Feasibility and safety of a novel magnetic-assisted capsule endoscope system in a preliminary examination for upper gastrointestinal tract

Authors: Gi-Shih Lien, Ming-Shun Wu, Chun-Nan Chen, Chih-Wen Liu, Fat-Moon Suk

Published in: Surgical Endoscopy | Issue 4/2018

Login to get access

Abstract

Background and study aim

Current capsule endoscopy procedures are ineffective for upper gastrointestinal (GI) tract examination because they do not allow for operator-controlled navigation of the capsule. External controllability of a capsule endoscope with an applied magnetic field is a possible solution to this problem. We developed a novel magnetic-assisted capsule endoscope (MACE) system to visualize the entire upper GI tract. The present study evaluated the safety and feasibility of the MACE system for the examination of the upper GI tract, including the esophagus, stomach, and duodenum.

Methods

The present open clinical study enrolled ten healthy volunteers. All participants swallowed a MACE, and an external magnetic field navigator was used for magnetic capsule manipulation in the upper GI tract. We assessed the maneuverability of the magnetic capsule and completeness of the MACE examination as well as the safety and tolerability of the procedure.

Results

The present study enrolled ten healthy volunteers with a mean age and body mass index of 47.7 years and 25.6 kg/m2, respectively. One volunteer withdrew because of difficulty in swallowing the capsule. In total, nine volunteers underwent the MACE examination. The average examination time was 27.1 min. The maneuverability of the capsule was assessed as good and fair in 55.6 and 44.4% of the participants, respectively. The overall completeness of the examination in the esophagus, stomach, and duodenum was 100, 85.2, and 86.1%, respectively. No severe adverse events occurred during this study. All participants exhibited satisfactory tolerance of the MACE examination.

