Published in: Graefe's Archive for Clinical and Experimental Ophthalmology 6/2014

01-06-2014 | Cataract

Optical quality of aspheric toric intraocular lenses at different degrees of decentering

Authors: Cari Pérez-Vives, Teresa Ferrer-Blasco, David Madrid-Costa, Santiago García-Lázaro, Robert Montés-Micó

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 6/2014

Login to get access

Abstract

Purpose

To analyze the optical quality of aspheric toric intraocular lenses (IOLs) at different degrees of decentering.

Methods

Wavefront aberrations of Acrysof IQ Toric IOLs (SN6AT3, SN6AT4, and SN6AT5; Alcon Laboratories Inc, Fort Worth, TX, USA) for different powers (15.00, 20.00 and 23.50 diopters [D]) and for different degrees of decentering (diagonal, horizontal, and vertical decentering of 0.3 and 0.6mm) were measured in vitro at 3- and 5-mm pupils. The Zernike coefficients of trefoil, coma, tetrafoil, secondary astigmatism, and spherical aberration were evaluated. The point spread functions (PSFs) and modulation transfer function (MTF) of each IOL evaluated were calculated from the wavefront aberrations.

Results

Coma aberration increased significantly with IOL decentration. Statistically significant differences were found between centered and all decentered positions in coma aberration (p < 0.05). Although, we only found visible differences between centered and decentered positions in PSF images and MTF curves for horizontal and vertical 0.6 mm of decentering at 3- and 5-mm pupil.

