Open Access 01-08-2019 | Original Article

Diaphragm electromyography results at different high flow nasal cannula flow rates

Authors: Eleanor Jeffreys, Katie A Hunt, Theodore Dassios, Anne Greenough

Published in: European Journal of Pediatrics | Issue 8/2019

Login to get access

Abstract

Heated, humidified, high-flow nasal cannula (HHHFNC) is increasingly being used, but there is a paucity of evidence as to the optimum flow rates in prematurely born infants. We have determined the impact of three flow rates on the work of breathing (WOB) assessed by transcutaneous diaphragm electromyography (EMG) amplitude in infants with respiratory distress or bronchopulmonary dysplasia (BPD). Flow rates of 4, 6 and 8 L/min were delivered in random order. The mean amplitude of the EMG trace and mean area under the EMG curve (AEMGC) were calculated and the occurrence of bradycardias and desaturations recorded. Eighteen infants were studied with a median gestational age of 27.8 (range 23.9–33.5) weeks and postnatal age of 54 (range 3–122) days. The median flow rate prior to the study was 5 (range 3–8) L/min and the fraction of inspired oxygen (FiO2) was 0.29 (range 0.21–0.50). There were no significant differences between the mean amplitude of the diaphragm EMG and the AEGMC and the number of bradycardias or desaturations between the three flow rates.
Conclusions: In infants with respiratory distress or BPD, there was no advantage of using high (8 L/min) compared with lower flow rates (4 or 6 L/min) during support by HHHFNC.
What is known:
• Humidified high flow nasal cannulae (HHHFNC) is increasingly being used as a non-invasive form of respiratory support for prematurely born infants.
• There is a paucity of evidence regarding the optimum flow rate with 1 to 8 L/min being used.
What is new:
• We have assessed the work of breathing using the amplitude of the electromyogram of the diaphragm at three HHHFNC flow rates in infants with respiratory distress or BPD.
• No significant differences were found in the EMG amplitude results or the numbers of bradycardias or desaturations at 4, 6 and 8 L/min.
Literature
1.
go back to reference Collins CL, Hoberton JR, Konig K (2013) Comparison of the pharyngeal pressure provided by two heated humidified high flow nasal cannulae devices in premature infants. J Pediatr Child Health 49:554–556CrossRef Collins CL, Hoberton JR, Konig K (2013) Comparison of the pharyngeal pressure provided by two heated humidified high flow nasal cannulae devices in premature infants. J Pediatr Child Health 49:554–556CrossRef
2.
go back to reference de Waal CG, Hutten GJ, Kraaijenga JV, de Jongh FH, van Kaam AH (2017) Electrical activity of the diaphragm during nCPAP and high flow nasal cannula. Arch Dis Child Fetal Neonatal Ed 102:F434–F438CrossRefPubMed de Waal CG, Hutten GJ, Kraaijenga JV, de Jongh FH, van Kaam AH (2017) Electrical activity of the diaphragm during nCPAP and high flow nasal cannula. Arch Dis Child Fetal Neonatal Ed 102:F434–F438CrossRefPubMed
3.
go back to reference Hasan RA, Habib RH (2011) Effects of flow rate and airleak at the nares and mouth opening on positive distending pressure delivery using commercially available high flow nasal cannula systems: a lung model study. Pediatr Crit Care Med 12:e29–e33CrossRefPubMed Hasan RA, Habib RH (2011) Effects of flow rate and airleak at the nares and mouth opening on positive distending pressure delivery using commercially available high flow nasal cannula systems: a lung model study. Pediatr Crit Care Med 12:e29–e33CrossRefPubMed
4.
