Published in: Graefe's Archive for Clinical and Experimental Ophthalmology 10/2020

Open Access 01-10-2020 | Basic Science

Comparison of waveform-derived corneal stiffness and stress-strain extensometry-derived corneal stiffness using different cross-linking irradiances: an experimental study with air-puff applanation of ex vivo porcine eyes

Authors: Robert Herber, Mathew Francis, Eberhard Spoerl, Lutz E. Pillunat, Frederik Raiskup, Abhijit Sinha Roy

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 10/2020

Login to get access

Abstract

Purpose

To assess corneal stiffening of standard (S-CXL) and accelerated (A-CXL) cross-linking protocols by dynamic corneal response parameters and corneal bending stiffness (Kc[mean/linear]) derived from Corvis (CVS) Scheimpflug-based tonometry. These investigations were validated by corneal tensile stiffness (K[ts]), derived from stress-strain extensometry in ex vivo porcine eyes.

Methods

Seventy-two fresh-enucleated and de-epithelized porcine eyes were soaked in 0.1% riboflavin solution including 10% dextran for 10 min. The eyes were separated into four groups: controls (n = 18), S-CXL (intensity in mW/cm2*time in min; 3*30) (n = 18), A-CXL (9*10) (n = 18), and A-CXL (18*5) (n = 18), respectively. CXL was performed using CCL Vario. CVS measurements were performed on all eyes. Subsequently, corneal strips were extracted by a double-bladed scalpel and used for stress-strain measurements. K[ts] was calculated from a force-displacement curve. Mean corneal stiffness (Kc[mean]) and constant corneal stiffness (Kc[linear]) were calculated from raw CVS data.

Results

In CVS, biomechanical effects of cross-linking were shown to have a significantly decreased deflection amplitude as well as integrated radius, an increased IOP, and SP A1 (P < 0.05). Kc[mean]/Kc[linear] were significantly increased after CXL (P < 0.05). In the range from 2 to 6% strain, K[ts] was significantly higher in S-CXL (3*30) compared to A-CXL (9*10), A-CXL (18*5), and controls (P < 0.05). At 8% to 10% strain, all protocols induced a higher stiffness than controls (P < 0.05).

Conclusion

Several CVS parameters and Kc[mean] as well as Kc[linear] verify corneal stiffening effect after CXL on porcine eyes. S-CXL seems to have a higher tendency of stiffening than A-CXL protocols have, which was demonstrated by Scheimpflug-based tonometry and stress-strain extensometry.
Literature
2.
go back to reference Spoerl E, Seiler T (1999) Techniques for stiffening the cornea. J Refract Surg 15:711–713PubMed Spoerl E, Seiler T (1999) Techniques for stiffening the cornea. J Refract Surg 15:711–713PubMed
3.
go back to reference Wollensak G, Spoerl E, Seiler T (2003) Riboflavin/ultraviolet-a-induced collagen crosslinking for the treatment of keratoconus. Am J Ophthalmol 135:620–627CrossRefPubMed Wollensak G, Spoerl E, Seiler T (2003) Riboflavin/ultraviolet-a-induced collagen crosslinking for the treatment of keratoconus. Am J Ophthalmol 135:620–627CrossRefPubMed
10.
go back to reference Wollensak G, Spoerl E, Seiler T (2003) Stress-strain measurements of human and porcine corneas after riboflavin-ultraviolet-A-induced cross-linking. J Cataract Refract Surg 29:1780–1785CrossRefPubMed Wollensak G, Spoerl E, Seiler T (2003) Stress-strain measurements of human and porcine corneas after riboflavin-ultraviolet-A-induced cross-linking. J Cataract Refract Surg 29:1780–1785CrossRefPubMed
19.
go back to reference Sinjab MM, Cummings AB (2017) Corneal collagen cross linking. Springer International Publishing, Imprint, Springer, ChamCrossRef Sinjab MM, Cummings AB (2017) Corneal collagen cross linking. Springer International Publishing, Imprint, Springer, ChamCrossRef
23.
27.
Metadata
Title
Comparison of waveform-derived corneal stiffness and stress-strain extensometry-derived corneal stiffness using different cross-linking irradiances: an experimental study with air-puff applanation of ex vivo porcine eyes
Authors
Robert Herber
Mathew Francis
Eberhard Spoerl
Lutz E. Pillunat
Frederik Raiskup
Abhijit Sinha Roy
Publication date
01-10-2020
Publisher
Springer Berlin Heidelberg
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 10/2020
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-020-04792-8

Other articles of this Issue 10/2020

Graefe's Archive for Clinical and Experimental Ophthalmology 10/2020 Go to the issue