Published in: Surgical Endoscopy 5/2013

01-05-2013 | Review

Instrument Failures for the da Vinci Surgical System: a Food and Drug Administration MAUDE Database Study

Authors: Diana C. W. Friedman, Thomas S. Lendvay, Blake Hannaford

Published in: Surgical Endoscopy | Issue 5/2013

Login to get access

Abstract

Background

Our goal was to analyze reported instances of the da Vinci robotic surgical system instrument failures using the FDA's MAUDE (Manufacturer and User Facility Device Experience) database. From these data we identified some root causes of failures as well as trends that may assist surgeons and users of the robotic technology.

Methods

We conducted a survey of the MAUDE database and tallied robotic instrument failures that occurred between January 2009 and December 2010. We categorized failures into five main groups (cautery, shaft, wrist or tool tip, cable, and control housing) based on technical differences in instrument design and function.

Results

A total of 565 instrument failures were documented through 528 reports. The majority of failures (285) were of the instrument’s wrist or tool tip. Cautery problems comprised 174 failures, 76 were shaft failures, 29 were cable failures, and 7 were control housing failures. Of the reports, 10 had no discernible failure mode and 49 exhibited multiple failures.

Conclusions

The data show that a number of robotic instrument failures occurred in a short period of time. In reality, many instrument failures may go unreported, thus a true failure rate cannot be determined from these data. However, education of hospital administrators, operating room staff, surgeons, and patients should be incorporated into discussions regarding the introduction and utilization of robotic technology. We recommend institutions incorporate standard failure reporting policies so that the community of robotic surgery companies and surgeons can improve on existing technologies for optimal patient safety and outcomes.
Literature
1.
go back to reference D’Annibale A, Morpurgo E, Fiscon V, Trevisan P, Sovernigo G, Orsini C, Guidolin D (2004) Robotic and laparoscopic surgery for treatment of colorectal diseases. Dis Colon Rectum 47(12):2162–2168PubMedCrossRef D’Annibale A, Morpurgo E, Fiscon V, Trevisan P, Sovernigo G, Orsini C, Guidolin D (2004) Robotic and laparoscopic surgery for treatment of colorectal diseases. Dis Colon Rectum 47(12):2162–2168PubMedCrossRef
2.
go back to reference Hubens G, Coveliers H, Balliu L, Ruppert M, Vaneerdeweg W (2003) A performance study comparing manual and robotically assisted laparoscopic surgery using the da Vinci system. Surg Endosc 17(10):1595–1599PubMedCrossRef Hubens G, Coveliers H, Balliu L, Ruppert M, Vaneerdeweg W (2003) A performance study comparing manual and robotically assisted laparoscopic surgery using the da Vinci system. Surg Endosc 17(10):1595–1599PubMedCrossRef
3.
go back to reference Kornprat P, Werkgartner G, Cerwenka H, Bacher H, El-Shabrawi A, Rehak P, Mischinger HJ (2006) Prospective study comparing standard and robotically assisted laparoscopic cholecystectomy. Langenbecks Arch Surg 391(3):216–221PubMedCrossRef Kornprat P, Werkgartner G, Cerwenka H, Bacher H, El-Shabrawi A, Rehak P, Mischinger HJ (2006) Prospective study comparing standard and robotically assisted laparoscopic cholecystectomy. Langenbecks Arch Surg 391(3):216–221PubMedCrossRef
4.
go back to reference Ham WS, Park SY, Yu HS, Choi YD, Hong SJ, Rha KH (2009) Malfunction of da Vinci Robotic System—disassembled surgeon’s console hand piece: case report and review of the literature. Urology 73(1):209.e7–209.e8CrossRef Ham WS, Park SY, Yu HS, Choi YD, Hong SJ, Rha KH (2009) Malfunction of da Vinci Robotic System—disassembled surgeon’s console hand piece: case report and review of the literature. Urology 73(1):209.e7–209.e8CrossRef
5.
go back to reference Lavery HJ, Thaly R, Albala D, Ahlering T, Shalhav A, Lee D, Fagin R, Wiklund P, Dasgupta P, Costello AJ, Tewari A, Coughlin G, Patel VR (2008) Robotic equipment malfunction during robotic prostatectomy: a multi-institutional study. J Endourol 22(9):2165–2168PubMedCrossRef Lavery HJ, Thaly R, Albala D, Ahlering T, Shalhav A, Lee D, Fagin R, Wiklund P, Dasgupta P, Costello AJ, Tewari A, Coughlin G, Patel VR (2008) Robotic equipment malfunction during robotic prostatectomy: a multi-institutional study. J Endourol 22(9):2165–2168PubMedCrossRef
6.