Conclusion

The MACE system has satisfactory maneuverability and visualization completeness with excellent acceptance and tolerance.
Appendix
Available only for authorised users
Literature
2.
go back to reference Carpi F, Galbiati S, Carpi A (2006) Magnetic shells for gastrointestinal endoscopic capsules as a means to control their motion. Biomed Pharmacother 60:370–374CrossRefPubMed Carpi F, Galbiati S, Carpi A (2006) Magnetic shells for gastrointestinal endoscopic capsules as a means to control their motion. Biomed Pharmacother 60:370–374CrossRefPubMed
3.
go back to reference Rahman I, Pioche M, Shim CS, Lee SP, Sung IK, Saurin JC, Patel P (2016) Magnetic-assisted capsule endoscopy in the upper GI tract by using a novel navigation system (with video). Gastrointest Endosc 83:889–895 e881CrossRefPubMed Rahman I, Pioche M, Shim CS, Lee SP, Sung IK, Saurin JC, Patel P (2016) Magnetic-assisted capsule endoscopy in the upper GI tract by using a novel navigation system (with video). Gastrointest Endosc 83:889–895 e881CrossRefPubMed
4.
go back to reference Keller J, Fibbe C, Volke F, Gerber J, Mosse AC, Reimann-Zawadzki M, Rabinovitz E, Layer P, Swain P (2010) Remote magnetic control of a wireless capsule endoscope in the esophagus is safe and feasible: results of a randomized, clinical trial in healthy volunteers. Gastrointest Endosc 72:941–946CrossRefPubMed Keller J, Fibbe C, Volke F, Gerber J, Mosse AC, Reimann-Zawadzki M, Rabinovitz E, Layer P, Swain P (2010) Remote magnetic control of a wireless capsule endoscope in the esophagus is safe and feasible: results of a randomized, clinical trial in healthy volunteers. Gastrointest Endosc 72:941–946CrossRefPubMed
5.
go back to reference Keller J, Fibbe C, Rosien U, Layer P (2012) Recent advances in capsule endoscopy: development of maneuverable capsules. Expert Rev Gastroenterol Hepatol 6:561–566CrossRefPubMed Keller J, Fibbe C, Rosien U, Layer P (2012) Recent advances in capsule endoscopy: development of maneuverable capsules. Expert Rev Gastroenterol Hepatol 6:561–566CrossRefPubMed
6.
go back to reference Keller J, Fibbe C, Volke F, Gerber J, Mosse AC, Reimann-Zawadzki M, Rabinovitz E, Layer P, Schmitt D, Andresen V, Rosien U, Swain P (2011) Inspection of the human stomach using remote-controlled capsule endoscopy: a feasibility study in healthy volunteers (with videos). Gastrointest Endosc 73:22–28CrossRefPubMed Keller J, Fibbe C, Volke F, Gerber J, Mosse AC, Reimann-Zawadzki M, Rabinovitz E, Layer P, Schmitt D, Andresen V, Rosien U, Swain P (2011) Inspection of the human stomach using remote-controlled capsule endoscopy: a feasibility study in healthy volunteers (with videos). Gastrointest Endosc 73:22–28CrossRefPubMed
7.
go back to reference Swain P, Toor A, Volke F, Keller J, Gerber J, Rabinovitz E, Rothstein RI (2010) Remote magnetic manipulation of a wireless capsule endoscope in the esophagus and stomach of humans (with videos). Gastrointest Endosc 71:1290–1293CrossRefPubMed Swain P, Toor A, Volke F, Keller J, Gerber J, Rabinovitz E, Rothstein RI (2010) Remote magnetic manipulation of a wireless capsule endoscope in the esophagus and stomach of humans (with videos). Gastrointest Endosc 71:1290–1293CrossRefPubMed
8.
go back to reference Rey JF, Ogata H, Hosoe N, Ohtsuka K, Ogata N, Ikeda K, Aihara H, Pangtay I, Hibi T, Kudo SE, Tajiri H (2012) Blinded nonrandomized comparative study of gastric examination with a magnetically guided capsule endoscope and standard video endoscope. Gastrointest Endosc 75:373–381CrossRefPubMed Rey JF, Ogata H, Hosoe N, Ohtsuka K, Ogata N, Ikeda K, Aihara H, Pangtay I, Hibi T, Kudo SE, Tajiri H (2012) Blinded nonrandomized comparative study of gastric examination with a magnetically guided capsule endoscope and standard video endoscope. Gastrointest Endosc 75:373–381CrossRefPubMed
9.
go back to reference Rey JF, Ogata H, Hosoe N, Ohtsuka K, Ogata N, Ikeda K, Aihara H, Pangtay I, Hibi T, Kudo S, Tajiri H (2010) Feasibility of stomach exploration with a guided capsule endoscope. Endoscopy 42:541–545CrossRefPubMed Rey JF, Ogata H, Hosoe N, Ohtsuka K, Ogata N, Ikeda K, Aihara H, Pangtay I, Hibi T, Kudo S, Tajiri H (2010) Feasibility of stomach exploration with a guided capsule endoscope. Endoscopy 42:541–545CrossRefPubMed
10.
go back to reference Denzer UW, Rosch T, Hoytat B, Abdel-Hamid M, Hebuterne X, Vanbiervielt G, Filippi J, Ogata H, Hosoe N, Ohtsuka K, Ogata N, Ikeda K, Aihara H, Kudo SE, Tajiri H, Treszl A, Wegscheider K, Greff M, Rey JF (2015) Magnetically guided capsule versus conventional gastroscopy for upper abdominal complaints: a prospective blinded study. J Clin Gastroenterol 49:101–107CrossRefPubMed Denzer UW, Rosch T, Hoytat B, Abdel-Hamid M, Hebuterne X, Vanbiervielt G, Filippi J, Ogata H, Hosoe N, Ohtsuka K, Ogata N, Ikeda K, Aihara H, Kudo SE, Tajiri H, Treszl A, Wegscheider K, Greff M, Rey JF (2015) Magnetically guided capsule versus conventional gastroscopy for upper abdominal complaints: a prospective blinded study. J Clin Gastroenterol 49:101–107CrossRefPubMed
11.
go back to reference Liao Z, Duan XD, Xin L, Bo LM, Wang XH, Xiao GH, Hu LH, Zhuang SL, Li ZS (2012) Feasibility and safety of magnetic-controlled capsule endoscopy system in examination of human stomach: a pilot study in healthy volunteers. J Interv Gastroenterol 2:155–160CrossRefPubMedPubMedCentral Liao Z, Duan XD, Xin L, Bo LM, Wang XH, Xiao GH, Hu LH, Zhuang SL, Li ZS (2012) Feasibility and safety of magnetic-controlled capsule endoscopy system in examination of human stomach: a pilot study in healthy volunteers. J Interv Gastroenterol 2:155–160CrossRefPubMedPubMedCentral
12.
go back to reference Zou WB, Hou XH, Xin L, Liu J, Bo LM, Yu GY, Liao Z, Li ZS (2015) Magnetic-controlled capsule endoscopy vs. gastroscopy for gastric diseases: a two-center self-controlled comparative trial. Endoscopy 47:525–528CrossRefPubMed Zou WB, Hou XH, Xin L, Liu J, Bo LM, Yu GY, Liao Z, Li ZS (2015) Magnetic-controlled capsule endoscopy vs. gastroscopy for gastric diseases: a two-center self-controlled comparative trial. Endoscopy 47:525–528CrossRefPubMed
13.
go back to reference Lien GS, Liu CW, Jiang JA, Chuang CL, Teng MT (2012) Magnetic control system targeted for capsule endoscopic operations in the stomach–design, fabrication, and in vitro and ex vivo evaluations. IEEE Trans Biomed Eng 59:2068–2079CrossRefPubMed Lien GS, Liu CW, Jiang JA, Chuang CL, Teng MT (2012) Magnetic control system targeted for capsule endoscopic operations in the stomach–design, fabrication, and in vitro and ex vivo evaluations. IEEE Trans Biomed Eng 59:2068–2079CrossRefPubMed
14.
go back to reference Chang CC, Chen SH, Lin CP, Hsieh CR, Lou HY, Suk FM, Pan S, Wu MS, Chen JN, Chen YF (2007) Premedication with pronase or N-acetylcysteine improves visibility during gastroendoscopy: an endoscopist-blinded, prospective, randomized study. World J Gastroenterol 13:444–447CrossRefPubMedPubMedCentral Chang CC, Chen SH, Lin CP, Hsieh CR, Lou HY, Suk FM, Pan S, Wu MS, Chen JN, Chen YF (2007) Premedication with pronase or N-acetylcysteine improves visibility during gastroendoscopy: an endoscopist-blinded, prospective, randomized study. World J Gastroenterol 13:444–447CrossRefPubMedPubMedCentral
15.
go back to reference Liu S, Li J, Lv Y (2012) Gastrointestinal damage caused by swallowing multiple magnets. Front Med 6:280–287CrossRefPubMed Liu S, Li J, Lv Y (2012) Gastrointestinal damage caused by swallowing multiple magnets. Front Med 6:280–287CrossRefPubMed
Metadata
Title
Feasibility and safety of a novel magnetic-assisted capsule endoscope system in a preliminary examination for upper gastrointestinal tract
Authors
Gi-Shih Lien
Ming-Shun Wu
Chun-Nan Chen
Chih-Wen Liu
Fat-Moon Suk
Publication date
01-04-2018
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 4/2018
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-017-5887-0

Other articles of this Issue 4/2018

Surgical Endoscopy 4/2018 Go to the issue