suction calibration system for effects on operative duration, staple-line corkscrewing, and esophageal perforation during laparoscopic sleeve gastrectomy. Surg Endosc 30(4):1648–1655CrossRef
4.
go back to reference Theodorou D, Doulami G, Larentzakis A, Almpanopoulos K, Stamou K, Zografos G, Menenakos E (2012) Bougie insertion: a common practice with underestimated dangers. Int J Surg Case Rep 3(2):74–77CrossRef Theodorou D, Doulami G, Larentzakis A, Almpanopoulos K, Stamou K, Zografos G, Menenakos E (2012) Bougie insertion: a common practice with underestimated dangers. Int J Surg Case Rep 3(2):74–77CrossRef
5.
go back to reference Zhang LP, Chang R, Matthews BD, Awad M, Meyers B, Eagon JC, Brunt LM (2014) Incidence, mechanisms, and outcomes of esophageal and gastric perforation during laparoscopic foregut surgery: a retrospective review of 1223 foregut cases. Surg Endosc 28(1):85–90CrossRef Zhang LP, Chang R, Matthews BD, Awad M, Meyers B, Eagon JC, Brunt LM (2014) Incidence, mechanisms, and outcomes of esophageal and gastric perforation during laparoscopic foregut surgery: a retrospective review of 1223 foregut cases. Surg Endosc 28(1):85–90CrossRef
6.
go back to reference Martines G, Digennaro R, De Fazio M, Capuano P (2017) Endoscopic management of esophageal perforation caused by calibrating bougie during laparoscopic sleeve gastrectomy. Case Stud Surg 3(2):13CrossRef Martines G, Digennaro R, De Fazio M, Capuano P (2017) Endoscopic management of esophageal perforation caused by calibrating bougie during laparoscopic sleeve gastrectomy. Case Stud Surg 3(2):13CrossRef
7.
go back to reference Schauer PR, Meyers WC, Eubanks S, Norem RF, Franklin M, Pappas TN (1996) Mechanisms of gastric and esophageal perforations during laparoscopic Nissen fundoplication. Ann Surg 223(1):43CrossRef Schauer PR, Meyers WC, Eubanks S, Norem RF, Franklin M, Pappas TN (1996) Mechanisms of gastric and esophageal perforations during laparoscopic Nissen fundoplication. Ann Surg 223(1):43CrossRef
Metadata
Title
Safety of orogastric tubes in foregut and bariatric surgery
Authors
Kulvir Nandra
Richard Ing
Publication date
01-10-2018
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 10/2018
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-018-6269-y

Other articles of this Issue 10/2018

Surgical Endoscopy 10/2018 Go to the issue