go back to reference Ruurda JP, Draaisma WA, van Hillegersberg R, Borel Rinkes IHM, Gooszen HG, Janssen LWM, Simmermacher RKJ, Broeders IAMJ (2005) Robot-assisted endoscopic surgery: a four-year single-center experience. Dig Surg 22(5):313–320PubMedCrossRef Ruurda JP, Draaisma WA, van Hillegersberg R, Borel Rinkes IHM, Gooszen HG, Janssen LWM, Simmermacher RKJ, Broeders IAMJ (2005) Robot-assisted endoscopic surgery: a four-year single-center experience. Dig Surg 22(5):313–320PubMedCrossRef
7.
go back to reference Zorn KC, Gofrit ON, Orvieto MA, Mikhail AA, Galocy RM, Shalhav AL, Zagaja GP (2007) Da Vinci robot error and failure rates: single institution experience on a single three-arm robot unit of more than 700 consecutive robot-assisted laparoscopic radical prostatectomies. J Endourol 21(11):1341–1344PubMedCrossRef Zorn KC, Gofrit ON, Orvieto MA, Mikhail AA, Galocy RM, Shalhav AL, Zagaja GP (2007) Da Vinci robot error and failure rates: single institution experience on a single three-arm robot unit of more than 700 consecutive robot-assisted laparoscopic radical prostatectomies. J Endourol 21(11):1341–1344PubMedCrossRef
8.
go back to reference Sorenson MD, Johnson MH, Delostrinos C, Bice JB, Grady RW, Lendvay TS (2010) Initiation of a pediatric robotic surgery program: institutional challenges and realistic outcomes. Surg Endosc 24(11):2803–2808CrossRef Sorenson MD, Johnson MH, Delostrinos C, Bice JB, Grady RW, Lendvay TS (2010) Initiation of a pediatric robotic surgery program: institutional challenges and realistic outcomes. Surg Endosc 24(11):2803–2808CrossRef
9.
go back to reference Mues AC, Boxa GN, Abazaa R (2011) Robotic instrument insulation failure: initial report of a potential source of patient injury. J Urol 77(1):104–107CrossRef Mues AC, Boxa GN, Abazaa R (2011) Robotic instrument insulation failure: initial report of a potential source of patient injury. J Urol 77(1):104–107CrossRef
10.
go back to reference Kim WT, Ham WS, Jeong W, Song HJ, Rha KH, Choi YD (2009) Failure and malfunction of da Vinci Surgical Systems during various robotic surgeries: experience from six departments at a single institute. Urology 74(6):1234–1237PubMedCrossRef Kim WT, Ham WS, Jeong W, Song HJ, Rha KH, Choi YD (2009) Failure and malfunction of da Vinci Surgical Systems during various robotic surgeries: experience from six departments at a single institute. Urology 74(6):1234–1237PubMedCrossRef
11.
go back to reference Nayyar R, Gupta NP (2010) Critical appraisal of technical problems with robotic urological surgery. Br J Urol Int 105(12):1710–1713CrossRef Nayyar R, Gupta NP (2010) Critical appraisal of technical problems with robotic urological surgery. Br J Urol Int 105(12):1710–1713CrossRef
12.
go back to reference Andonian S, Okeke Z, Okeke DA, Rastinehad A, Vanderbrink BA, Richstone L, Lee BR (2008) Device failures associated with patient injuries during robot-assisted laparoscopic surgeries: a comprehensive review of FDA MAUDE database. Can J Urol 15(1):3912–3916PubMed Andonian S, Okeke Z, Okeke DA, Rastinehad A, Vanderbrink BA, Richstone L, Lee BR (2008) Device failures associated with patient injuries during robot-assisted laparoscopic surgeries: a comprehensive review of FDA MAUDE database. Can J Urol 15(1):3912–3916PubMed
13.
go back to reference Kaushik D, High R, Clark CJ, LaGrange CA (2010) Malfunction of the da Vinci robotic system during robot-assisted laparoscopic prostatectomy: an international survey. J Endourol 24(4):571–575PubMedCrossRef Kaushik D, High R, Clark CJ, LaGrange CA (2010) Malfunction of the da Vinci robotic system during robot-assisted laparoscopic prostatectomy: an international survey. J Endourol 24(4):571–575PubMedCrossRef
14.
go back to reference Park SY, Ahn JJ, Jeong W, Ham WS, Rha KH (2010) A unique instrumental malfunction during robotic prostatectomy. Yonsei Med J 51(1):148–150PubMedCrossRef Park SY, Ahn JJ, Jeong W, Ham WS, Rha KH (2010) A unique instrumental malfunction during robotic prostatectomy. Yonsei Med J 51(1):148–150PubMedCrossRef
15.
go back to reference Park SY, Cho KS, Lee SW, Soh BH, Rha KH (2008) Intraoperative breakage of needle driver jaw during robotic-assisted laparoscopic radical prostatectomy. J Urol 71(1):168.e5–168.e6CrossRef Park SY, Cho KS, Lee SW, Soh BH, Rha KH (2008) Intraoperative breakage of needle driver jaw during robotic-assisted laparoscopic radical prostatectomy. J Urol 71(1):168.e5–168.e6CrossRef
Metadata
Title
Instrument Failures for the da Vinci Surgical System: a Food and Drug Administration MAUDE Database Study
Authors
Diana C. W. Friedman
Thomas S. Lendvay
Blake Hannaford
Publication date
01-05-2013
Publisher
Springer-Verlag
Published in
Surgical Endoscopy / Issue 5/2013
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-012-2659-8

Other articles of this Issue 5/2013

Surgical Endoscopy 5/2013 Go to the issue