01-10-2014 | Original Article

Bipolar vessel sealing: instrument contamination and wear have little effect on seal quality and success in a porcine in vitro model

Authors: Christian W. Wallwiener, Susanna H. Junginger, Wolfgang Zubke, Sara Y. Brucker, Markus D. Enderle, Alexander Neugebauer, Birgitt Schönfisch, Markus Wallwiener

Published in: Langenbeck's Archives of Surgery | Issue 7/2014

Login to get access

Abstract

Purpose

The aim of this study is to systematically investigate under standardized experimental conditions the effects of instrument contamination and wear on the quality of bipolar vessel sealing (BVS) achieved using a reusable instrument.

Methods

The study was designed as a prospective, randomized, and controlled in vitro study and conducted in an academic research environment. Reusable bipolar coagulation forceps (BiClamp® 200 C, ERBE Elektromedizin) were used to apply sealing pressures of 300–1,100 mN/mm2 to 239 renal arteries from commercially slaughtered female pigs (Swabian-Hall Swine). Forceps jaws were coated with porcine blood, blood and collagen, or blood, collagen, and fat to simulate instrument contamination with biological material during surgery. Clinical wear was mimicked by sandpaper abrasion. The main outcome measures were seal success (resistance to 250 mmHg intraluminal pressure for 2 min) and seal stability (burst pressure).

Results

Sealing pressure had a significant impact, with 800 mN/mm2 producing the best sealing results. Seal success increased with total energy applied to the tissue, a higher maximum temperature, and longer coagulation as indicated by desiccation time. Experimental contamination had no significant impact on seal success and only a limited effect on seal stability. Similarly, abrasive wear also had no significant effect on either seal quality or seal strength.

