Published in: Graefe's Archive for Clinical and Experimental Ophthalmology 1/2019

01-01-2019 | Glaucoma

Anterior segment Scheimpflug imaging for detecting primary angle closure disease

Authors: Andrew Winegarner, Atsuya Miki, Miho Kumoi, Yuichiro Ishida, Taku Wakabayashi, Susumu Sakimoto, Shinichi Usui, Kenji Matsushita, Kohji Nishida

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 1/2019

Login to get access

Abstract

Purpose

To evaluate the capability of anterior segment Scheimpflug imaging for detecting primary angle closure disease (PACD): primary angle closure suspect, primary angle closure, and primary angle closure glaucoma, using cutoff points derived from reference databases of healthy subjects.

Methods

Eighty-seven patients with PACD and 49 age-matched control subjects were included. We evaluated the sensitivity and specificity of anterior chamber depth (ACD), anterior chamber volume (ACV), and anterior chamber angle (ACA) to differentiate patients with PACD from controls. Additionally, the study’s raw data was analyzed via receiver operating characteristic curves for comparison.

Results

One standard deviation from the normative data’s mean values was used as the cutoff point and yielded a sensitivity and specificity of 96.2% and 92.6% for ACD, 97.1% and 75.9% for ACV, and 93.3% and 72.2% for ACA, respectively. Receiver operating characteristic analysis of the raw data showed the area under the curve to be 0.984, 0.975, and 0.931 for ACD, ACV, and ACA, respectively.

