Computer-assessed performance of psychomotor skills in endoscopic otolaryngology surgery: construct validity of the Dundee Endoscopic Psychomotor Otolaryngology Surgery Trainer (DEPOST)

Authors: Peter D. Ross, Richard Steven, Dong Zhang, Heng Li, Eric W. Abel

Published in: Surgical Endoscopy | Issue 11/2015

Login to get access

Abstract

Background

This study was undertaken to introduce and establish the value of the Dundee Endoscopic Psychomotor Otolaryngology Surgery Trainer (DEPOST) as a customisable, objective real-time scoring system for trainee assessment. The construct validity of the system was assessed by comparing the performance of experienced otolaryngologists with that of otolaryngology trainees, junior doctors and medical students.

Methods

Forty two subjects (13 Consultants, 8 senior trainees, 13 junior trainees and 8 junior doctors/medical students) completed a single test on DEPOST. The test involved using a 30° rigid endoscope and a probe with position sensor, to identify a series of lights in a complex 3-dimensional model. The system scored subjects for time, success rate, and economy of movement (distance travelled). An analysis of variance and correlation analysis were used for the data analysis, with statistical significance set at 0.05.

Results

Increasing experience led to significantly improved performance with the DEPOST (p < 0.01). Senior trainees’ results were significantly better than those of consultant otolaryngologists in success rate and time (p < 0.05 & p < 0.05). Consultants were the most efficient in their movement (p = 0.051)

