Published in: Surgical Endoscopy 2/2008

01-02-2008

A novel method to localize antibody-targeted cancer deposits intraoperatively using handheld PET beta and gamma probes

Authors: Vivian E. Strong, John Humm, Paul Russo, Achim Jungbluth, W. Douglas Wong, Farhad Daghighian, Lloyd Old, Yuman Fong, Steven M. Larson

Published in: Surgical Endoscopy | Issue 2/2008

Login to get access

Abstract

Background

Assessing cancer margins, lymph nodes, and small cancer deposits intraoperatively can be challenging. A new device has become available that allows the detection of positron emission tomography (PET) radiotracers through both high-energy gamma and short-range beta emissions. These PET probes are handheld, allowing for real-time evaluation of cancer using a tool that provides surgeons with better intraoperative assessment of tumor sites.

Methods

Within the context of two institutional review board (IRB)-approved protocols investigating new applications of antibody-labeled PET scanning, 124I-labeled humanized monoclonal antibodies specific for colorectal cancer (huA33) and renal tumors (cG250) were constructed. Patients underwent preoperative PET scans, approximately seven days post-tracer infusion, when tumor-to-nontumor ratios were high. Suspected tumor deposits were evaluated intraoperatively with handheld beta and gamma PET probes.

Results

Handheld PET probes detected emissions from all tumors. Count rates from the gamma probe on tumor ranged from 48 to 306 cps, and for the beta probe ranged from 18 to 190 cps. Gamma and beta emissions exhibited a strong positive correlation. The ratio of gamma and beta counts was at least twice that of the background counts for all tumors evaluated.

Conclusions

This study is the first to demonstrate the utility of beta probes for the intraoperative detection of radiolabeled antibodies targeting cancer. Importantly, the recorded beta count rates from the beta probe correlate with the count rates from the high-energy gamma probe. Furthermore, the beta probe may offer superior specificity for real-time localization of small tumor deposits, compared to gamma probes. The intraoperative portable PET probe may prove a valuable bridge to combining tumor biology and PET technology to guide surgical therapy.
Literature
1.
go back to reference Alazraki NP, Mishkin FS (1991) Fundamentals of nuclear medicine, 2nd edition. The Society of Nuclear Medicine, Inc., ISBM? 0-932004-29-6, p 85 Alazraki NP, Mishkin FS (1991) Fundamentals of nuclear medicine, 2nd edition. The Society of Nuclear Medicine, Inc., ISBM? 0-932004-29-6, p 85
2.
go back to reference Daghighian F, Mazziotta JC, Hoffman EJ, Siegel S, Shenderov P, Eshaghian B, Phelps ME (1994) Intraoperative beta probe: A device for detecting tissue labeled with positron- or electron-emitting isotopes during surgery. Med Phys 21:153–157CrossRefPubMed Daghighian F, Mazziotta JC, Hoffman EJ, Siegel S, Shenderov P, Eshaghian B, Phelps ME (1994) Intraoperative beta probe: A device for detecting tissue labeled with positron- or electron-emitting isotopes during surgery. Med Phys 21:153–157CrossRefPubMed
3.
go back to reference Essner R, Daghighian F, Giuliano A (2002) Advances in FDG PET probes in surgical oncology. Cancer J8:100–108 Essner R, Daghighian F, Giuliano A (2002) Advances in FDG PET probes in surgical oncology. Cancer J8:100–108
4.
go back to reference Gulec SA, Dahgighian F, Essner R (2006) PET Probe: Evaluation of technical performance and clinical utility of a handheld high-energy gamma probe in oncologic surgery. Ann Surg Oncol 1–8 Gulec SA, Dahgighian F, Essner R (2006) PET Probe: Evaluation of technical performance and clinical utility of a handheld high-energy gamma probe in oncologic surgery. Ann Surg Oncol 1–8
5.
go back to reference Essner R, Hsueh E, Haigh P, Glass E, Huynh Y, Daghighian F (2001) Application of an 18F fluorodeoxyglucose-sensitive probe for the intraoperative detection of malignancy. J Surg Res 96:120–126CrossRefPubMed Essner R, Hsueh E, Haigh P, Glass E, Huynh Y, Daghighian F (2001) Application of an 18F fluorodeoxyglucose-sensitive probe for the intraoperative detection of malignancy. J Surg Res 96:120–126CrossRefPubMed
6.
go back to reference Franc B, Mari C, Johnson D, Leong S (2005) The role of a positron and high-energy gamma photon probe in intraoperative localization of recurrent melanoma. Clin Nucl Med 30:787–791CrossRefPubMed Franc B, Mari C, Johnson D, Leong S (2005) The role of a positron and high-energy gamma photon probe in intraoperative localization of recurrent melanoma. Clin Nucl Med 30:787–791CrossRefPubMed
7.
go back to reference Divgi CR, Pandit-Taskar N, Jungbluth AA, Reuter VE, Gonen M, Ruan S, Pierre C, Nagel A, Pryma DA, Humm J, Larson SM, Old LJ, Russo P (2007) Preoperative characterisation of clear-cell carcinoma using iodine-124-labelled antibody chimeric G250 (124I-cG250) and PET in patients with renal masses: a phase I trial. Lancet Oncol 8:304–310CrossRefPubMed Divgi CR, Pandit-Taskar N, Jungbluth AA, Reuter VE, Gonen M, Ruan S, Pierre C, Nagel A, Pryma DA, Humm J, Larson SM, Old LJ, Russo P (2007) Preoperative characterisation of clear-cell carcinoma using iodine-124-labelled antibody chimeric G250 (124I-cG250) and PET in patients with renal masses: a phase I trial. Lancet Oncol 8:304–310CrossRefPubMed
Metadata
Title
A novel method to localize antibody-targeted cancer deposits intraoperatively using handheld PET beta and gamma probes
Authors
Vivian E. Strong
John Humm
Paul Russo
Achim Jungbluth
W. Douglas Wong
Farhad Daghighian
Lloyd Old
Yuman Fong
Steven M. Larson
Publication date
01-02-2008
Publisher
Springer-Verlag
Published in
Surgical Endoscopy / Issue 2/2008
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-007-9611-3

Other articles of this Issue 2/2008

Surgical Endoscopy 2/2008 Go to the issue