go back to reference Hochwald O, Osiovich H (2010) The use of high flow nasal cannulae in neonatal intensive care units: is clinical practice consistent with the evidence? J Neonatal-Perinatal Med 3:187–191 Hochwald O, Osiovich H (2010) The use of high flow nasal cannulae in neonatal intensive care units: is clinical practice consistent with the evidence? J Neonatal-Perinatal Med 3:187–191
5.
go back to reference Iyer NP, Mhanna MJ (2016) Association between high flow nasal cannula and end expiratory esophageal pressures in premature infants. Respir Care 61:285–290CrossRefPubMed Iyer NP, Mhanna MJ (2016) Association between high flow nasal cannula and end expiratory esophageal pressures in premature infants. Respir Care 61:285–290CrossRefPubMed
6.
go back to reference Jobe AH, Bancalari E (2001) Bronchopulmonary dysplasia. Am J Respir Crit Care Med 163:1723–1729CrossRef Jobe AH, Bancalari E (2001) Bronchopulmonary dysplasia. Am J Respir Crit Care Med 163:1723–1729CrossRef
7.
go back to reference Kraaijenga JV, Hutten GJ, de Jongh FH, van Kaam AH (2015) Transcutaneous electromyography of the diaphragm: a cardio-respiratory monitor for preterm infants. Pediatr Pulmonol 50:889–895CrossRefPubMed Kraaijenga JV, Hutten GJ, de Jongh FH, van Kaam AH (2015) Transcutaneous electromyography of the diaphragm: a cardio-respiratory monitor for preterm infants. Pediatr Pulmonol 50:889–895CrossRefPubMed
8.
go back to reference Kraaijenga JV, Hutten GJ, de Jongh FH, van Kaam AH (2015) The effect of caffeine on diaphragmatic activity and tidal volume in preterm infants. J Pediatr 167:70–75CrossRefPubMed Kraaijenga JV, Hutten GJ, de Jongh FH, van Kaam AH (2015) The effect of caffeine on diaphragmatic activity and tidal volume in preterm infants. J Pediatr 167:70–75CrossRefPubMed
9.
go back to reference Kubicka ZJ, Limauro J, Darnall RA (2008) Heated, humidified high-flow nasal cannula therapy: yet another way to deliver continuous positive airway pressure? Pediatrics 121:82–88CrossRefPubMed Kubicka ZJ, Limauro J, Darnall RA (2008) Heated, humidified high-flow nasal cannula therapy: yet another way to deliver continuous positive airway pressure? Pediatrics 121:82–88CrossRefPubMed
10.
go back to reference Lampland AL, Plumm B, Meyers PA, Worwa CT, Mammel MC (2009) Observational study of humidified high flow nasal cannula compared with nasal continuous positive airway pressure. J Pediatr 154:177–182CrossRefPubMed Lampland AL, Plumm B, Meyers PA, Worwa CT, Mammel MC (2009) Observational study of humidified high flow nasal cannula compared with nasal continuous positive airway pressure. J Pediatr 154:177–182CrossRefPubMed
11.
go back to reference Lavizzari A, Veneroni C, Colnaghi M, Ciuffini F, Zannin E, Fumagalli M, Mosca F, Dellaca RL (2014) Respiratory mechanics during nCPAP and HHHFNC at equal distending pressures. Arch Dis Child Fetal Neonatal Ed 99:F315–F320CrossRefPubMed Lavizzari A, Veneroni C, Colnaghi M, Ciuffini F, Zannin E, Fumagalli M, Mosca F, Dellaca RL (2014) Respiratory mechanics during nCPAP and HHHFNC at equal distending pressures. Arch Dis Child Fetal Neonatal Ed 99:F315–F320CrossRefPubMed
12.
go back to reference Maarsingh EJ, van Eykern L, Sprikkelman AB, Hoekstra MO, van Aalderen WM (2000) Respiratory muscle activity measured with a noninvasive EMG technique: technical aspects and reproducibility. J Appl Physiol 88:1955–1961CrossRefPubMed Maarsingh EJ, van Eykern L, Sprikkelman AB, Hoekstra MO, van Aalderen WM (2000) Respiratory muscle activity measured with a noninvasive EMG technique: technical aspects and reproducibility. J Appl Physiol 88:1955–1961CrossRefPubMed
13.