Conclusions

The system provides an accurate and customisable assessment of endoscopic skill in otolaryngologists. The DEPOST system has construct validity, with master surgeons and senior trainees completing the tasks more accurately without sacrificing execution time, success rate or efficiency of movement.
Literature
3.
4.
go back to reference Tansley P, Kakar S, Withey S, Butler P (2007) Visuospatial and technical ability in the selection and assessment of higher surgical trainees in the London Deanery. Ann R Coll Surg Engl 89:591–595PubMedCentralCrossRefPubMed Tansley P, Kakar S, Withey S, Butler P (2007) Visuospatial and technical ability in the selection and assessment of higher surgical trainees in the London Deanery. Ann R Coll Surg Engl 89:591–595PubMedCentralCrossRefPubMed
5.
go back to reference Frances NK, Hanna GB, Cuschieri A (2001) Reliability of the advanced dundee endoscopic psychomotor tester for bimanual tasks. Arch Surg 136(1):40–44CrossRef Frances NK, Hanna GB, Cuschieri A (2001) Reliability of the advanced dundee endoscopic psychomotor tester for bimanual tasks. Arch Surg 136(1):40–44CrossRef
6.
go back to reference Arora H, Uribe J, Ralph W, Zeltsan M, Cuellar H, Gallagher A, Fried MP (2005) Assessment of construct validity of the endoscopic sinus surgery simulator. Arch Otolaryngol Head Neck Surg 131(3):217–221CrossRefPubMed Arora H, Uribe J, Ralph W, Zeltsan M, Cuellar H, Gallagher A, Fried MP (2005) Assessment of construct validity of the endoscopic sinus surgery simulator. Arch Otolaryngol Head Neck Surg 131(3):217–221CrossRefPubMed
7.
go back to reference Weghorst S, Airola C, Oppenheimer P, Edmond CV, Patience T, Heskamp D, Miller J (1998) Validation of the madigan ESS simulator. Stud Health Technol Inform 50:399–405PubMed Weghorst S, Airola C, Oppenheimer P, Edmond CV, Patience T, Heskamp D, Miller J (1998) Validation of the madigan ESS simulator. Stud Health Technol Inform 50:399–405PubMed
8.
go back to reference Hanna GB, Drew T, Clinch P, Hunter B, Shimi S, Dunkley MP, Cuschieri A (1996) A microrpocessor-controlled psychomotor tester for minimal access surgery. Surg Endoscopy 10:965–969CrossRef Hanna GB, Drew T, Clinch P, Hunter B, Shimi S, Dunkley MP, Cuschieri A (1996) A microrpocessor-controlled psychomotor tester for minimal access surgery. Surg Endoscopy 10:965–969CrossRef
9.
go back to reference Francis NK, Hanna GB, Cuschieri A (2002) The performance of master surgeons on the advanced dundee endoscopic psychomotor tester. Contrast validity study. Arch Surg 137(7):841–844CrossRefPubMed Francis NK, Hanna GB, Cuschieri A (2002) The performance of master surgeons on the advanced dundee endoscopic psychomotor tester. Contrast validity study. Arch Surg 137(7):841–844CrossRefPubMed
10.
go back to reference Macmillan AIM, Cuschieri A (1999) Assessment of innate ability and skills for endoscopic manipulations by the Advanced Dundee Endoscopic Psychomotor Tester: predictive and concurrent validity. Am J Surg 177(3):274–277CrossRefPubMed Macmillan AIM, Cuschieri A (1999) Assessment of innate ability and skills for endoscopic manipulations by the Advanced Dundee Endoscopic Psychomotor Tester: predictive and concurrent validity. Am J Surg 177(3):274–277CrossRefPubMed
11.
go back to reference Schijven MP, Jakimowicz J, Schot C (2002) The Advanced Dundee Endoscopic Psychomotor Tester (ADEPT) objectifying subjective psychomotor test performance. Surg Endosc 16(6):943–948CrossRefPubMed Schijven MP, Jakimowicz J, Schot C (2002) The Advanced Dundee Endoscopic Psychomotor Tester (ADEPT) objectifying subjective psychomotor test performance. Surg Endosc 16(6):943–948CrossRefPubMed
12.
go back to reference Breatnach E, Abbott GC, Fraser RG (1984) Dimensions of the normal human trachea. Am J Roentgenol 141:903–906CrossRef Breatnach E, Abbott GC, Fraser RG (1984) Dimensions of the normal human trachea. Am J Roentgenol 141:903–906CrossRef
13.
go back to reference Meskers CGM, Fraterman H, van der Helm FC, Vermeulen HM, Rozing PM (1999) Calibration of the “Flock of Birds’’ electromagnetic tracking device. J Biomech 32:629–633CrossRefPubMed Meskers CGM, Fraterman H, van der Helm FC, Vermeulen HM, Rozing PM (1999) Calibration of the “Flock of Birds’’ electromagnetic tracking device. J Biomech 32:629–633CrossRefPubMed
14.
go back to reference Ramsey PG, Carline JD, Inui TS, Larson EB, LoGerfo JP, Norcini JJ, Wenrich MD (1991) Changes over time in the knowledge base of practicing internists. JAMA 266:1103–1107CrossRefPubMed Ramsey PG, Carline JD, Inui TS, Larson EB, LoGerfo JP, Norcini JJ, Wenrich MD (1991) Changes over time in the knowledge base of practicing internists. JAMA 266:1103–1107CrossRefPubMed
15.
go back to reference Stolley PD, Becker MH, Lasagna L, McEvilla JD, Sloane LM (1972) The relationship between physician characteristics and prescribing appropriateness. Med Care 10:17–28CrossRefPubMed Stolley PD, Becker MH, Lasagna L, McEvilla JD, Sloane LM (1972) The relationship between physician characteristics and prescribing appropriateness. Med Care 10:17–28CrossRefPubMed
16.
go back to reference Waljee JF, Greenfield LJ, Dimick JB, Birkmeyer JD (2006) Surgeon age and operative mortality in the United States. Ann Surg 244(3):353–362PubMedCentralPubMed Waljee JF, Greenfield LJ, Dimick JB, Birkmeyer JD (2006) Surgeon age and operative mortality in the United States. Ann Surg 244(3):353–362PubMedCentralPubMed
17.
go back to reference Duclos A, Peix J, Colin C, Kraimps J, Menegaux F, Pattou F (2012) Influence of experience on performance of individual surgeons in thyroid surgery: prospective cross sectional multicentre study. BMJ 34:344 Duclos A, Peix J, Colin C, Kraimps J, Menegaux F, Pattou F (2012) Influence of experience on performance of individual surgeons in thyroid surgery: prospective cross sectional multicentre study. BMJ 34:344
18.
go back to reference Abel EW, Zhuo Y, Ross PD, White PS (2013) Automatic glare removal in endoscopic imaging. Surg Endosc 28(2):584–591CrossRefPubMed Abel EW, Zhuo Y, Ross PD, White PS (2013) Automatic glare removal in endoscopic imaging. Surg Endosc 28(2):584–591CrossRefPubMed
19.
go back to reference White PS, Nassif R, Saleh H, Drew T (2004) Pilot study of a device for measuring instrument forces during endoscopic sinus surgery. Acta Otolaryngol 124(2):176–178CrossRefPubMed White PS, Nassif R, Saleh H, Drew T (2004) Pilot study of a device for measuring instrument forces during endoscopic sinus surgery. Acta Otolaryngol 124(2):176–178CrossRefPubMed
20.
go back to reference Joice P, Ross PD, Wang D, Abel EW, White PS (2012) Measurement of osteotomy force during endoscopic sinus surgery. Allergy Rhinol (Providence) 3(2):e61–e65CrossRef Joice P, Ross PD, Wang D, Abel EW, White PS (2012) Measurement of osteotomy force during endoscopic sinus surgery. Allergy Rhinol (Providence) 3(2):e61–e65CrossRef
21.
go back to reference Ross PD, Joice P, Nassif RG, White PS (2010) Surgical instrument force exerted during endoscopic sinus surgery: differences within the ethmoid sinus complex. Am J Rhinol Allergy 24(1):76–80CrossRefPubMed Ross PD, Joice P, Nassif RG, White PS (2010) Surgical instrument force exerted during endoscopic sinus surgery: differences within the ethmoid sinus complex. Am J Rhinol Allergy 24(1):76–80CrossRefPubMed
22.
go back to reference Sheedy MB, Bergin M, Wylie G, Ross P, Dove R, Bird P (2012) Development of a surgical instrument for measuring forces applied to the ossicles of the middle ear. Aust Phys Eng Sci Med 35(4):503–510CrossRef Sheedy MB, Bergin M, Wylie G, Ross P, Dove R, Bird P (2012) Development of a surgical instrument for measuring forces applied to the ossicles of the middle ear. Aust Phys Eng Sci Med 35(4):503–510CrossRef
Metadata
Title
Computer-assessed performance of psychomotor skills in endoscopic otolaryngology surgery: construct validity of the Dundee Endoscopic Psychomotor Otolaryngology Surgery Trainer (DEPOST)
Authors
Peter D. Ross
Richard Steven
Dong Zhang
Heng Li
Eric W. Abel
Publication date
01-11-2015
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 11/2015
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-014-4036-2

Other articles of this Issue 11/2015

Surgical Endoscopy 11/2015 Go to the issue