Conclusions

The impact of bipolar forceps contamination and wear on seal success and quality was negligible in our in vitro model. To achieve high-quality seals, it is essential to use adequate sealing pressures. Our findings could have direct implications for the design and clinical handling of BVS instruments.
Literature
1.
go back to reference Philosophe R (2003) Avoiding complications of laparoscopic surgery. Fertil Steril 80(Suppl 4):30–39, quiz 54-36PubMedCrossRef Philosophe R (2003) Avoiding complications of laparoscopic surgery. Fertil Steril 80(Suppl 4):30–39, quiz 54-36PubMedCrossRef
2.
go back to reference Presthus JB, Brooks PG, Kirchhof N (2003) Vessel sealing using a pulsed bipolar system and open forceps. J Am Assoc Gynecol Laparosc 10(4):528–533PubMedCrossRef Presthus JB, Brooks PG, Kirchhof N (2003) Vessel sealing using a pulsed bipolar system and open forceps. J Am Assoc Gynecol Laparosc 10(4):528–533PubMedCrossRef
3.
go back to reference Spivak H, Richardson WS, Hunter JG (1998) The use of bipolar cautery, laparosonic coagulating shears, and vascular clips for hemostasis of small and medium-sized vessels. Surg Endosc 12(2):183–185PubMedCrossRef Spivak H, Richardson WS, Hunter JG (1998) The use of bipolar cautery, laparosonic coagulating shears, and vascular clips for hemostasis of small and medium-sized vessels. Surg Endosc 12(2):183–185PubMedCrossRef
4.
go back to reference Kennedy JS, Stranahan PL, Taylor KD, Chandler JG (1998) High-burst-strength, feedback-controlled bipolar vessel sealing. Surg Endosc 12(6):876–878PubMedCrossRef Kennedy JS, Stranahan PL, Taylor KD, Chandler JG (1998) High-burst-strength, feedback-controlled bipolar vessel sealing. Surg Endosc 12(6):876–878PubMedCrossRef
5.
go back to reference Isaacson K (2002) New developments in radiofrequency technology for laparoscopic surgery. Contemp Ob Gyn 47:26–42 Isaacson K (2002) New developments in radiofrequency technology for laparoscopic surgery. Contemp Ob Gyn 47:26–42
8.
go back to reference Katkhouda N, Mavor E, Friedlander MH, Mason RJ, Kiyabu M, Grant SW, Achanta K, Kirkman EL, Narayanan K, Essani R (2001) Use of fibrin sealant for prosthetic mesh fixation in laparoscopic extraperitoneal inguinal hernia repair. Ann Surg 233(1):18–25PubMedCrossRefPubMedCentral Katkhouda N, Mavor E, Friedlander MH, Mason RJ, Kiyabu M, Grant SW, Achanta K, Kirkman EL, Narayanan K, Essani R (2001) Use of fibrin sealant for prosthetic mesh fixation in laparoscopic extraperitoneal inguinal hernia repair. Ann Surg 233(1):18–25PubMedCrossRefPubMedCentral
9.
go back to reference Harold KL, Pollinger H, Matthews BD, Kercher KW, Sing RF, Heniford BT (2003) Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries. Surg Endosc 17(8):1228–1230. doi:10.1007/s00464-002-8833-7 PubMedCrossRef Harold KL, Pollinger H, Matthews BD, Kercher KW, Sing RF, Heniford BT (2003) Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries. Surg Endosc 17(8):1228–1230. doi:10.​1007/​s00464-002-8833-7 PubMedCrossRef
10.
go back to reference Novitsky YW, Rosen MJ, Harrell AG, Sing RF, Kercher KW, Heniford BT (2005) Evaluation of the efficacy of the electrosurgical bipolar vessel sealer (LigaSure) devices in sealing lymphatic vessels. Surg Innov 12(2):155–160PubMedCrossRef Novitsky YW, Rosen MJ, Harrell AG, Sing RF, Kercher KW, Heniford BT (2005) Evaluation of the efficacy of the electrosurgical bipolar vessel sealer (LigaSure) devices in sealing lymphatic vessels. Surg Innov 12(2):155–160PubMedCrossRef
11.
go back to reference Pietrow PK, Weizer AZ, L'Esperance JO, Auge BK, Silverstein A, Cummings T, Preminger GM, Albala DM (2005) PlasmaKinetic bipolar vessel sealing: burst pressures and thermal spread in an animal model. J Endourol 19(1):107–110. doi:10.1089/end.2005.19.107 PubMedCrossRef Pietrow PK, Weizer AZ, L'Esperance JO, Auge BK, Silverstein A, Cummings T, Preminger GM, Albala DM (2005) PlasmaKinetic bipolar vessel sealing: burst pressures and thermal spread in an animal model. J Endourol 19(1):107–110. doi:10.​1089/​end.​2005.​19.​107 PubMedCrossRef
13.
go back to reference Levy B, Emery L (2003) Randomized trial of suture versus electrosurgical bipolar vessel sealing in vaginal hysterectomy. Obstet Gynecol 102(1):147–151PubMedCrossRef Levy B, Emery L (2003) Randomized trial of suture versus electrosurgical bipolar vessel sealing in vaginal hysterectomy. Obstet Gynecol 102(1):147–151PubMedCrossRef
14.