Conclusions

Our study demonstrated that the parameters of anterior segment Scheimpflug imaging, particularly ACD, accurately discriminate PACD. This was the first study to validate the device’s normative data in a separate population. With its high reproducibility, ease of use, non-invasiveness, and speed, anterior segment Scheimpflug imaging is a potentially powerful screening tool for PACD.
Literature
1.
go back to reference Cheng J, Zong Y, Zeng Y, Wei R (2014) The prevalence of primary angle closure glaucoma in adult Asians: a systematic review and meta-analysis. PLoS One 24(9):7 Cheng J, Zong Y, Zeng Y, Wei R (2014) The prevalence of primary angle closure glaucoma in adult Asians: a systematic review and meta-analysis. PLoS One 24(9):7
2.
go back to reference Day AC, Baio G, Gazzard G, Bunce C, Azuara-Blanco A, Munoz B, Friedman DS, Foster PJ (2012) The prevalence of primary angle closure glaucoma in European derived populations: a systemic review. Br J Ophthalmol 96(9):1162–1167CrossRefPubMed Day AC, Baio G, Gazzard G, Bunce C, Azuara-Blanco A, Munoz B, Friedman DS, Foster PJ (2012) The prevalence of primary angle closure glaucoma in European derived populations: a systemic review. Br J Ophthalmol 96(9):1162–1167CrossRefPubMed
3.
go back to reference Aung T, Friedman DS, Chew PT et al (2004) Long-term outcomes in Asians after acute primary angle closure. Ophthalmology 111:1464–1469CrossRefPubMed Aung T, Friedman DS, Chew PT et al (2004) Long-term outcomes in Asians after acute primary angle closure. Ophthalmology 111:1464–1469CrossRefPubMed
4.
go back to reference Weinreb RN, Healey PR, Topouzis F (2008) Glaucoma screening. World Glaucoma Association consensus series 5. Kugler, The Hague Weinreb RN, Healey PR, Topouzis F (2008) Glaucoma screening. World Glaucoma Association consensus series 5. Kugler, The Hague
5.
go back to reference Foster PJ, Buhrmann R, Quigley HA, Johnson GJ (2002) The definition and classification of glaucoma in prevalence surveys. Br J Ophthalmol 86(2):238–242CrossRefPubMedPubMedCentral Foster PJ, Buhrmann R, Quigley HA, Johnson GJ (2002) The definition and classification of glaucoma in prevalence surveys. Br J Ophthalmol 86(2):238–242CrossRefPubMedPubMedCentral
6.
go back to reference Thomas R, George R, Parikh R, Muliyil J, Jacob A (2003) Five year risk of progression of primary angle closure suspects to primary angle closure: a population based study. Br J Ophthalmol 87:450–454CrossRefPubMedPubMedCentral Thomas R, George R, Parikh R, Muliyil J, Jacob A (2003) Five year risk of progression of primary angle closure suspects to primary angle closure: a population based study. Br J Ophthalmol 87:450–454CrossRefPubMedPubMedCentral
7.
go back to reference Thomas R, Parikh R, Muliyil J, Kumar RS (2003) Five-year risk of progression of primary angle closure to primary angle closure glaucoma: a population-based study. Acta Ophthalmol Scand 81:480–485CrossRefPubMed Thomas R, Parikh R, Muliyil J, Kumar RS (2003) Five-year risk of progression of primary angle closure to primary angle closure glaucoma: a population-based study. Acta Ophthalmol Scand 81:480–485CrossRefPubMed
8.
go back to reference Jain R, Grewal SPS (2009) Pentacam: principle and clinical application. J Curr Glaucoma Pract 3(2):20–32CrossRef Jain R, Grewal SPS (2009) Pentacam: principle and clinical application. J Curr Glaucoma Pract 3(2):20–32CrossRef
9.
go back to reference Labiris G, Gkika M, Katsanos A, Fanariotis M, Alvanos E, Kozobolis V (2009) Anterior chamber volume measurements with Visante optical coherence tomography and Pentacam: repeatability and level of agreement. Clin Exp Ophthalmol 37:772–774CrossRefPubMed Labiris G, Gkika M, Katsanos A, Fanariotis M, Alvanos E, Kozobolis V (2009) Anterior chamber volume measurements with Visante optical coherence tomography and Pentacam: repeatability and level of agreement. Clin Exp Ophthalmol 37:772–774CrossRefPubMed