Conclusions

Despite the coma increment, these values were clinically negligible and have no effect on visual performance, except for vertical and horizontal 0.6 mm of decentering, which could have a negative effect on visual quality.
Literature
1.
go back to reference Rocha KM, Soriano ES, Chalita MR, Yamada AC, Bottós K, Bottós J, Morimoto L, Nosé W (2006) Wavefront analysis and contrast sensitivity of aspheric and spherical intraocular lenses: a randomized prospective study. Am J Ophthalmol 142:750–756PubMedCrossRef Rocha KM, Soriano ES, Chalita MR, Yamada AC, Bottós K, Bottós J, Morimoto L, Nosé W (2006) Wavefront analysis and contrast sensitivity of aspheric and spherical intraocular lenses: a randomized prospective study. Am J Ophthalmol 142:750–756PubMedCrossRef
2.
go back to reference Rocha KM, Soriano ES, Chamon W, Chalita MR, Nosé W (2007) Spherical aberration and depth of focus in eyes implanted with aspheric and spherical intraocular lenses; a prospective randomized study. Ophthalmology 114:2050–2054PubMedCrossRef Rocha KM, Soriano ES, Chamon W, Chalita MR, Nosé W (2007) Spherical aberration and depth of focus in eyes implanted with aspheric and spherical intraocular lenses; a prospective randomized study. Ophthalmology 114:2050–2054PubMedCrossRef
3.
go back to reference Cadarso L, Iglesias A, Ollero A, Pita B, Montés-Micó R (2008) Postoperative optical aberrations in eyes implanted with Acrysof spherical and aspheric intraocular lenses. J Refract Surg 24:811–816PubMed Cadarso L, Iglesias A, Ollero A, Pita B, Montés-Micó R (2008) Postoperative optical aberrations in eyes implanted with Acrysof spherical and aspheric intraocular lenses. J Refract Surg 24:811–816PubMed
4.
go back to reference Montés-Micó R, Ferrer-Blasco T, Cerviño A (2009) Analysis of the possible benefits of aspheric intraocular lenses: Review of the literature. J Cataract Refract Surg 35:172–181PubMedCrossRef Montés-Micó R, Ferrer-Blasco T, Cerviño A (2009) Analysis of the possible benefits of aspheric intraocular lenses: Review of the literature. J Cataract Refract Surg 35:172–181PubMedCrossRef
5.
go back to reference Trueb PR, Albach C, Montés-Micó R, Ferrer-Blasco T (2009) Visual acuity and contrast sensitivity in eyes implanted with aspheric and spherical intraocular lenses. Ophthalmology 116:890–895PubMedCrossRef Trueb PR, Albach C, Montés-Micó R, Ferrer-Blasco T (2009) Visual acuity and contrast sensitivity in eyes implanted with aspheric and spherical intraocular lenses. Ophthalmology 116:890–895PubMedCrossRef
6.
go back to reference Madrid-Costa D, Ruiz-Alcocer J, Pérez-Vives C, Ferrer-Blasco T, Montés-Micó R (2012) Visual simulation through different intraocular lenses using adaptive optics: effect of tilt and decentration. J Cataract Refract Surg 38:947–958PubMedCrossRef Madrid-Costa D, Ruiz-Alcocer J, Pérez-Vives C, Ferrer-Blasco T, Montés-Micó R (2012) Visual simulation through different intraocular lenses using adaptive optics: effect of tilt and decentration. J Cataract Refract Surg 38:947–958PubMedCrossRef
7.
go back to reference Madrid-Costa D, Pérez-Vives C, Ruiz-Alcocer J, Albarrán-Diego C, Montés-Micó R (2012) Visual simulation through different intraocular lenses in patients with previous myopic corneal ablation using adaptive optics: effect of tilt and decentration. J Cataract Refract Surg 38:774–786PubMedCrossRef Madrid-Costa D, Pérez-Vives C, Ruiz-Alcocer J, Albarrán-Diego C, Montés-Micó R (2012) Visual simulation through different intraocular lenses in patients with previous myopic corneal ablation using adaptive optics: effect of tilt and decentration. J Cataract Refract Surg 38:774–786PubMedCrossRef
8.
go back to reference Ruiz-Alcocer J, Pérez-Vives C, Madrid-Costa D, López-Gil N, Montés-Micó R (2012) Effect of simulated tilt and decentration on spherical aberration after hyperopic LASIK for different intraocular lenses. J Refract Surg 28:327–334PubMedCrossRef Ruiz-Alcocer J, Pérez-Vives C, Madrid-Costa D, López-Gil N, Montés-Micó R (2012) Effect of simulated tilt and decentration on spherical aberration after hyperopic LASIK for different intraocular lenses. J Refract Surg 28:327–334PubMedCrossRef
9.
go back to reference Ruiz-Alcocer J, Pérez-Vives C, Madrid-Costa D, García-Lázaro S, Montés-Micó R (2012) Depth of focus through different intraocular lenses in patients with different corneal profiles using adaptive optics visual simulation. J Refract Surg 28:406–412PubMedCrossRef Ruiz-Alcocer J, Pérez-Vives C, Madrid-Costa D, García-Lázaro S, Montés-Micó R (2012) Depth of focus through different intraocular lenses in patients with different corneal profiles using adaptive optics visual simulation. J Refract Surg 28:406–412PubMedCrossRef
10.