go back to reference Song C, Tang B, Campbell PA, Cuschieri A (2009) Thermal spread and heat absorbance differences between open and laparoscopic surgeries during energized dissections by electrosurgical instruments. Surg Endosc 23(11):2480–2487. doi:10.1007/s00464-009-0421-7 PubMedCrossRef Song C, Tang B, Campbell PA, Cuschieri A (2009) Thermal spread and heat absorbance differences between open and laparoscopic surgeries during energized dissections by electrosurgical instruments. Surg Endosc 23(11):2480–2487. doi:10.​1007/​s00464-009-0421-7 PubMedCrossRef
15.
17.
18.
go back to reference Vallfors B, Erlandson BE, Hansson HA, Wieck BO (1983) Current leakage in bipolar electrocoagulation. Neurosurgery 13(2):111–118PubMedCrossRef Vallfors B, Erlandson BE, Hansson HA, Wieck BO (1983) Current leakage in bipolar electrocoagulation. Neurosurgery 13(2):111–118PubMedCrossRef
19.
go back to reference Bergdahl B, Stenquist B (1993) An automatic computerized bipolar coagulator for dermatologic surgery. J Dermatol Surg Oncol 19(3):225–227PubMedCrossRef Bergdahl B, Stenquist B (1993) An automatic computerized bipolar coagulator for dermatologic surgery. J Dermatol Surg Oncol 19(3):225–227PubMedCrossRef
21.
go back to reference Menovsky T (2004) Bipolar forceps. J Neurosurg 101(6):1082, author reply 1082-1083PubMed Menovsky T (2004) Bipolar forceps. J Neurosurg 101(6):1082, author reply 1082-1083PubMed
22.
go back to reference Sigel B, Dunn MR (1965) The mechanism of blood vessel closure by high frequency electrocoagulation. Surg Gynecol Obstet 121(4):823–831PubMed Sigel B, Dunn MR (1965) The mechanism of blood vessel closure by high frequency electrocoagulation. Surg Gynecol Obstet 121(4):823–831PubMed
23.
go back to reference Sigel B, Hatke FL (1967) Physical factors in electrocoaptation of blood vessels. Arch Surg 95(1):54–58PubMedCrossRef Sigel B, Hatke FL (1967) Physical factors in electrocoaptation of blood vessels. Arch Surg 95(1):54–58PubMedCrossRef
24.
go back to reference Wallwiener CW, Rajab TK, Zubke W, Isaacson KB, Enderle M, Schaller D, Wallwiener M (2008) Thermal conduction, compression, and electrical current—an evaluation of major parameters of electrosurgical vessel sealing in a porcine in vitro model. J Minim Invasive Gynecol 15(5):605–610. doi:10.1016/j.jmig.2008.05.003 PubMedCrossRef Wallwiener CW, Rajab TK, Zubke W, Isaacson KB, Enderle M, Schaller D, Wallwiener M (2008) Thermal conduction, compression, and electrical current—an evaluation of major parameters of electrosurgical vessel sealing in a porcine in vitro model. J Minim Invasive Gynecol 15(5):605–610. doi:10.​1016/​j.​jmig.​2008.​05.​003 PubMedCrossRef
27.
go back to reference Aksan A, McGrath JJ, Nielubowicz DS Jr (2005) Thermal damage prediction for collagenous tissues part I: a clinically relevant numerical simulation incorporating heating rate dependent denaturation. J Biomech Eng 127(1):85–97PubMedCrossRef Aksan A, McGrath JJ, Nielubowicz DS Jr (2005) Thermal damage prediction for collagenous tissues part I: a clinically relevant numerical simulation incorporating heating rate dependent denaturation. J Biomech Eng 127(1):85–97PubMedCrossRef
28.
go back to reference Wilson MJ, Lopez M, Vargas M, Julian C, Tellez W, Rodriguez A, Bigham A, Armaza JF, Niermeyer S, Shriver M, Vargas E, Moore LG (2007) Greater uterine artery blood flow during pregnancy in multigenerational (Andean) than shorter-term (European) high-altitude residents. Am J Physiol Regul Integr Comp Physiol 293(3):R1313–1324. doi:10.1152/ajpregu.00806.2006 PubMedCrossRef Wilson MJ, Lopez M, Vargas M, Julian C, Tellez W, Rodriguez A, Bigham A, Armaza JF, Niermeyer S, Shriver M, Vargas E, Moore LG (2007) Greater uterine artery blood flow during pregnancy in multigenerational (Andean) than shorter-term (European) high-altitude residents. Am J Physiol Regul Integr Comp Physiol 293(3):R1313–1324. doi:10.​1152/​ajpregu.​00806.​2006 PubMedCrossRef
Metadata
Title
Bipolar vessel sealing: instrument contamination and wear have little effect on seal quality and success in a porcine in vitro model
Authors
Christian W. Wallwiener
Susanna H. Junginger
Wolfgang Zubke
Sara Y. Brucker
Markus D. Enderle
Alexander Neugebauer
Birgitt Schönfisch
Markus Wallwiener
Publication date
01-10-2014
Publisher
Springer Berlin Heidelberg
Published in
Langenbeck's Archives of Surgery / Issue 7/2014
Print ISSN: 1435-2443
Electronic ISSN: 1435-2451
DOI
https://doi.org/10.1007/s00423-014-1234-2

Other articles of this Issue 7/2014

Langenbeck's Archives of Surgery 7/2014 Go to the issue