10.
go back to reference Shankar H, Taranath D, Santhirathelagan CT, Pesudovs K (2008) Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements. J Cataract Refract Surg 34:103–113CrossRefPubMed Shankar H, Taranath D, Santhirathelagan CT, Pesudovs K (2008) Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements. J Cataract Refract Surg 34:103–113CrossRefPubMed
11.
go back to reference Kurita N, Mayama C, Tomidokoro A, Aihara M, Araie M (2009) Potential of the pentacam in screening for primary angle closure and primary angle closure suspect. J Glaucoma 18(7):506–512CrossRefPubMed Kurita N, Mayama C, Tomidokoro A, Aihara M, Araie M (2009) Potential of the pentacam in screening for primary angle closure and primary angle closure suspect. J Glaucoma 18(7):506–512CrossRefPubMed
12.
go back to reference Hong S, Yi J, Kang SY, Seong GJ, Kim CY (2009) Detection of occludable angles with the Pentacam and the anterior segment optical coherence tomography. Yonsei Med J 50(4):525–528CrossRefPubMedPubMedCentral Hong S, Yi J, Kang SY, Seong GJ, Kim CY (2009) Detection of occludable angles with the Pentacam and the anterior segment optical coherence tomography. Yonsei Med J 50(4):525–528CrossRefPubMedPubMedCentral
13.
go back to reference Grewal DS, Brar GS, Jain R, Grewal SP (2011) Comparison of Scheimpflug imaging and spectral domain anterior segment optical coherence tomography for detection of narrow anterior chamber angles. Eye (Lond) 25(5):603–611CrossRef Grewal DS, Brar GS, Jain R, Grewal SP (2011) Comparison of Scheimpflug imaging and spectral domain anterior segment optical coherence tomography for detection of narrow anterior chamber angles. Eye (Lond) 25(5):603–611CrossRef
14.
go back to reference Koç M, Özülken K, Ayar O, Karakurt A (2013) Measurement of the anterior chamber angle according to quadrants and age groups using Pentacam Scheimpflug camera. J Glaucoma 22(3):226–229CrossRefPubMed Koç M, Özülken K, Ayar O, Karakurt A (2013) Measurement of the anterior chamber angle according to quadrants and age groups using Pentacam Scheimpflug camera. J Glaucoma 22(3):226–229CrossRefPubMed
15.
go back to reference Dabasia PL, Edgar DF, Murdoch IE, Lawrenson JG (2015) Noncontact screening methods for the detection of narrow anterior chamber angles. Invest Ophthalmo Vis Sci 56(6):3929–3935CrossRef Dabasia PL, Edgar DF, Murdoch IE, Lawrenson JG (2015) Noncontact screening methods for the detection of narrow anterior chamber angles. Invest Ophthalmo Vis Sci 56(6):3929–3935CrossRef
16.
go back to reference Rabsilber TM, Khoramnia R, Auffarth GU (2006) Anterior chamber measurements using Pentacam rotating Scheimpflug camera. J Cataract Refract Surg 32(3):456–459CrossRefPubMed Rabsilber TM, Khoramnia R, Auffarth GU (2006) Anterior chamber measurements using Pentacam rotating Scheimpflug camera. J Cataract Refract Surg 32(3):456–459CrossRefPubMed
17.
go back to reference Pakravan M, Sharifipour F, Yazdani S, Koohestani N, Yaseri M (2012) Scheimpflug imaging criteria for identifying eyes at high risk of acute angle closure. J Ophthalmic Vis Res 7(2):111–117PubMedPubMedCentral Pakravan M, Sharifipour F, Yazdani S, Koohestani N, Yaseri M (2012) Scheimpflug imaging criteria for identifying eyes at high risk of acute angle closure. J Ophthalmic Vis Res 7(2):111–117PubMedPubMedCentral
18.
go back to reference Cho YK, Chang HS, La TY et al (2010) Anterior segment parameters using Pentacam and prediction of corneal endothelial cell loss after cataract surgery. Korean J Ophthalmol 24:284CrossRefPubMedPubMedCentral Cho YK, Chang HS, La TY et al (2010) Anterior segment parameters using Pentacam and prediction of corneal endothelial cell loss after cataract surgery. Korean J Ophthalmol 24:284CrossRefPubMedPubMedCentral
20.