go back to reference Barbero S, Marcos S, Jiméno-Alfaro I (2003) Optical aberrations of intraocular lenses measured in vivo and in vitro. J Opt Soc Am A Opt Image Sci Vis 20:1841–1851PubMedCrossRef Barbero S, Marcos S, Jiméno-Alfaro I (2003) Optical aberrations of intraocular lenses measured in vivo and in vitro. J Opt Soc Am A Opt Image Sci Vis 20:1841–1851PubMedCrossRef
11.
go back to reference Ferrer-Blasco T, Montés-Micó R, Peixoto-de-Matos SC, González-Méijome JM, Cerviño A (2009) Prevalence of corneal astigmatism before cataract surgery. J Cataract Refract Surg 35:70–75PubMedCrossRef Ferrer-Blasco T, Montés-Micó R, Peixoto-de-Matos SC, González-Méijome JM, Cerviño A (2009) Prevalence of corneal astigmatism before cataract surgery. J Cataract Refract Surg 35:70–75PubMedCrossRef
12.
go back to reference Joannes L, Hough T, Hutsebaut X, Dubois X, Ligot R, Saoul B, Donink PV, De Coninck K (2010) The reproducibility of a new power mapping instrument based on the phase shifting schlieren method for the measurement of spherical and toric contact lenses. Contact Lens Anterior Eye 33:3–8PubMedCrossRef Joannes L, Hough T, Hutsebaut X, Dubois X, Ligot R, Saoul B, Donink PV, De Coninck K (2010) The reproducibility of a new power mapping instrument based on the phase shifting schlieren method for the measurement of spherical and toric contact lenses. Contact Lens Anterior Eye 33:3–8PubMedCrossRef
13.
go back to reference Joannes L, Dubois F, Legros JC (2003) Phase-shifting schlieren: high-resolution quantitative schlieren that uses the phase-shifting technique principle. Appl Opt 42:5046–5053PubMedCrossRef Joannes L, Dubois F, Legros JC (2003) Phase-shifting schlieren: high-resolution quantitative schlieren that uses the phase-shifting technique principle. Appl Opt 42:5046–5053PubMedCrossRef
14.
go back to reference López-Gil N, Howland HC, Howland B, Charman N, Applegate R (1998) Generation of third-order spherical and coma aberration using radially symmetric fourth-order lenses. J Opt Soc Am A 15:2563–2571CrossRef López-Gil N, Howland HC, Howland B, Charman N, Applegate R (1998) Generation of third-order spherical and coma aberration using radially symmetric fourth-order lenses. J Opt Soc Am A 15:2563–2571CrossRef
15.
go back to reference Guirao A, Williams DR, Cox IG (2001) Effect of rotation and translation on the expected benefit of an ideal method to correct the eye’s higher-order aberrations. J Opt Soc Am A 18:1003–1015CrossRef Guirao A, Williams DR, Cox IG (2001) Effect of rotation and translation on the expected benefit of an ideal method to correct the eye’s higher-order aberrations. J Opt Soc Am A 18:1003–1015CrossRef
16.
go back to reference López-Gil N, Castejón-Mochón JF, Fernández-Sánchez V (2009) Limitations of the ocular wavefront correction with contact lenses. Vis Res 49:1729–1737PubMedCrossRef López-Gil N, Castejón-Mochón JF, Fernández-Sánchez V (2009) Limitations of the ocular wavefront correction with contact lenses. Vis Res 49:1729–1737PubMedCrossRef
17.
go back to reference Rocha KM, Vabre L, Harms F, Chateau N, Krueger RR (2007) Effects of Zernike wavefront aberrations on visual acuity measured using electromagnetic adaptive optics technology. J Refract Surg 23:953–959PubMed Rocha KM, Vabre L, Harms F, Chateau N, Krueger RR (2007) Effects of Zernike wavefront aberrations on visual acuity measured using electromagnetic adaptive optics technology. J Refract Surg 23:953–959PubMed
18.
go back to reference Altmann GE, Nichamin LD, Lane SS, Pepose JS (2005) Optical performance of 3 intraocular lens designs in the presence of decentration. J Cataract Refract Surg 31:574–585PubMedCrossRef Altmann GE, Nichamin LD, Lane SS, Pepose JS (2005) Optical performance of 3 intraocular lens designs in the presence of decentration. J Cataract Refract Surg 31:574–585PubMedCrossRef
19.
go back to reference Mckelvie J, McArdle B, McGhee C (2011) The influence of tilt, decentration, and pupil size on the higher order aberration profile of aspheric intraocular lens. Ophthalmology 118:1724–1731PubMedCrossRef Mckelvie J, McArdle B, McGhee C (2011) The influence of tilt, decentration, and pupil size on the higher order aberration profile of aspheric intraocular lens. Ophthalmology 118:1724–1731PubMedCrossRef
Metadata
Title
Optical quality of aspheric toric intraocular lenses at different degrees of decentering
Authors
Cari Pérez-Vives
Teresa Ferrer-Blasco
David Madrid-Costa
Santiago García-Lázaro
Robert Montés-Micó
Publication date
01-06-2014
Publisher
Springer Berlin Heidelberg
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 6/2014
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-014-2629-z

Other articles of this Issue 6/2014

Graefe's Archive for Clinical and Experimental Ophthalmology 6/2014 Go to the issue