Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurements

Authors: Joshua M. Drouin, Tamara C. Valovich-mcLeod, Sandra J. Shultz, Bruce M. Gansneder, David H. Perrin

Published in: European Journal of Applied Physiology | Issue 1/2004

Login to get access

Abstract

This study quantitatively assessed the mechanical reliability and validity of position, torque and velocity measurements of the Biodex System 3 isokinetic dynamometer. Trial-to-trial and day-to-day reliability were assessed during three trials on two separate days. To assess instrument validity, measurement of each variable using the Biodex System 3 dynamometer was compared to a criterion measure of position, torque and velocity. Position was assessed at 5° increments across the available range of motion of the dynamometer. Torque measures were assessed isometrically by hanging six different calibrated weights from the lever arm. Velocity was assessed (30°/s to 500°/s) across a 70° arc of motion by manually accelerating the weighted lever arm. With the exception of a systematic decrease in velocity at speeds of 300°/s and higher, the Biodex System 3 performed with acceptable mechanical reliability and validity on all variables tested.
Literature
go back to reference Bemben MG, Grump KJ, Massey BH (1988) Assessment of technical accuracy of the Cybex II isokinetic dynamometer and analog recording system. J Orthop Sports Phys Ther 10:12–17 Bemben MG, Grump KJ, Massey BH (1988) Assessment of technical accuracy of the Cybex II isokinetic dynamometer and analog recording system. J Orthop Sports Phys Ther 10:12–17
go back to reference Brown LE, Whitehurst M, Bryant JR, Buchalter DN (1993) Reliability of the Biodex System 2 isokinetic dynamometer concentric mode. Isokin Exerc Sci 3:106–163 Brown LE, Whitehurst M, Bryant JR, Buchalter DN (1993) Reliability of the Biodex System 2 isokinetic dynamometer concentric mode. Isokin Exerc Sci 3:106–163
go back to reference Denegar CR, Ball DW (1993) Assessing reliability and precision of measurement: an introduction to intraclass correlation and standard error of measurement. J Sport Rehabil 2:35–42 Denegar CR, Ball DW (1993) Assessing reliability and precision of measurement: an introduction to intraclass correlation and standard error of measurement. J Sport Rehabil 2:35–42
go back to reference Farrell M, Richards JG (1986) Analysis of the reliability and validity of the kinetic communicator exercise device. Med Sci Sports Exerc 18:44–49PubMed Farrell M, Richards JG (1986) Analysis of the reliability and validity of the kinetic communicator exercise device. Med Sci Sports Exerc 18:44–49PubMed
go back to reference Feiring DC, Ellenbecker TS, Dersheid GL (1990) Test-retest reliability of the Biodex isokinetic dynamometer. J Orthop Sports Phys Ther 11:298–300 Feiring DC, Ellenbecker TS, Dersheid GL (1990) Test-retest reliability of the Biodex isokinetic dynamometer. J Orthop Sports Phys Ther 11:298–300
go back to reference Gleeson NP, Mercer TH (1996) The utility of isokinetic dynamometry in the assessment of human muscle function. Sports Med 21:18–34PubMed Gleeson NP, Mercer TH (1996) The utility of isokinetic dynamometry in the assessment of human muscle function. Sports Med 21:18–34PubMed
go back to reference Handel M, Dickhuth HH, Mayer F, Gulch RW (1996) Prerequisites and limitations to isokinetic measurements in humans: Investigations on a servomotor-controlled dynamometer. Eur J Appl Physiol 73:225–230 Handel M, Dickhuth HH, Mayer F, Gulch RW (1996) Prerequisites and limitations to isokinetic measurements in humans: Investigations on a servomotor-controlled dynamometer. Eur J Appl Physiol 73:225–230
go back to reference Iossifidou AN, Baltzopoulos V (2000) Inertial effects on moment development during isokinetic concentric knee extension testing. J Orthop Sports Phys Ther 30:317–327PubMed Iossifidou AN, Baltzopoulos V (2000) Inertial effects on moment development during isokinetic concentric knee extension testing. J Orthop Sports Phys Ther 30:317–327PubMed
go back to reference Murray DA, Harrison E (1986) Constant velocity dynamometer: an appraisal using mechanical loading. Med Sci Sports Exerc 18:612–624PubMed Murray DA, Harrison E (1986) Constant velocity dynamometer: an appraisal using mechanical loading. Med Sci Sports Exerc 18:612–624PubMed
go back to reference Patterson LA, Spivey WE (1992) Validity and reliability of the LIDO active isokinetic system. J Orthop Sports Phys Ther 15:32–36 Patterson LA, Spivey WE (1992) Validity and reliability of the LIDO active isokinetic system. J Orthop Sports Phys Ther 15:32–36
go back to reference Portney LG, Watkins MP (2000) Statistical measures of reliability. In: Foundations of clinical research: applications to practice, 2nd edn. Prentice Hall, New Jersey, pp 557–584 Portney LG, Watkins MP (2000) Statistical measures of reliability. In: Foundations of clinical research: applications to practice, 2nd edn. Prentice Hall, New Jersey, pp 557–584
go back to reference Sapega AA, Nicholas JA, Sokolow D, Saraniti A (1982) The nature of torque "overshoot" in Cybex isokinetic dynamometry. Med Sci Sports Exerc 14:368–375PubMed Sapega AA, Nicholas JA, Sokolow D, Saraniti A (1982) The nature of torque "overshoot" in Cybex isokinetic dynamometry. Med Sci Sports Exerc 14:368–375PubMed
go back to reference Shrout PE, Fleiss JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86:420–428CrossRef Shrout PE, Fleiss JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86:420–428CrossRef
go back to reference Taylor NAS, Sanders RH, Howick EI, Stanley SN (1991) Static and dynamic assessment of the Biodex dynamometer. Eur J Appl Physiol 62:180–188 Taylor NAS, Sanders RH, Howick EI, Stanley SN (1991) Static and dynamic assessment of the Biodex dynamometer. Eur J Appl Physiol 62:180–188
go back to reference Timm KE, Gennrich P, Burns R, Fyke D (1992) The mechanical and physiological performance reliability of selected isokinetic dynamometers. Isokin Exerc Sci 2:182–190 Timm KE, Gennrich P, Burns R, Fyke D (1992) The mechanical and physiological performance reliability of selected isokinetic dynamometers. Isokin Exerc Sci 2:182–190
Metadata
Title
Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurements
Authors
Joshua M. Drouin
Tamara C. Valovich-mcLeod
Sandra J. Shultz
Bruce M. Gansneder
David H. Perrin
Publication date
01-01-2004
Publisher
Springer-Verlag
Published in
European Journal of Applied Physiology / Issue 1/2004
Print ISSN: 1439-6319
Electronic ISSN: 1439-6327
DOI
https://doi.org/10.1007/s00421-003-0933-0

Other articles of this Issue 1/2004

European Journal of Applied Physiology 1/2004 Go to the issue