Abstract

Background

The Gore suture passer instrument is an ideal reusable device for various laparoscopic procedures albeit with a few drawbacks. The needle is not self-retracting inside the sheath and this leads to incomplete (manual) retraction of the needle, causing bending and breakage of the needle tip. The intracorporeal orientation of the suture grasper is difficult to maintain because the needle and suture grasper is fully rotational inside the handle.

Methods

We rectified these drawbacks by inserting a spring below the connecting nut in the handle component to make the needle self-retracting. The needle and suture grasper was stabilized by a screw in the main body of the handle to fit in a longitudinal slot cut on the brass connecting nut.

Results

Our modification made the needle and suture grasper self-retracting inside the sheath and stabilize the needle during the insertion. We have used this modified suture passer on 80 patients without breakage and bending of the needle.

Conclusion

This modification has made this device user friendly, steady, and has reduced the operation time.
Literature
1.
go back to reference Costanza MJ, Heniford BT, Arca MJ, Mayes JT, Gagner M (1998) Laparoscopic repair of recurrent ventral hernias. Am Surg 64: 1121–1127PubMed Costanza MJ, Heniford BT, Arca MJ, Mayes JT, Gagner M (1998) Laparoscopic repair of recurrent ventral hernias. Am Surg 64: 1121–1127PubMed
2.
go back to reference Leblanc KA (2004) Laparoscopic incisional and ventral hernia repair: complications—how to avoid and handle. Hernia 8: 323–331PubMedCrossRef Leblanc KA (2004) Laparoscopic incisional and ventral hernia repair: complications—how to avoid and handle. Hernia 8: 323–331PubMedCrossRef
3.
go back to reference Phillips E, Dardano AN, Saxe A (1997) Laparoscopic repair of abdominal hernias using an ePTFE patch—a modification of a previously described technique. J Soc Laparoendosc Surg 1: 277–279 Phillips E, Dardano AN, Saxe A (1997) Laparoscopic repair of abdominal hernias using an ePTFE patch—a modification of a previously described technique. J Soc Laparoendosc Surg 1: 277–279
4.
go back to reference Salameh JR (2005) Suture passer tip breakage during laparoscopic ventral hernia repair. Surg Laparosc Endosc Percut Tech 15: 112–114CrossRef Salameh JR (2005) Suture passer tip breakage during laparoscopic ventral hernia repair. Surg Laparosc Endosc Percut Tech 15: 112–114CrossRef
5.
go back to reference Varghese TK, Denham DW, Dawes LG, Murayama KM, Prystowsky JB, Joehl RJ (2002) Laparoscopic ventral hernia repair: an initial institutional experience. J Surg 105: 115–118CrossRef Varghese TK, Denham DW, Dawes LG, Murayama KM, Prystowsky JB, Joehl RJ (2002) Laparoscopic ventral hernia repair: an initial institutional experience. J Surg 105: 115–118CrossRef
Metadata
Title
Modification of Gore suture passer instrument
Author
V. Golash
Publication date
01-10-2006
Publisher
Springer-Verlag
Published in
Surgical Endoscopy / Issue 10/2006
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-005-0673-9

Other articles of this Issue 10/2006

Surgical Endoscopy 10/2006 Go to the issue