go back to reference O’Brien MJ, van Eykern LA, Prechtl H (1983) Monitoring respiratory activity in infants: a nonintrusive diaphragm EMG technique. Noninvasive Physiological Measurements, London, Academic 2:131–177 O’Brien MJ, van Eykern LA, Prechtl H (1983) Monitoring respiratory activity in infants: a nonintrusive diaphragm EMG technique. Noninvasive Physiological Measurements, London, Academic 2:131–177
14.
go back to reference Oda A, Parikka V, Lehtonen L, Porres I, Soukka H (2019) Nasal high flow therapy decreased electrical activity of the diaphragm in preterm infants during the weaning phase. Acta Paediatr 108:253–257CrossRefPubMed Oda A, Parikka V, Lehtonen L, Porres I, Soukka H (2019) Nasal high flow therapy decreased electrical activity of the diaphragm in preterm infants during the weaning phase. Acta Paediatr 108:253–257CrossRefPubMed
15.
go back to reference Saslow JG, Aghai ZH, Nakhla TA, Hart JJ, Lawrysh R, Stahl GE, Pyon KH (2006) Work of breathing using high-flow nasal cannula in preterm infants. J Perinatol 26:476–480CrossRefPubMed Saslow JG, Aghai ZH, Nakhla TA, Hart JJ, Lawrysh R, Stahl GE, Pyon KH (2006) Work of breathing using high-flow nasal cannula in preterm infants. J Perinatol 26:476–480CrossRefPubMed
16.
go back to reference Shetty S, Sundaresan A, Hunt K, Desai P, Greenough A (2016) Changes in the use of humidified high flow nasal cannula oxygen. Arch Dis Child Fetal Neonatal Ed 101:F371–F372CrossRefPubMed Shetty S, Sundaresan A, Hunt K, Desai P, Greenough A (2016) Changes in the use of humidified high flow nasal cannula oxygen. Arch Dis Child Fetal Neonatal Ed 101:F371–F372CrossRefPubMed
17.
go back to reference Sivieri EM, Gerdes JS, Abbasi S (2013) Effect of HFNC flow rate, cannula size and nares diameter on generated airway pressures: an in vitro study. Pediatr Pulmonol 48:506–514CrossRefPubMed Sivieri EM, Gerdes JS, Abbasi S (2013) Effect of HFNC flow rate, cannula size and nares diameter on generated airway pressures: an in vitro study. Pediatr Pulmonol 48:506–514CrossRefPubMed
18.
go back to reference Spence KL, Murphy D, Kilian C, McGonigle R, Kilani RA (2007) High flow nasal cannula as a device to provide continuous positive airway pressure in infants. J Perinatol 27:772–775CrossRefPubMed Spence KL, Murphy D, Kilian C, McGonigle R, Kilani RA (2007) High flow nasal cannula as a device to provide continuous positive airway pressure in infants. J Perinatol 27:772–775CrossRefPubMed
20.
go back to reference Wilkinson D, Andersen C, O’Donnell CP, De Paoli AG, Manley BJ (2016) High flow nasal cannula for respiratory support in preterm infants. Cochrane Database Syst Rev (2):CD006405 Wilkinson D, Andersen C, O’Donnell CP, De Paoli AG, Manley BJ (2016) High flow nasal cannula for respiratory support in preterm infants. Cochrane Database Syst Rev (2):CD006405
21.
go back to reference Wilkinson DJ, Andersen CC, Smith K, Holberton J (2008) Pharyngeal pressure with high-flow nasal cannulae in premature infants. J Perinatol 28:42–47CrossRefPubMed Wilkinson DJ, Andersen CC, Smith K, Holberton J (2008) Pharyngeal pressure with high-flow nasal cannulae in premature infants. J Perinatol 28:42–47CrossRefPubMed
Metadata
Title
Diaphragm electromyography results at different high flow nasal cannula flow rates
Authors
Eleanor Jeffreys
Katie A Hunt
Theodore Dassios
Anne Greenough
Publication date
01-08-2019
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Pediatrics / Issue 8/2019
Print ISSN: 0340-6199
Electronic ISSN: 1432-1076
DOI
https://doi.org/10.1007/s00431-019-03401-z

Other articles of this Issue 8/2019

European Journal of Pediatrics 8/2019 Go to the issue