go back to reference Herman CR, Gill HK, Eng J, Fajardo LL (2002) Screening for preclinical disease: test and disease characteristics. AJR Am J Roentgenol 179(4):825–831CrossRefPubMed Herman CR, Gill HK, Eng J, Fajardo LL (2002) Screening for preclinical disease: test and disease characteristics. AJR Am J Roentgenol 179(4):825–831CrossRefPubMed
21.
go back to reference Omura T, Tanito M, Doi R, Ishida R, Ohira A (2012) Anterior chamber parameters measured using the Pentacam Scheimpflug imaging device before and after cataract surgery in eyes with primary angle closure. Acta Ophthalmol 90(8):e654–e655CrossRefPubMed Omura T, Tanito M, Doi R, Ishida R, Ohira A (2012) Anterior chamber parameters measured using the Pentacam Scheimpflug imaging device before and after cataract surgery in eyes with primary angle closure. Acta Ophthalmol 90(8):e654–e655CrossRefPubMed
22.
go back to reference Liang J, Liu W, Xing X, Liu H, Zhao S, Ji J (2010) Evaluation of the agreement between Pentacam and ultrasound biomicroscopy measurements of anterior chamber depth in Chinese patients with primary angle-closure glaucoma. Jpn J Ophthalmol 54(4):361–362CrossRefPubMed Liang J, Liu W, Xing X, Liu H, Zhao S, Ji J (2010) Evaluation of the agreement between Pentacam and ultrasound biomicroscopy measurements of anterior chamber depth in Chinese patients with primary angle-closure glaucoma. Jpn J Ophthalmol 54(4):361–362CrossRefPubMed
23.
go back to reference Zou X, Duan XC, Zhong Q (2010) Alteration of anterior chamber in 81 glaucomatous eyes using Pentacam Scheimpflug system. Int J Ophthalmol 3(4):349–351PubMedPubMedCentral Zou X, Duan XC, Zhong Q (2010) Alteration of anterior chamber in 81 glaucomatous eyes using Pentacam Scheimpflug system. Int J Ophthalmol 3(4):349–351PubMedPubMedCentral
24.
go back to reference Li X, Wang Z, Cao Q, Hu L, Tian F, Dai H (2014) Pentacam could be a useful tool for evaluating and qualifying the anterior chamber morphology. Int J Clin Exp Med 7(7):1878–1882PubMedPubMedCentral Li X, Wang Z, Cao Q, Hu L, Tian F, Dai H (2014) Pentacam could be a useful tool for evaluating and qualifying the anterior chamber morphology. Int J Clin Exp Med 7(7):1878–1882PubMedPubMedCentral
25.
go back to reference Zhang Y, Li SZ, Li L, Thomas R, Wang NL (2014) The Handan Eye Study: comparison of screening methods for primary angle closure suspects in a rural Chinese population. Ophthalmic Epidemiol 21(4):268–275CrossRefPubMed Zhang Y, Li SZ, Li L, Thomas R, Wang NL (2014) The Handan Eye Study: comparison of screening methods for primary angle closure suspects in a rural Chinese population. Ophthalmic Epidemiol 21(4):268–275CrossRefPubMed
26.
go back to reference Sawaguchi S, Sakai H, Iwase A, Yamamoto T, Abe H, Tomita G, Tomidokoro A, Araie M (2012) Prevalence of primary angle closure and primary angle-closure glaucoma in a southwestern rural population of Japan: the Kumejima Study. Ophthalmology 119(6):1134–1142CrossRefPubMed Sawaguchi S, Sakai H, Iwase A, Yamamoto T, Abe H, Tomita G, Tomidokoro A, Araie M (2012) Prevalence of primary angle closure and primary angle-closure glaucoma in a southwestern rural population of Japan: the Kumejima Study. Ophthalmology 119(6):1134–1142CrossRefPubMed
27.
go back to reference Guzman CP, Gong T, Nongpiur ME, Perera SA, How AC, Lee HK, Cheng L, He M, Baskaran M, Aung T (2013) Anterior segment optical coherence tomography parameters in subtypes of primary angle closure. Invest Ophthalmol Vis Sci 54(8):5281–5286CrossRefPubMed Guzman CP, Gong T, Nongpiur ME, Perera SA, How AC, Lee HK, Cheng L, He M, Baskaran M, Aung T (2013) Anterior segment optical coherence tomography parameters in subtypes of primary angle closure. Invest Ophthalmol Vis Sci 54(8):5281–5286CrossRefPubMed
28.
go back to reference Nongpiur ME, He M, Amerasinghe N, Friedman DS, Tay WT, Baskaran M, Smith SD, Wong TY, Aung T (2011) Lens vault, thickness, and position in Chinese subjects with angle closure. Ophthalmology 118(3):474–479CrossRefPubMed Nongpiur ME, He M, Amerasinghe N, Friedman DS, Tay WT, Baskaran M, Smith SD, Wong TY, Aung T (2011) Lens vault, thickness, and position in Chinese subjects with angle closure. Ophthalmology 118(3):474–479CrossRefPubMed
29.
go back to reference Tan GS, He M, Zhao W, Sakata LM, Li J, Nonqpiur ME, Lavanya R, Friedman DS, Aung T (2012) Determinants of lens vault and association with narrow angles in patients from Singapore. Am J Ophthalmol 154(1):39–46CrossRefPubMed Tan GS, He M, Zhao W, Sakata LM, Li J, Nonqpiur ME, Lavanya R, Friedman DS, Aung T (2012) Determinants of lens vault and association with narrow angles in patients from Singapore. Am J Ophthalmol 154(1):39–46CrossRefPubMed
30.
go back to reference Nongpiur ME, Haaland BA, Perera SA, Friedman DS, He M, Sakata LM, Baskaran M, Aung T (2014) Development of a score and probability estimate for detecting angle closure based on anterior segment optical coherence tomography. Am J Ophthalmol 157(1):32–38CrossRefPubMed Nongpiur ME, Haaland BA, Perera SA, Friedman DS, He M, Sakata LM, Baskaran M, Aung T (2014) Development of a score and probability estimate for detecting angle closure based on anterior segment optical coherence tomography. Am J Ophthalmol 157(1):32–38CrossRefPubMed
31.
go back to reference Han S, Sung KR, Lee KS, Hong JW (2014) Outcomes of laser peripheral iridotomy in angle closure subgroups according to anterior segment optical coherence tomography parameters. Invest Ophthalmol Vis Sci 55(10):6795–6801CrossRefPubMed Han S, Sung KR, Lee KS, Hong JW (2014) Outcomes of laser peripheral iridotomy in angle closure subgroups according to anterior segment optical coherence tomography parameters. Invest Ophthalmol Vis Sci 55(10):6795–6801CrossRefPubMed
32.
go back to reference Baskaran M, Iyer JV, Narayanaswamy AK et al (2015) Anterior segment imaging predicts incident gonioscopic angle closure. Ophthalmology 122(12):2380–2384CrossRefPubMed Baskaran M, Iyer JV, Narayanaswamy AK et al (2015) Anterior segment imaging predicts incident gonioscopic angle closure. Ophthalmology 122(12):2380–2384CrossRefPubMed
33.
go back to reference Smith S, Singh K, Lin S, Chen P (2013) Evaluation of the anterior chamber angle in glaucoma: a report by the american academy of ophthalmology. Ophthalmology 120:1985–1997CrossRefPubMed Smith S, Singh K, Lin S, Chen P (2013) Evaluation of the anterior chamber angle in glaucoma: a report by the american academy of ophthalmology. Ophthalmology 120:1985–1997CrossRefPubMed
34.
go back to reference Wang D, Qi M, He M, Wu L, Lin S (2012) Ethnic difference of the anterior chamber area and volume and its association with angle width. Invest Ophthalmol Vis Sci 53:3139–3144CrossRefPubMedPubMedCentral Wang D, Qi M, He M, Wu L, Lin S (2012) Ethnic difference of the anterior chamber area and volume and its association with angle width. Invest Ophthalmol Vis Sci 53:3139–3144CrossRefPubMedPubMedCentral
35.
go back to reference Chansangpetch S, Tran B, Perez CI et al (2018) Comparison of anterior segment optical coherence tomography parameters among Vietnamese, Chinese, and Caucasians. Am J Ophthalmol 195:72–82CrossRefPubMed Chansangpetch S, Tran B, Perez CI et al (2018) Comparison of anterior segment optical coherence tomography parameters among Vietnamese, Chinese, and Caucasians. Am J Ophthalmol 195:72–82CrossRefPubMed
Metadata
Title
Anterior segment Scheimpflug imaging for detecting primary angle closure disease
Authors
Andrew Winegarner
Atsuya Miki
Miho Kumoi
Yuichiro Ishida
Taku Wakabayashi
Susumu Sakimoto
Shinichi Usui
Kenji Matsushita
Kohji Nishida
Publication date
01-01-2019
Publisher
Springer Berlin Heidelberg
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 1/2019
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-018-4171-x

Other articles of this Issue 1/2019

Graefe's Archive for Clinical and Experimental Ophthalmology 1/2019 Go to the issue